← Product Code [GEX](/productcode/GEX) · K261821

# 1565nm Laser Therapy Device (PZ-JG86-01,PZ-JG86-02) (K261821)

_Zhengzhou PZ Laser Slim Technology Co., Ltd. · GEX · Aug 27, 2026 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K261821

## Device Facts

- **Applicant:** Zhengzhou PZ Laser Slim Technology Co., Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Aug 27, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The 1565nm Laser Therapy Device is indicated for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue, and improving the appearance of scalp hair in adult males and females with Fitzpatrick skin type I to IV who are seeking treatment for hair loss.

## Device Story

1565nm Laser Therapy Device (models PZ-JG86-01, PZ-JG86-02) is an Er:Glass fiber-laser system with scanner; generates 1565nm laser beams for fractional skin resurfacing and soft tissue coagulation. Device components include host, treatment handle, two treatment tips, foot switch, and power cord. Operates via selective photothermolysis and fractional laser technology. Used in clinical settings by trained professionals. Treatment tips contact patient skin; made of gold-plated medical-grade aluminum alloy. Output consists of laser micro-beams (up to 500/cm2) with energy 10-70mJ per beam. Healthcare providers use the scanner to deliver energy in various shapes (square, circle, ring, hexagon, lines) to target tissue. Benefits include non-invasive skin resurfacing and hair loss treatment.

## Clinical Evidence

No clinical data. Bench testing only, including electrical safety (IEC 60601-1, IEC 60601-2-22), laser safety (IEC 60825-1), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10, ISO 10993-23).

## Technological Characteristics

Er:Glass fiber-laser; 1565nm wavelength; 650nm aiming beam. Energy: 10-70mJ/micro-beam; max pulse width 10ms; beam density up to 500/cm2. Materials: medical-grade aluminum alloy with gold-plated surface (treatment tips). Standards: IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2, ISO 10993. Scanner shapes: square, circle, ring, hexagon, lines, rectangles.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- F65 Laser System ([K222790](/device/K222790.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**U.S. FOOD & DRUG**
ADMINISTRATION

August 27, 2026

Zhengzhou PZ Laser Slim Technology Co., Ltd.

Xiaopei Zhao

Regulatory Affairs Specialist

3rd Floor, Bldg. 1, # 101 Jinbai Rd., High-Tech Development Zone, 450001 Zhengzhou City, Henan Province, People'S Republic Of China.

Zhengzhou,

China

Re: K261821

Trade/Device Name: 1565nm Laser Therapy Device (PZ-JG86-01,PZ-JG86-02)

Regulation Number: 21 CFR 878.4810

Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology

Regulatory Class: Class II

Product Code: GEX

Dated: May 30, 2026

Received: June 1, 2026

Dear Xiaopei Zhao:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K261821 - Xiaopei Zhao

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K261821 - Xiaopei Zhao

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L.
HITHE -S

Digitally signed by
TANISHA L. HITHE -S
Date: 2026.08.27 18:13:56
-04'00'

Tanisha Hithe

Assistant Director

DHT4A: Division of General Surgery Devices

OHT4: Office of Surgical and Infection Control Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261821 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  1565nm Laser Therapy Device (PZ-JG86-01,PZ-JG86-02)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The 1565nm Laser Therapy Device is indicated for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue, and improving the appearance of scalp hair in adult males and females with Fitzpatrick skin type I to IV who are seeking treatment for hair loss.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |
|  Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ?  |

{4}

510(k) Summary

The assigned 510(k) Number: K261821

### 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

**1. Date of preparation:** 05/27/2026

**2. Sponsor Identification**

Zhengzhou PZ Laser Slim Technology Co., Ltd.

3rd Floor, Building 1, No. 101 Jinbai Road, High-tech Development Zone, 450001 Zhengzhou City, Henan Province, People's Republic of China.

Contact Person: Xiaojie Shi

Position: RA

Tel:+86-18703657355

Fax:+86-371-55677886

Email: shixiaojie1225@163.com

**3. Designated Submission Correspondent**

Zhengzhou PZ Laser Slim Technology Co., Ltd.

3rd Floor, Building 1, No. 101 Jinbai Road, High-tech Development Zone, 450001 Zhengzhou City, Henan Province, People's Republic of China.

Contact Person: Xiaopei Zhao

Tel:+86-15225108693

Fax:+86-371-55677886

Email: 1340918289@qq.com

1 / 6

{5}

510(k) Summary---

#### 4. Identification of Proposed Device

Trade Name: 1565nm Laser Therapy Device

Models: PZ-JG86-01, PZ-JG86-02

Common Name: Powered Laser Surgical Instrument

Regulatory Information

**Classification Name:** Powered Laser Surgical Instrument

**Classification:** II

**Product Code:** GEX

**Regulation Number:** 21 CFR 878.4810

**Regulation Name:** Laser surgical instrument for use in general and plastic surgery and in dermatology.

**Review Panel:** General& Plastic Surgery

**Intended Use:** The 1565nm Laser Therapy Device is indicated for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue, and improving the appearance of scalp hair in adult males and females with Fitzpatrick skin type I to IV who are seeking treatment for hair loss.

#### 5. Device Description

1565nm Laser Therapy Device is Er:Glass Fiber-laser with scanner laser system that generate laser beams at 1565nm wavelengths. The 1565nm Laser Therapy Device is mainly composed of a host, a treatment handle, two treatment tips, a foot switch and a power cord. The treatment tip, intended for direct patient contact, is made of medical-grade aluminum alloy with gold-plated surface.

Its principle of operation is based on the theory of Selective Photothermolysis and Fractional Laser Technology.

These two models are completely identical in terms of their product principles, structure and composition, intended use, key components, and electrical structure, with only the Scanner shapes are different.

2 / 6

{6}

510(k) Summary---

# **6. Identification of Predicate Device**

510(k) Number: K222790

Product Name: F65 Laser System

Manufacturer: Lumenis Be Ltd.

# **7. Comparison of technological characteristics with the predicate devices**

Table 1 General Comparison

|  Description | Proposed Device | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  Device Name | 1565nm Laser Therapy Device | F65 Laser System (F1565nm module) K222790 | /  |
|  Classification regulation | 21CFR 878.4810 | 21CFR 878.4810 | Same  |
|  Classification Panel | General&Plastic Surgery | General&Plastic Surgery | Same  |
|  class | II | II | Same  |
|  Product Code | GEX | GEX | Same  |
|  Common name | Powered laser surgical instrument | Powered laser surgical instrument | Same  |

3 / 6

{7}

510(k) Summary

|  Indication for use | The 1565nm Laser Therapy Device is indicated for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue, and improving the appearance of scalp hair in adult males and females with Fitzpatrick skin type I to IV who are seeking treatment for hair loss. | The F65 module and hand piece, with wavelength of 1565 nm, are indicated for: Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue ; Improving the appearance of scalp hair in adult males and females with Fitzpatrick skin type I to IV who are seeking treatment for hair loss. | Same  |
| --- | --- | --- | --- |
|  Principle of Operation | Selective photothermolysis, fractional laser technology | Selective photothermolysis, fractional laser technology | Same  |
|  Prescription use or not | Prescription use | Prescription use | Same  |

Table 2 Performance comparison

|  Description | Proposed device | Predicate device | Comparison  |
| --- | --- | --- | --- |
|  Device Name | 1565nm Laser Therapy Device | F65 Laser System (F1565nm module) K222790 | /  |
|  Type of Laser | Er:Glass Fiber-laser with scanner | Er:Glass Fiber-laser with scanner | Same  |
|  Laser Wavelength | 1565nm | 1565nm | Same  |
|  Beam density | Up to 500 micro beams per cm2 | up to 500 micro beams per cm2 | Same  |
|  Energy | 10mJ-70mJ per micro beam | Up to 70 mJ per micro beam | Same  |

4 / 6

{8}

510(k) Summary

|  Tip treatment width | Up to 18mm | Up to 18mm | Same  |
| --- | --- | --- | --- |
|  Max pulse width | 10ms | 10ms | Same  |
|  Scanner shapes | Square, circle, ring, regular hexagon, horizontal line, vertical line, horizontal rectangle, vertical rectangle. | Line, square, rectangle, circle, donut, hexagon, vertical line, and vertical rectangle | Same  |
|  Aiming beam Laser Wavelength | 650nm | 650nm | Same  |

#### SE Analysis:

proposed device is compared only with the F1565nm module of the predicate device (K222790). The comparative analysis results show that the proposed device is consistent with the F1565nm module of the predicate device (K222790) in key performance metrics such as intended use, laser wavelength, laser type, beam density, energy density, tip treatment width, and scanner shape. Although the predicate device contains multiple modules, each module has a different intended use and can operate independently without interfering with one another. Therefore, the proposed device is equivalent to the F1565nm module of the predicate device (K222790); the proposed device is substantially equivalent to the predicate device.

## 8. Performance data

The following performance data were provided in support of the substantial equivalence determination:

1) Electrical Safety, Electromagnetic Compatibility Testing

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards.

5 / 6

{9}

510(k) Summary

IEC 60601-1:2020 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.

IEC 60601-2-22:2019, Medical Electrical Equipment - Part 2-22: Particular

Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment.

IEC 60825-1: 2014, Safety of laser products - Part 1: Equipment classification and requirements.

IEC 60601-1-2:2020, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic disturbances- Requirements and tests.

## 2) Biocompatibility

Biocompatibility tests were conducted to verify if the proposed device meets safety and effectiveness criteria. The test results indicate that the proposed device complies with the following standards:

ISO 10993-5:2009, Biological Evaluation of Medical Devices, Part 5-Tests for in Vitro cytotoxicity.

ISO 10993-10:2021, Biological Evaluation of Medical Devices, Part 10-Tests for skin sensitization.

ISO 10993-23:2021/Amd1: 2025, Biological Evaluation of Medical Devices-Part 23: Tests for irritation.

## 8. Clinical Test Conclusion

No clinical study is included in this submission.

## 9. Conclusions

Proposing differences between the device and the predicate device will not raise new safety and effectiveness-related issues. The proposed device is identical to the legally marketed predicate device (K222790) in terms of intended use, key technical characteristics (such as laser wavelength, energy, beam density), etc. Furthermore, the proposed device has been verified through electromagnetic compatibility, electrical safety, and biocompatibility testing. Therefore, the safety and effectiveness of the proposed device are equivalent to those of predicate device K222790.

6 / 6

---

**Source:** [https://fda-staging.innolitics.com/device/K261821](https://fda-staging.innolitics.com/device/K261821)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).