← Product Code [DQY](/productcode/DQY) · K261766

# CPS Locator 3D Plus Delivery Catheter (K261766)

_CenterPoint Systems, LLC · DQY · Jul 29, 2026 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K261766

## Device Facts

- **Applicant:** CenterPoint Systems, LLC
- **Product Code:** [DQY](/productcode/DQY.md)
- **Decision Date:** Jul 29, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Delivery Catheter is indicated for the introduction of various types of pacing or defibrillator leads to the right heart, inclusive of the Left Bundle Branch Area.

## Device Story

Single-use percutaneous catheter; facilitates introduction of pacing or defibrillator leads to right heart; includes Left Bundle Branch Area. Device includes dilator for vascular access; proximal hemostatic valve; distal soft, rounded, radiopaque tip for fluoroscopic imaging. Designed to be slittable for removal after lead placement. Available in various curves and lengths to accommodate patient anatomy. Operated by clinicians in clinical settings. Benefits include precise lead delivery and simplified removal post-placement.

## Clinical Evidence

Bench testing only. Testing included sterilization validation (ISO 11137), packaging validation (ISO 11607-1), visual inspection, flush testing, tensile testing, simulated use with ancillary devices, valve liquid leak testing, and dimensional verification.

## Technological Characteristics

Percutaneous catheter; 10.3F outer diameter; 8.5F inner diameter; 42-48cm working lengths. Materials: PEBAX with Barium Sulfate and colorants. Features: Braid reinforcement, hydrophilic liner, hemostatic valve, radiopaque distal tip. Sterilization: E-Beam (SAL 10^-6). Compatible with 0.035" guidewires.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- CPS Locator 3D Plus Delivery Catheter ([K250828](/device/K250828.md))

## Reference Devices

- CPS Locator 3D Delivery Catheter ([K250147](/device/K250147.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD & DRUG ADMINISTRATION

July 29, 2026

CenterPoint Systems, LLC

Lauren Marold

Senior Regulatory Specialist

3338 Pkwy. Blvd.

West Valley City, Utah 84119

Re: K261766

Trade/Device Name: CPS Locator 3D Plus Delivery Catheter

Regulation Number: 21 CFR 870.1250

Regulation Name: Percutaneous Catheter

Regulatory Class: Class II

Product Code: DQY

Dated: June 29, 2026

Received: June 30, 2026

Dear Lauren Marold:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food & Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993

www.fda.gov

{1}

K261766 - Lauren Marold

Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K261766 - Lauren Marold

Page 3

Sincerely,

JESSICA L.

BATISTA -S

Digitally signed by JESSICA

L. BATISTA -S

Date: 2026.07.29 13:03:51

-04'00'

Jessica Batista Bertolini

Acting Assistant Director

Division of Cardiac Electrophysiology,

Diagnostics, and Monitoring Devices

Office of Cardiovascular Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

# **Indications for Use**

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K261766

Device Name
CPS Locator 3D Plus Delivery Catheter

Indications for Use (Describe)
The Delivery Catheter is indicated for the introduction of various types of pacing or defibrillator leads to the right heart, inclusive of the Left Bundle Branch Area.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)

☐ Over-The-Counter Use (21 CFR 801 Subpart C)

**CONTINUE ON A SEPARATE PAGE IF NEEDED.**

This section applies only to requirements of the Paperwork Reduction Act of 1995.

**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*

FORM FDA 3881 (8/23)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}

# **510(k) Summary**

# **1.1 Submitter**

Name CenterPoint Systems LLC

Address 3338 Parkway Blvd

West Valley City UT

Phone 860-707-3892

Contact Person: Lauren Marold, Senior Regulatory Specialist

Date Prepared: 11 June 2026

# **1.2 Device**

Name of Device: CPS Locator 3D Plus Delivery Catheter

Classification Name: Percutaneous Catheter

Regulatory Class: Class II per 21 CFR 870.1250

Product Code: DQY

# **1.3 Predicate Device**

**Predicate Name and 510(k) Number:** CPS Locator 3D Plus Delivery Catheter, K250828

This predicate has not been subject to a design-related recall. The reference device, CPS Locator 3D Delivery Catheter, K250147, has not been subject to a design-related recall.

# **1.4 Device Description**

The modified Delivery Catheter (CPS Locator 3D Delivery Catheter) is a single-use percutaneous catheter indicated for the introduction of various types of pacing or defibrillation leads to the right heart, inclusive of the Left Bundle Branch Area.

The modified Delivery Catheter is packaged with a dilator for introduction into the vasculature. Proximally, the modified Delivery Catheter is equipped with a hemostatic valve, and the distal soft, rounded, radiopaque tip facilitates imaging under fluoroscopy. The modified Delivery Catheter is designed to be slittable, thereby allowing its removal after device placement. A variety of curves and lengths are available to accommodate various anatomies and different locations.

K261766

Page 1 of 6

{5}

## 1.5 Indications for Use

The Delivery Catheter is indicated for the introduction of various types of pacing or defibrillation leads to the right heart, inclusive of the Left Bundle Branch Area.

## 1.6 Comparison of Technological Characteristics with the Predicate Devices

The Proposed Device and Predicate Device are similar in indications for use, intended use, technological characteristics, and principles of operation. The Proposed Device and the Reference Device have the same indications for use and intended use, and are similar in technological characteristics and principles of operation.

The differences between the Proposed Device and the Predicate Device are minor and raise no different questions of safety and effectiveness. In accordance with 21CFR807.92(a)(6) a summary of how the technological characteristics of the Proposed Device compares to the Predicate Device and Reference Device is provided below.

|  Feature | CPS Locator 3D Plus Delivery Catheter (subject device) | Primary Predicate: CPS Locator 3D Plus Delivery Catheter (K250828) | Reference Device: CPS Locator 3D Delivery Catheter (K250147) | Same / Different between Proposed & Predicates  |
| --- | --- | --- | --- | --- |
|  Intended Use | Percutaneous catheter for the delivery of leads | Percutaneous catheter for the delivery of leads | Percutaneous catheter for the delivery of leads | Same  |
|  Indications for Use | The Delivery Catheter is indicated for the introduction of various types of pacing or defibrillator leads to the right heart, inclusive of the Left Bundle Branch Area. | The Delivery Catheter is indicated for the introduction of various types of catheters and pacing or defibrillator leads. | The Delivery Catheter is indicated for the introduction of various types of pacing or defibrillator leads to the right heart, inclusive of the Left Bundle Branch Area. | Substantially Equivalent to Predicate Device Same as Reference Device  |
|  Product Code | DQY | DQY | DQY | Same  |
|  Regulation number | 21 CFR 870.1250 | 21 CFR 870.1250 | 21 CFR 870.1250 | Same  |
|  Prescription Device | Yes | Yes | Yes | Same  |
|  Catheter Type | Percutaneous Catheter | Percutaneous Catheter | Percutaneous Catheter | Same  |
|  Guidewire Compatibility | 0.035' | 0.035' | 0.035' | Same  |
|  Outer Diameter | 10.3F | 10.2F | 9.0F | Substantially Equivalent  |

K261766  
Page 2 of 6

{6}

|  Feature | CPS Locator 3D Plus Delivery Catheter (subject device) | Primary Predicate: CPS Locator 3D Plus Delivery Catheter (K250828) | Reference Device: CPS Locator 3D Delivery Catheter (K250147) | Same / Different between Proposed & Predicates  |
| --- | --- | --- | --- | --- |
|  Inner Diameter | 8.5F | 8.4F | 7.0F | Substantially Equivalent  |
|  Working Length | 42cm, 45cm, 48cm | 42cm, 45cm | 42cm, 45cm | Substantially Equivalent  |
|  Components Provided | Catheter, Dilator, Luer cap | Catheter, Dilator, Luer cap | Catheter, Dilator, Luer cap | Same  |
|  Hydrophilic Liner | Yes | Yes | Yes | Same  |
|  Radiopaque Distal Tip | Yes | Yes | Yes | Same  |
|  Valve | Yes | Yes | Yes | Substantially Equivalent  |
|  Shaft Materials | PEBAX with Barium Sulfate and colorants | PEBAX with Barium Sulfate and colorants | PEBAX with Barium Sulfate and colorants | Same  |
|  Braid Reinforcement | Yes | Yes | Yes | Same  |
|  Dilator | Yes | Yes | Yes | Same  |
|  Multiple Distal End Shapes Available | Yes | Yes | Yes | Same  |
|  Sterility | E-Beam Sterilization, SAL 10^{-6} | E-Beam Sterilization, SAL 10^{-6} | E-Beam Sterilization, SAL 10^{-6} | Same  |
|  Number of Uses | Single patient use | Single patient use | Single patient use | Same  |

K261766

Page 3 of 6

{7}

|  Feature | CPS Locator 3D Plus Delivery Catheter (subject device) | Primary Predicate: CPS Locator 3D Plus Delivery Catheter (K250828) | Reference Device: CPS Locator 3D Delivery Catheter (K250147) | Same / Different between Proposed & Predicates  |
| --- | --- | --- | --- | --- |
|  Principles of Operation | After venous access is gained, the catheter and dilator are advanced over a guidewire to the desired location. The dilator is removed and a catheter or lead is placed through the Delivery Catheter. The Delivery Catheter may be removed by slitting. | After venous access is gained, the catheter and dilator are advanced over a guidewire to the desired location. The dilator is removed and a catheter or lead is placed through the Delivery Catheter. The Delivery Catheter may be removed by slitting. | After venous access is gained, the catheter and dilator are advanced over a guidewire to the desired location. The dilator is removed and a catheter or lead is placed through the Delivery Catheter. The Delivery Catheter may be removed by slitting. | Same  |

The modified Delivery Catheter is used for the same intended use in the same anatomical location using the same principles of operation as the predicate device.

K261766

Page 4 of 6

{8}

### 1.7 Performance Data

All necessary performance testing has been conducted on the CPS Locator 3D Plus Delivery Catheter to assure substantial equivalence to the predicate devices and to demonstrate the device performs as intended. All testing was performed on test units representative of finished devices.

The device passed the following tests, which were conducted in accordance with noted standards:

- Sterilization validation per ISO 11137
- Packaging validation per ANSI/AAMI/ISO 11607-1
- Visual Inspection
- Flush test
- Tensile tests

- Simulated use testing, including use/compatibility with ancillary devices
- Valve liquid leak test
- Sheath and Dilator Dimensional verification, including OD/ID, working length

K261766

Page 5 of 6

{9}

### 1.8 Conclusions

Based on the similarity of the subject and predicate devices in terms of the intended use, principle of operation and overall technological characteristics, and the results of the testing summarized in section 1.7, the CPS Locator 3D Plus Delivery Catheter is substantially equivalent to the predicate device (K250828).

K261766

Page 6 of 6

---

**Source:** [https://fda-staging.innolitics.com/device/K261766](https://fda-staging.innolitics.com/device/K261766)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).