K261706 · KARL STORZ SE & CO. KG · HIG · Jul 16, 2026 · Obstetrics/Gynecology
Device Facts
Record ID
K261706
Device Name
Endomat Select (UP220)
Applicant
KARL STORZ SE & CO. KG
Product Code
HIG · Obstetrics/Gynecology
Decision Date
Jul 16, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1700
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Endomat Select is intended to:·provide the infusion of the sterile irrigation solution into the ureter and upper urinary tract, as well as to suction irrigation fluids, bodily fluids, secretions, tissue and gas during diagnostic and operative endoscopic urological procedures.·provide the infusion of the sterile irrigation solution into the uterus, as well as to suction irrigation fluids, bodily fluids, secretions, tissue and gas during diagnostic and operative endoscopic hysteroscopic procedure.·provide the infusion of the sterile irrigation solution into organs and operating fields during diagnostic and operative endoscopic procedures in laparoscopic and open general surgery.·provide sustained liquid irrigation and distention of joint or intra-articular spaces as well as to suction irrigation fluids, bodily fluids, secretions, tissue and gas during all phases of arthroscopic surgery.·provide the infusion of the sterile irrigation solution in order to enable cleaning of the lens during endoscopically assisted Functional Endoscopic Sinus Surgery.
Device Story
Endomat Select (UP220) is a multi-functional, pressure-controlled, peristaltic irrigation and suction pump for endoscopic procedures. Device inputs include encoded tubing sets that trigger automatic mode selection; user inputs via LCD touch screen. Device transforms inputs into controlled fluid infusion and suction; operates in various modes (HYS, ART, URO, SURG, ENT). Used in OR/clinic settings by physicians/nurses. Features network connectivity (HIVE protocol) for remote servicing. Output provides irrigation/distention and suction of fluids/tissue; enables lens clarity. Clinical benefit: maintains surgical field visibility and distention, facilitating diagnostic and operative interventions. Bench testing confirms performance of new suction/irrigation modes.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including verification of new suction/irrigation modes, electrical safety (IEC 60601 series), EMC, and software/cybersecurity validation.
Technological Characteristics
Microprocessor-controlled peristaltic roller pump. Materials: medical-grade components per ISO 13485. Connectivity: Networked (HIVE protocol) for remote/on-site service. User interface: LCD touch screen. Max suction: -0.96 bar; Max irrigation: 500 mmHg. Software: Modular packages (UP601-UP610). Sterilization: Compatible with sterile tubing sets.
Indications for Use
Indicated for patients undergoing diagnostic and operative endoscopic procedures in urology, hysteroscopy, laparoscopic/open general surgery, arthroscopy, and Functional Endoscopic Sinus Surgery. Applicable to all age groups from neonates to adults.
Regulatory Classification
Identification
A hysteroscopic insufflator is a device designed to distend the uterus by filling the uterine cavity with a liquid or gas to facilitate viewing with a hysteroscope.
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
July 16, 2026
Karl Storz Se & Co. Kg
Emily Rhiel
Senior Regulatory Affairs Specialist
Dr.-Karl-Storz-Strabe 34
Tuttlingen, Baden-Wurttemberg 78532
GERMANY
Re: K261706
Trade/Device Name: Endomat Select (UP220)
Regulation Number: 21 CFR 884.1700
Regulation Name: Hysteroscopic Insufflator
Regulatory Class: II
Product Code: HIG, LJH, BTA, EOB, HRX, OCX
Dated: May 22, 2026
Received: May 22, 2026
Dear Emily Rhiel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices an Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261706 - Emily Rhiel
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K261706 - Emily Rhiel
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# JASON ROBERTS -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261706 | ? |
| Please provide the device trade name(s). | | ? |
| Endomat Select (UP220) | | |
| Please provide your Indications for Use below. | | ? |
| Endomat Select is intended to:·provide the infusion of the sterile irrigation solution into the ureter and upper urinary tract, as well as to suction irrigation fluids, bodily fluids, secretions, tissue and gas during diagnostic and operative endoscopic urological procedures.·provide the infusion of the sterile irrigation solution into the uterus, as well as to suction irrigation fluids, bodily fluids, secretions, tissue and gas during diagnostic and operative endoscopic hysteroscopic procedure.·provide the infusion of the sterile irrigation solution into organs and operating fields during diagnostic and operative endoscopic procedures in laparoscopic and open general surgery.·provide sustained liquid irrigation and distention of joint or intra-articular spaces as well as to suction irrigation fluids, bodily fluids, secretions, tissue and gas during all phases of arthroscopic surgery.·provide the infusion of the sterile irrigation solution in order to enable cleaning of the lens during endoscopically assisted Functional Endoscopic Sinus Surgery. | | |
| Please select the types of uses (select one or both, as applicable). | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | Neonates/Newborns (Birth to < 29 days old)Infants (29 days old to < 2 years old)Children (2 years old to < 12 years old)Adolescents (12 years old to < 22 years old)Adults (22 years old and greater) | ? |
{4}
STORZ
KARL STORZ – ENDOSKOPE
KARL STORZ Premarket Notification
Endomat Select (UP220)
K261706 510(k) Summary
K261706, Page 1 of 5
### 510(k) Summary
| Submitter: | KARL STORZ SE & Co. KG Dr.-Karl-Storz-Straße 34 78532 Tuttlingen, Germany |
| --- | --- |
| Contact: | Emily Rhiel Senior Regulatory Affairs Specialist Tel.: (774) 318-2820 Email: Emily.Rhiel@karlstorz.com |
| Date of Preparation: | July 16, 2026 |
| Type of 510(k) Submission: | Traditional 510(k) |
| Device Identification: | Trade Name: Endomat Select (UP220) Common Name: Hysteroscopic Insufflator Regulation Number: 21 CFR 884.1700 |
| Regulatory Class: | II |
| Product Code: | HIG, BTA, LJH, HRX, OCX, EOB |
| Guidance Document: | Hysteroscopic and Laparoscopic Insufflators: Submission Guidance for a 510(K) issued on August 1, 1995 for Product Code HIG |
| Predicate Devices: | KARL STORZ Endomat Select (K201355) The predicate has not been subject to a design related recall. |
| Device Description: | The Endomat Select is a multi-functional, pressure-controlled, combined irrigation and suction pump. It can be used for irrigation and, where appropriate dilation, during Hysteroscopic (HYS), Arthroscopic (ART), Urological (URO) and General Surgical or Laparoscopic (SURG) interventions. The device can be used for suction during Urological (URO) and Hysteroscopic (IBS) interventions. In addition, the device can function in an oscillatory mode (ENT) providing a fluid means to maintain lens clarity during use in Transnasal procedures. The device has a modern LCD display with touch screen user interface. The tubing sets are encoded so that they can be identified by the device, which selects the appropriate operating modes based on the tubing set connected. The device also has network connectivity which can be used for remote servicing and to connect to other KARL STORZ devices proprietary network communication protocol (HIVE). |
| Indications For Use: | Endomat Select is intended to:. • provide the infusion of the sterile irrigation solution into the ureter and upper urinary tract, as well as to suction irrigation fluids, bodily fluids, secretions, tissue and gas during diagnostic and operative endoscopic urological procedures. |
{5}
STORZ
KARL STORZ – ENDOSKOPE
KARL STORZ Premarket Notification
Endomat Select (UP220)
K261706 510(k) Summary
K261706, Page 2 of 5
| | - provide the infusion of the sterile irrigation solution into the uterus, as well as to suction irrigation fluids, bodily fluids, secretions, tissue and gas during diagnostic and operative endoscopic hysteroscopic procedure. - provide the infusion of the sterile irrigation solution into organs and operating fields during diagnostic and operative endoscopic procedures in laparoscopic and open general surgery. - provide sustained liquid irrigation and distention of joint or intra-articular spaces as well as to suction irrigation fluids, bodily fluids, secretions, tissue and gas during all phases of arthroscopic surgery. - provide the infusion of the sterile irrigation solution in order to enable cleaning of the lens during endoscopically assisted Functional Endoscopic Sinus Surgery. | | |
| --- | --- | --- | --- |
| **Predicate Indications for Use** | Endomat Select is intended to: - provide the infusion of the sterile irrigant solutions into the ureter and upper urinary tract, as well as to suction off the irrigation fluids, bodily fluids, secretions, tissue and gas during diagnostic and operative endoscopic urological procedures - provide the infusion of the sterile irrigant solutions into the uterus, as well as to suction off the irrigation fluids, bodily fluids, secretions, tissue and gas during diagnostic and operative endoscopic hysteroscopic procedures - provide the infusion of the sterile irrigant solutions into organs and operating fields during diagnostic and operative procedures in laparoscopic and open general surgery - provide sustained liquid irrigation and distention of joint or intra-articular spaces during all phases of arthroscopic surgery - provide the infusion of the sterile irrigant solutions in order to enable the Lens Cleaning during endoscopically assisted Functional Endoscopic Sinus Surgery and endoscopically assisted transnasal pituitary gland surgery | | |
| **Indications for Use Comparison** | The subject device extends the arthroscopy indication to include suction, in addition to the cleared arthroscopic irrigation indications. The subject device also removes the indications for use during “endoscopically assisted Transnasal Pituitary Gland Surgery”. The modifications to not introduce a new intended use. | | |
| **Technological Characteristics:** | | **Subject device** | **Predicate Device** **K201355** |
| | Intended Use | Suction/irrigation pumps and their accessories are used for the introduction of irrigation fluids into organs, joints and operating fields as well as the suctioning off of irrigation fluids and bodily fluids, secretions, tissue and gas during diagnostic or therapeutic interventions. | Same as subject |
| | Device Design | Microprocessor controlled, pressure monitoring peristaltic pump for irrigation and flow controlled peristaltic pump for suction | Same as subject |
{6}
STORZ
KARL STORZ – ENDOSKOPE
KARL STORZ Premarket Notification
Endomat Select (UP220)
K261706 510(k) Summary
K261706, Page 3 of 5
| | Device Type | Roller-pump | Same as Subject |
| --- | --- | --- | --- |
| | Input/Output devices | LED touch screen | Same as subject |
| | Maximum Suction pressure | -0.96 bar | Same as subject |
| | Maximum Irrigation pressure | 500 mmHg | Same as subject |
| | Irrigation Pressure | Gen/Lap surgery: 100, 300 and 500 mmHg Hyst: 20-150 mmHg Uro: 20-150 mmHg Uro (irrigation RES): 20 – 150 mmHg Arth:20-150 mmHg ENT: N/A | Gen/Lap surgery: 100, 300 and 500 mmHg Hyst: 20-150 mmHg Uro: 20-150 mmHg Arth:20-150 mmHg ENT/Neuro: N/A |
| | Irrigation Flow | Gen/Lap Surgery: 100-3500 ml/min Hyst: 200, 400, 600 ml/min Uro: 200, 400, 600 ml/min **Uro (irrigation RES): 200, 400, 600 ml/min** Arth: 1500, 2000, 2500 ml/min ENT: 50, 65, 80, 95, 110, 130 ml/min | Gen/Lap Surgery: 100-3500 ml/min Hyst: 200, 400, 600 ml/min Uro: 200, 400, 600 ml/min Arth: 1500, 2000, 2500 ml/min ENT/Neuro: 50, 65, 80, 95, 110, 130 ml/min |
| | Suction Pressure | Hyst (suction): N/A IBS (suction): N/A Uro (suction RES): N/A **Arth (suction): N/A** | Hyst: N/A IBS (suction): N/A Uro (suction RES): N/A **Uro (Calcuson): N/A** |
| | Suction Flow | **Hyst (suction): 10 – 180 ml/min** IBS (suction): 100 – 300 ml/min Uro (suction RES): 100 – 1000 ml/min **Arth (suction): 100 – 1000 ml/min** | Hyst (suction): N/A IBS (suction): 100 – 300 ml/min Uro (suction RES): 100 – 1000 ml/min **Uro (Calcuson): 300 – 1000 ml/min** |
| | Service/update function | Remote or on-site | On-site |
| | Software Packages | UP601 – SURGERY; includes procedures "LAP", "THOR" and "PROCTO" UP602 – HYSTEROSCOPY, includes procedures "HYS" (**suction** and irrigation) | UP601 – Same as subject UP602 – HYSTEROSCOPY Software, license, allows selection of the procedure "HYS" (**irrigation only**) |
{7}
STORZ
KARL STORZ – ENDOSKOPE
KARL STORZ Premarket Notification
Endomat Select (UP220)
K261706 510(k) Summary
K261706, Page 4 of 5
| | | UP603 – GYN Shaver, includes procedure "GYN Shaver" (suction and irrigation) UP604 - UROLOGY, includes procedures "CYST", "PCN", "URS" and "RES (suction and Irrigation) UP605 – ARTHROSCOPY, includes procedures "KNEE", "HIP", "SHOULDER" and "SMALL JOINTS" (suction and irrigation) UP606 – ENT includes procedure "CLEARVISION" UP610 – ADVANCED BOOST Available with UP604 and UP605 irrigation modes | UP603 – Same as subject UP604 – UROLOGY, includes procedures "CYST", "PCN", "URS", "CALCUSON" and "RES (suction only) UP605 – ARTHROSCOPY, includes procedures "KNEE", "HIP", "SHOULDER" and "SMALL JOINTS" (irrigation only) UP606 – ENT/NEURO, includes procedures "CLEARVISION" UP610 – same as subject BOOST available for UP604 and UP605 irrigation mode |
| --- | --- | --- | --- |
| The subject device has the same intended use as the predicate. The differences between the subject device and predicate device include new modes (Uro irrigation RES, Hys suction, Arth suction), remote servicing options, removal of Calcuson option for urological suction, and removal of NEURO as a procedure type for software package UP606. These differences do not raise different questions of safety and effectiveness and have been confirmed through testing. | | | |
| Non-Clinical Performance Data: | The Endomat Select follows the FDA recognized consensus standards and is tested according to the following standards and FDA Guidances: ❖ Electrical Safety and EMC ➢ IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 ➢ IEC 60601-1-6:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 ➢ IEC 60601-1-8: 2006, AMD1:2012, AMD2:2020 ➢ IEC 60601-1-2:2014, IEC 60601-1-2:2014/AMD1:2020 ➢ IEC 60601-2-2: 2017 ➢ “Electromagnetic Compatibility (EMC) of Medical Devices” (June 2022) ❖ Software and Cybersecurity Verification and Validation ➢ Software verification and validation testing were conducted and documentation was provided as recommended by the FDA’s guidance, “Content of Premarket Submissions for Device Software Functions” issued in June 2023. ➢ Cybersecurity was evaluated according to the FDA guidance “Cybersecurity in Medical Device: Quality System Considerations and Content of Premarket Submissions” issued in February 20263 ➢ IEC 62304 Edition 1.1 2015-06 ❖ Stability and Shelf Life ➢ ASTM F88/F88M-23 ➢ ASTM F2096-11 ➢ ASTM F1929-23 | | |
{8}
STORZ
KARL STORZ – ENDOSKOPE
KARL STORZ Premarket Notification
Endomat Select (UP220)
K261706 510(k) Summary
K261706, Page 5 of 5
| | ➤ ASTM D4169-22 | |
| --- | --- | --- |
| | Bench testing was performed to ensure the device met its design specifications for the additional procedures not applicable to the predicate: | |
| | Verification Test | Conclusion |
| | HYS Suction | Pass |
| | Flow URO RES | Pass |
| | ART Suction | Pass |
| | BOOST feature for URO RES | Pass |
| The bench testing verified and validated that the Endomat Select has met all its design specifications and is substantially equivalent to the predicate device. | | |
| Clinical Performance Data: | Clinical testing was not required to demonstrate the substantial equivalence to the predicate device. Non-clinical bench testing was sufficient to establish the substantial equivalence of the modifications. | |
| Conclusion: | The conclusions drawn from the nonclinical tests demonstrate that the subject device, the Endomat Select (UP220) is as safe and effective as the predicate device (UP210), which is legally marketed for the same intended use. The data support the conclusion that the subject device is substantially equivalent to the predicate device. | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.