K261610 · Instylla, Inc. · KRA · Jun 11, 2026 · Cardiovascular
Device Facts
Record ID
K261610
Device Name
Verge™ Microcatheter
Applicant
Instylla, Inc.
Product Code
KRA · Cardiovascular
Decision Date
Jun 11, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1210
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Verge™ Microcatheter is intended for use in small vessel or subselective anatomy for peripheral vascular diagnostic and interventional procedures. The Verge™ Microcatheter can be used for the infusion of diagnostic, embolic, or therapeutic materials into vessels.
Device Story
Single lumen, multipurpose microcatheter for peripheral vascular access; tracks coaxially over steerable guidewire (≤0.014" for 1.7 Fr; ≤0.010" for 1.2 Fr) to target site; used for infusion of diagnostic, embolic, or therapeutic materials. Operated by clinicians in clinical settings. Features radiopaque distal marker for fluoroscopic visualization; includes Tuohy-Borst adapter, extensions, and check valve. Gradual internal hub-to-lumen transition improves guidewire insertion. Benefits include precise delivery of therapeutic agents to small or subselective vessels.
Clinical Evidence
Bench testing only. Testing performed per ISO 10555-1 and FDA guidance for peripheral specialty catheters. Evaluated: visual/dimensional inspection, trackability, kink resistance, pushability, tip stability, fluid injection, static burst pressure, leakage (fluid/air), tensile strength, and compatibility with embolic agents/accessories. Biocompatibility testing (ISO 10993-1) included cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemolysis, complement activation, PTT, and in vivo thromboresistance. Sterilization validated per ISO 11135; packaging validated per ISO 11607-1 and ISTA 3A.
Technological Characteristics
Single lumen catheter; 1.2 Fr (0.40 mm) or 1.7 Fr (0.56 mm) OD. Materials: Polycarbonate hub, PEBAX 72D/63D shaft, Stainless Steel, PTFE, Platinum/Iridium marker band, PEBAX 63D distal tip. Manual operation; steerable guidewire compatible. Sterilized via Ethylene Oxide (EO) to SAL 10^-6. Includes Tuohy-Borst adapter, extensions, and duckbill check valve.
Indications for Use
Indicated for use in small vessel or subselective peripheral vascular diagnostic and interventional procedures, including infusion of diagnostic, embolic, or therapeutic materials. Contraindicated for coronary or neurovasculature use. Safety and effectiveness in pediatric patients not established.
Regulatory Classification
Identification
A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 11, 2026
Instylla, Inc.
Easwar Prabakaran
Regulatory Specialist II
201 Burlington Rd.
North Bldg.
Bedford, Massachusetts 01730
Re: K261610
Trade/Device Name: Verge™ Microcatheter
Regulation Number: 21 CFR 870.1210
Regulation Name: Continuous flush catheter
Regulatory Class: Class II
Product Code: KRA
Dated: May 14, 2026
Received: May 14, 2026
Dear Easwar Prabakaran:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261610 - Easwar Prabakaran
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261610 - Easwar Prabakaran
Page 3
Sincerely,
GREGORY W.
O'CONNELL -S
Digitally signed by GREGORY W.
O'CONNELL -S
Date: 2026.06.11 12:59:26 -04'00'
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K261610
Device Name
Verge™ Microcatheter
Indications for Use (Describe)
The Verge™ Microcatheter is intended for use in small vessel or subselective anatomy for peripheral vascular diagnostic and interventional procedures. The Verge™ Microcatheter can be used for the infusion of diagnostic, embolic, or therapeutic materials into vessels.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Special 510(k) - K261610
Verge™ Microcatheter
Instylla, Inc.


# 510(k) Summary
Submitter Information:
Instylla, Inc.
201 Burlington Rd
North Building
Bedford, MA 01730
Contact Person:
Easwar Prabakaran
Regulatory Specialist II
Phone: 813-471-7585
E-mail: easwarp@instylla.com
Date Prepared:
June 11, 2026
Subject Device:
| Proprietary Name: | Verge™ Microcatheter |
| --- | --- |
| Common Name: | Catheter, Continuous Flush |
| Classification Name: | Continuous Flush Catheter |
| Classification Regulation: | 21 CFR 870.1210 |
| Product Code: | KRA |
| Device Classification: | Class II |
| Classification Panel: | Cardiovascular |
Primary Predicate Device:
| Proprietary Name: | Instylla 1.2 Microcatheter |
| --- | --- |
| Manufacturer: | Instylla, Inc. |
| 510(k) Number: | K210808 |
Secondary Predicate Device:
| Proprietary Name: | Instylla Microcatheter |
| --- | --- |
| Manufacturer: | Instylla, Inc. |
| 510(k) Number: | K200744 |
Device Description:
The Verge™ Microcatheter device is a single lumen, multipurpose catheter intended for use in the peripheral vasculature. The basic operating principle is to advance the microcatheter through a guiding
{5}
Special 510(k) - K261610
Verge™ Microcatheter
Instylla, Inc.
catheter and track coaxially over a steerable guidewire to access the treatment site. The microcatheter lumen can accommodate steerable guidewires that are $\leq 0.014''$ (0.36 mm) in diameter for the 1.7 Fr size and $\leq 0.010''$ (0.25 mm) for the 1.2 Fr size. Once the target region has been accessed, the microcatheter can be used to deliver diagnostic, embolic, or therapeutic materials into vessels.
The Verge™ Microcatheter has a 1.2 Fr (0.40 mm) or 1.7 Fr (0.56 mm) outside diameter (OD) with constant flexibility along its length. The inner diameter (ID) of the microcatheter is $0.012''$ (0.30 mm) or $0.016''$ (0.41 mm) along its length. The proximal end of the microcatheter incorporates a standard luer hub to enable the attachment of accessories and a strain relief that allows for flexibility and securement inside a Tuohy-Borst adapter for maintaining position inside a guiding catheter as needed. The Verge™ Microcatheter has a radiopaque marker at the distal tip to aid in fluoroscopic visualization. A 4Fr Tuohy-Borst with a side-port adapter, a short catheter extension, a long catheter extension adapter, and a duckbill check valve are also included. The Verge™ Microcatheter System is available in 122 cm, 142 cm, and 162 cm usable lengths.
#### Indications for Use:
The Verge™ Microcatheter is intended for use in small vessel or sub selective anatomy for peripheral vascular diagnostic and interventional procedures. The Verge™ Microcatheter can be used for the infusion of diagnostic, embolic, or therapeutic materials into vessels.
#### Comparison of Technological Characteristics to the Predicate Device:
The Verge™ Microcatheter is substantially equivalent in intended use and fundamental technological characteristics to the legally marketed predicate devices. The below table summarizes the similarities in design and configuration of the Verge™ Microcatheter compared with the predicate devices.
| Attribute | Subject Device: Verge™ Microcatheter | Primary Predicate Device: Instylla 1.2 Microcatheter (K210808) | Secondary Predicate Device: Instylla Microcatheter (K200744) | Substantial Equivalence |
| --- | --- | --- | --- | --- |
| Indications for Use | The Verge™ Microcatheter is intended for use in small vessel or sub selective anatomy for peripheral vascular diagnostic and interventional procedures. The Verge™ Microcatheter | The Instylla Microcatheter 1.2 is intended for use in small vessel or super selective anatomy for peripheral diagnostic and interventional procedures. The Instylla | The Instylla Microcatheter is intended for use in small vessel or super selective anatomy for peripheral diagnostic and interventional procedures. The Instylla Microcatheter can be used for | Similar |
{6}
Special 510(k) - K261610
Verge™ Microcatheter
Instylla, Inc.
| Attribute | Subject Device: Verge™ Microcatheter | Primary Predicate Device: Instylla 1.2 Microcatheter (K210808) | Secondary Predicate Device: Instylla Microcatheter (K200744) | Substantial Equivalence |
| --- | --- | --- | --- | --- |
| | can be used for the infusion of diagnostic, embolic, or therapeutic materials into vessels. | Microcatheter 1.2 can be used for the infusion of diagnostic, embolic, or therapeutic materials into vessels. | the infusion of diagnostic, embolic, or therapeutic materials into vessels. | |
| Basic Design/ Components | Microcatheter, 4Fr Tuohy Borst with Side Port-Adaptor, Short Extension Adapter, Long Extension Adapter and Duckbill Check-Valve Connector | Microcatheter, 4Fr Tuohy Borst with Side Port-Adaptor, Short Extension Adapter, Long Extension Adapter and Duckbill Check-Valve Connector | Microcatheter, 4Fr Tuohy Borst with Side Port-Adaptor, Short Extension Adapter, Long Extension Adapter and Duckbill Check-Valve Connector | Same |
| Microcatheter Materials | Hub – Polycarbonate Shaft –PEBAX 72D/63D, Stainless Steel, PTFE, Platinum/Iridium Marker Band Distal Tip– PEBAX 63D | Hub – Polycarbonate Shaft –PEBAX 72D/63D, Stainless Steel, PTFE, Platinum/Iridium Marker Band Distal Tip– PEBAX 72D/40D | Hub – Polycarbonate Shaft –PEBAX 72D/63D, Stainless Steel, PTFE, Platinum/Iridium Marker Band Distal Tip– PEBAX 72D/40D | The subject and predicate device’s materials remain the same with the exception of the distal tip material which was modified to improve the Microcatheter manufacturability. |
| Principle of Operation | Manually tracked over a steerable guidewire to access vasculature. | Manually tracked over a steerable guidewire to access vasculature. | Manually tracked over a steerable guidewire to access vasculature. | Same |
{7}
Special 510(k) -K261610
Verge™ Microcatheter
Instylla, Inc.
| Attribute | Subject Device: Verge™ Microcatheter | Primary Predicate Device: Instylla 1.2 Microcatheter (K210808) | Secondary Predicate Device: Instylla Microcatheter (K200744) | Substantial Equivalence |
| --- | --- | --- | --- | --- |
| Contraindications | The Verge™ Microcatheter is not intended to be used in coronary or neurovasculature. The safety and effectiveness of the Verge™ Microcatheter in pediatric patients has not been established. | The Verge™ Microcatheter is not intended to be used in coronary or neurovasculature. | The Verge™ Microcatheter is not intended to be used in coronary or neurovasculature. | The subject device includes a pediatric contraindication. |
| Inner Diameter (ID) | 1.7 Fr: 0.016 in (0.41 mm) 1.2 Fr: 0.012 in (0.30 mm) | 0.012 in (0.30 mm) | 0.016 in (0.41 mm) | Same |
| Catheter Shaft Outer Diameter (OD) | 1.7 Fr (0.56 mm) 1.2 Fr (0.40 mm) | 1.2 Fr (0.40 mm) | 1.7 Fr (0.56 mm) | Same |
| Catheter Lengths | 122 cm, 142 cm, 162 cm | 122 cm, 142 cm, 162 cm | 122 cm, 142 cm, 162 cm | Same |
| Guidewire Compatibility (Size) | 1.7 Fr: 0.014 in (0.36 mm) 1.2 Fr: 0.010 in (0.25 mm) | 0.010 in (0.25 mm) | 0.014 in (0.36 mm) | Same |
| Radiopaque Marker | Yes | Yes | Yes | Same |
| Packaging | Tyvek Pouches | Tyvek Pouches | Tyvek Pouches | Same |
| Sterilization Method | Ethylene Oxide (EO) to a SAL of $$10^{-6}$$ | Ethylene Oxide (EO) to a SAL of $$10^{-6}$$ | Ethylene Oxide (EO) to a SAL of $$10^{-6}$$ | Same |
| Shelf Life | 6 months | 6 months | 6 months | Same |
| Hub Proximal End – Transition | Smooth transition from the hub into | Small “step” in the hub. Tapers | Small “step” in the hub. Taper | The more gradual internal transition |
{8}
Special 510(k) -K261610
Verge™ Microcatheter
Instylla, Inc.
![Instylla logo]()
| Attribute | Subject Device: Verge™ Microcatheter | Primary Predicate Device: Instylla 1.2 Microcatheter (K210808) | Secondary Predicate Device: Instylla Microcatheter (K200744) | Substantial Equivalence |
| --- | --- | --- | --- | --- |
| from Hub to Microcatheter Lumen | the catheter lumen | down to the ID of the lumen. | down to the ID of the lumen. | from hub to catheter shaft ID improves guidewire insertion |
The subject and primary predicate device differ from one another in the packaging, labeling, distal tip material, and design of hub to microcatheter transition.
The indications for use, principle of operations, and technological characteristics are similar between the subject and predicate devices.
#### Performance Data
Performance testing of the sterilized Verge™ Microcatheter included bench testing and functional testing to verify specifications fundamental to the design of the device in accordance with the FDA Guidance for Industry *Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters – Premarket Notification (510(k)) Submissions* and ISO 10555-1 *Intravascular catheters – Sterile and single-use catheters – Part 1: General Requirements*. Testing included the following:
- • Visual Inspection of Components
- • Dimensional Verification of Components
- • Trackability
- • Kink Resistant
- • Pushability
- • Tip Stability
- • Injection of Fluids
- • Static Burst Pressure
- • Freedom from Leakage – Fluids
- • Freedom from Leakage – Air
- • Catheter Tensile Strength
- • Microcatheter Compatibility with Microcatheters, Embolic Agents
- • Accessory Compatibility and Functionality
- • Torque Strength
The Verge™ Microcatheter met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing.
#### Biocompatibility Testing
A biocompatibility evaluation was conducted in accordance with the FDA Guidance for Industry, *Use of International Standard ISO 10993-1, “Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process,”* consistent with an external communicating circulating blood
{9}
Special 510(k) - K261610
Verge™ Microcatheter
Instylla, Inc.
instylla
device with limited body contact. The following biocompatibility tests were successfully completed on the Verge™ Microcatheter:
- • Cytotoxicity
- • Sensitization
- • Irritation
- • Acute Systemic Toxicity
- • Material Mediated Pyrogenicity
- • Hemolysis
- • Complement Activation Assay
- • Partial Thromboplastin Time Assay
- • *In Vivo* Thromboresistance Study
### Sterility
The Verge™ Microcatheter is sterilized via an Ethylene Oxide (EO) process to a Sterility Assurance Level (SAL) of 10$^{-6}$. The sterilization process was validated per ISO 11135 *Sterilization of health-care products – Ethylene oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices*. The Verge™ Microcatheter was evaluated and adopted into the validated process per AAMI TIR28 *Product adoption and process equivalence for ethylene oxide sterilization*. The EO and ECH levels were determined to be acceptable in accordance with ISO 10993-7 *Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals*.
A bacterial endotoxin test (BET), also known as the Limulus amebocyte lysate (LAL) test, was also validated to establish that the Verge™ Microcatheter endotoxin level will be <20 endotoxin units (EU)/device.
### Packaging
Packaging testing was conducted on sterilized Verge™ Microcatheter to evaluate the effectiveness of the packaging configuration to maintain the sterile barrier in accordance with ISO 11607-1 *Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems*. The subject device and packaging were subjected to simulated distribution challenge conditions in accordance with International Safe Transit Association (ISTA) 3A and evaluated for performance of the sterile barrier and product performance.
### Shelf Life
The Verge™ Microcatheter has a 6 month shelf life. Shelf life studies have been conducted to demonstrate that the device maintains its performance and the packaging will maintain its sterile barrier over the entirety of the intended shelf life.
### Conclusion
Instylla has demonstrated that the Verge™ Microcatheter is substantially equivalent in fundamental design, function, device materials, packaging, sterilization, operating principle, intended use/indication for use and fundamental technology as the legally marketed predicate devices, the Instylla 1.2 Microcatheter
{10}
Special 510(k) - K261610
Verge™ Microcatheter
Instylla, Inc.
which was cleared under 510(k) Premarket Notification K210808 on April 15, 2021, and the Instylla Microcatheter which was cleared under 510(k) Premarket Notification K200744 on April 21, 2020.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.