The Microcatheter is indicated to provide vascular access for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature, including large intracranial vessels (e.g., internal carotid artery [ICA] and middle cerebral artery [MCA]), during diagnostic and/or therapeutic procedures.
Device Story
Single-lumen, braided shaft, variable stiffness microcatheter; includes radiopaque distal markers and luer hub; hydrophilic coating on distal 100 cm reduces friction. Used in neurovascular procedures; advanced over guidewire to target location; facilitates delivery of diagnostic/therapeutic devices or fluids. Operated by physicians in clinical settings. Provides vascular access; enables selective placement within intracranial vessels; aids fluoroscopic visualization via radiopaque components. Supplied sterile for single use with rotating hemostasis valve.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing and design/material similarity to the predicate device.
Indicated for patients requiring vascular access for selective placement of devices or fluids (e.g., contrast media) within the neurovasculature, including large intracranial vessels like the ICA and MCA, during diagnostic or therapeutic procedures.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 1, 2026
Stryker Neurovascular
Jesse Nelson
Staff Regulatory Affairs Specialist
47900 Bayside Parkway
Fremont, California 94538
Re: K261889
Trade/Device Name: Trevo Trak 21 Microcatheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: QJP, DQO
Dated: June 5, 2026
Received: June 5, 2026
Dear Jesse Nelson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261889 - Jesse Nelson
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261889 - Jesse Nelson
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NAIRA MURADYAN -S
Naira Muradyan, PhD
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261889
Device Name
Trevo Trak 21 Microcatheter
Indications for Use (Describe)
The Microcatheter is indicated to provide vascular access for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature, including large intracranial vessels (e.g., internal carotid artery [ICA] and middle cerebral artery [MCA]), during diagnostic and/or therapeutic procedures.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
510(k) Summary – K261889
Premarket Notification, Special 510(k)
Trevo Trak 21 Microcatheter
Date Prepared: June 29, 2026
Submitter: Stryker Neurovascular
47900 Bayside Parkway
Fremont, CA 94538
Facility Registration #3008853977
Contact: Jesse Nelson
Staff Regulatory Affairs Specialist
Tel (801) 809-3895
E-mail: Jesse.Nelson@stryker.com
Trade Name: Trevo Trak™ 21 Microcatheter
Common Name: Catheter, Percutaneous, Neurovasculature
Regulation Name: Percutaneous Catheter, 21 CFR 870.1250 – Class II
Product Code: QJP
Trade Name: Trevo Trak™ 21 Microcatheter
Common Name: Catheter, intravascular, diagnostic
Regulation Name: Catheter, intravascular, diagnostic, 21 CFR 870.1200 – Class II
Product Code: DQO
Legally Marketed Predicate Device:
| Name of Predicate Device | 510(k) Number |
| --- | --- |
| Trevo Trak 21 Microcatheter | K211594 (Cleared 26-Nov-2021) |
Stryker Neurovascular
Page 1 of 5
{5}
510(k) Summary – K261889
Premarket Notification, Special 510(k)
Trevo Trak 21 Microcatheter
## Device Description
The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating on the distal 100 cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. Device dimensions and configuration are shown on the product label. A rotating hemostasis valve with side-arm adapter is provided with each microcatheter.
## Indications for Use
The Microcatheter is indicated to provide vascular access for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature, including large intracranial vessels (e.g., internal carotid artery [ICA] and middle cerebral artery [MCA]), during diagnostic and/or therapeutic procedures.
## Technological Characteristics and Product Feature Comparison
Stryker Neurovascular has demonstrated the Trevo Trak™ 21 Microcatheter is substantially equivalent to the Predicate device, Trevo Trak 21 Microcatheter (K211594), based on the similar intended use, the same materials, same design, and the same fundamental operating principles. A comparison of the Subject device with the Predicate device is summarized in Table 1 below.
Table 1: Product Feature Comparison of Subject Device to Predicate Device
| Detail | Subject Device (K261889) | Predicate Device (K211594) |
| --- | --- | --- |
| Manufacturer | Stryker Neurovascular | Same |
| 510(k) Number | K261889 | K211594 |
| Device Trade Name | **Trevo Trak 21 Microcatheter** | Same |
| Regulation Number | 21 CFR 870.1250 21 CFR 870.1200 | Same |
| Regulation Name | Percutaneous Catheter Diagnostic Intravascular Catheter | Same |
Stryker Neurovascular
Page 2 of 5
{6}
510(k) Summary – K261889
Premarket Notification, Special 510(k)
Trevo Trak 21 Microcatheter
| Classification | II | Same |
| --- | --- | --- |
| Product Code | QJP, DQO | DQY, QJP, DQO |
| Intended Use/Indication for Use | INTENDED USEThe Microcatheter is intended for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature during diagnostic and/or therapeutic procedures.INDICATIONS FOR USEThe Microcatheter is indicated to provide vascular access for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature, including large intracranial vessels (e.g., internal carotid artery [ICA] and middle cerebral artery [MCA]), during diagnostic and/or therapeutic procedures. | Intended Use/Indication for UseThe Microcatheter is indicated for use in the selective placement of devices and/or fluids, such as contrast media, into the peripheral, coronary, and neuro vasculature during diagnostic and/or therapeutic procedures. |
| Device Description | The Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coatings on the distal 100cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. Device dimensions and configuration are shown on the product label. | The Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating on the distal 100cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. Device dimensions and configuration are shown on the product label. A rotating hemostatic |
Stryker Neurovascular
Page 3 of 5
{7}
510(k) Summary – K261889
Premarket Notification, Special 510(k)
Trevo Trak 21 Microcatheter
| | A rotating hemostatic valve with side-arm adapter is provided with each microcatheter. The Trevo Trak 21 Microcatheter provides vascular access to the neurovascular for selective placement of devices and/or the delivery of diagnostic agents into the neurovasculature during diagnostic and/or therapeutic endovascular procedures when used as intended. | valve with side-arm adapter is provided with each microcatheter. |
| --- | --- | --- |
| Accessory Devices | Rotating Hemostasis Valve (RHV) packaged with device | Same |
| Outer Jacket | Polymeric microcatheter | Same |
| Shaft Braid | Stainless Steel | Same |
| Strain Relief | Polyolefin | Same |
| Inner Layer | PTFE | Same |
| Catheter Hub | Polyurethane | Same |
| Marker Band | Platinum/Iridium | Same |
| Adhesive | Acrylic (Acrylated Urethane) | Same |
| Outer Jacket Coating | Hydrophilic Coating | Same |
| Labeled Shaft Outer Diameter | 2.4F/2.0F | Same |
| Labeled Shaft Inner Diameter | 0.021” | Same |
| Effective Length | 162 cm | Same |
| Packaging Materials | HDPE Packaging Hoop, Tyvek/Film Pouch, SBS Carton | Same |
Stryker Neurovascular
Page 4 of 5
{8}
## 510(k) Summary – K261889
Premarket Notification, Special 510(k)
Trevo Trak 21 Microcatheter
| and Configuration | | |
| --- | --- | --- |
| Sterilization Method | EO Sterilization | Same |
| How Supplied | Single Use/Sterile | Same |
| Principles of Operation | The device is advanced into the neurovascular vasculature over an appropriately sized guide wire. Once the microcatheter is inserted, the catheter can be advanced through the neurovasculature to the desired location. | The device is advanced into the vasculature over an appropriately sized guide wire. Once the microcatheter is inserted, the catheter can be advanced through the vasculature to the desired location |
| Shelf life | 2-year | Same |
## Performance Testing
The K261889 submission does not propose any modifications to the subject device design, materials, or manufacturing processes. The proposed changes to the intended use and indications for use, limiting device use to neurovasculature only, did not warrant new testing.
## Shelf-Life Testing
Shelf life testing previously conducted for the Trevo Trak 21 Microcatheter in K192122 was leveraged since there are no changes to device materials, design, or manufacturing processes as a result of the intended use, indications for use and labeling changes. As with the Predicate device, the Subject device is labeled with a 2-year shelf life.
## Conclusion
Stryker Neurovascular has demonstrated the Trevo Trak 21 Microcatheter is substantially equivalent to the Predicate device (**K211594**) based on similar intended use, same materials, same fundamental design, and the same operating principles. The amended intended use and indications for use described in this premarket notification limit the device use to the neurovasculature, compared to prior/predicate’s use in peripheral, coronary and neurovasculature, and do not introduce any new risks and have no impact on the existing risks associated with the device use.
Stryker Neurovascular
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.