Hemolase Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200nm have received regulatory clearance. Hemolase Fibers are intended for use with any cleared surgical laser with an SMA 905 connector.
Device Story
Hemolase Fiber is a sterile, single-use radial-type surgical laser fiber; distal end is flat with a conical glass tube. Device connects to commercially available diode laser systems (500nm-2200nm) via SMA 905 connector; maximum output power 30W. Used in surgical settings for soft tissue procedures including incision, excision, vaporization, ablation, hemostasis, and coagulation. Operated by surgeons in contact or non-contact modes. Fiber delivers laser energy to target tissue; clinical benefit derived from precise tissue interaction enabled by laser system compatibility. Device is EO gas sterilized.
Clinical Evidence
No clinical studies performed. Evidence consists of bench testing (product appearance, dimensions, transmission power/efficiency, tensile strength, bending radius) and biocompatibility assessment based on identity to predicate device materials.
Technological Characteristics
Radial-type surgical laser fiber; silica glass core; SMA 905 connector. Dimensions: 0.40mm/0.60mm core, 0.9mm/1.3mm outer diameter. Wavelength range 500nm-2200nm; max power 30W. EO gas sterilized. Biocompatibility per ISO 10993.
Indications for Use
Indicated for use in all surgical specialties requiring laser-based incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissues in adults. Compatible with laser systems (500nm-2200nm) using SMA 905 connectors.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 18, 2026
Micro-Energy Medical Technology Co., Ltd.
Danning Zhu
Registered Engineer
Section B4301, B4701 And B4901, # 13, Baolan Rd.
Kengzi St., Pingshan District
Shenzhen, Guangdong 518118
China
Re: K261701
Trade/Device Name: Hemolase Fiber (Hemo-R-0.40-2.5/Hemo-R-0.60-2.5/Hemo-R-0.40-1.8/Hemo-R-0.60-1.8)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: May 22, 2026
Received: May 22, 2026
Dear Danning Zhu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}
K261701 - Danning Zhu
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K261701 - Danning Zhu
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2026.06.18
16:54:23 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K261701
?
Please provide the device trade name(s).
?
Hemolase Fiber (Hemo-R-0.40-2.5/Hemo-R-0.60-2.5/Hemo-R-0.40-1.8/Hemo-R-0.60-1.8)
Please provide your Indications for Use below.
?
Hemolase Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200nm have received regulatory clearance. Hemolase Fibers are intended for use with any cleared surgical laser with an SMA 905 connector.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
Please select the age group(s) for which the device(s) is to be used.
☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☐ Children (2 years old to < 12 years old)
☐ Adolescents (12 years old to < 22 years old)
☐ Adults (22 years old and greater)
?
{4}
# 510(k) Summary:K261701
Date Prepared:May.22nd,2026
This summary of 510(k) information is submitted as required by requirements of 21 CFR §807.92.
## 1. Submitter's Information
Name of Sponsor: Micro-Energy Medical Technology Co., Ltd.
Address: Section B4301,B4701 and B4901 No.13, Baolan Road,
Kengzi Street, Pingshan District,Shenzhen, China
Contact Name: Elly Xu
Telephone No.: +86-755-28336782
Fax No.: +86-755-28336782
Email Address.: sha.x@micro-energy.com
## 2. Correspondent's Information
Company Name: Micro-Energy Medical Technology Co., Ltd.
Correspondent Name: Danning Zhu
Telephone No.: +86-755-28339524
Email Address: danning.z@micro-energy.com
## 3. Trade Name, Common Name, Classification
| Trade Name: | Hemolase Fiber |
| --- | --- |
| Common Name: | Sterile Single-use Radial Fiber |
| Model: | Hemo-R-0.40-2.5/Hemo-R-0.60-2.5/Hemo-R-0.40-1.8/Hemo-R-0.60-1.8 |
| Classification Name: | Powered Laser Surgical Instrument |
| Product Code: | GEX |
| Classification Panel: | General & Plastic Surgery |
| Device Class: | II |
| Classification Regulation: | 21 CFR 878.4810 |
1
{5}
# 4. Identification of Predicate Device(s)
The identified predicates within this submission are as follows:
| Predicate Device : | 510(k) number:K252745 Trade name:Halo Sterile Single-use Radial Fiber Product code:GEX |
| --- | --- |
# 5. Description of the Device
The Hemolase Fibers (Model:Hemo-R-0.60-2.5/Hemo-R-0.40-2.5/Hemo-R-0.40-1.8/Hemo-R-0.60-1.8), as surgical laser fibers, are suitable for use in combination with commercially available diode laser systems with a wavelength ranging from 500nm to 2200nm and equipped with SMA905 connectors. The output power should not exceed 30 W. This device is single use only and sterilized by EO gas.The Hemolase Fibers are intended for general surgical operations such as incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissues in either contact or non - contact mode. The Fiber is radial-typed fiber.The fiber has a flat distal end, and its distal glass tube is conical .
# 6. Intended Use/Indication for Use
Hemolase Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200nm have received regulatory clearance. Hemolase Fibers are intended for use with any cleared surgical laser with an SMA 905 connector.
# 7. Technological Characteristics
The Hemolase Fibers (Model:Hemo-R-0.40-2.5/Hemo-R-0.60-2.5/Hemo-R-0.40-1.8/Hemo-R-0.60-1.8) uses the same technology that is utilized in the predicate device. A comparison of technological characteristics is provided in the following table:
2
{6}
**SUBSTANTIAL EQUIVALENCE COMPARISON TABLE**
| Item | Subject device | Predicate device (K252745) | Comparison |
| --- | --- | --- | --- |
| Product Code | GEX | GEX | Same |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Class | II | II | Same |
| Device Name | Hemolase Fiber | Halo Sterile Single-use Radial Fiber | NA |
| Device Model | Hemo-R-0.40-2.5/Hemo-R-0.60-2.5/Hemo-R-0.40-1.8/Hemo-R-0.60-1.8 | Halo-R-0.40-2.5/Halo-R-0.60-2.5/Novo-R-0.40-1.8/Novo-R-0.60-1.8 | NA |
| Manufacturer | Micro-Energy Medical Technology Co., Ltd. | Micro-Energy Medical Technology Co., Ltd. | NA |
| Intended Use/Indication for Use | Hemolase Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200nm have received regulatory clearance. Hemolase Fibers are intended for use with any cleared surgical laser with an SMA 905 connector | Halo Sterile Single-use Radial Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200nm have received regulatory clearance. Halo Sterile Single-use Radial Fibers are intended for use with any cleared surgical laser with an SMA 905 connector. | Same Only the trade names are different, while their intended use and indications for use are exactly the same.. |
3
{7}
| Target Population | Adult | Adult | Same |
| --- | --- | --- | --- |
| Wavelength Range | 500nm-2200mm | 500nm-2200mm | Same |
| Fiber Core Diameter | 0.40mm 0.60mm | 0.40mm 0.60mm | Same |
| Outer Diameter | 0.9mm 1.3mm | 0.9mm 1.3mm | Same |
| Maximum Output Power | 1-30Watts | 1-30Watts | Same |
| Fiber Core Material | Silica Glass | Silica Glass | Same |
| Connectors | SMA 905 | SMA 905 | Same |
| Single Use | Yes | Yes | Same |
| Sterility | EO gas sterilized | EO gas sterilized | Same |
| Biocompatibility | ISO10993 | ISO10993 | Same |
4
{8}
As seen in the comparison tables, The Hemolase Fibers that are the subject of this premarket notification use same technology as that of the Halo Sterile Single-use Radial Fiber of the K252745 510(k). And performance testing demonstrates that the Hemolase Fibers can be used safely and effectively for the proposed indications for use.. The Hemolase Fibers are considered to be substantially equivalent to the predicate K252745.
5
{9}
# 8. Performance testing
To evaluate the minor structural changes to the distal tip of the fiber transmission body and the distal glass tube of the four new models of Hemolase Fibers covered by this submission, design verification was performed to support the safe and effective use of the fibers.
The following tests were conducted on the four new models:
-Product Appearance
-Product Dimensions
-Maximum Transmission Power
-Transmission Efficiency
-Tensile Strength
-Minimum Bending Working Radius of the Fiber
From a biocompatibility standpoint, the Hemolase Fibers covered by this submission are identical to the predicate device Halo Sterile Single-use Radial Fibers (cleared under K252745), with the same overall product structure and materials.
A biological evaluation was completed for the subject fibers, and it was concluded that there are no biocompatibility risks associated with the product.
# 9. Brief discussion of clinical tests
No clinical studies were considered necessary and performed.
# 10. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Micro-Energy Medical Technology Co., Ltd concludes that:
The indications for use of Hemolase Fibers (Model:Hemo-R-0.40-2.5/Hemo-R-0.60-2.5/Hemo-R-0.40-1.8/Hemo-R-0.60-1.8) are the same as those of the predicate device(K252745). The technological characteristics are the same or similar to those of the predicate device(K252745).
- Hemolase Fibers (Model:Hemo-R-0.40-2.5/Hemo-R-0.60-2.5/Hemo-R-0.40-1.8/Hemo-R-0.60-1.8) is substantially equivalent to the predicate device(K252745).
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.