HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens; HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens; HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens
Applicant
HOYA Lamphun , Ltd.
Product Code
MVN · Ophthalmic
Decision Date
Jul 7, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -12.00D, exhibiting astigmatism of 1.00 diopter or less that does not interfere with visual acuity. The HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -10.00D. The lenses may be worn by persons who exhibit astigmatism of up to 2.75 diopters that does not interfere with visual acuity. The HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) and presbyopia in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -10.00D with add power ranging from +0.50D to +2.50D. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopter or less that does not interfere with visual acuity. The sorafilcon A contact lenses are to be prescribed for single-use disposable wear and are to be discarded upon each removal.
Device Story
Daily disposable soft contact lenses; made of sorafilcon A (silicone hydrogel); refractive medium focusing light on retina. Prescribed by eye care professionals for single-use wear; discarded after removal. Subject device adds Reactive Blue 246 visibility tint to previously cleared predicate lens family. Used by patients for vision correction; no complex operation or software involved.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing (lens parameters, water content, refractive index, light/UV transmittance, mechanical properties) and biocompatibility testing (cytotoxicity, sensitization, ocular irritation, systemic toxicity) per ISO 10993 and ANSI Z80.20 standards.
Technological Characteristics
Material: sorafilcon A (silicone hydrogel, 48% water). Visibility tint: Reactive Blue 246. Manufacturing: cast molded. Sterilization: moist heat. UV-blocking monomer included. Oxygen permeability: 112 x 10^-11 (cm2/sec)(cm3O2)/(mL x mmHg) @ 35 °C. Standards: ANSI Z80.20:2016, ISO 18369, ISO 10993-1:2018.
Indications for Use
Indicated for correction of refractive ametropia (myopia, hyperopia, astigmatism, presbyopia) in non-aphakic persons with non-diseased eyes. Applicable to adolescents (12-22 years) and adults (22+ years).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens (K223397)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 7, 2026
HOYA Lamphun, Ltd.
Timothy Ting
Vice President of Quality Assurance/Regulatory Affairs
Northern Region Industrial Estate, 75/2 Moo 4
Tambol Banklang, Amphur Muang
Lamphun, 51000
Thailand
Re: K261812
Trade/Device Name: HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens;
HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens;
HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens
Regulation Number: 21 CFR 886.5925
Regulation Name: Soft (hydrophilic) contact lens
Regulatory Class: Class II
Product Code: LPL, MVN
Dated: May 27, 2026
Received: June 8, 2026
Dear Timothy Ting:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261812 - Timothy Ting
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K261812 - Timothy Ting
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
J Angelo Green -S
J. Angelo Green, Ph.D.
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K261812 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens; HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens; HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens | | | |
| Please provide your Indications for Use below. | | | ? |
| The HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -12.00D, exhibiting astigmatism of 1.00 diopter or less that does not interfere with visual acuity. The HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -10.00D. The lenses may be worn by persons who exhibit astigmatism of up to 2.75 diopters that does not interfere with visual acuity. The HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) and presbyopia in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -10.00D with add power ranging from +0.50D to +2.50D. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopter or less that does not interfere with visual acuity. The sorafilcon A contact lenses are to be prescribed for single-use disposable wear and are to be discarded upon each removal. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? | |
{4}
| HOYA | HOYA Lamphun Ltd. Traditional 510(k) Premarket Notification HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens | Submission Date: 27-May-2026 |
| --- | --- | --- |
| | | Submission No.: K261812 |
### 510(k) Summary
K261812
### SUBMITTER INFORMATION
Date Prepared: May 27, 2026
Name: HOYA Lamphun Ltd.
Address: Northern Region Industrial Estate, 75/2 Moo 4, Tambol Banklang, Amphur Muang,
Lamphun 51000 Thailand.
Contact Person: Timothy Ting
Vice President of Quality Assurance/Regulatory Affairs
Phone Number: (66) 93-132-1699
Email: chongchai.ting@hoya.com
### DEVICE INFORMATION
Trade Name: HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens
HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens
HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens
Brand Name: ILLUMINA 1 DAY Soft silicone hydrogel contact lens
ILLUMINA 1 DAY Toric Soft silicone hydrogel contact lens
ILLUMINA 1 DAY Multifocal Soft silicone hydrogel contact lens
Common Name: Soft contact lens, daily disposable wear.
Device Classification: Class II (21 CFR 886.5925)
Classification Name: Lens, Contact, (Disposable)
Product Code: LPL, MVN.
Classification Panel: Ophthalmic.
### PREDICATE DEVICES
HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens with Reactive Blue 246 as visibility handling tint is substantially equivalent to the following predicate devices cleared under K223397:
HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens
HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens
HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens
### DEVICE DESCRIPTION
The HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lenses are made from sorafilcon A material, a hydrophilic siloxane copolymer of N-vinyl pyrrolidone and is 48% water by weight when immersed in a sterile phosphate buffered saline packaging solution. A UV-absorbing monomer is used to
Page 1 of 5
{5}
| HOYA | HOYA Lamphun Ltd. Traditional 510(k) Premarket Notification HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens | Submission Date: 27-May-2026 |
| --- | --- | --- |
| | | Submission No.: K261812 |
block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280 nm to 315 nm and less than 50% in the UVA range of 316 nm to 380 nm. This lens with Reactive Blue 246 dye to provide visibility handling tint. The color additive conforms to 21 CFR Part 73.3106.
The HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lenses are to be prescribed for single-use disposable wear.
The physical properties of the lens are:
| Refractive Index | 1.408 |
| --- | --- |
| Light Transmittance | ≥ 95% |
| Water Content | 48% |
| Specific Gravity | 1.131 |
| Surface Character | Hydrophilic |
| Oxygen Permeability | 112 x 10^{-11} (cm^{2}/sec)(cm^{3}O_{2})/(mL x mmHg) @ 35 °C (polarographic method). |
The lenses will be manufactured with the following parameters:
| Diameter | 14.0 mm to 14.7 mm |
| --- | --- |
| Center Thickness | 0.062 mm to 0.330 mm |
| Base Curve | 8.40 mm to 9.20 mm |
| Power Range | +5.00D to -12.00D |
| Cylinder Power (Toric) | -0.75D to -2.75D |
| Cylinder Axis | 0° to 180° in 10° increments |
| Add Power Range | +0.50D to +2.50D |
## INDICATIONS FOR USE
### HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens
The HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -12.00D, exhibiting astigmatism of 1.00 diopter or less that does not interfere with visual acuity.
### HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens
The HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -10.00D. The lenses may be worn by persons who exhibit astigmatism of up to 2.75 diopters that does not interfere with visual acuity.
### HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens
The HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lenses are indicated for the
Page 2 of 5
{6}
| HOYA | HOYA Lamphun Ltd. Traditional 510(k) Premarket Notification HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens | Submission Date: 27-May-2026 |
| --- | --- | --- |
| | | Submission No.: K261812 |
correction of refractive ametropia (myopia or hyperopia) and presbyopia in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -10.00D with add power ranging from +0.50D to +2.50D. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopter or less that does not interfere with visual acuity.
The HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lenses are to be prescribed for single-use disposable wear and are to be discarded upon each removal.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
A side-by-side comparison of technological characteristics of the predicate devices and the new device:
| Property | Predicate Device HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens Product Family K223397 | Subject Device HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens Product Family |
| --- | --- | --- |
| Intended Use | Daily wear and daily disposable cornea soft contact lens for myopia, hyperopia, astigmatism, presbyopia | Same as predicate devices |
| Functionality | The contact lenses act as a refractive medium that focus light rays from near and distant objects on the retina. | Same as predicate devices |
| Modality | Daily disposable wear contact lens | Same as predicate devices |
| Manufacturing Method | Cast Molded | Same as predicate devices |
| Material Group | Silicone Hydrogel FDA Group I (low water content, nonionic polymer) | Same as predicate devices |
| USAN Name | sorafilcon A | Same as predicate devices |
| Water Content | 48% | Same as predicate devices |
| Oxygen Permeability* | 112 | Same as predicate devices |
| Refractive Index | 1.408 | Same as predicate devices |
| Specific Gravity | 1.123 | 1.131 |
| Visibility Tint | No color pigment was added | Reactive Blue 246 |
| Sterilization Method | Moist Heat | Same as predicate devices |
Note: *Oxygen Permeability shown was determined by polarographic method:
x 10⁻¹¹ [(cm²/sec)x(cm³O₂)/ (ml x mmHg)] @ 35 °C
## SUMMARY OF NON-CLINICAL TESTING
Non-clinical studies for HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lenses with Reactive Blue 246 were conducted in accordance with FDA's Class II Daily Wear Contact Lenses - Premarket Notification [510(k)] Guidance Document issued date June 27, 1994 and 21 CFR Part 58 Good Laboratory
Page 3 of 5
{7}
| HOYA | HOYA Lamphun Ltd. Traditional 510(k) Premarket Notification HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens | Submission Date: 27-May-2026 |
| --- | --- | --- |
| | | Submission No.: K261812 |
Practice for Nonclinical Laboratory Studies. Each test was conducted based on the ANSI, ISO, and ASTM standard as indicated below:
- Lens shape, edge thickness, inclusion and surface imperfection per ANSI Z80.20:2016 (R2021) and ISO 18369-3:2017.
- Lens parameters per ANSI Z80.20: 2016 and ISO 18369-3:2017.
- Water content per ANSI Z80.20-2016 and ISO 18369-4:2017.
- Refractive index per ANSI Z80.20-2016 and ISO 18369-4:2017.
- Light transmittance per ANSI Z80.20-2016 and ISO 18369-4:2017.
- UV radiation transmittance per ANSI Z80.20-2016 and ISO 18369-4:2017.
- Extractables per ANSI Z80.20-2016 (R2021) and ISO 18369-4:2017.
- Contact angle per ANSI Z80.20-2016.
- Mechanical properties per ANSI Z80.20-2016 (R2021) and ASTM D882-18.
- Sterile blister packaging integrity per ASTM F1929-23.
All test results met the pre-established acceptance criteria per predicate device which demonstrated the lenses are safe and effective.
## SUMMARY OF BIOCOMPATIBILITY TESTING
Biocompatibility testing for HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lenses with Reactive Blue 246 was conducted in accordance with the FDA's Class II Daily Wear Contact Lenses - Premarket Notification [510(k)] Guidance Document issued date June 27, 1994, ISO 10993-1:2018 Biological Evaluation of Medical Devices and 21 CFR Part 58 Good Laboratory Practice for Nonclinical Laboratory Studies as listed below:
- Cytotoxicity per ISO 10993-5: 2009
- Guinea Pig Maximization (GPMT) Skin Sensitization per ISO 10993-10: 2021
- 22 day contact lens wear study in rabbits per ISO 9394: 2012
- Acute Ocular Irritation per ISO 10993-23:2021
- Systemic Toxicity per ISO 10993-11: 2017
The above studies were conducted using the final finished sterilized subject device lenses as the test article. No separate biocompatibility testing was conducted on the primary packaging or packaging solution, since these two device components are identical (in all materials, formulation, processing, etc.) to the primary packaging and packaging solution used for the K223397 predicate. Please note that while that the 22-day contact lens study in rabbits per ISO 9394:2012 and the GPMT skin sensitization study per ISO 10993-10:2021 were performed, these two studies were not needed to support the substantial equivalence of the subject device lenses, since the only change was the addition of Reactive Blue 246 as a handling tint to the lenses.
HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lenses with Reactive Blue 246 were demonstrated to be biocompatible with the mucosal membrane (cornea) and the device performs as
Page 4 of 5
{8}
|  | HOYA Lamphun Ltd. Traditional 510(k) Premarket Notification HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens | Submission Date: 27-May-2026 |
| --- | --- | --- |
| | | Submission No.: K261812 |
expected and is equivalent to the predicates without creating additional risk to the user.
### SUBSTANTIAL EQUIVALENCE CONCLUSION
The cumulative results of the bench performance testing, in vitro and in vivo testing, and biocompatibility testing on the HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lenses with Reactive Blue 246 sponsored by HOYA Lamphun Ltd. demonstrated that the safety, effectiveness, and performance of the subject device is substantially equivalent to the predicate devices.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.