The equipment is intended to produce two-dimensional (2D) or three-dimensional (3D) images of the human dental, maxillofacial regions and/or skull as diagnostic support for adult and pediatric patients.
Device Story
Digital X-ray imaging unit utilizing cone beam computed tomography (CBCT) for 3D maxillofacial, TMJ, and ENT imaging; also performs 2D panoramic (PANO) imaging via tomosynthesis and optional 2D cephalometric/carpus radiography. System comprises X-ray tube, detector, column, patient supports, and workstation software. Operated by healthcare professionals in clinical environments. Input: X-ray projections acquired during patient rotation. Output: 2D/3D radiographic images for diagnostic support. Clinical utility: aids visualization of dental/maxillofacial anatomy for diagnosis when conventional radiography is insufficient. Benefits: high-quality diagnostic imaging with radiation dose management.
Clinical Evidence
Bench testing only. Performance evaluated per IEC 61223-3-7 and IEC 61223-3-4; all models met acceptance criteria for spatial resolution, CNR, uniformity, and radiation output. Biocompatibility confirmed per ISO 10993. Clinical performance supported by independent review of sample images by an ABR-certified radiologist, confirming clinically acceptable image quality.
Indicated for adult and pediatric patients requiring 2D or 3D diagnostic imaging of dental, maxillofacial, and skull anatomy.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
[LOGO]
FDA U.S. FOOD & DRUG ADMINISTRATION
July 23, 2026
Hefei Meyer Optoelectronic Technology, Inc.
% Kyra Kang
Director
Landlink Healthcare Technology (Shanghai) Co., Ltd.
Rm. 1308, Baohua International Plz.,
West Guangzhong Rd. 555, Jingan District
SHANGHAI, CHINA
Re: K261748
Trade/Device Name: Dental Cone-beam Computed Tomography (mDX-15P Series)
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: OAS
Dated: May 27, 2026
Received: May 28, 2026
Dear Kyra Kang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K261748 - Kyra Kang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261748 - Kyra Kang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K261748
Please provide the device trade name(s).
Dental Cone-beam Computed Tomography (mDX-15P Series)
Please provide your Indications for Use below.
The equipment is intended to produce two-dimensional (2D) or three-dimensional (3D) images of the human dental, maxillofacial regions and/or skull as diagnostic support for adult and pediatric patients.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
{4}
# K261748
# 510(k) Summary
# I. Submitter
Hefei Meyer Optoelectronic Technology Inc.
No.668 Wangjiang West Road, High & New Tech. Industry Development Zone, 230088
Hefei, PEOPLE'S REPUBLIC OF CHINA
Contact person: Mr. Zixuan Zhao
Position: RA manager
Tel.: +86-18756559593
E-mail: zhaozixuan@meyerop.com
Preparation date: May 14, 2026
Submission Correspondent
Ms. Kyra Kang
Landlink Healthcare Technology (Shanghai) Co., Ltd.
E-mail: kyra.kang@landlink-health.com
# II. Proposed Device
| Device Trade Name: | Dental Cone-beam Computed Tomography (mDX-15P Series) |
| --- | --- |
| Common name: | Computed tomography x-ray system |
| Regulation Number: | 21 CFR 892.1750 |
| Regulatory Class: | Class II |
| Product code: | OAS |
| Review Panel: | Radiology |
# III. Predicate Devices
510(k) Number: K230985
Trade name: Planmeca Viso
Common name: Computed tomography x-ray system.
Classification: Class II
Product Code: OAS
Manufacturer PLANMECA OY
Page 1 of 9
{5}
#### IV. Device description
The device is an advanced digital X-ray imaging unit that producing high quality digital images of dentitionTMJ and skull.
The device uses cone beam computed tomography (CBCT) to produce three-dimensional (3D)radiographic images of the maxilofacial, TMJ and ENT anatomies. The device can also producetwo-dimensional (2D) panoramic (PANO) images with the tomosynthesis method as well as optionalcephalometric (CEPH) images and carpus images with conventional 2D radiography.
The device includes the following models:
- mDX-15Pro-K10
- mDX-15Pro-Y30
- mDX-15Pro-B50
- mDX-15Plus-Y50S
#### V. Indications for use
The equipment is intended to produce two-dimensional (2D) or three-dimensional (3D) images of the human dental, maxillofacial regions and/or skull as diagnostic support for adult and pediatric patients.
#### VI. Comparison of technological characteristics with the predicate devices
The comparison and discussion between the subject device and the predicate devices are listed in below table
Table 5.2-1 General Comparison
| Characteristics | Proposed device | Predicate device (K230985) | Discussion |
| --- | --- | --- | --- |
| Trade name | Dental Cone-beam Computed Tomography mDX-15P Series | Dental Computed Tomography X-ray System | / |
| Classification Name | Computed tomography x-ray system | Computed tomography x-ray system | Same |
| Product Code | OAS | OAS | Same |
| Regulation Number | 21 CFR 892.1750 | 21 CFR 892.1750 | Same |
Page 2 of 9
{6}
| Indications for use | The equipment is intended to produce two-dimensional (2D) or three-dimensional (3D) images of the human dental, maxillofacial regions and/or skull as diagnostic support for adult and pediatric patients. | Planmeca Viso is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital x-ray images as well as three-dimensional (3D) optical images of the dentomaxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients | Same |
| --- | --- | --- | --- |
| Device Structure | The product consists of X-ray tube assembly, X-ray detector (for CBCT/PANO), X-ray detector (for CEPH/Carpus), column, touch control panel, exposure switch, exposure control box, chin cup, chin support, temple support, nasion support, ear support, TMJ nose support, handle, bite block, workstation, and software (Work Terminal, Data Center, Image Processing). | The product consists of X-ray unit, sensor with digital cameras, patient supports, patient handle, touch screen, moving column, stationary column, 3D reconstruction PC, Planmeca Romexis program, Ethernet switch. | Similar |
| Power supply | 100 ~ 240 VAC 50 / 60 Hz | AC 100-220V 50/60Hz, 230-240V, 50Hz | Analysis 1 |
| Patient population | pediatric and adult patients | pediatric and adult patients | Same |
| Clinical conditions | The X-ray images can be used for the examination of dental, maxillofacial and/or skull anatomy. The 3D images from CBCT shall only be used as an | This X-ray unit is intended to be used in a professional healthcare environment such as dental offices, clinics and similar environments. | Same |
Page 3 of 9
{7}
| | | imaging technique where conventional radiography failed to support diagnosis | | | |
| --- | --- | --- | --- | --- | --- |
| Dimenson | | 1320 (L) × 1818 (W) × 1640-2340 (H) (mm) | | 1540 (L) × 2160 (W) × 2340 (H) (mm) | Similar |
| Tube voltage | | 60 – 100 kV | | 60-120kV | Analysis 2 |
| Tube current | | 2-10mA | | 2-16mA | Analysis 3 |
| X-ray generator rated output power | | 1kW | | 1.6kW | Analysis 4 |
| Focus nominal value | | 0.5 mm | | 0.5 mm | Same |
| Voxel size (mm) | | 0.07mm, 0.08mm, 0.20mm, 0.25mm | | 0.6-0.075 | Same |
| SID (Source Image Distance) | mDX-15Plus-Y50S | CBCT/PANO : 660 mm CEPH : 1700 mm | CT photography: 700mm | Analysis 5 | |
| | | mDX-15Pro-B50 mDX-15Pro-K10 mDX-15Pro-Y30 | CBCT/PANO : 565 mm CEPH : 1700 mm | | Panorama photography: 700mm |
| | | | | | Cephalometric photography: 1700mm |
| Active area & Pixel size | CT | mDX-15Pro-K10: 151*201mm, 100um mDX-15Pro-Y30: 179.2*179.2mm, 100um mDX-15Pro-B50: 157×159.4mm, 120um mDX-15Plus-Y50S: 204.8cm×256mm, 100um | | 247.7*301.1mm, 139um | Analysis 6 |
| | PANO | mDX-15Pro-K10: 11*169 mm, 100um | | 8.896/17.792*166.8mm, 139um | |
Page 4 of 9
{8}
| | | mDX-15Pro-Y30: 11*169 mm, 100um mDX-15Pro-B50: 11*159 mm, 120um mDX-15Plus-Y50S: 11*169 mm, 100um | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | CE PH | mDX-15Pro-K10: 7.2*233.1mm, 100um mDX-15Pro-Y30: 6.8cm*225.2 mm, 100um mDX-15Pro-B50: 6*223.2 mm, 100um mDX-15Plus-Y50S: 300*250 mm, 125um | | | | | 301.82*248.9mm, 131um | |
| FOV | Scan Model | mD X- 15Pr o- K10 | mD X- 15Pr o- Y30 | mD X- 15Pr o- B50 | mD X- 15Pl us- Y50 S | 26*30, 16*16, 14*10, 10*10, 5*5 | Analysis 7 | |
| | | Skul l Sca n | not exist | not exist | not exist | | | 22 × 23* |
| | | Max illof acia l Sca n | 16 × 18* | 18 × 18* | 17 × 17* | | | 22 × 16 |
| | | Den titi o n Sca n | 16 × 12 | 18 × 11 | 17 × 10 | | | 18 × 11 |
| | | Dual Arc h Sca n | 10 × 10 | 11 × 11 | 10 × 10 | | | 11 × 11 |
| | | Sing le Arc h Sca n | 10 × 6 | 10 × 6 | 10 × 6 | | | 10 × 6 |
| | | Teet h Sca n | 5 × 8 | 5 × 8 | 5 × 8 | | | 5 × 8 |
Page 5 of 9
{9}
| | | **Endo Scan** | 4 × 4* | 4 × 4* | 4 × 4* | 4 × 4* | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | **PANO** | 20 × 11 | 20 × 11 | 20 × 11 | 20 × 11 | | |
| | | **CEPH** | 23 × 19 | 23 × 19 | 23 × 19 | 26 × 21 | | |
| | | * The scans with this mark are optional. | | | | | | |
| Scan time | | 0.8 – 16 s (depending on scan mode) | | | | | Max 36s | Analysis 8 |
| Software | Compatibility | Dicom 3.0 format compatible | | | | | Dicom 3.0 format compatible | Similar |
| | Name | Workstation System Software, Control System Software | | | | | Planmeca Romexis | |
| | Function | The storage and management of all data including patient information and images. The operator to add patients, acquire images, and analyze images daily. Process the images produced by the equipment for the analysis and diagnosis of diseases. | | | | | 2D&3D image view, Import/Export, DIR Media Storage, Print SCU, Storage SCU/SCP, Worklist SCU, Query/Retrieve SCU/SCP, Storage Commitment SCU, MPPS | |
### Analysis 1 Power supply
Both devices are designed to operate within standard voltage and frequency ranges commonly available in the US market. The Proposed device's power supply parameters are fully compatible with the electrical infrastructure in these regions, ensuring reliable and safe operation. Additionally, the device has been tested and validated to meet all relevant electrical safety standards, confirming that the specified power supply range does not compromise its performance or safety.
### Analysis 2 Tube voltage
The tube voltage range of the proposed device is comparable to that of the predicate device for the intended use. The minor difference does not affect image quality, radiation dose, or device performance. Therefore, this variation does not pose any risk to the safety or effectiveness of the device.
### Analysis 3 Tube current
The Proposed device's tube current range is sufficient to achieve the intended
Page 6 of 9
{10}
diagnostic image quality while maintaining patient safety. The device has been validated to ensure it meets all applicable performance and safety standards, demonstrating that the specified tube current range is appropriate for its intended use. Therefore, this variation does not pose any risk to the safety or effectiveness of the device.
# Analysis 4 X-ray generator rated output power
The proposed device's rated output power is lower than that of the predicate device. However, image quality testing confirmed that the system produces diagnostic image quality across all intended scan modes at this power level. The lower rated power contributes to radiation dose management while maintaining clinical utility. This is further supported by an independent clinical image quality assessment, in which an ABR-certified radiologist reviewed sample images from the device and confirmed that the overall image quality was clinically acceptable. Therefore, this difference does not adversely affect safety or effectiveness.
# Analysis 5 SID
SID is a design parameter that, together with focal spot size and detector pixel size, influences geometric unsharpness and spatial resolution. The proposed device uses a focal spot size equivalent to that of the predicate device and incorporates pixel sizes that are favorable for spatial resolution to compensate for the SID difference. Measured spatial resolution (IEC 61223-3-7) meets or exceeds clinical requirements. Radiation dose (Kerma/DAP) was also verified and complies with applicable limits. Therefore, the SID difference does not adversely affect safety or effectiveness.
# Analysis 6 Active area & Pixel size
The active area of each detector is selected based on the intended FOV for each specific model and is verified to fully cover the corresponding imaging field. The proposed device's pixel sizes are favorable for spatial resolution compared to the predicate device. Measured spatial resolution data (IEC 61223-3-7) confirmed that all models meet the clinical requirements. Therefore, the differences in detector active area and pixel size do not adversely affect imaging performance or clinical utility.
# Analysis 7 FOV
The system has been rigorously tested and validated to confirm that the FOV specifications provide sufficient coverage and resolution for its intended diagnostic applications. These differences do not compromise the device's ability to produce high-quality images or its compliance with applicable performance and safety standards. Therefore, the proposed device remains safe and effective for its intended use. Analysis 8 Scan time
Page 7 of 9
{11}
This minor variation in scan time is not clinically significant and does not compromise the safety or effectiveness of the device. The difference in scan time is attributed to variations in hardware design and image acquisition algorithms, which do not affect the overall performance, patient safety, or diagnostic efficacy of the device. Therefore, this variation does not pose any risk to the safety or effectiveness of the device.
### VII. Non-Clinical Testing
The device described in this summary, were tested and demonstrated to be in conformance with the following standards:
Performance testing :
IEC 61223-3-7
All models met the acceptance criteria for spatial resolution (Spatial Resolution Index 10 > 1 lp/mm)CNR (retention ≥60% after constancy test), acceptance index (≥100), uniformity (>5), and artifacts (no ring or geometric artifacts visible). Kiso values were below the 50 mGy limit across all models. After 3,600 exposures simulating 6 months of use, all constancy test results remained within the specified limits.
IEC 61223-3-4
Acceptance tests were performed on all four models (mDX-15Pro-K10, mDX-15Pro-Y30, mDX-15Pro-B50, and mDX-15Plus-Y50S) in both panoramic and lateral cephalometric modes. All models met the acceptance criteria for X-ray tube voltage deviation (±10%), total filtration (≥3.5 mm Al), focal spot size (width 0.5-0.75 mm, length 0.7-1.1 mm), radiation output reproducibility (coefficient of variation ≤0.05), and spatial resolution (≥3.1 lp/mm). Low contrast resolution (four holes of 2.5 mm to 1 mm distinguishable), image uniformity (no visible stripe fluctuation), patient positioning lasers, and panoramic focal trough (steel pins clearly visible with left-right symmetry) also passed visual and functional evaluations. The test results confirm that all models comply with the performance requirements of IEC 61223-3-4:2000.
IEC 60336
The X-ray tube (Canon D-054SB) used in all models was certified by the manufacturer to comply with IEC 60336:2005.
ISO 12052
The workstation software (mDX-15P_WorkstationSystem_V1.0) was tested for network communication (C-ECHO, C-STORE, C-FIND services), media storage and exchange (file format, compression format, CD/DVD and USB media), and data format integrity (core DICOM tags, metadata association, special character support). All test
Page 8 of 9
{12}
items passed. The conformance statement was verified to be complete and consistent with actual performance. The software fully complies with the DICOM interoperability requirements of ISO 12052:2017.
### Biocompatibility testing
Biocompatibility of patient-contacting components was evaluated per ISO 10993-1, -5, -10, and -23. The test article showed no evidence of cytotoxicity, sensitization, or irritation. All test results passed.
### Software
Software verification and validation was performed as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance of Premarket Submissions for Software Contained in Medical Devices.". No issues have been identified during the verification and validation Process.
### Cybersecurity
The device cybersecurity was evaluated in accordance with FDA Guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions." No high-risk vulnerabilities were identified.
### Electrical Safety and EMC
The device was tested for compliance with IEC 60601-1 / ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-4-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-63, and IEC 62366-1 through third-party testing. All applicable tests passed.
### VIII. Clinical Performance Data
Sample clinical images obtained from the subject device were provided to demonstrate its imaging performance. An ABR-certified radiologist with over 5 years of experience post-certification independently reviewed a number of sample scans and diagnostic images to evaluate different essential image quality related items. The results were documented in the Clinical Performance Evaluation Report. The overall image quality was acceptable.
### IX. Conclusion
The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device.
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.