Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.
Device Story
Home-use IPL hair removal device; utilizes xenon arc flashlamp to emit light pulses (530-1200nm) to reduce hair growth; device includes main unit, power adapter, lamp tube, filter, display, buttons, thermoelectric cooler, and fan; skin sensor prevents light emission unless in full contact with skin; operated by user via manual finger switch or auto-emission mode; cooling function (sapphire) enhances comfort; output intensity adjustable (5 levels); intended for small areas (underarm, facial hair below chin) and large areas (arms, legs); non-invasive; no cutting or pulling.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-83), and photobiological safety (IEC 62471).
Technological Characteristics
IPL-based hair removal; Xenon arc flashlamp source; 530-1200nm wavelength; 1.4-3.85 J/cm² energy density; 3.9±0.25cm² spot size; 6-11ms pulse duration; sapphire cooling; external power supply (100-240V, 50/60Hz); microprocessor-controlled; biocompatible materials per ISO 10993; home-use environment.
Indications for Use
Indicated for removal of unwanted body and facial hair in individuals with Fitzpatrick Skin Types I-V.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
August 10, 2026
Shenzhen Jianchao Intelligent Technology Co., Ltd.
% Renne Qin
Primary Contact
Shenzhen Reanny Medical Devices Management Consulting Co., Ltd.
Rm. 1509, Jingting Bldg., Dongzhou Community, Guangming St., Guangming District
Shenzhen, Guangdong 518107
China
Re: K261705
Trade/Device Name: Hair Removal Device (R3C16-G Pro, R3C16-V Pro)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHT
Dated: July 14, 2026
Received: July 14, 2026
Dear Renne Qin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261705 - Renne Qin
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-
{2}
K261705 - Renne Qin
Page 3
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE-S
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261705
Device Name
Name of Device: Hair Removal Device
Model(s): R3C16-G Pro, R3C16-V Pro
Indications for Use (Describe)
Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Shenzhen Jianchao Intelligent Technology Co., Ltd.
510(k)s – K261705\_510(k) Summary
File No.:July 2, 2026---
# K261705
## 510(k) Summary
**'510(k) Summary" as required by 21 CFR Part 807.92.**
### I. Submitter
Shenzhen Jianchao Intelligent Technology Co., Ltd.
Rm101, 201, 301, Bldg.25, No.68 Hexiu West Road, Fuhai St., Baoan, Shenzhen,
Guangdong, 518103, CHINA
Phone:+86 17521227703
Email:ce@jianchaosmart.com
Contact person:Renne Qin
Title:RA Specialist
Date of preparation: July 24, 2026
### II. Subject Device
Name of Device: Hair Removal Device
Model(s): R3C16-G Pro, R3C16-V Pro
Common or Usual Name: Light Based Over-The-Counter Hair Removal
Regulation Name: Laser surgical instrument for use in general and plastic surgery and
in dermatology
Regulatory Class: II
Product Code: OHT
Regulation Number: 21 CFR 878.4810
1 / 8
{5}
Shenzhen Jianchao Intelligent Technology Co., Ltd.
510(k)s – K261705\_510(k) Summary
File No.:July 2, 2026
### III. Predicate Device
#### Predicate device:
| Predicate Device | Manufacturer | 510(k) Number | Product code |
| --- | --- | --- | --- |
| Hair Removal Device (R3C16-P, R3C16-W, R3C16-G, R3C16-P Pro, R3C16-W Pro, R3C16-G Pro, R3505-W, R3505-B, R3505-W Pro, R3505-B Pro) | Shenzhen Jianchao Intelligent Technology Co.,Ltd. | K242710 | OHT |
| IPL Hair Removal Device(JM-843, JM-865, JM-638) | Shenzhen Junmei Technology Co., Ltd. | K241881 | OHT |
### IV. Device Description
Hair Removal Device (Model: R3C16-G Pro, R3C16-V Pro), is an over-the-counter, home-use and personal device for hair reduction based on Intense Pulsed Light (IPL). It works below the skin’s surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.
Hair Removal Device includes two models with identical intended use, performance and operation, consisting of IPL main unit and power adapter, which the main unit is mainly composed of lamp tube, filter, display screen, buttons, thermoelectric cooler, fan, and DC socket.
The device is only powered by the external power adapter. The device is fitted with a skin sensor to detect appropriate skin contact, if the treatment window of the device is not in full contact with the skin, the device cannot emit light pulses, and the IPL emission activation is by manual finger switch or auto light emission. All models of the Hair Removal Device have a cooling function for a better hair removal experience. The difference of design between models are listed below:
2 / 8
{6}
Shenzhen Jianchao Intelligent Technology Co., Ltd.
510(k)s – K261705_510(k) Summary
File No.:July 2, 2026
| Model (s) | Enclosure color | Cooling function | Product drawing |
| --- | --- | --- | --- |
| R3C16-G Pro | Golden | Sapphire cooling | |
| R3C16-V Pro | Purple | Sapphire cooling | |
## V. Indications for Use
Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.
Comparison of Indications for Use: The indications for use of the subject device is comparable to the indications for use of the predicate device.
## VI. Comparison of Technological Characteristics With the Predicate Device
The Hair Removal Device has the same indications for use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use. Therefore, the Hair Removal Device may be found substantially equivalent to its predicate devices.
Hair Removal Device is compared with the following predicate devices and reference devices in terms of intended use, design, specifications and performance:
3 / 8
{7}
Shenzhen Jianchao Intelligent Technology Co., Ltd.
510(k)s – K261705\_510(k) Summary
File No.:July 2, 2026
| Item | Subject Device | Predicated device 1 | Predicated device 2 |
| --- | --- | --- | --- |
| 510(k) number | K261705 | K242710 | K241881 |
| Trade name | Hair Removal Device | Hair Removal Device | IPL Hair Removal Device |
| Model(s) | R3C16-G Pro, R3C16-V Pro | R3C16-P, R3C16-W, R3C16-G, R3C16-P Pro, R3C16-W Pro, R3C16-G Pro, R3505-W, R3505-B, R3505-W Pro, R3505-B Pro | JM-843, JM-865, JM-638 |
| Manufacturer | Shenzhen Jianchao Intelligent Technology Co.,Ltd. | Shenzhen Jianchao Intelligent Technology Co.,Ltd. | Shenzhen Junmei Technology Co., Ltd. |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Product code | OHT | OHT | OHT |
| Device classification | Class II | Class II | Class II |
| Indication for use/Intended use | Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair | Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair | The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair |
| Prescription or OTC | OTC | OTC | OTC |
| Treatment area | Small areas such as underarm and facial hair below the chin | Small areas such as underarm and facial hair below the chin line, large | Information not available in 510(K) |
4 / 8
{8}
Shenzhen Jianchao Intelligent Technology Co., Ltd.
510(k)s – K261705\_510(k) Summary
File No.:July 2, 2026
| | line, large areas such as arms and legs | areas such as arms and legs | summary |
| --- | --- | --- | --- |
| Applicable skin tone | Fitzpatrick Skin Types I-V | Fitzpatrick Skin Types I-V | Fitzpatrick Skin Types I-V |
| Source energy | An external power supply | An external power supply | An external power adaptor |
| Power supply | 100~240V, 50/60Hz | 100~240V, 50/60Hz | 100~240V, 50/60Hz |
| Dimension(mm) | 184.4*65.4*43.9mm | R3C16 series: 185*70*40mm R3505 series: 198*117*51mm | JM-843,JM-865:194*91*55 mm JM-638:132*80*52 mm |
| Sterilization | Not required | Not required | Not required |
| Light source | Intense Pulse Light | Intense Pulse Light | Intense Pulse Light |
| Energy medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc lamp |
| Wavelength range | 530~1200nm | 530~1200nm | JM-843,JM-865:530~1200nm JM-638:500~1200nm |
| Energy density | **R3C16-G Pro, R3C16-V Pro:** 1.4J/cm²~3.85J/cm² | **R3C16-P, R3C16-W, R3C16-G:** 1.56J/cm²~4.69J/cm² **R3C16-P Pro, R3C16-W Pro, R3C16-G Pro:** 1.39J/cm²~4.17J/cm² **R3505-W, R3505-B:** | JM-843,JM-865:1.94~5.45 J/cm² JM-638:1.20~4.00J/cm² |
5 / 8
{9}
Shenzhen Jianchao Intelligent Technology Co., Ltd.
510(k)s – K261705\_510(k) Summary
File No.:July 2, 2026
| | | 3.57J/cm²~7.5J/cm² **R3505-W Pro, R3505-B Pro:** 3.03J/cm²~6.36J/cm² | |
| --- | --- | --- | --- |
| Spot size | 3.9±0.25cm² | **R3C16-P, R3C16-W, R3C16-G:** 3.2±0.25cm² **R3C16-P Pro, R3C16-W Pro, R3C16-G Pro:** 3.6±0.25cm² **R3505-W, R3505-B:** 2.8±0.25cm² **R3505-W Pro, R3505-B Pro:** 3.3±0.25cm² | JM-843,JM-865:3.3cm²±0.25cm² JM-638:3.0cm²±0.25cm² |
| Pulse duration | 6-11ms | R3C16 series: 6-11ms R3505 series: 11-12ms | JM-843,JM-865:10.2±1ms JM-638:9.0±1ms |
| Pulsing control | Finger switch | Finger switch | Finger switch |
| Delivery device | Direct illumination to tissue | Direct illumination to tissue | Information not available in 510(K) summary |
6 / 8
{10}
Shenzhen Jianchao Intelligent Technology Co., Ltd.
510(k)s – K261705\_510(k) Summary
File No.:July 2, 2026
| Output intensity level | 5 levels | 9 levels | 1-5 levels |
| --- | --- | --- | --- |
| Software/Firmware /Microprocessor control? | Yes | Yes | Yes |
| Electrical safety | IEC 60601-1 | IEC 60601-1 | EC 60601-1 |
| | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 |
| | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 |
| | IEC 60601-2-83 | IEC 60601-2-83 | IEC 60601-2-83 |
| Eye safety | IEC 62471 | IEC 62471 | IEC 62471 |
| Biocompatibility | ISO 10993-5 | ISO 10993-5 | ISO 10993-5 |
| | ISO 10993-10 | ISO 10993-10 | ISO 10993-10 |
| | ISO 10993-23 | ISO 10993-23 | ISO 10993-23 |
## VII. Non-Clinical Testing
Non-clinical testing have been conducted to verify that the subject device meets all design specifications which supports the conclusion that it's substantially equivalent to the predicate device. The testing results demonstrate that the subject device complies with the following standards:
- ➤ ISO 10993-5: 2009, Biological evaluation of medical devices –Part 5: Tests for in vitro cytotoxicity
7 / 8
{11}
Shenzhen Jianchao Intelligent Technology Co., Ltd.
510(k)s – K261705_510(k) Summary
File No.:July 2, 2026
➤ ISO 10993-10: 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
➤ ISO 10993-23: 2021, Biological evaluation of medical devices - Part 23: Tests for irritation
➤ IEC 60601-1: 2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
➤ IEC 60601-1-2: 2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
➤ IEC 60601-1-11: 2020, Medical Electrical Equipment –Part 1-11: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
➤ IEC 60601-2-83: 2022, Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
➤ IEC 62471: 2006, Photobiological safety of lamps and lamp systems
### **VIII. Clinical Testing**
Not applicable.
### **IX. Conclusions**
Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.
8 / 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.