K261833 · Neurent Medical , Ltd. · GEI · Jun 16, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261833
Device Name
NEUROMARK System (NMK00301)
Applicant
Neurent Medical , Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 16, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NEUROMARK System is indicated for use in otorhinolaryngology (ENT) surgery for creation of radiofrequency (RF) lesions through destruction of soft tissue to disrupt posterior nasal nerves in patients with chronic rhinitis.
Device Story
NEUROMARK System comprises hand-held, single-use, bipolar RF device and RF console; used in ENT surgery for chronic rhinitis. Device features malleable shaft and deployable End Effector with bipolar electrode array. Operator positions device in nasal cavity; retracts sheath via slider; initiates bio-impedance check via activation button to confirm tissue contact. Console monitors feedback, controls RF energy delivery, and provides GUI-based procedural guidance. System creates thermal lesions to disrupt posterior nasal nerves. Intended for use by clinicians in surgical settings.
Clinical Evidence
No new performance data required; previously submitted verification and validation data remain applicable.
Technological Characteristics
Bipolar RF electrosurgical device; malleable shaft; End Effector with bipolar electrode array. Feedback control via tissue bio-impedance monitoring. Console-based RF delivery with GUI. Sterile (EO). Identical to predicate K250048.
Indications for Use
Indicated for patients with chronic rhinitis undergoing ENT surgery to disrupt posterior nasal nerves via RF lesion creation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 16, 2026
Neurent Medical , Ltd.
Fay Dalton
Senior Regulatory Affairs Specialist
Oranpoint, Main St., Oranmore
Galway, H91 D7X2
Ireland
Re: K261833
Trade/Device Name: NEUROMARK System (NMK00301)
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: June 2, 2026
Received: June 2, 2026
Dear Fay Dalton:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261833 - Fay Dalton
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261833 - Fay Dalton
Page 3
Sincerely,
SHUCHEN PENG -S
Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K261833
Device Name
NEUROMARK System
Indications for Use (Describe)
The NEUROMARK System is indicated for use in otorhinolaryngology (ENT) surgery for creation of radiofrequency (RF) lesions through destruction of soft tissue to disrupt posterior nasal nerves in patients with chronic rhinitis.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Neurent Medical
# **510(k) Summary**
**Date Prepared:** June 2, 2026
**Submitter Information:** Neurent Medical
Oranpoint, Main Street
Oranmore, Co. Galway, H91 D7X2
Ireland
**Establishment Registration:** 3016813690
**Contact Information:** Fay Dalton
Senior Regulatory Affairs Specialist
Neurent Medical
+353 85 7289921
fay@neurentmedical.com
# **Device Information:**
**Trade Name:** NEUROMARK System
**Common Name:** Radiofrequency Probe
**Classification Name:** Electrosurgical cutting and coagulation device and accessories
**Product Code:** GEI
**Classification:** Class II
**Regulation Number:** 21 CFR 878.4400
**Predicate Device:** NEUROMARK System [K250048]
# **Device Description:**
The NEUROMARK System is intended for the application of radiofrequency (RF) energy to create lesions in mucosal tissue in otolaryngological [also known as Ear, Nose and Throat (ENT)] procedures in patients with chronic rhinitis.
The NEUROMARK System is composed of the NEUROMARK Device and the NEUROMARK Radiofrequency (RF) Console.
The NEUROMARK Device is a hand-held, single-use, bipolar radiofrequency device which comprises a handle, shaft, and treatment tip. The treatment tip, which is referred to as the End Effector, consists of an array of bipolar electrodes that deliver RF energy while monitoring feedback on tissue bio-impedance changes allowing for controlled RF energy delivery. The shaft of the device is pre-shaped to aid access and delivery to the nasal cavity but is malleable to allow the user to bend or shape it to accommodate variations in anatomy to access the desired treatment area. The NEUROMARK Device is operated via handle, slider and activation button. Once in the desired position within the nasal cavity,
Page 1 of 3
{5}
Neurent Medical
the operator moves the slider backwards which retracts the outer sheath, deploying the End Effector. Using the activation button, the user initiates a bio-impedance check to assess and confirm contact between the End Effector and the treatment area. Once the System confirms contact has been achieved, a subsequent press of the activation button initiates the RF energy delivery cycle. The NEUROMARK Device is intended for single use and provided sterile (EO).
The NEUROMARK Device is designed for use with the NEUROMARK Radiofrequency (RF) Console only and is connected via a flexible interface cable.
The NEUROMARK Console delivers, monitors and controls RF energy to the Device. The Console includes a Graphical User Interface (GUI) which provides operational instructions for the procedure, directs the user to select nasal cavities for treatment, indicates when the device is in contact with tissue and ready to start treatment, provides status of therapy and indicates when the procedure is complete. The NEUROMARK Console works in conjunction with software.
#### **Indication for Use:**
The NEUROMARK System is indicated for use in otorhinolaryngology (ENT) surgery for creation of radiofrequency (RF) lesions through destruction of soft tissue to disrupt posterior nasal nerves in patients with chronic rhinitis.
#### **Technological Characteristics:**
The subject NEUROMARK System is identical to the predicate NEUROMARK System [K250048] with respect to all technological characteristics, including but not limited to:
- • Fundamental scientific technology and principle of operation;
- • Anatomical location of use;
- • Device design and configuration;
- • Energy type, waveform, frequency, and maximum power output;
- • Feedback control mechanism;
- • Thermal lesion characteristics and thermal safety;
- • Software, electrical safety, and electromagnetic compatibility;
- • Patient-contacting materials and biocompatibility;
- • Sterilization method, sterile barrier packaging, and shelf-life.
There are no new or modified technological characteristics of the subject device in comparison to those of the predicate device [K250048].
Page 2 of 3
{6}
Neurent Medical
# **Performance Data:**
The modification is limited to clarification of the indications for use only and does not affect device design, performance, or technological characteristics. Accordingly, no new performance data were required. Previously submitted verification and validation data remain applicable and continue to demonstrate that the device meets all acceptance criteria and supports a determination of substantial equivalence to the predicate device.
# **Substantial Equivalence:**
The NEUROMARK System has the same intended use and fundamental scientific technology as the predicate device. An evaluation of the impact of the modification was conducted and confirmed that the change to the indications for use does not affect the device design, performance characteristics, or risk profile, and does not raise different questions of safety and effectiveness. As no changes impacting device performance were implemented, no new performance testing was required; previously submitted verification and validation data remain applicable and continue to demonstrate that the device meets its specifications and intended use. The NEUROMARK System is substantially equivalent to the predicate device.
# **Conclusion:**
In conclusion, the intended use and fundamental technological characteristics are the same as the predicate device. No new performance testing was required, and previously submitted data remain applicable. Accordingly, the device is as safe and effective as the predicate device and is substantially equivalent.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.