K261932 · Elekta Solutions AB · HAW · Aug 5, 2026 · Neurology
Device Facts
Record ID
K261932
Device Name
Leksell® Vantage Stereotactic System
Applicant
Elekta Solutions AB
Product Code
HAW · Neurology
Decision Date
Aug 5, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended purpose of Leksell® Vantage Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures; for example deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.
Device Story
Leksell® Vantage Stereotactic System is a mechanical neurosurgical instrument for coordinate referencing and skull fixation. Device comprises a head frame and semi-circular arc (190 mm radius). It uses Cartesian (x, y, z) coordinates to triangulate brain targets. Fiducial markers are registered to intracranial images (CT, MR, X-ray) or brain atlases. Used in access-controlled healthcare facilities by neurosurgeons. Provides support/guidance for surgical instruments (biopsy needles, electrodes, etc.) to reach targets. X-ray accessories (fiducial box, table fixations) enable target localization during angiography. System ensures mechanical accuracy of 0.9 mm. Benefits include precise, minimally invasive access to intracranial targets for therapeutic or diagnostic interventions.
Clinical Evidence
No clinical or animal testing performed. Substantial equivalence established via bench testing, including design verification, usability validation, and biocompatibility assessment of new X-ray accessories. Performance data demonstrates the device is as safe and effective as the predicate devices.
Technological Characteristics
Mechanical stereotactic instrument; 190 mm arc radius; Cartesian coordinate system. Materials assessed per ISO 10993-1. Includes head frame, fixation pins, and X-ray fiducial accessories. Imaging compatibility: CT, MR, X-ray. Mechanical accuracy: 0.9 mm. No electronic or software-based sensing/actuation; purely mechanical guidance system.
Indications for Use
Indicated for patients requiring stereotactic neurosurgical procedures, including deep brain stimulation, lesioning, biopsies, targeted injections, aspirations, and minimally invasive tumor treatments, requiring head fixation and target localization.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
August 5, 2026
Elekta Solutions AB
Jesper Bruske
Regulatory Affairs Engineer
Hagaplan 4
Stockholm, 113 68
Sweden
Re: K261932
Trade/Device Name: Leksell® Vantage Stereotactic System
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: HAW
Dated: June 9, 2026
Received: June 9, 2026
Dear Jesper Bruske:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261932 - Jesper Bruske
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance).
{2}
K261932 - Jesper Bruske
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# JAIME RABEN -S
Jaime Raben, PhD
Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K261932 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| Leksell® Vantage Stereotactic System | | | |
| Please provide your Indications for Use below. | | | ? |
| The intended purpose of Leksell® Vantage Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures; for example deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☐ Adults (22 years old and greater) | | ? |
{4}
K261932
Page 1 of 5
[LOGO]
### Traditional 510(k) Summary (21 CFR § 807.92)
### I. Submitter
| Address | Elekta Solutions AB Hagaplan 4 113 68 Stockholm Sweden |
| --- | --- |
| Contact | Jesper Bruske Regulatory Affairs Engineer +46707336376 jesper.bruske@elekta.com |
| 510(k) Number | TBD |
| Date Prepared | 2025-06-09 |
### II. Device
| Trade Name | Leksell® Vantage Stereotactic System |
| --- | --- |
| Product Classification | Class II |
| Common Name | Neurological Stereotaxic Instrument |
| Regulation Number | 882.4560 |
| Product Code | HAW |
### III. Predicate Device
| Predicate # | K190887 (Primary Predicate) K152558 (Predicate) |
| --- | --- |
| Predicate Trade Name | Leksell® Vantage Stereotactic System (Primary Predicate) Leksell® Stereotactic System (Predicate) |
| Predicate Product Code | HAW |
### IV. Device Description Summary
The Leksell® Vantage Stereotactic System is a medical device used for minimally invasive neurosurgical procedures. It enables coordinate referencing of the brain and fixation of the patient's skull during image acquisition and treatment. The coordinate referencing enables target localization and accurate stereotactic treatment of brain targets; such as aspirations, biopsies, electrode placements, injections of cells & drugs, heamatoma evacuations as well as lesioning using a number of methodologies.
### V. Intended Use/ Indications for Use
The intended purpose of the Leksell® Vantage Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures; for example deep brain stimulation, lesioning, biopsies, target injections, aspirations and minimal invasive tumor treatments.
Elekta Solutions AB
LVSS Traditional 510(k)
June 2026
Elekta Solutions AB | Box 7593 | SE-103 93 Stockholm | Sweden
Tel: +46 8 587 254 00 | Fax: +46 8 587 255 00 | www.elekta.com | Reg No. 559157-5286
{5}
Elekta
K261932
Page 2 of 5
## VI. Indications for Use Comparison
The intended use of the subject device and the primary predicate device (K190887) are the same. The addition of the X-ray accessories and Starburst Adapters does not affect the intended use.
The intended use of X-ray Fiducial Box and X-ray Table Fixations is the same as the predicate device (K152558)
| Subject Device | Primary Predicate Device K190887 | Predicate Device K152558 |
| --- | --- | --- |
| The intended purpose of Leksell® Vantage Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments. | The Intended Purpose of Leksell® Vantage™ Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments. | The Leksell Stereotactic System® with Reusable Fixation Screws is a system intended for localization and diagnosis of intracranial disorders and their surgical treatment, including radiosurgery and stereotactic radiation therapy. |
## VII. Technological Comparison
The Leksell® Vantage Stereotactic System is a medical device used for minimally invasive neurosurgical procedures. It enables coordinate referencing of the brain and fixation of the patient's skull during image acquisition and treatment. The coordinate referencing enables target localization and accurate stereotactic treatment of brain targets; such as aspirations, biopsies, electrode placements, injections of cells & drugs, hematoma evacuations as well as lesioning using a number of methodologies.
The major change of the Leksell® Vantage Stereotactic System (subject device), as compared to the primary predicate device Leksell® Vantage Stereotactic System (K190887) is addition of X-ray accessories.
The fundamental technical characteristics of the subject device Leksell® Vantage Stereotactic System have not changed and are substantially equivalent to its predicate device Leksell® Vantage Stereotactic System (K190887). The subject device is also substantially equivalent to predicate device (K152558) regarding usage of X-ray accessories.
Elekta Solutions AB
LVSS Traditional 510(k)
June 2026
{6}
K261932
Page 3 of 5

| Item category description | Item identifier | Subject device | Primary Predicate device K190887 | Predicate device K152558 |
| --- | --- | --- | --- | --- |
| Trade Name | / | Leksell® Vantage Stereotactic System | Leksell® Vantage™ Stereotactic System | Leksell Stereotactic System® |
| Class | / | Class II | Class II | Class II |
| Regulation | / | 21 CFR 882.4560 | 21 CFR 882.4560 | 21 CFR 882.4560 |
| Product Code | / | HAW | HAW | HAW |
| Description | / | Head Frame and semi-circular arc system used for the coordinate referencing of targets in the brain and the guidance of neurosurgical instruments to these brain targets. | Head Frame and semi-circular arc system used for the coordinate referencing of targets in the brain and the guidance of neurosurgical instruments to these brain targets. | Head Frame and semi-circular arc system used for the coordinate referencing of targets in the brain and the guidance of neurosurgical instruments to these brain targets. |
| Use Environment | / | Access-controlled Healthcare facilities | Access-controlled Healthcare facilities | Access-controlled Healthcare facilities |
| Indications for Use / Intended Use | / | The intended purpose of Leksell® Vantage Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments. | The Intended Purpose of Leksell® Vantage™ Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments. | The Leksell Stereotactic System® with Reusable Fixation Screws is a system intended for localization and diagnosis of intracranial disorders and their surgical treatment, including radiosurgery and stereotactic radiation therapy. |
| Principles of Use | / | Center of the arc principle using an arc radius of 190 mm. Cartesian coordinates applied to the brain of the patient by the use of a head frame fixated to the skull. • comprises a headframe and arc, • is target centered • uses Cartesian (x,y,z) coordinates to triangulate target position in frame • utilizes fiducial markers that can be registered to a stereotactic brain atlas and/or intracranial image, • provides support and accessories for surgical instruments to the target | Center of the arc principle using an arc radius of 190 mm. Cartesian coordinates applied to the brain of the patient by the use of a head frame fixated to the skull. • comprises a headframe and arc, • is target centered • uses Cartesian (x,y,z) coordinates to triangulate target position in frame • utilizes fiducial markers that can be registered to a stereotactic brain atlas and/or intracranial image, • provides support and accessories for surgical instruments to the target | Center of the arc principle using an arc radius of 190 mm. Cartesian coordinates applied to the brain of the patient by the use of a head frame fixated to the skull. • comprises a headframe and arc, • is target centered • uses Cartesian (x,y,z) coordinates to triangulate target position in frame • utilizes fiducial markers that can be registered to a stereotactic brain atlas and/or intracranial image, • provides support and accessories for surgical instruments to the target, |
| Technical Features | Coordinate System Fixation | Coordinate System: Leksell Coordinate System using Cartesian coordinates Fixation: Head Frame applied to patient skull with four fixation pins | Coordinate System: Leksell Coordinate System using Cartesian coordinates Fixation: Head Frame applied to patient skull with four fixation pins | Coordinate System: Leksell Coordinate System using Cartesian coordinates Fixation: Head Frame applied to patient skull with four fixation pins. |
Elekta Solutions AB
LVSS Traditional 510(k)
June 2026
{7}
K261932
Page 4 of 5

| Technical Features | Fixation pins | Fixation Pins: Single use Fixation Pins and Inserts that secures the Vantage head frame to the patient skull using Leksell Vantage Keys | Fixation Pins: Single use Fixation Pins and Inserts that secures the Vantage head frame to the patient skull using Leksell Vantage Keys | Fixation Pins: Titanium Fixation Pins that secure the Leksell Coordinate Frame G to the patient skull using Instrument Screw Drivers |
| --- | --- | --- | --- | --- |
| Imaging modality | / | CT, MR and X-ray | CT and MR | CT, MR and X-ray |
| Technical Features | X-ray accessories | The X-ray Fiducial box is used during stereotactic X-ray imaging to impose fiducials on images of the patient. The fiducials are used for determining target coordinates during angiography. The X-Ray fiducial box consists of four labeled fiducial plates (RIGHT and LEFT, ANTERIOR and POSTERIOR). The fiducials are made of high-density material to impose markers that are clearly visible on angiograms. X-ray table fixations are provided to prevent motions and keep the patient attached to the table during imaging. | / | The X-Ray Indicator with Markers is an accessory for target localization in stereotactic neurosurgery and radiosurgery. It is used in conjunction with Leksell Coordinate Frame G. The purpose of the indicator is to impose fiducials on X-ray films of the patient. The fiducials are used for target localization X-ray table fixations are provided to prevent motions and keep the patient attached to the table during imaging. |
| Accuracy | / | Mechanical accuracy 0.9 mm | Mechanical accuracy 0.9 mm | Mechanical accuracy 0.9 mm |
| Biocompatibility | / | Materials have been assessed based on ISO 10993-1 | Materials have been assessed based on ISO 10993-1 | Materials have been assessed based on ISO 10993-1 |
| Accessories | / | Neurosurgical Instruments with working length of 190 mm | Neurosurgical Instruments with working length of 190 mm | Neurosurgical Instruments with working length of 190 mm |
Elekta Solutions AB
LVSS Traditional 510(k)
June 2026
{8}
K261932
Page 5 of 5
[LOGO]
### VIII. Summary of Performance Testing (Non-Clinical)
Verification activities have been performed to demonstrate that the X-ray accessories work as intended according to the system requirements for the Leksell Vantage Stereotactic System and its relevant subsystem requirements and that there is no unforeseen side-effect to the existing product base of Leksell Vantage Stereotactic System resulting in the changes made.
Performance testing, including design and usability validation of the system, has been performed by competent and professionally qualified personnel to ensure that the product fulfils the intended use and user needs.
Results from verification and validation testing demonstrate that conformance to applicable technical requirement specification, user needs have been met and that the device functions as intended.
The new X-ray accessories have also been evaluated for biocompatibility.
### IX. Summary of Performance Testing (Clinical)
No animal or clinical tests were performed to establish substantial equivalence with the predicate devices. The performance data demonstrate that the subject device is as safe and effective and performs as well as the predicate devices (K190887 and K152558).
### X. Substantial Equivalence Conclusion
The intended use, the basic functionality, and the fundamental technological features of the Leksell Vantage Stereotactic System did not change.
The subject device remains as safe and effective as the primary predicate device (K190877) and predicate device (K152558).
The subject device does not raise different questions of safety and effectiveness than the predicate devices and safety and effectiveness is demonstrated with the same methods (FDA guidance and recognized consensus standards) and with the same results of the predicate devices. This shows that the subject device is substantially equivalent to the primary predicate device (K190877) in intended use, principles of operation, technological characteristics, performance, and labelling. Also, the subject device is also substantially equivalent to predicate device (K152558) regarding usage of X-ray accessories.
Elekta Solutions AB
LVSS Traditional 510(k)
June 2026
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.