Non-randomized, on-body ambulatory performance study
The sponsor conducted an on-body performance study in an ambulatory setting to validate the wear-life and ECG signal quality of the UbiqVue™ 2AYe Wearable Biosensor under normal use conditions.
The UbiqVue™ 2AYe Cardiac System is intended to capture, record for analysis and report both symptomatic and asymptomatic cardiac events, along with continuous electrocardiogram (ECG) information, for long-term monitoring. Following patient monitoring using the UbiqVue™ 2AYe Wearable Biosensor, a comprehensive final report is generated from the entire ECG recording. Optionally, intended users may choose to generate an interim report from periodically or intermittently transferred ECG recording wirelessly from the UbiqVue™ 2AYe Wearable Biosensor. Final & interim reporting is based on the beat-to-beat information extracted from the ECG recording.The UbiqVue™ 2AYe Cardiac System is intended for non-critical, adult population, who are 18 years of age or older, at home and in healthcare settings, who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety. The reports are provided for review by the intended user to render a diagnosis based on their clinical judgment and experience.
Device Story
System captures, records, and reports continuous ECG data for long-term monitoring. Input: two-channel ECG signals from disposable wearable biosensor (8-day wear-life). Biosensor buffers data locally if wireless connection to relay (mobile app or bridge) is unavailable. Data transmitted via BLE (5.1) or Wi-Fi to cloud-based Central Server. Web Server generates interim/final raw ECG files (.EDF) for export to third-party analysis software. Web Portal allows clinical staff (physicians, nurses, technicians) to manage procedures, view status, and review reports. Output: ECG reports for clinical diagnosis. Benefits: enables remote cardiac monitoring, symptom correlation, and flexible reporting workflows.
Clinical Evidence
Bench testing only. Includes electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), wireless coexistence (ANSI/IEEE C63.27), and usability (IEC 60601-1-6). Wear-life performance validated via non-randomized, on-body study (192 hours). Biocompatibility testing (cytotoxicity, irritation, sensitization, intracutaneous reactivity) performed per ISO 10993-1.
Technological Characteristics
Two-channel ECG biosensor; 16-bit ADC; 244.14 Hz sampling rate. Materials: hydrogel electrodes, long-term wear acrylic adhesive. Connectivity: BLE 5.2, WLAN (802.11b). Power: LiMnO2 battery. Dimensions: 116.5 x 91.5 x 15.5 mm. Sterilization: not specified. Software: cloud-based server/portal architecture; WPA2-PSK/CCMP AES-CCM security. Type CF applied part; defibrillation-proof; HF surgical equipment compatible.
Indications for Use
Indicated for non-critical adults (18+ years) at home or in healthcare settings, who are asymptomatic or experiencing transient symptoms (palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety) requiring long-term ECG monitoring.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 9, 2026
Lifesignals, Inc.
Saravanan Balasubramanian
Vice President - Global, Medical Technology & Regulatory Operations
426 S Hillview Dr.
Milpitas, California 95035
Re: K261569
Trade/Device Name: UbiqVue™ 2AYe Cardiac System (UX2253)
Regulation Number: 21 CFR 870.2910
Regulation Name: Radiofrequency Physiological Signal Transmitter And Receiver
Regulatory Class: Class II
Product Code: DRG, QYX, MWJ, DXH
Dated: May 11, 2026
Received: May 12, 2026
Dear Saravanan Balasubramanian:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261569 - Saravanan Balasubramanian
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261569 - Saravanan Balasubramanian
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261569 | ? |
| Please provide the device trade name(s). | | ? |
| UbiqVue™ 2AYe Cardiac System (UX2253) | | |
| Please provide your Indications for Use below. | | ? |
| The UbiqVue™ 2AYe Cardiac System is intended to capture, record for analysis and report both symptomatic and asymptomatic cardiac events, along with continuous electrocardiogram (ECG) information, for long-term monitoring. Following patient monitoring using the UbiqVue™ 2AYe Wearable Biosensor, a comprehensive final report is generated from the entire ECG recording. Optionally, intended users may choose to generate an interim report from periodically or intermittently transferred ECG recording wirelessly from the UbiqVue™ 2AYe Wearable Biosensor. Final & interim reporting is based on the beat-to-beat information extracted from the ECG recording.The UbiqVue™ 2AYe Cardiac System is intended for non-critical, adult population, who are 18 years of age or older, at home and in healthcare settings, who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety. The reports are provided for review by the intended user to render a diagnosis based on their clinical judgment and experience. | | |
| Please select the types of uses (select one or both, as applicable). | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | Neonates/Newborns (Birth to < 29 days old)Infants (29 days old to < 2 years old)Children (2 years old to < 12 years old)Adolescents (12 years old to < 22 years old)Adults (22 years old and greater) | ? |
{4}
[Non-Text]
LifeSignals
K261569 510(k) Summary
# 510(k) Summary for
# UbiqVue™ 2AYe Cardiac System
1 Background : LifeSignals, Inc.
Address : 426 S Hillview Drive
Milpitas, CA 95035.
USA
Contact : Saravanan Balasubramanian
Vice President – Global, Medical Technology & Regulatory Operations
Email: saravanan@lifesignals.com
Tel: 510.770.6412 Ext. 4
2 Date prepared : May 11, 2026
3 Device
Trade Name : UbiqVue™
Model Name : 2AYe Cardiac System
Common Name : Transmitters and Receivers, Physiological Signal, Radio Frequency
4 Classification Product code
| | Regulation Classification | Product Code | Description |
| --- | --- | --- | --- |
| Cardiovascular | 21 CFR 870.2910 Class II | DRG | Transmitters and Receivers, Physiological Signal, Radio frequency |
5 Subsequent Product code
| | Regulation Classification | Product Code | Description |
| --- | --- | --- | --- |
| Cardiovascular | 21 CFR 870.2920 Class II | DXH | Transmitter and Receivers, Electrocardiograph, Telephone |
| Cardiovascular | 21 CFR 870.2800 Class II | MWJ | Medical magnetic tape recorder |
| Cardiovascular | 21 CFR 870.1025 Class II | QYX | Outpatient cardiac telemetry |
6 Predicate & Reference Devices
Primary Predicate Device:
Rev C
UbiqVue 2AYe Cardiac System
Page 1 of 15
{5}
[LOGO]
LifeSignals
K261569 510(k) Summary
| 510(k) Number | : K200690 |
| --- | --- |
| Model | : LifeSignals ECG Remote Monitoring Patch Platform |
| Manufacturer | : LifeSignals, Inc. USA |
Secondary Predicate Device:
| 510(k) Number | : K163512 |
| --- | --- |
| Model | : Zio AT/QX ECG Monitoring System |
| Manufacturer | : iRhythm Technologies, Inc. USA |
Reference Device (for Biosensor Technology & Software Architecture):
| 510(k) Number | : K242018 |
| --- | --- |
| Model | : UbiqVue™ 2A Multiparameter System |
| Manufacturer | : LifeSignals, Inc. USA |
### 7 Device description
UbiqVue™ 2AYe Cardiac System consist of below components:
- UbiqVue™ 2AYe Wearable Biosensor
- UbiqVue™ Holter Mobile App
- UbiqVue™ Relay Bridge
- UbiqVue™ Holter Central Server
- UbiqVue™ Holter Web Server
- UbiqVue™ Holter Web Portal
UbiqVue™ 2AYe Wearable Biosensor acquires ECG signals from the body, pre-processes as two channels of ECG data which is wirelessly transmitted to UbiqVue Central Server via any UbiqVue™ Relay devices (UbiqVue™ Holter Mobile App or UbiqVue™ Relay Bridge). If the receiver device is not available or not communicating, the data shall be buffered locally in Biosensor till the wireless connection is re-established and resume communication. The UbiqVue™ 2AYe Wearable Biosensor is completely disposable and has a battery life, storage memory & intended for wear-life of 8 days. Patients can also record any symptom events by pressing the ON/Event button on the Biosensor.
UbiqVue™ 2AYe Wearable Biosensor is technologically similar & subset of UbiqVue™ 2A Wearable Biosensor used in UbiqVue™ 2A Multiparameter system (Reference device - K242018).
UbiqVue™ Holter Mobile App is intended to be used by patients to retrieve the data from Biosensor and transfer to the UbiqVue™ Holter Central Server. Registered patients can download the App in smart phone (iOS or Android) and establish connection with the Biosensor. The App will automatically communicate (sync) with UbiqVue 2AYe wearable Biosensor through BLE (5.1) for every 10 minutes (approximately) throughout the Holter procedure, retrieving the acquired ECG data from Biosensor and transferring to UbiqVue Holter Central Server. For any symptom events (button presses on the Biosensor), patients can enter description of symptoms & associated activity electronically through UbiqVue™ Holter Mobile App.
Rev C
UbiqVue 2AYe Cardiac System
Page 2 of 15
{6}
LifeSignals
K261569 510(k) Summary
**UbiqVue™ Relay Bridge Software** which is already regulatory cleared (Reference device K242018) is intended to be used by Clinical personnel as an alternate method of ECG data retrieval from the Biosensor after the Holter procedure is completed. This software shall be installed on an Open WRT based Access Point (Router) and shall wirelessly connect to one or more Biosensor at a time via Wi-Fi, retrieving the acquired ECG data and transferring to UbiqVue™ Holter Central Server.
**UbiqVue™ Holter Central Server Software** installed in a cloud server, receives, stores & provide access to ECG data retrieved from Biosensors after the Holter Procedure (or during the Holter procedure). The stored ECG data is identified with the Ubique Biosensor ID associated with the patient. But no patient specific information is stored in UbiqVue™ Holter Central Server, except the patient diary events. The UbiqVue™ Holter Central Server allows UbiqVue™ Web Server to access these stored ECG data with appropriate authentication via a suitable Application Specific Interface (API).
**UbiqVue™ Holter Web Server Software** installed in a cloud server, communicates with UbiqVue Central Server, UbiqVue Holter Web Portal, EMR (Electronic Medical Record) and stores the Biosensor ECG data by linking to appropriate patient demographic & procedure information. It also provides Biosensor, Holter Mobile App, Relay Bridge status information to UbiqVue Holter Web Portal by accessing UbiqVue Holter Center Server.
On demand (by users), UbiqVue Holter Web Server may also generate an interim raw ECG file (.EDF file format) based on the available ECG data, during the cardiac procedure. The Holter Web Server also generates the final ECG file(.EDF file format) once the cardiac procedure is completed. These raw ECG file (edf file format) generated by the Web server can be transferred (offline or online) to a compatible regulatory cleared 3rd party analysis software for doing necessary automated arrhythmia analysis (with or without manual over-read) as per the approved cardiac workflow.
**UbiqVue™ Holter Web Portal** is a browser-based application that provides user interface to various users of clinical facility (Admin, Nurse, Prescribing Physician, Interpreting Physician, ECG Technician, etc.) to manage Cardiac (Holter) workflow for their patients such as registering a Cardiac (Holter) procedure, linking the Biosensor ID to the procedure, entering patient demographic information, manage & configure Holter reports, view Holter Mobile App data retrieval status & Biosensor connection status.
If a compatible regulatory cleared 3rd party analysis software is integrated with the Web server, then web server could also receive the ECG analysis report (typically in .pdf format) generated by the software (with or without a manual over-read) for Interpreting Physician to view, review, interpret, eSign & approve the reports using Web portal. These finalized reports shall be available for prescribing physician to view, download and print through web portal.
Rev C
UbiqVue 2AYe Cardiac System
Page 3 of 15
{7}
LifeSignals
K261569 510(k) Summary
# 8 Indications for Use
- The UbiqVue™ 2AYe Cardiac System is intended to capture, record for analysis and report both symptomatic and asymptomatic cardiac events, along with continuous electrocardiogram (ECG) information, for long-term monitoring. Following patient monitoring using the UbiqVue™ Wearable Biosensor, a comprehensive final report is generated from the entire ECG recording. Optionally, intended users may choose to generate an interim report from periodically or intermittently transferred ECG recording wirelessly from the UbiqVue™ Wearable Biosensor. Final & interim reporting is based on the beat-to-beat information extracted from the ECG recording.
- The UbiqVue™ 2AYe Cardiac system is intended for non-critical, adult population, who are 18 years of age or older, at home and in healthcare settings, who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety. The reports are provided for review by the intended user to render a diagnosis based on their clinical judgment and experience.
Rev C
UbiqVue 2AYe Cardiac System
Page 4 of 15
{8}
LifeSignals
K261569 510(k) Summary
9 Substantial equivalence comparison with Predicate device(s)
| Comparison | Primary Predicate (K200690) | Secondary Predicate (K163512) | Subject Device |
| --- | --- | --- | --- |
| Manufacturer | LifeSignals, Inc. USA | iRhythm Technologies, Inc. | LifeSignals, Inc. USA |
| Product Codes | DRG, MWJ | QYX, DSH, DQK, DSI, DXH | DRG, QYX, MWJ, DXH |
| Regulation Classification (Primary) | 21 CFR 870.2910 Class II | 21 CFR 870.1025 Class II & 21 CFR 870.2800 Class II | 21 CFR 870.2910 Class II |
| Intended use / Indications for use | The LifeSignals ECG Remote Monitoring Patch Platform is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of Electrocardiography (ECG) and Heart Rate monitoring in home and healthcare settings. Data is transmitted wirelessly from LifeSignals Biosensor Patch to Remote Secure Server for storage and analysis. The LifeSignals ECG Remote Monitoring Patch Platform is intended for non-critical, adult population, who are 18 years of age or older. The LifeSignals ECG Remote Monitoring Patch Platform includes an ability | The Zio AT/QX ECG Monitoring System is intended to capture, analyze and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram (ECG) information for long-term monitoring. While continuously recording patient ECG, both patient-triggered and automatically detected arrhythmia events are transmitted to a monitoring center for reporting. After wear, a final report is generated based on beat-to-beat information from the entire ECG recording. It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of | The UbiqVue™ 2AYe Cardiac System is intended to capture, record for analysis and report both symptomatic and asymptomatic cardiac events, along with continuous electrocardiogram (ECG) information, for long-term monitoring. Following patient monitoring using the UbiqVue™ Wearable Biosensor, a comprehensive final report is generated from the entire ECG recording. Optionally, intended users may choose to generate an interim report from periodically or intermittently transferred ECG recording wirelessly from the UbiqVue™ Wearable Biosensor. Final & interim reporting is based on the |
Rev C
UbiqVue 2AYe Cardiac System
Page 5 of 15
{9}
LifeSignals
K261569 510(k) Summary
| Comparison | Primary Predicate (K200690) | Secondary Predicate (K163512) | Subject Device |
| --- | --- | --- | --- |
| | to notify healthcare professionals when Heart Rate falls outside the set limits. | breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety. The reports are provided for review by the intended user to render a diagnosis based on clinical judgment and experience. It is not intended for use on critical care patients. | beat-to-beat information extracted from the ECG recording. The UbiqVue™ 2AYe Cardiac system is intended for non-critical, adult population, who are 18 years of age or older, at home and in healthcare settings, who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety. The reports are provided for review by the intended user to render a diagnosis based on their clinical judgment and experience. |
| Intended Population | Non-critical, adult population, 18 years or older | Non-pediatric, non-critical care patient | Non-critical, adult population, 18 years or older |
| Intended Use Environment | Home & Healthcare settings | Home & Healthcare settings | Home & Healthcare settings |
| Supported Workflow | Remote Patient Monitoring, Holter, Cardiac Event Monitor, Mobile Cardiac Telemetry (through | Holter and Mobile Cardiac Telemetry | Holter, Cardiac Event Monitor, Mobile Cardiac Telemetry (through integration with another ECG analysis software) |
Rev C
UbiqVue 2AYe Cardiac System
Page 6 of 15
{10}
LifeSignals
K261569 510(k) Summary
| Comparison | Primary Predicate (K200690) | Secondary Predicate (K163512) | Subject Device |
| --- | --- | --- | --- |
| | integration with another ECG analysis software) | | |
| **Biosensor** | | | |
| Single Use | ✓ | ✓ (recycle) | ✓ |
| Wear / Battery Life | 03 days | 14 days | 8 days |
| Recording Format | Continuous | Continuous | Continuous |
| Data Storage capacity (max) | 03 days | 14 days | 8 days |
| Data Transfer | Continuous or periodic during the procedure, with data buffered and transmitted when connectivity is available | Periodic during the procedure (AT) and after procedure completion (QX) | Continuous or Periodic during the procedure or after procedure completion. |
| Data Transfer method | Wireless | Mailed (Zio QX) Wireless (Zio AT) | Wireless |
| Number of ECG Channels | Two Channel | Single Channel | Two Channel |
| ECG Sampling Rate | 244.14 Hz | 200 Hz | 244.14 Hz |
| ADC Resolution | 16 bits | ~15.5 bits | 16 bits |
| Dimension | 112 × 79 × 6 mm | 139.7 × 55.8 × 10.6 mm | 116.5 x 91.5 x 15.5 mm |
| Weight | 24.5 g | 10 g | 32 g |
| Location of Body | Upper Left Chest | Upper Left Chest | Upper Left Chest |
| Battery | Zinc-air | LiMnO₂ | LiMnO₂ |
| ECG electrode | Wet Electrode (hydrogel) | Wet Electrode (hydrogel) | Wet Electrode (hydrogel) |
| Applied part category | Type CF | Type BF | Type CF |
| Defibrillation Proof | ✓ | × | ✓ |
Rev C
UbiqVue 2AYe Cardiac System
Page 7 of 15
{11}
LifeSignals
K261569 510(k) Summary
| Comparison | Primary Predicate (K200690) | Secondary Predicate (K163512) | Subject Device |
| --- | --- | --- | --- |
| HF Surgical equipment compatibility | ✓ | ✗ | ✓ |
| Communication protocol | WLAN (802.11b) | BLE (only for Zio AT) | WLAN (802.11b) & BLE (5.2) |
| Wireless Radio Frequency | 2.4 – 2.485 GHz | 2.4 – 2.485 GHz | 2.4 - 2.4835 GHz |
| Communication Security | WPA2-PSK/AES-CCM | AES-CCM | WPA2-PSK/CCMP AES-CCM |
| **Patient Mobile App.** | | | |
| Data transferred to Server | ✓ | ✓ | ✓ |
| Symptoms & activity Recording | ✓ | ✓ | ✓ |
| **ECG Analysis Software** | No in-built arrhythmia analysis; intended to be used with any regulatory cleared third-party arrhythmia analysis software. | Arrhythmia analysis software integrated into the system for Holter Report (Zio QX) Automatic Arrhythmia analysis integrated in Biosensor (Zio AT) | No in-built arrhythmia analysis; intended to be used with any regulatory cleared third-party arrhythmia analysis software. |
| **Other components** | Secure Server, Web UI (Portal) | Patch, Relay Device and Monitoring Center (ZEUS system) | Relay Bridge, Central Server, Web Server, Web Portal, Mobile App |
Note: ✓ Feature is identical
✗ Feature not available
- Feature Unknown
Rev C
UbiqVue 2AYe Cardiac System
Page 8 of 15
{12}
[Non-Text]
LifeSignals
K261569 510(k) Summary
### Differences and Risks associated:
- The Subject device is intended for any cardiac workflow based on integration of the third-party ECG analysis software (similar to primary predicate), while the secondary predicate device is intended for Holter and Mobile Cardiac Telemetry based on the Biosensor used. However, this change does not add any risk.
- For meeting the intended use, the secondary predicate device Zio QX patch (K163512) transfers the stored ECG data only after the end of the procedure. However, the subject device has the ability to do a periodic data transfer during the procedure through “Holter Mobile App” in addition to the option of transferring the data after end of the procedure, similar to primary predicate (K200690) and secondary predicate Zio AT patch (K163512).
This additional feature does not add risk, as it is verified & validated (Usability) by LifeSignals, Inc. This feature helps clinical personnel an option to generate the interim (ECG) report when required during the procedure, giving additional clinical benefits. Further, it expedites the generation of ECG file (e.g., EDF format) immediately after the end of procedure as the data is immediately retrieved.
- The Subject device has the following additional system components compared to Predicate (which may have similar components):
○ Relay Bridge Software. This is one of the components of UbiqVue™ 2A Multiparameter system that is cleared under K242018 (Reference Device). In Subject device this component is intended for transferring the ECG data after the end of the procedure at clinical facility by connecting one or more Biosensor at a time using Wi-Fi protocol.
- Web Server Software. This is technically equivalent to Central Server, which is also one of the components of UbiqVue™ 2A Multiparameter system that is cleared under K242018 (Reference device).
These additional system components in the subject device do not add risk as these components are either part of the already cleared system and the performance of these components are verified & validated for usability & performance.
Rev C
UbiqVue 2AYe Cardiac System
Page 9 of 15
{13}
LifeSignals
K261569 510(k) Summary
# 10 Comparison with Reference device (For Biosensor Technology & Software component architecture)
| Comparison | Reference Device (K242018) | Subject Device |
| --- | --- | --- |
| Manufacturer | LifeSignals, Inc., USA | LifeSignals, Inc., USA |
| Product Codes | DRG, MHX, FLL, DQA | DRG, QYX, MWJ, DXH |
| Regulation Classification (Primary) | 21 CFR 870.2910 Class II | Identical |
| Intended Population | Non-critical, adult population, 18 years or older | Identical |
| Intended Use Environment | Home & Healthcare settings | Identical |
| Supported Workflow | Near real-time continuous patient monitoring | Holter, Cardiac Event Monitor, Mobile Cardiac Telemetry (through integration with another ECG analysis software) |
| Intended use / Indications for use | The UbiqVue™ 2A Multiparameter System is a wireless remote patient monitoring system intended for use by healthcare professionals for continuous collection of physiological data at home and in healthcare settings. This shall include electrocardiography, heart rate, SpO2%, respiration rate, pulse rate, photoplethysmography waveform, skin temperature, body temperature, body posture, body motion, R-R Interval, heart rate variability (HRV) and Blood Pressure (optional). Data is transmitted wirelessly near real time from UbiqVue 2A Wearable Biosensor and 3rd party device (for Blood Pressure only) to Remote Central server for display, storage, and analysis. The UbiqVue™ 2A Multiparameter System is intended for non-critical, adult population. | The UbiqVue™ 2AYe Cardiac System is intended to capture, record for analysis and report both symptomatic and asymptomatic cardiac events, along with continuous electrocardiogram (ECG) information, for long-term monitoring. Following patient monitoring using the UbiqVue™ Wearable Biosensor, a comprehensive final report is generated from the entire ECG recording. Optionally, intended users may choose to generate an interim report from periodically or intermittently transferred ECG recording wirelessly from the UbiqVue™ Wearable Biosensor. Final & interim reporting is based on the beat-to-beat information extracted from the ECG recording. |
Rev C
UbiqVue 2AYe Cardiac System
Page 10 of 15
{14}
[Non-Text]
LifeSignals
K261569 510(k) Summary
| Comparison | Reference Device (K242018) | Subject Device |
| --- | --- | --- |
| | The UbiqVueTM 2A Multiparameter System shall include the ability to notify healthcare professionals through alerts when Physiological parameters fall outside the set limits, manual trigger by patient and to display multiple patient's physiological data for remote monitoring at home and with visual alarm for active monitoring at hospitals and out of hospital patient care settings (such as clinics, outpatient surgery facilities, long-term care facilities and physician offices) in which care is administered by healthcare professionals. | The UbiqVueTM 2AYe Cardiac system is intended for non-critical, adult population, who are 18 years of age or older, at home and in healthcare settings, who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety.The reports are provided for review by the intended user to render a diagnosis based on their clinical judgment and experience. |
| Wearable Biosensor | | |
| Single Use Device | Yes | Identical |
| ECG Electrodes | Four Integrated wet gel | Identical |
| Thermistor | For Skin temperature | Not Applicable |
| Optical Sensor | For SpO2 & PR | Not Applicable |
| Accelerometer | For Posture & Motion | Available & but not Used |
| Ambient sensor | For Body temperature | Not Applicable |
| Body contact material | Hydrogel & Long-term wear acrylic adhesive | Identical |
| Battery | Li-MnO2 (1600 mAh) | Identical |
| Applied part category | Type CF | Identical |
| Defibrillation Proof | Yes | Identical |
| HF Surgical equipment compatibility | Yes | Identical |
| MRI Safe | No | Identical |
| Communication protocol | WLAN (802.11b), BLE 5.2 | Identical |
| Wireless Radio Frequency | 2.4 - 2.4835 GHz | Identical |
| Communication Security | WPA2-PSK /CCMP AES-CCM 128 | Identical |
| Status LED indicator | Green, Red | Identical |
Rev C
UbiqVue 2AYe Cardiac System
Page 11 of 15
{15}
LifeSignals
K261569 510(k) Summary
| Comparison | Reference Device (K242018) | Subject Device |
| --- | --- | --- |
| Patient Event Logging Button | Yes | Identical |
| Wear-life | 120 hours | 192 hours |
| Shelf life (claimed) | 13 months | Identical |
| Length | 116 x 91 x17 mm | Identical |
| Width | 91 mm | Identical |
| Thickness | 17 mm | 15.5 mm |
| Weight | 35 g | 32 g |
| Patient Relay Device | | |
| Supported OS | Android | Android & iOS |
| Authentication from Server | ✓ | ✓ |
| Data transferred to Server | ✓ | ✓ |
| Connectivity with Biosensor | Wi-Fi (802.11b) | BLE (5.2) |
| Symptom & Activity Recording | ✓ | ✓ |
| Relay Bridge Application | | |
| Supported OS | Open WRT | Identical |
| Supported Hardware | Off-the Shelf Router | Identical |
| Connectivity with Biosensor | Wi-Fi (802.11b) | Identical |
| Support for Multiple Biosensor connectivity | ✓ | ✓ |
| Central Server Software | | |
| Architecture | Linux based platform (including cloud-based hardware platform) | Identical |
| Data Access | Data is stored for access by Active Monitoring Portal | Data is stored for access by Web Server |
| Alert Notification | (Clinical, Technical, Manual) | Not Applicable |
Rev C
UbiqVue 2AYe Cardiac System
Page 12 of 15
{16}
[Non-Text]
LifeSignals
K261569 510(k) Summary
| Comparison | Reference Device (K242018) | Subject Device |
| --- | --- | --- |
| Data encryption & cybersecurity control | Conformance to guidance | Identical |
| Web Server | | |
| Architecture | - | Equivalent to Central Server. |
| Data Access | - | Authorized Users of Web Portal |
| Web (Monitoring) Portal | | |
| Architecture | Browser based application | Identical |
| User Interface | Designed for Multiparameter workflow | Designed for Cardiac workflow |
| Active Monitoring | Yes | Not Applicable |
Rev C
UbiqVue 2AYe Cardiac System
Page 13 of 15
{17}
LifeSignals
K261569 510(k) Summary
## 11 Summary of Performance Testing
Verification & Validation activities were performed on UbiqVue™ Cardiac System to demonstrate substantial equivalence to the predicate devices:
- Electrical Safety (EMT) and electromagnetic compatibility (EMC) testing were conducted on UbiqVue™ 2AYe Cardiac System with UbiqVue™ 2AYe Wearable Biosensors for conformance to IEC 60601-1(Ed.3.2), IEC 60601-1-2 (Ed. 4.1) and IEC 60601-1-11 (Ed. 2.1)
- Wear-life performance of ECG quality of UbiqVue™ 2AYe Cardiac System using UbiqVue 2AYe Wearable Biosensor under normal use condition (Ambulatory use) was validated using non-randomized, on-body performance study for the wear period of 192 hours.
- Usability study was conducted on UbiqVue™ 2AYe Cardiac System for conformance to IEC 60601-1-6 (Ed. 3.2) and “USFDA Guidance- Applying Human Factors and Usability Engineering to Medical Devices.”
- Wireless performance & coexistence testing conducted as per ANSI/IEEE C63.27:2017: American National Standard for Evaluation of Wireless coexistence on UbiqVue™ 2AYe Cardiac System.
- Software components in UbiqVue™ 2AYe Cardiac System were designed, documented, verified & validated as per the IEC 62304: Medical device Software - Software Life Cycle Process and USFDA Guidance for the content of premarket submissions for Software contained in medical devices.
- Software components in UbiqVue™ 2AYe Cardiac System were securely designed by following IEC 81001-5-1 Health software and health IT systems safety, effectiveness and security – Activities in the product life cycle and USFDA guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”
- Shelf-life & Reliability testing were conducted as per MIL-STD-810H, IEC 60068-206, RTCA DO-160 & IEC 60068-2-27 and verified on UbiqVue™ 2AYe Wearable Biosensor, as per the acceptance criteria.
- Biocompatibility testing of In-vitro cytotoxicity, skin irritation, skin sensitization and intracutaneous reactivity were conducted on UbiqVue™ 2A Wearable Biosensor (Reference Device K242018), according to ISO 10993-1: 2018 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process and submitted for the clearance. These reports are applicable for UbiqVue™ 2AYe Wearable Biosensor also. In addition to the above, Intracutaneous reactivity tests
Rev C
UbiqVue 2AYe Cardiac System
Page 14 of 15
{18}
LifeSignals
K261569 510(k) Summary
were also conducted on UbiqVue 2AYe Wearable Biosensor and its body contact (direct and in-direct) components.
## 12 Conclusion
The UbiqVue™ 2AYe Cardiac System is substantially equivalent in indications for use & intended use to the legally marketed Predicate Device(s). The UbiqVue™ 2AYe Wearable Biosensor technology & all the software components architecture of UbiqVue 2AYe Cardiac System is substantially equivalent to the technology adopted in the Reference Device. Minor differences between UbiqVue™ 2AYe Cardiac System and the Predicate & Reference devices have no significant effect on safety or effectiveness, as established through various performance tests. Thus, the proposed device/system is substantially equivalent to the legally marketed predicate & reference devices.
Rev C
UbiqVue 2AYe Cardiac System
Page 15 of 15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.