Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. Target Detachable Coils are indicated for endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolizations in the peripheral vasculature
Device Story
Target Xtra Detachable Coil is a stretch-resistant, electrolytically detachable platinum-tungsten alloy coil attached to a stainless-steel delivery wire. Designed for endovascular embolization of neurovascular and peripheral vascular abnormalities; used with Stryker Neurovascular InZone Detachment System. Device is a line extension of the predicate, combining helical and tetrahedral secondary shapes and a modified stretch-resistant fiber. Operated by physicians in clinical settings; delivered via microcatheter to the target site. Electrolytic detachment allows controlled release of the coil to obstruct blood flow. Benefits include precise occlusion of aneurysms and vascular malformations. No clinical or animal studies performed; substantial equivalence supported by bench testing.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing only, including simulated use in neurovascular anatomy models for coil softness, junction tensile strength, durability, particulate release, and frictional force.
Technological Characteristics
Materials: Platinum-tungsten alloy coil, stainless-steel delivery wire. Principle: Electrolytic detachment. Form factor: 1.5–3 mm secondary coil diameter, 8–32 cm length. Connectivity: Compatible with InZone Detachment System. Sterilization: Ethylene Oxide (EtO).
Indications for Use
Indicated for endovascular embolization of intracranial aneurysms, neurovascular abnormalities (arteriovenous malformations, arteriovenous fistulae), and peripheral arterial/venous embolizations in adults.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 24, 2026
Stryker Neurovascular
Zach Borla
Senior Staff Regulatory Affairs Specialist
47900 Bayside Pkwy.
Fremont, California 94538
Re: K261772
Trade/Device Name: Target Xtra Detachable Coil
Regulation Number: 21 CFR 882.5950
Regulation Name: Neurovascular Embolization Device
Regulatory Class: Class II
Product Code: HCG, KRD
Dated: May 28, 2026
Received: May 29, 2026
Dear Zach Borla:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261772 - Zach Borla
Page 2
(https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261772 - Zach Borla
Page 3
Sincerely,
SARA S. THOMPSON -S
Sara S. Thompson, D.V.M.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K261772 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| Target Xtra Detachable Coil | | | |
| Please provide your Indications for Use below. | | | ? |
| Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. | | | |
| Target Detachable Coils are indicated for endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolizations in the peripheral vasculature | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
K261772
# **Stryker Neurovascular**
510(k) Premarket Notification, Special 510(K)
Target Xtra Detachable Coil---
**Date Prepared:** June 24, 2026
**Submitter:** Stryker Neurovascular
47900 Bayside Parkway
Fremont, CA 94538
Facility Registration #3008853977
**Contact:** Mr. Zach Borla
Senior Staff Regulatory Affairs Specialist
Tel (385) 210-5620
E-mail: zach.borla@stryker.com
**Device Trade Name:** Target Xtra Detachable Coil
**Classification Name:** Target Xtra Detachable Coils are vascular and neurovascular embolization devices under 21 CFR 870.3300 (KRD) and 21 CFR 882.5950 (HCG), respectively, and are Class II devices (special controls).
The special control for the device is FDA’s guidance document, *Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices* (issued 29 Dec 2004).
**Legally Marketed Predicate Device:** K252694 (Cleared 11 Dec 2025)
---Page 1 of 5
{5}
K261772
# Stryker Neurovascular
510(k) Premarket Notification, Special 510(K)
Target Xtra Detachable Coil
# DEVICE DESCRIPTION
Target Xtra Detachable Coil is a stretch resistant, electrolytically detachable coil consisting of a platinum-tungsten alloy coil attached to a stainless-steel delivery wire.
Target Xtra Detachable Coil is specifically designed for use with Stryker Neurovascular InZone® Detachment System (sold separately).
Target Xtra Detachable Coil is compatible with Stryker Neurovascular 2-tip marker microcatheters; refer to Instructions for Use (IFU) for the compatible microcatheter size.
# ACCESSORIES
Target Xtra Detachable Coil is not packaged with any accessories.
# INTENDED USE / INDICATION FOR USE:
Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Target Detachable Coils are indicated for endovascular embolization of:
- Intracranial aneurysms
- Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- Arterial and venous embolizations in the peripheral vasculature
# COMPARISON TO THE PREDICATE DEVICE: TARGET DETACHABLE COIL
Target Xtra Detachable Coil, is a line extension of the predicate device, Target Detachable Coil (family), that combines existing helical and tetrahedral secondary shapes in a single coil configuration and incorporates a modified stretch resistant fiber into the main coil.
The Target Xtra Detachable Coil has the same intended use and indications for use, and the same fundamental scientific technology as the predicate device. A comparison of the subject device with the predicate device is summarized in Table 1 below.
Table 1. Substantial Equivalence Comparison
| Characteristic | Predicate Device K252694 | Subject Device |
| --- | --- | --- |
| Manufacturer | Stryker | Same |
| Trade Name | Target Detachable Coil | Target Xtra Detachable Coil |
| Device Type | Vascular Embolization Device; Neurovascular | Same |
Page 2 of 5
{6}
K261772
# Stryker Neurovascular
510(k) Premarket Notification, Special 510(K)
Target Xtra Detachable Coil
| Characteristic | Predicate Device K252694 | Subject Device |
| --- | --- | --- |
| | Embolization Device | |
| Classification Regulation (21 CFR) | 870.3300, Class 2 882.5950, Class 2 | Same |
| Product Code: | KRD, HCG | Same |
| Intended Use/Indication for Use | Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. Target Detachable Coils are indicated for endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolization's in the peripheral vasculature | Same |
| How Supplied | Single Use/Sterile | Same |
| Method of Sterilization | Ethylene Oxide (EtO) Gas | Same |
| Coil Material | Platinum/Tungsten alloy | Same |
| Delivery Wire Material | Stainless steel | Same |
| **Features** | | |
| Secondary Coil Outer Diameter | 1 mm – 24 mm | Target Xtra fits within the range of currently offered Target Detachable Coils and utilizes a Secondary Coil Outer Diameter of 1.5–3 mm. |
| Coil Length | 1 cm – 50 cm | Target Xtra fits within the range of currently offered Target Detachable Coils and utilizes a Coil Length of 8–32 cm. |
| Coil Secondary Shape Types | 360, 3D, Helical, Tetrahedral | Helical-Tetrahedral Combination |
| Power Supply Compatibility | InZone Detachment System | Same |
| Packaging Configuration and Materials | **Pouch:** Tyvek/Film pouch **Carton:** Chipboard carton **Dispenser coil (Hoop):** HDPE tubing and clips | Same |
Page 3 of 5
{7}
K261772
# Stryker Neurovascular
510(k) Premarket Notification, Special 510(K)
Target Xtra Detachable Coil
### RISK ASSESSMENT
Risk assessment has been conducted for the subject device in accordance with EN ISO 14971:2019, “Medical devices — Application of risk management to medical devices.” Stryker Neurovascular has determined that no new questions of safety or effectiveness are raised when compared to the predicate device. This line extension does not result in any new or modified risks.
### BIOCOMPATIBILITY
No changes have been made to the materials or manufacturing processes used for the subject device. No new or additional biological risks have been identified. Therefore, the previously completed biocompatibility testing for the predicate device remains applicable and supports the biological safety of the subject device.
### PERFORMANCE TESTING – BENCH TESTING
Performance bench testing has demonstrated that the Target Xtra Detachable Coil is substantially equivalent to the predicate device and does not raise new questions of safety or effectiveness.
Performance bench testing of the Target Xtra Detachable Coil consisted of the following tests listed in Table 2.
Table 2. Performance Bench Testing
| Test | Test Method Summary/Purpose | Conclusions |
| --- | --- | --- |
| Main Coil Softness (Coil Conformability) | Simulated use testing was conducted in a neurovascular anatomy model. | Met acceptance criteria |
| Main Junction Tensile Strength | Use tensile tester to measure applied peak tensile force of the main junction. | Met acceptance criteria |
| Durability | Simulated use testing was conducted in a neurovascular anatomy model. The coil was visually inspected for fractures and damage after simulated deployment and retraction. | Met acceptance criteria |
| Particulate | Particulate release due to delivery and detachment of the coil was evaluated in a neurovascular anatomy model. | Particulate characterization was acceptable |
| Coil Friction through Microcatheter | Frictional force through a compatible microcatheter was evaluated in a neurovascular anatomy model. | Met acceptance criteria |
Page 4 of 5
{8}
K261772
# Stryker Neurovascular
510(k) Premarket Notification, Special 510(K)
Target Xtra Detachable Coil
# PERFORMANCE TESTING – ANIMAL STUDY
No animal study was conducted because bench testing was determined sufficient to support substantial equivalence to the predicate device.
# PERFORMANCE TESTING – CLINICAL STUDY
No clinical study was conducted because bench testing was determined sufficient to support substantial equivalence to the predicate device.
# CONCLUSION
The Target Xtra Detachable Coil shares the same intended use, indications for use, and fundamental scientific technology as the predicate device, Target Detachable Coil (family). Additionally, risk assessment of the modifications raises no new questions of safety and effectiveness and successful verification and validation testing demonstrated that the device functions as intended. Therefore, Stryker Neurovascular has determined the Target Xtra Detachable Coil to be substantially equivalent to the predicate device.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.