Plum Vagiceuticals™ Vaginal Moisturizing Suppositories are a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
Device Story
Plum Vagiceuticals™ are non-hormonal, glyceride-based vaginal suppositories for OTC use. Device consists of 1.8g solid, off-white suppositories containing 1% sodium hyaluronate, lipid excipients, and mucoadhesive agents. Upon intravaginal insertion, suppository melts at body temperature (30°C-37°C); viscous lipid matrix disperses over mucosal surface. In presence of vaginal fluid, formulation undergoes self-emulsification to form a viscous emulsion that adheres to mucosa, providing lubrication and alleviating dryness. Cleared naturally by gravity and vaginal self-cleaning mechanisms. Intended to enhance comfort during intimate sexual activity. Not compatible with natural rubber latex, polyurethane, or polyisoprene condoms.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), vaginal irritation (ISO 10993-23), and acute systemic toxicity (ISO 10993-11). Results confirmed device is non-cytotoxic, non-irritating, non-sensitizing, and not systemically toxic. Shelf-life validated via accelerated aging per ASTM F1980-16.
Indicated for individuals experiencing vaginal dryness requiring a personal lubricant to moisturize, lubricate, and enhance comfort during intimate sexual activity. Not compatible with natural rubber latex, polyurethane, or polyisoprene condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
July 6, 2026
PCCA, Inc.
% Dave Yungvirt
CEO
Third Party Review Group, LLC
1887 Whitney Mesa Dr.
Suite 1960
Henderson, Nevada 89014
Re: K261909
Trade/Device Name: Plum Vagiceuticals™ Vaginal Moisturizing Suppositories
Regulation Number: 21 CFR 884.5300
Regulation Name: Condom
Regulatory Class: II
Product Code: NUC
Received: June 8, 2026
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261909 - Dave Yungvirt
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K261909 - Dave Yungvirt
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
VASUDHA C. SHUKLA -S
For
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261909
Device Name
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories
Indications for Use (Describe)
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories are a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
Type of Use (Select one or both, as applicable)
☐
Prescription Use (Part 21 CFR 801 Subpart D)
☑
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
PCCA, Inc
K261909
Page 1 of 6
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories
### 510(k) Summary
### Plum Vagiceuticals™ Vaginal Moisturizing Suppositories
### K261909
### I. General Information
Submitter/Owner:
PCCA
9901 South Wilcrest Drive
Houston, Texas, 77099
United States
Contact Person:
Daniel Banov
Vice President of Research and Development
PCCA
Phone: 832-295-4426
Email: DBanov@pccarx.com
July 6, 2026
Preparation Date:
### II. Device Information
Trade Name:
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories
Common Name:
Personal Lubricant
Regulation Number:
21 CFR 884.5300
Regulation Name:
Condom
Regulatory Class:
Class II
Product Code:
NUC (Lubricant, Personal)
### III. Predicate Device
Revaree Plus Vaginal Suppositories (K213220)
The predicate device has not been subject to a design-related recall.
Page 1
{5}
K261909
Page 2 of 6
PCCA, Inc
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories
#### IV. Device Description
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories are non-hormonal, non-sterile, glyceride-based personal lubricant, designed for over-the-counter use to alleviate vaginal dryness, enhance comfort during intimate activity, and support the body's natural lubrication.
The suppositories are solid, off-white, 1.8 g each, and the shape is designed for easy intravaginal administration. Each vaginal suppository is individually packaged in a polyvinyl chloride (PVC) /polyethylene (PE) blister and marketed in cartons containing 10 vaginal suppositories (two blister packs of 5 vaginal suppositories each).
Each vaginal suppository contains 1% sodium hyaluronate in combination with lipid excipients, self-emulsifying agent, and additional hydrating and mucoadhesive agents. The suppositories do not contain hormones or pharmacologically active drug substances.
Upon insertion into the vaginal cavity, the suppository melts at body temperature. The viscous lipid matrix disperses over the mucosal surface and provides lubrication in a dry environment. In the presence of vaginal fluid, the formulation undergoes self-emulsification, forming a viscous emulsion. The emulsion system remains in contact with the vaginal mucosa for the intended period of action and is ultimately cleared by gravity and the natural self-cleaning mechanisms of the vagina.
The specifications for Plum Vagiceuticals™ Vaginal Moisturizing Suppositories are provided in Table 1.
Page 2
{6}
K261909
Page 3 of 6
PCCA, Inc
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories
Table 1. Device specifications
| Parameter | Test Method | Specification |
| --- | --- | --- |
| Appearance | Visual | Solid |
| Color | Visual | White to off-white |
| Odor | Organoleptic | Practically odorless |
| Osmolality | USP <785> Osmolality/ after dilution | 100-300 mOsm/kg (after taking account of dilution factor of 1:10) |
| Disintegration Time | Eu.Ph 2.9.2 | ≤ 30 minutes |
| Melting Point | USP <741> Class III | 30°C - 37°C |
| Water Activity | USP <1112> | < 0.600 |
| Average Weight | Gravimetry | 1.8 g |
| Content of Sodium Hyaluronate | HPLC | 10 mg/g |
| TAMC | USP <61> | ≤ 100 cfu/g |
| TYMC | USP <61> | ≤ 10 cfu/g |
| Pseudomonas Aeruginosa, Staphylococcus Aureus, Candida Albicans | USP <62> | Absent /g |
### V. Indications for Use
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories are a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
### VI. Substantial Equivalence Discussion
A comparison of the intended use and key technological characteristics of the subject and predicate device is provided in Table 2 below.
Page 3
{7}
K261909
Page 4 of 6
PCCA, Inc
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories
Table 2. Intended Use and Technological Characteristics of Subject Device Compared to Predicate Device
| | Plum VagiceuticalsTM Vaginal Moisturizing Suppositories | Revaree Plus Vaginal Suppositories (K213220) | Comparison |
| --- | --- | --- | --- |
| Indications for Use | Plum VagiceuticalsTM Vaginal Moisturizing Suppositories are a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms. | Revaree plus vaginal suppositories are a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms. | Similar |
| Over the counter | Yes | Yes | Same |
| Base type | Fat-based | Glyceride | Similar |
| Ingredients | Hard fat, stearoyl polyoxyl-32 glycerides, polyoxyl 20 cetostearyl ether, sodium hyaluronate, Hippophae Rhamnoides Fruit Oil, sodium polyglutamate, sh-polypeptide-123, tocopheryl acetate, zinc hydrolyzed hyaluronate | HA, glycerides, and sweet almond oil | Different |
| Appearance | Off-white opaque vaginal suppositories | Off-white opaque vaginal suppositories | Same |
| Average Weight | 1.8 g | 2 g | Different |
Page 4
{8}
K261909
Page 5 of 6
PCCA, Inc
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories
| TAMC | ≤100 cfu/g | ≤100 cfu/g | Same |
| --- | --- | --- | --- |
| TYMC | ≤10 cfu/g | ≤10 cfu/g | Same |
| Absence of Pseudomonas Aeruginosa, Staphylococcus Aureus, Candida Albicans | Absent/g | Absent/g | Same |
| Condom Compatibility | Not compatible with natural rubber latex, polyisoprene, and polyurethane condoms | Not compatible with natural rubber latex, polyisoprene, and polyurethane condoms | Same |
| Packaging | PVC/PE blister of 5 vaginal suppositories | PVC/PE blister of 3 or 5 vaginal suppositories | Same |
The subject and predicate device have the similar indications for use and have the same intended use – to provide lubrication during intimate sexual activity. The subject and predicate devices have different technological characteristics, including different formulations and device specifications. The different technological characteristics do not raise different questions of safety and effectiveness.
### VII. Summary of Non-Clinical Performance Data
# Biocompatibility
Biocompatibility testing on the subject lubricant was performed in accordance with the 2023 FDA guidance document Use of International Standard ISO 10993-1, “Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process” and ISO 10993-1:2018 as follows:
- Cytotoxicity (ISO 10993-5:2009)
- Guinea Pig Maximization Sensitization Test (ISO 10993-10: 2021)
- Vaginal Irritation (ISO 10993-12: 2021; ISO 10993-23: 2021)
- Acute Systemic Toxicity (ISO 10993-11:2017)
Page 5
{9}
K261909
Page 6 of 6
PCCA, Inc
Plum Vagiceuticals™ Vaginal Moisturizing Suppositories
The results of this testing demonstrate that the subject device is non-cytotoxic, non-irritating, non-sensitizing, and not systemically toxic.
# Shelf-Life
The shelf-life of Plum Vagiceuticals™ Vaginal Moisturizing Suppositories is 12 months. This is based on the results of an intermediate accelerated aging study conducted at 30°C in accordance with ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
# Condom Compatibility
Condom compatibility testing was not conducted for the subject device. Therefore, Plum Vagiceuticals™ Vaginal Moisturizing Suppositories are not compatible with natural rubber latex, polyisoprene, and polyurethane condoms
# VIII. Conclusions
The results of the testing described above demonstrate that Plum Vagiceuticals™ Vaginal Moisturizing Suppositories are as safe and effective as the predicate device and supports a determination of substantial equivalence.
Page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.