Total Hip Arthroplasty (THA) surgical planning and navigation
Fixed/static machine learning algorithms
—
—
—
—
—
—
Indications for Use
OTS Hip is indicated to enable planning of orthopedic surgical procedures based on CT and X-Ray medical imaging data of the patient anatomy. It is an intraoperative image-guided localization system that enables navigated surgery. It links a freehand probe, tracked by a passive marker sensor system, to virtual computer image space on a patient's preoperative image data being processed by the OTS platform.The system is indicated for orthopedic hip surgical procedures where a reference to a rigid anatomical structure, such as the pelvis, can be identified relative to a system generated model of the anatomy. The system aids the surgeon to accurately navigate a compatible prosthesis to the preoperatively planned position.The system is designed for orthopedic surgical procedures including:- Pre-operative planning of total hip arthroplasty (THA)- Intraoperative navigated surgery for THA using a posterior approach
Device Story
OTS Hip is an image-guided stereotaxic system for orthopedic hip replacement. Input: patient CT/X-ray DICOM images. Operation: software (OTS Hip Plan) creates 3D anatomical models; intraoperative navigation (OTS Hip Guide) uses optical tracking of passive markers on surgical instruments and patient anatomy to align prosthetic components with the preoperative plan. Used in clinics and ORs by orthopedic surgeons. Output: real-time visual guidance on a display, showing instrument/implant position relative to the planned target. Benefits: improved accuracy in prosthesis placement during THA. System includes camera, computer, footswitch, and various instrument adapters.
Clinical Evidence
Bench testing only. Includes mechanical robustness testing (simulated surgical use/impaction/reaming), compatibility validation for new instruments, electrical/mechanical safety testing per IEC 60601-1, and software verification/validation per IEC 62304.
Technological Characteristics
Stereotaxic system; optical tracking (NDI camera); software-based planning and navigation. Materials: biocompatible per ISO 10993. Sterilization: steam (reusable), gamma (single-use). Connectivity: DICOM-compliant. Software: fixed/static ML algorithms for landmark detection and segmentation. Power: mains.
Indications for Use
Indicated for orthopedic hip surgical procedures, specifically total hip arthroplasty (THA) using a posterior approach, in adult patients. Contraindicated for patients with implants in the treatment side.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 1, 2026
Ortoma AB
% John Smith
Partner
Hogan Lovells US Llp
Columbia Sq.
555 Thirteenth St., NW
Washington, District of Columbia 20004
Re: K261852
Trade/Device Name: OTS Hip
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: June 3, 2026
Received: June 3, 2026
Dear John Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261852 - John Smith
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261852 - John Smith
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tejen D. Soni -S
For
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261852 | ? |
| Please provide the device trade name(s). | | ? |
| OTS Hip | | |
| Please provide your Indications for Use below. | | ? |
| OTS Hip is indicated to enable planning of orthopedic surgical procedures based on CT and X-Ray medical imaging data of the patient anatomy. It is an intraoperative image-guided localization system that enables navigated surgery. It links a freehand probe, tracked by a passive marker sensor system, to virtual computer image space on a patient's preoperative image data being processed by the OTS platform.The system is indicated for orthopedic hip surgical procedures where a reference to a rigid anatomical structure, such as the pelvis, can be identified relative to a system generated model of the anatomy. The system aids the surgeon to accurately navigate a compatible prosthesis to the preoperatively planned position.The system is designed for orthopedic surgical procedures including:- Pre-operative planning of total hip arthroplasty (THA)- Intraoperative navigated surgery for THA using a posterior approach | | |
| Please select the types of uses (select one or both, as applicable). | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | Neonates/Newborns (Birth to < 29 days old)Infants (29 days old to < 2 years old)Children (2 years old to < 12 years old)Adolescents (12 years old to < 22 years old)Adults (22 years old and greater) | ? |
Page 1 of 1
{4}
[LOGO]
ORTOMA
Special 510(k) Submission
### Special 510(k) SUMMARY K261852
### Ortoma AB OTS Hip
**Submitter**
Ortoma AB
Falkenbergsgatan 3
412 85 Göteborg
Sweden
Phone: +46 73 688 88 70
Contact Person:
John Smith
Hogan Lovells, LLC
Columbia Square, 555 Thirteenth Street, NW
D.C., 20004
+1 (202) 637-3638
john.smith@hoganlovells.com
Date Prepared: June 30, 2026
Name of Device: OTS Hip
Common or Usual Name: Stereotaxic Instrument
Classification Name: 21 CFR Section 882.4560, Stereotaxic Instrument
Regulatory Class: II
Product Code: OLO
Predicate Devices
Ortoma AB, OTS Hip (K250086)
K261852 - Page 1 of 9
{5}
[LOGO]
ORTOMA
Special 510(k) Submission
## Device Description
OTS Hip is a system to support a surgeon with preoperative planning and intraoperative guidance during orthopedic hip joint replacement surgery.
The OTS Hip device is a modified device from the company's previously cleared OTS Hip (K250086).
OTS Hip is comprised of software systems and hardware components that work together to form a stereotaxic system. The system uses medical imaging data in DICOM format that is loaded into the system for access in the software that are part of the system.
OTS Hip software consists of OTS Hip Plan (OHP), which is a 3D preoperative planning software, and OTS Hip Guide (OHG) that provides intraoperative real-time navigation for the guidance of surgical tools and prosthetic components in relation to the preoperatively determined goal positions.
OHP is a software for preoperative planning prior to a THA (Total Hip Arthroplasty) surgery. OHP enables the orthopedic surgeon to prepare surgery by analyzing the patient anatomy in a 3D environment based on medical imaging data.
OHG imports the result from the preceding planning stage, a released plan, with the 3D model and planned data, from the database of the OTS system. In addition, OHG monitors the real-time information of the position of instruments and prosthetic components in a 3D environment by means of medical imaging data.
The components of the OHG device include a camera and computer stand with an electrical system to which a camera and a medical panel PC are attached, a footswitch, a keyboard, Tracers (passive markers), adapters that hold the Tracers and can be mounted to compatible surgical instruments and that are used for calibration, and tools and instruments that are used during surgery.
The OTS Hip is compatible with the following components:
- PINN GB OFFSET GRATER HANDLE, DePuy Synthes 2550-00-100
- EMPHASYS Offset reamer, DePuy Synthes 4811-00-510
- Greatbatch Offset Cup Impactor, DePuy Synthes 2550-00-115
- PINNACLE straight impactor, DePuy Synthes 2217-50-041
- EMPHASYS Straight impactor, DePuy Synthes 4812-01-150
- KINCISE™ PINNACLE Straight Shell Impactor, DePuy Synthes 2000-02-002 (long)
- KINCISE™ PINNACLE Straight Shell Impactor, DePuy Synthes 2000-02-012 (short)
- KINCISE™ EMPHASYS Straight Shell Impactor, DePuy Synthes 2000-03-001 (long)
- KINCISE™ EMPHASYS Straight Shell Impactor, DePuy Synthes 2000-03-012 (short)
- KINCISE™ EMPHASYS Offset Shell Impactor, DePuy Synthes 2000-03-002
- KINCISE™ PINNACLE Offset Shell Impactor, DePuy Synthes 2000-02-003
- Crossback Acetabular Reamer Head, DePuy Synthes, 2550-00-0XX (Size 36mm to 80mm)
- QUICKSET Acetabular Grater Head, DePuy Synthes 2440-00-5XX (Size 36mm to 80mm)
K261852 - Page 2 of 9
{6}
[LOGO]
ORTOMA
Special 510(k) Submission
### Indications for Use
OTS Hip is indicated to enable planning of orthopedic surgical procedures based on CT and X-Ray medical imaging data of the patient anatomy. It is an intraoperative image-guided localization system that enables navigated surgery. It links a freehand probe, tracked by a passive marker sensor system, to virtual computer image space on a patient's preoperative image data being processed by the OTS platform.
The system is indicated for orthopedic hip surgical procedures where a reference to a rigid anatomical structure, such as the pelvis, can be identified relative to a system generated model of the anatomy. The system aids the surgeon to accurately navigate a compatible prosthesis to the preoperatively planned position.
The system is designed for orthopedic surgical procedures including:
- Pre-operative planning of total hip arthroplasty (THA)
- Intraoperative navigated surgery for THA using a posterior approach
### Additional Considerations for Use
The device should not be used for patients with implants in the treatment side.
### Summary of Technological Characteristics
Both the predicate device and the subject device enable pre-operative planning and navigation of prosthetic components. OTS Hip is comprised of software systems and hardware components that work together to form a stereotaxic system.
Like the predicate device, the subject device includes software for pre-operative planning of orthopedic prosthetic components. The predicate device and the subject device have the same workflow for anatomical landmarks and planning of implant size and positions. Both devices detect landmarks and perform segmentation using fixed/static machine learning (ML) algorithms and generate a 3D model based on the segmentation.
Like the predicate device, the subject device enables intraoperative image-guided navigated surgery using the OHG software and hardware components. The subject device and the predicate device include hardware components and a software for real-time navigation of surgical instruments and implants relative to the patient. Both devices use similar tracking arrays (tracers). The software included in OHG of the subject device and the predicate device is workflow based, where the user is guided to perform various steps in the workflow.
The OTS Hip updates include minor hardware and instrument changes, as well as expanded compatibility with KINCISE™ Offset shell impactors and QUICKSET™ reamer heads. Across the OTS
K261852 - Page 3 of 9
{7}
[LOGO]
ORTOMA
Special 510(k) Submission
software modules, updates consist of incremental user interface and logging enhancements, ongoing software maintenance, bug fixes, security updates, labeling updates, and limited functional improvements.
The performance testing demonstrates that the performance characteristics of the OTS Hip are equivalent to those of the predicate device and therefore supports a determination of Substantial Equivalence for the proposed indications for use.
Differences between the subject and predicate device do not impact or affect the safety or effectiveness of the device or raise different questions of safety and effectiveness.
### Performance Testing
To support the changes to OTS Hip, the following performance testing has been completed for the subject device in support of the substantial equivalence decision:
- Mechanical robustness testing:
Simulated surgical-use testing (including repeated impaction and reaming scenarios) was performed to confirm structural integrity, durability, and maintenance of calibration accuracy. All tests met predefined acceptance criteria.
- Compatibility validation with new instruments:
Integration and functional compatibility were verified through simulated use and system-level testing to ensure correct performance, calibration stability, and safe interaction with newly supported instruments.
- Electrical and mechanical safety of system updates:
IEC 60601-1 testing (including stability, mechanical transport, electrical safety, leakage current, and grounding) was conducted with all requirements successfully met.
- Software verification and validation
Software V&V activities were executed in accordance with IEC 62304. All results were as expected.
K261852 - Page 4 of 9
{8}
[LOGO]
ORTOMA
Special 510(k) Submission
# Substantial Equivalence Comparison
| Characteristic | OTS Hip - Subject Device | OTS Hip - Predicate Device | Equivalence Assessment |
| --- | --- | --- | --- |
| 510(k) Number | K261852 | K250086 | N/A |
| Manufacturer | Ortoma AB | Ortoma AB | N/A |
| Regulation | 21 CFR 882.4560 | 21 CFR 882.4560 | Same |
| Product Code | OLO | OLO | Same |
| Intended Use | To enable planning of orthopedic surgical procedures and enable intraoperative image-guided surgery. | To enable planning of orthopedic surgical procedures and enable intraoperative image-guided surgery. | Same |
| Indications for Use | OTS Hip is indicated to enable planning of orthopedic surgical procedures based on CT and X-Ray medical imaging data of the patient anatomy. It is an intraoperative image-guided localization system that enables navigated surgery. It links a freehand probe, tracked by a passive marker sensor system, to virtual computer image space on a patient's preoperative image data being processed by the OTS platform. The system is indicated for orthopedic hip surgical procedures where a reference to a rigid anatomical structure, such as the pelvis, can be identified relative to a system generated model of the anatomy. The system aids the surgeon to accurately navigate a compatible prosthesis to the preoperatively planned position. | OTS Hip is indicated to enable planning of orthopedic surgical procedures based on CT and X-Ray medical imaging data of the patient anatomy. It is an intraoperative image-guided localization system that enables navigated surgery. It links a freehand probe, tracked by a passive marker sensor system, to virtual computer image space on a patient's preoperative image data being processed by the OTS platform. The system is indicated for orthopedic hip surgical procedures where a reference to a rigid anatomical structure, such as the pelvis, can be identified relative to a system generated model of the anatomy. The system aids the surgeon to accurately navigate a compatible prosthesis to the preoperatively planned position. | Same |
K261852 - Page 5 of 9
{9}
ORTOMA
Special 510(k) Submission
| Characteristic | OTS Hip - Subject Device | OTS Hip - Predicate Device | Equivalence Assessment |
| --- | --- | --- | --- |
| | The system is designed for orthopedic surgical procedures including: - Pre-operative planning of Total Hip Arthroplasty (THA) - Intraoperative navigated surgery for THA using a posterior approach | The system is designed for orthopedic surgical procedures including: - Pre-operative planning of Total Hip Arthroplasty (THA) - Intraoperative navigated surgery for THA using a posterior approach | |
| User Population | Orthopedic surgeon | Orthopedic surgeon | Same |
| Anatomical Site | Hip | Hip | Same |
| Where Used | Office of user and Operating room | Office of user and Operating room | Same |
| Technological Principle of Operation | Intraoperative image-guided localization system allowing user to plan surgery using premeasurements of patient anatomy. Software tracks anatomy, implants and surgical tools in real-time. | Intraoperative image-guided localization system allowing user to plan surgery using premeasurements of patient anatomy. Software tracks anatomy, implants and surgical tools in real-time. | Same |
| Main Technology of MIS | The technology of Minimally Invasive Surgery (MIS) is based on Image Guided Surgery (IGS) devices. | The technology of Minimally Invasive Surgery (MIS) is based on Image Guided Surgery (IGS) devices. | Same |
| Principle of Operation Flow | Preoperative image > surgical planning > surgical guiding > recording | Preoperative image > surgical planning > surgical guiding > recording | Same |
| Major Components | Software for planning and guiding Minor SW updates and bug fixes | Software for planning and guiding | Similar |
| | Calibration Adapter Unit OTD | Calibration Adapter Unit OTD | Same |
| | Calibration Adapter Unit OTD | Calibration Adapter Unit OTD Emphasys | Same |
| | Inserter Adapter OTD – design and manufacturing | Inserter Adapter OTD | Similar |
K261852 - Page 6 of 9
{10}
[LOGO]
ORTOMA
Special 510(k) Submission
| Characteristic | OTS Hip - Subject Device | OTS Hip - Predicate Device | Equivalence Assessment |
| --- | --- | --- | --- |
| | to multiple milled sub-components | | |
| | Reamer Adapter OTD – design and manufacturing to multiple milled sub-components | Reamer Adapter OTD | Similar |
| | Inserter Adapter OTD Straight - design and manufacturing to multiple milled sub-components | Inserter Adapter OTD Straight | Similar |
| | Inserter Adapter OTD Pinnacle Straight design and manufacturing to multiple milled sub-components | Inserter Adapter OTD Pinnacle Straight | Similar |
| | Attachment Adapter Fix OT | Attachment Adapter Fix OT | Same |
| | Attachment Adapter Twin OT | Attachment Adapter Twin OT | Same |
| | Stylet & Pin | Stylet & Pin | Same |
| | Pointer OT | Pointer OT | Same |
| | Pointer Holder | Pointer Holder | Same |
| | Tracers | Tracers | Same |
| | OTS Instrumentation – minor design change to threaded shaft of measurement screw driver | OTS Instrumentation | Similar |
| | Camera (NDI) | Camera (NDI) | Same |
| | Computer (Baaske, e-medic Silence TP2) | Computer (Baaske, e-medic Silence TP2) | Same |
| | Computer and Camera Stand (with integrated Power inlet and MSO (Jansen Medicars, Flexx one 2G 180 - Ortoma, 3012.00.00.071)) Small dimension update of chart | Computer and Camera Stand (with integrated Power inlet and MSO (Jansen Medicars, Flexx one 2G 180 - Ortoma, 3012.00.00.071)) | Same |
K261852 - Page 7 of 9
{11}
[LOGO]
ORTOMA
Special 510(k) Submission
| Characteristic | OTS Hip - Subject Device | OTS Hip - Predicate Device | Equivalence Assessment |
| --- | --- | --- | --- |
| | Keyboard (ProKeys e.K., K10 MED Compact-LS-USB-US/JP) | Keyboard (ProKeys e.K., K10 MED Compact-LS-USB-US/JP) | Same |
| | Footswitch (Herga, MD3G-DGA-GZ1-AAA-001) | Footswitch (Herga, MD3G-DGA-GZ1-AAA-001) | Same |
| Tracking/Navigation Technology | Real-time Optical Tracking System (OPS) | Real-time Optical Tracking System (OPS) | Same |
| Input Image Planning | Computer Tomography (CT), X-Ray | Computer Tomography (CT), X-Ray | Same |
| Input Image Guiding | 3D image of the unique patient's anatomy | 3D image of the unique patient's anatomy | Same |
| DICOM compliance | Yes | Yes | Same |
| Save/load planning | Yes | Yes | Same |
| Accessories | None | None | Same |
| Power Source | Mains | Mains | Same |
| Biocompatibility | Tested per ISO 10993 | Tested per ISO 10993 | Same |
| Software | SW application for pre-operative planning and navigation Minor SW updates and bug fixes | SW application for pre-operative planning and navigation | Similar |
| Sterilization | Steam sterilization for reusable components Gamma sterilization for single use components | Steam sterilization for reusable components Gamma sterilization for single use components | Same |
K261852 - Page 8 of 9
{12}
ORTOMA
Special 510(k) Submission
## Conclusions
The OTS Hip is as safe and effective as the Ortoma Treatment Solution system (K250086). The OTS Hip has the same intended use and indications for use, similar technological characteristics, and principles of operation as its predicate device. The minor differences in the indications do not alter the intended therapeutic use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the OTS Hip and the predicate OTS Hip device raise no new issues of safety or effectiveness. Performance data demonstrate that the OTS Hip is as safe and effective as the predicate OTS Hip (K250086). Thus, the OTS Hip (K261852) is substantially equivalent.
K261852 - Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.