K261734 · Total Joint Orthopedics, Inc. · JWH · Jul 23, 2026 · Orthopedic
Device Facts
Record ID
K261734
Device Name
Klassic Infection Spacer System
Applicant
Total Joint Orthopedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Jul 23, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Klassic® Knee System is intended for prosthetic replacement in treatment of the following: - Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular necrosis and osteoarthritis - Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis - Patients with failed previous surgery where pain, deformity, or dysfunction persists - Correctable varus-valgus deformity and moderate flexion contracture - Revision of a previously failed knee arthroplasty - Patients who require a total knee replacement The Klassic® Knee System is indicated for cemented use only, except for the Klassic Femur with Cobalt 3D®, the Klassic Tibial Baseplate with Ti-Coat® and the Universal® Cone with Ti-Coat, which are indicated for cementless use. The Klassic® Knee System is also indicated for temporary use (maximum 180 days) as an adjunct to total knee replacement (TKR) in patients undergoing a two-stage procedure due to a septic process and where gentamicin is the most appropriate antibiotic based on the susceptibility pattern of the infecting micro-organism(s). The Infection Spacer Stem Dowel, Infection Spacer Femur Adapter and Infection Spacer Femur Adapter, Ti are only indicated for this temporary use. These devices are intended for implantation with FDA cleared gentamicin loaded bone cement following implant removal and radical debridement, and in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach for an infection). For the first stage revision, these devices are not intended for use for more than 180 days, at which time they must be explanted and permanent devices implanted, or another appropriate treatment performed (e.g. resection arthroplasty, fusion, etc.). These devices are only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period.
Device Story
Klassic Knee System provides prosthetic knee replacement components; includes infection spacers, dome extensions, augments, stem dowels, and femur adapters. Used in orthopedic surgery for primary or revision total knee arthroplasty; temporary infection spacers used in two-stage septic revision procedures. Components implanted via cementation or cementless fixation; infection spacers used with gentamicin-loaded bone cement. Surgeon selects components based on patient anatomy and pathology; output is a reconstructed knee joint or temporary spacer construct. Benefits include restoration of joint function, pain relief, and management of periprosthetic infection via antibiotic-loaded spacer. Operates in clinical/OR settings.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and engineering analyses: ASTM F1223 (constraint), ISO 21536 (contact stress), ISO 14243-3 (wear), range of motion evaluation, cement fixation analysis, and volumetric gentamicin content analysis.
Technological Characteristics
Materials: UHMWPE (GUR-1050/1020 per ASTM F648) and Titanium Alloy (Ti6Al4V per ASTM F136). Components include infection spacers, augments, and adapters. Fixation: Cemented or cementless. Sterilization: EO or Gamma. Design features include threaded interfaces for spacers/extensions.
Indications for Use
Indicated for patients requiring total knee replacement due to NIDJD (avascular necrosis, osteoarthritis), IJD (rheumatoid arthritis), failed previous surgery, varus-valgus deformity, flexion contracture, or revision arthroplasty. Temporary use (max 180 days) indicated for two-stage revision of septic knee processes using gentamicin-loaded bone cement and systemic antibiotics; requires consistent use of mobility aids.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
July 23, 2026
Total Joint Orthopedics, Inc.
% Chris Weaber
Founder, Principal Consultant
Arthroplasty Regulatory Consulting
1433 E. Mulberry Way
Sandy, Utah 84093
Re: K261734
Trade/Device Name: Klassic Knee System
Regulation Number: 21 CFR 888.3560
Regulation Name: Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis
Regulatory Class: Class II
Product Code: JWH, MBB
Dated: May 26, 2026
Received: May 26, 2026
Dear Chris Weaber:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261734 - Chris Weaber
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K261734 - Chris Weaber
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSE
MUIR -S
Digitally signed by
JESSE MUIR -S
Date: 2026.07.23
11:23:21 -04'00'
Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K261734
Device Name
The Klassic® Knee System
Indications for Use (Describe)
The Klassic® Knee System is intended for prosthetic replacement in treatment of the following:
- Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular necrosis and osteoarthritis
- Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis
- Patients with failed previous surgery where pain, deformity, or dysfunction persists
- Correctable varus-valgus deformity and moderate flexion contracture
- Revision of a previously failed knee arthroplasty
- Patients who require a total knee replacement
The Klassic® Knee System is indicated for cemented use only, except for the Klassic Femur with Cobalt 3D®, the Klassic Tibial Baseplate with Ti-Coat® and the Universal® Cone with Ti-Coat, which are indicated for cementless use.
The Klassic® Knee System is also indicated for temporary use (maximum 180 days) as an adjunct to total knee replacement (TKR) in patients undergoing a two-stage procedure due to a septic process and where gentamicin is the most appropriate antibiotic based on the susceptibility pattern of the infecting micro-organism(s). The Infection Spacer Stem Dowel, Infection Spacer Femur Adapter and Infection Spacer Femur Adapter, Ti are only indicated for this temporary use. These devices are intended for implantation with FDA cleared gentamicin loaded bone cement following implant removal and radical debridement, and in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach for an infection). For the first stage revision, these devices are not intended for use for more than 180 days, at which time they must be explanted and permanent devices implanted, or another appropriate treatment performed (e.g. resection arthroplasty, fusion, etc.). These devices are only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
K261734
# 510(k) Summary
| Manufacturer: | Total Joint Orthopedics, Inc. 1567 E. Stratford Avenue Salt Lake City, UT 84106 Phone: 801.486.6070 |
| --- | --- |
| Contact: | Mr. Chris Weaber Consultant |
| Prepared By: | Chris Weaber Arthroplasty Regulatory Consulting 1433 E. Mulberry Way Sandy, UT 84093 Phone: 650.922.7326 |
| Date Prepared: | July 21, 2026 |
| Device Trade Name: | Klassic® Knee System |
| Device Common Name: | Spacer, Dome Extension, Augments, Stem Dowel, Femur Adapter |
| Classification: | 21 CFR 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis 21 CFR 888.3027 Polymethylmethacrylate (PMMA) bone cement. Class II |
| Product Code: | JWH, MBB |
Primary Predicate: Signature Rx Knee System, K240683, 21 CFR 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
Additional Predicate: The Klassic® Knee System, K243991, 21 CFR 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
Additional Predicate: Osteoremedies Remedy Knee Spacer, K183017, 21 CFR 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
510(k) Summary – Klassic® Infection Spacer System
Page 1
{5}
K261734
# Indications for Use:
The Klassic® Knee System is intended for prosthetic replacement in treatment of the following:
- Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular necrosis and osteoarthritis
- Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis
- Patients with failed previous surgery where pain, deformity, or dysfunction persists
- Correctable varus-valgus deformity and moderate flexion contracture
- Revision of a previously failed knee arthroplasty
- Patients who require a total knee replacement
The Klassic® Knee System is indicated for cemented use only, except for the Klassic Femur with Cobalt 3D®, the Klassic Tibial Baseplate with Ti-Coat® and the Universal® Cone with Ti-Coat, which are indicated for cementless use.
The Klassic® Knee System is also indicated for temporary use (maximum 180 days) as an adjunct to total knee replacement (TKR) in patients undergoing a two-stage procedure due to a septic process and where gentamicin is the most appropriate antibiotic based on the susceptibility pattern of the infecting micro-organism(s). The Infection Spacer Stem Dowel, Infection Spacer Femur Adapter and Infection Spacer Femur Adapter, Ti are only indicated for this temporary use. These devices are intended for implantation with FDA cleared gentamicin loaded bone cement following implant removal and radical debridement, and in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach for an infection). For the first stage revision, these devices are not intended for use for more than 180 days, at which time they must be explanted and permanent devices implanted, or another appropriate treatment performed (e.g. resection arthroplasty, fusion, etc.). These devices are only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period.
# Device Description:
The Klassic® Knee System is being updated with modifications to include additional Infection Spacer, Infection Spacer Dome Extension, Infection Spacer Stem Dowel, Infection Spacer Femur Adapter, and Infection Spacer Femur Adapter, Ti implants.
These proposed geometries are fabricated from Standard Poly (GUR-1050 or GUR-1020 UHMWPE) per ASTM F648 or Titanium Alloy (Ti6Al4V) per ASTM F136.
# Comparison of Technological Characteristics:
The modified Klassic® Knee System features the same materials (GUR 1050 UHMWPE, GUR 1020 UHMWPE, and Ti6Al4V), same use in Total Knee Arthroplasty, and same EO or Gamma sterilization compared to the primary predicate Rx Knee System (K240683) and reference predicate Klassic® Knee System (K243991).
The modified Klassic® Knee System is substantially equivalent to the primary predicate Signature Rx Knee System which includes the same indications of Temporary Use (180-day max) as an adjunct to total knee replacement (TKR) in patients undergoing a two-stage procedure due to a septic process.
510(k) Summary – Klassic® Infection Spacer System
Page 2
{6}
K261734
The subject modified Klassic® Knee System includes components of a total knee system for temporary use similar to the primary predicate. The subject system includes similar but different design features, such as articulations, sizes, and geometries. The subject Stem Dowel is comprised entirely of UHMWPE which differs from the predicate Universal Stem component which is PMMA bone cement with gentamicin with metal reinforcement. The subject Femur Adapters affix to the femoral component with cement which differs from the threaded femoral fixation for the RX Knee System.
The modified Klassic® Knee System is substantially equivalent to the additional predicate device Klassic® Knee System (K243991) with respect to design, materials, function, packaging, and sterilization. The subject Infection Spacer features a threaded interface with the Infection Spacer Dome Extension which differs from the predicate Klassic® All Poly Tibia, Ultra-PS (K243991) which is a monoblock geometry. Additionally, the subject Augment introduces a modified design and material for the predicate Tibial Augment (K243991).
The subject components are substantially equivalent to the additional predicate Osteoremedies Remedy Knee Spacer (K183017) for total Gentamicin implant construct content and feature the same indications for Temporary Use (180-day max). The subject components include UHMWPE and Titanium materials, which differs compared to the predicate Osteoremedies Remedy Knee Spacer (K183017) which is composed entirely of Gentamicin Bone Cement with stainless steel reinforcement.
# Discussion of Non-Clinical Testing/Performance Data:
Engineering analyses were performed to evaluate the subject component equivalence to the primary and reference predicates. Additionally, the Klassic® Infection Spacer System implants are in compliance with LAL testing requirements for orthopedic implants.
Engineering analysis conducted to demonstrate substantial equivalence was as follows:
- Femoral/Tibial Constraint Analysis (ASTM F1223)
- Contact Area/Contact Stress Analysis (ISO 21536)
- Range of Motion Evaluation
- Wear Analysis (ISO 14243-3)
- Cement Fixation Analysis
- Subject Infection Spacer Component Thickness per FDA Guidance “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous Coated Uncemented Prostheses; Guidance for Industry and FDA”
- Volumetric Cement Analysis - Total content of Gentamicin side by side predicate comparison
- Temporary Use: Subject Infection Stem Dowel, Femur Adapter and Femur Adapter, Ti Component Strength and Fixation
# Conclusion:
Engineering analyses demonstrated that the subject components met the pre-determined acceptance criteria identified in the Design Control Activities, demonstrating that the subject components perform as safe and effective compared to the predicate components, and are substantially equivalent to the predicates.
510(k) Summary – Klassic® Infection Spacer System
Page 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.