Medical Thulium Fiber Laser Systems (UroFiber 150Q)
K261738 · Rhein Laser Technologies Co., Ltd. · GEX · Jul 21, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261738
Device Name
Medical Thulium Fiber Laser Systems (UroFiber 150Q)
Applicant
Rhein Laser Technologies Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 21, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Medical Thulium Fiber Laser Systems (UroFiber 150Q) is intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery and Arthroscopy.
Device Story
Medical Thulium Fiber Laser System (UroFiber 150Q) provides surgical laser energy for soft tissue management. System components include host unit (power supply, thulium fiber laser, safety control, cooling module), foot switch, and single-use optical fibers. Operated by physicians in clinical settings (OR, clinic). Laser energy (1940 nm) delivered via fiber to target tissue; enables incision, excision, resection, ablation, vaporization, coagulation, and hemostasis. Output controlled by user via foot switch; visual/audible feedback provided by system. Clinical benefit derived from precise tissue interaction and hemostatic capabilities during endoscopic, laparoscopic, or open procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (AAMI/ANSI ES60601-1, IEC 60601-1-2), laser safety (IEC 60825-1, IEC 60601-2-22), biocompatibility (ISO 10993-1 for cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity), and software verification/validation. Specification testing validated wavelength, pulse width, pulse energy, frequency, and power output.
Technological Characteristics
Thulium fiber laser; 1940 nm wavelength; CW/pulsed emission; 2500 Hz frequency; 150 W max average power; 200-1000 μm fiber core diameter. System includes host, foot switch, and optical fibers. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60825-1, and IEC 60601-2-22. Biocompatibility per ISO 10993-1. Software level of concern: basic.
Indications for Use
Indicated for patients requiring soft tissue surgical procedures including incision, excision, resection, ablation, vaporization, coagulation, and hemostasis across multiple specialties including Urology, Gastroenterology, Thoracic/Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery, and Arthroscopy.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
July 21, 2026
Rhein Laser Technologies Co., Ltd.
Na Wu
Quality Manager
801,8f, E2 Bldg., Future City, #999 High-Tech Ave., E. Lake High-Tech Development Zone, Wuhan
430206, China (Free Trade Zone Wuhan Area)
Wuhan, 430206
China
Re: K261738
Trade/Device Name: Medical Thulium Fiber Laser Systems (UroFiber 150Q)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: May 26, 2026
Received: May 27, 2026
Dear Na Wu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261738 - Na Wu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261738 - Na Wu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yan Fu-S Digitally signed by YAN
FU-S
Date: 2026.07.21 21:06:28
-04'00'
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261738 | ? |
| Please provide the device trade name(s). | | ? |
| Medical Thulium Fiber Laser Systems (UroFiber 150Q) | | |
| Please provide your Indications for Use below. | | ? |
| Medical Thulium Fiber Laser Systems (UroFiber 150Q) is intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery and Arthroscopy. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☐ Adults (22 years old and greater) | ? |
{4}
K261738
# Rhein Laser Technologies Co., Ltd.
# **510(k) Summary**
Medical Thulium Fiber Laser Systems
Date of Summary Modification: July 3rd, 2026
# **1. Applicant Information**
Rhein Laser Technologies Co., Ltd.
801,8F, E2 Building, Future City, No.999 High-Tech Avenue, East Lake
High-Tech Development Zone, Wuhan 430206, China ( Free Trade Zone
Wuhan Area)
Contact Person: Na Wu
Title: Quality Manager
Telephone Number: +86-027-65279157
Email: na.wu@rheinlaser.com
# **2. Proposed Device**
**Trade Name:** Medical Thulium Fiber Laser Systems (UroFiber 150Q)
**Common Name:** Powered Laser Surgical Instrument
**Classification Name:** Laser surgical instrument for use in general and
plastic surgery and in dermatology
**Model(s):** UroFiber 150Q
**Classification Regulation Number:** 21 CFR 878.4810
**Product Code:** GEX (Powered Laser Surgical Instrument)
**Device Class:** II
**Review Panel:** General & Plastic Surgery
# **3. Predicate Device 1**
**510(k) Number:** K253100
**Applicant:** Asclepion Laser Technologies GmbH
**Classification Name:** Laser surgical instrument for use in general and
plastic surgery and in dermatology
**Trade Name:** MultiPulse TFL
# **4. Predicate Device 2**
**510(k) Number:** K133891
**Applicant:** Asclepion Laser Technologies GmbH
**Classification Name:** Laser surgical instrument for use in general and
1 / 5
{5}
K261738
Rhein Laser Technologies Co., Ltd.
**Trade Name:**
plastic surgery and in dermatology MULTIPULSE TM+1470
# **5. Device Description**
Medical Thulium Fiber Laser Systems (UroFiber 150Q) consists of a host (including a power supply, thulium fiber laser, safety control module, cooling module), foot switch and optical fiber (single use, sterilized by the user before use. Model: RL-200, RL-272, RL-365, RL-550, RL-800, RL-1000, specification: 300) composition. Protective glasses not included.
# **6. Indications for Use**
Medical Thulium Fiber Laser Systems (UroFiber 150Q) is intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery and Arthroscopy.
# **7. Comparison to the Predicate Devices**
# **Indications for Use/Intended Use:**
The indications for use of the Medical Thulium Fiber Laser Systems (UroFiber 150Q) are within those of the predicate devices.
# **Technological Characteristics:**
The Medical Thulium Fiber Laser Systems (UroFiber 150Q) has the same intended use, principle of operation and technical characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do not raise any issues of safety or efficacy.
Performance data supports that the device is as safe and effective as the predicate devices for its intended use.
2 / 5
{6}
K261738
Rhein Laser Technologies Co., Ltd.
Table 1 Comparison between UroFiber 150Q and the Predicate Devices
| Feature | Proposed Device | Predicate device 1 (K253100) | Predicate Device 2 (K133891) | Remarks |
| --- | --- | --- | --- | --- |
| Manufacturer | Rhein Laser Technologies Co., Ltd. | Asclepion Laser Technologies GmbH | Asclepion Laser Technologies GmbH | / |
| Device Name | Medical Thulium Fiber Laser Systems | MultiPulse TFL | MULTIPULSE TM+1470 | / |
| Model(s) | UroFiber 150Q | / | / | / |
| Laser Source | Thulium Laser | Thulium Laser | Thulium Laser | Same |
| Wavelength | 1940 nm | 1940 nm | 1940 nm and 1470 nm | Same |
| Emission | CW/Pulsed | CW/Pulsed | CW/Pulsed | Same |
| Pulse Width | CW or 0.1 to 15 ms | CW or 0.1 to 12 ms | CW or 0.5 to 750 ms | Within Predicate Devices |
| Frequency | 2500 Hz | 2500 Hz | 2500 Hz | Same |
| Max Average Power | 150 W | 200 W | 150 W | Within Predicate Devices |
| Fiber core diameter | 200 – 1000 μm | 200 – 1000 μm | 200 – 1000 μm | Same |
3 / 5
{7}
K261738
Rhein Laser Technologies Co., Ltd.
8. **Performance Data**
The following performance data were provided in support of the substantial equivalence determination:
**Electrical safety and electromagnetic compatibility (EMC)**
The proposed device complies with the following recognized consensus standards:
- AAMI/ANSI ES60601-1:2005(R) 2012 and A1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- IEC 60601-1-2:2014+AMD1:2020 Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
**Laser Safety**
Testing was conducted to demonstrate that the proposed device complies with the following standards:
- IEC 60825-1:2014 Safety of Laser Products - Part 1: Equipment Classification and Requirements
- IEC 60601-2-22:2019 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
**Biocompatibility testing**
Biocompatibility evaluation for the Optical Fibers after sterilization was conducted in accordance with the FDA Guidance “Use of International Standard ISO-10993-1, ‘Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process’” September 2023. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation
- Acute Systemic Toxicity
- Material Mediated Pyrogenicity
The optical fibers are considered tissue contacting for a duration of less than 24 hours.
4 / 5
{8}
K261738
Rhein Laser Technologies Co., Ltd.---
# **Software Verification and Validation Testing**
Software verification and validation testing were conducted and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.” The software documentation level for this device was determined to be “basic.”
# **Specification Testing**
Testing was conducted to validate that the proposed device meets design specifications for wavelength, pulse width, pulse energy, frequency, and maximum average power.
# **9. Conclusions**
The above performance testing supports that the Medical Thulium Fiber Laser Systems (UroFiber 150Q) is as safe and as effective as the predicate devices currently marketed for the same intended use.
5 / 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.