# Trevo Trak 21 Microcatheter (K261889)

_Stryker Neurovascular · QJP · Jul 1, 2026 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K261889

## Device Facts

- **Applicant:** Stryker Neurovascular
- **Product Code:** [QJP](/productcode/QJP.md)
- **Decision Date:** Jul 1, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Microcatheter is indicated to provide vascular access for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature, including large intracranial vessels (e.g., internal carotid artery [ICA] and middle cerebral artery [MCA]), during diagnostic and/or therapeutic procedures.

## Device Story

Single-lumen, braided shaft, variable stiffness microcatheter; includes radiopaque distal markers and luer hub; hydrophilic coating on distal 100 cm reduces friction. Used in neurovascular procedures; advanced over guidewire to target location; facilitates delivery of diagnostic/therapeutic devices or fluids. Operated by physicians in clinical settings. Provides vascular access; enables selective placement within intracranial vessels; aids fluoroscopic visualization via radiopaque components. Supplied sterile for single use with rotating hemostasis valve.

## Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing and design/material similarity to the predicate device.

## Technological Characteristics

Single-lumen braided shaft (stainless steel); PTFE inner layer; polyurethane hub; platinum/iridium marker bands; polyolefin strain relief; hydrophilic outer coating. EO sterilized. 2.4F/2.0F outer diameter; 0.021" inner diameter; 162 cm length. Standalone device.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Trevo Trak 21 Microcatheter ([K211594](/device/K211594.md))

## Reference Devices

- Trevo Trak 21 Microcatheter ([K192122](/device/K192122.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD & DRUG ADMINISTRATION

July 1, 2026

Stryker Neurovascular
Jesse Nelson
Staff Regulatory Affairs Specialist
47900 Bayside Parkway
Fremont, California 94538

Re: K261889

Trade/Device Name: Trevo Trak 21 Microcatheter

Regulation Number: 21 CFR 870.1250

Regulation Name: Percutaneous Catheter

Regulatory Class: Class II

Product Code: QJP, DQO

Dated: June 5, 2026

Received: June 5, 2026

Dear Jesse Nelson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K261889 - Jesse Nelson

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K261889 - Jesse Nelson

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

NAIRA MURADYAN -S

Naira Muradyan, PhD

Assistant Director

DHT5A: Division of Neurosurgical,

Neurointerventional, and

Neurodiagnostic Devices

OHT5: Office of Neurological and

Physical Medicine Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

# **Indications for Use**

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K261889

Device Name
Trevo Trak 21 Microcatheter

Indications for Use (Describe)

The Microcatheter is indicated to provide vascular access for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature, including large intracranial vessels (e.g., internal carotid artery [ICA] and middle cerebral artery [MCA]), during diagnostic and/or therapeutic procedures.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)

☐ Over-The-Counter Use (21 CFR 801 Subpart C)

**CONTINUE ON A SEPARATE PAGE IF NEEDED.**

This section applies only to requirements of the Paperwork Reduction Act of 1995.

**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*

FORM FDA 3881 (8/23)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}

510(k) Summary – K261889

Premarket Notification, Special 510(k)

Trevo Trak 21 Microcatheter

Date Prepared: June 29, 2026

Submitter: Stryker Neurovascular
47900 Bayside Parkway
Fremont, CA 94538
Facility Registration #3008853977

Contact: Jesse Nelson
Staff Regulatory Affairs Specialist
Tel (801) 809-3895
E-mail: Jesse.Nelson@stryker.com

Trade Name: Trevo Trak™ 21 Microcatheter
Common Name: Catheter, Percutaneous, Neurovasculature
Regulation Name: Percutaneous Catheter, 21 CFR 870.1250 – Class II
Product Code: QJP

Trade Name: Trevo Trak™ 21 Microcatheter
Common Name: Catheter, intravascular, diagnostic
Regulation Name: Catheter, intravascular, diagnostic, 21 CFR 870.1200 – Class II
Product Code: DQO

Legally Marketed Predicate Device:

|  Name of Predicate Device | 510(k) Number  |
| --- | --- |
|  Trevo Trak 21 Microcatheter | K211594 (Cleared 26-Nov-2021)  |

Stryker Neurovascular

Page 1 of 5

{5}

510(k) Summary – K261889

Premarket Notification, Special 510(k)

Trevo Trak 21 Microcatheter

## Device Description

The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating on the distal 100 cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. Device dimensions and configuration are shown on the product label. A rotating hemostasis valve with side-arm adapter is provided with each microcatheter.

## Indications for Use

The Microcatheter is indicated to provide vascular access for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature, including large intracranial vessels (e.g., internal carotid artery [ICA] and middle cerebral artery [MCA]), during diagnostic and/or therapeutic procedures.

## Technological Characteristics and Product Feature Comparison

Stryker Neurovascular has demonstrated the Trevo Trak™ 21 Microcatheter is substantially equivalent to the Predicate device, Trevo Trak 21 Microcatheter (K211594), based on the similar intended use, the same materials, same design, and the same fundamental operating principles. A comparison of the Subject device with the Predicate device is summarized in Table 1 below.

Table 1: Product Feature Comparison of Subject Device to Predicate Device

|  Detail | Subject Device (K261889) | Predicate Device (K211594)  |
| --- | --- | --- |
|  Manufacturer | Stryker Neurovascular | Same  |
|  510(k) Number | K261889 | K211594  |
|  Device Trade Name | **Trevo Trak 21 Microcatheter** | Same  |
|  Regulation Number | 21 CFR 870.1250 21 CFR 870.1200 | Same  |
|  Regulation Name | Percutaneous Catheter Diagnostic Intravascular Catheter | Same  |

Stryker Neurovascular

Page 2 of 5

{6}

510(k) Summary – K261889

Premarket Notification, Special 510(k)

Trevo Trak 21 Microcatheter

|  Classification | II | Same  |
| --- | --- | --- |
|  Product Code | QJP, DQO | DQY, QJP, DQO  |
|  Intended Use/Indication for Use | INTENDED USEThe Microcatheter is intended for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature during diagnostic and/or therapeutic procedures.INDICATIONS FOR USEThe Microcatheter is indicated to provide vascular access for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature, including large intracranial vessels (e.g., internal carotid artery [ICA] and middle cerebral artery [MCA]), during diagnostic and/or therapeutic procedures. | Intended Use/Indication for UseThe Microcatheter is indicated for use in the selective placement of devices and/or fluids, such as contrast media, into the peripheral, coronary, and neuro vasculature during diagnostic and/or therapeutic procedures.  |
|  Device Description | The Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coatings on the distal 100cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. Device dimensions and configuration are shown on the product label. | The Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating on the distal 100cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. Device dimensions and configuration are shown on the product label. A rotating hemostatic  |

Stryker Neurovascular

Page 3 of 5

{7}

510(k) Summary – K261889

Premarket Notification, Special 510(k)

Trevo Trak 21 Microcatheter

|   | A rotating hemostatic valve with side-arm adapter is provided with each microcatheter. The Trevo Trak 21 Microcatheter provides vascular access to the neurovascular for selective placement of devices and/or the delivery of diagnostic agents into the neurovasculature during diagnostic and/or therapeutic endovascular procedures when used as intended. | valve with side-arm adapter is provided with each microcatheter.  |
| --- | --- | --- |
|  Accessory Devices | Rotating Hemostasis Valve (RHV) packaged with device | Same  |
|  Outer Jacket | Polymeric microcatheter | Same  |
|  Shaft Braid | Stainless Steel | Same  |
|  Strain Relief | Polyolefin | Same  |
|  Inner Layer | PTFE | Same  |
|  Catheter Hub | Polyurethane | Same  |
|  Marker Band | Platinum/Iridium | Same  |
|  Adhesive | Acrylic (Acrylated Urethane) | Same  |
|  Outer Jacket Coating | Hydrophilic Coating | Same  |
|  Labeled Shaft Outer Diameter | 2.4F/2.0F | Same  |
|  Labeled Shaft Inner Diameter | 0.021” | Same  |
|  Effective Length | 162 cm | Same  |
|  Packaging Materials | HDPE Packaging Hoop, Tyvek/Film Pouch, SBS Carton | Same  |

Stryker Neurovascular

Page 4 of 5

{8}

## 510(k) Summary – K261889

Premarket Notification, Special 510(k)

Trevo Trak 21 Microcatheter

|  and Configuration |  |   |
| --- | --- | --- |
|  Sterilization Method | EO Sterilization | Same  |
|  How Supplied | Single Use/Sterile | Same  |
|  Principles of Operation | The device is advanced into the neurovascular vasculature over an appropriately sized guide wire. Once the microcatheter is inserted, the catheter can be advanced through the neurovasculature to the desired location. | The device is advanced into the vasculature over an appropriately sized guide wire. Once the microcatheter is inserted, the catheter can be advanced through the vasculature to the desired location  |
|  Shelf life | 2-year | Same  |

## Performance Testing

The K261889 submission does not propose any modifications to the subject device design, materials, or manufacturing processes. The proposed changes to the intended use and indications for use, limiting device use to neurovasculature only, did not warrant new testing.

## Shelf-Life Testing

Shelf life testing previously conducted for the Trevo Trak 21 Microcatheter in K192122 was leveraged since there are no changes to device materials, design, or manufacturing processes as a result of the intended use, indications for use and labeling changes. As with the Predicate device, the Subject device is labeled with a 2-year shelf life.

## Conclusion

Stryker Neurovascular has demonstrated the Trevo Trak 21 Microcatheter is substantially equivalent to the Predicate device (**K211594**) based on similar intended use, same materials, same fundamental design, and the same operating principles. The amended intended use and indications for use described in this premarket notification limit the device use to the neurovasculature, compared to prior/predicate’s use in peripheral, coronary and neurovasculature, and do not introduce any new risks and have no impact on the existing risks associated with the device use.

Stryker Neurovascular

Page 5 of 5

---

**Source:** [https://fda-staging.innolitics.com/device/K261889](https://fda-staging.innolitics.com/device/K261889)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).