The Vertex Thrombectomy System is indicated for use in the pulmonary arteries for: •The non-surgical removal of emboli and thrombi. •Injection, infusion, and/or aspiration of contrast media and other fluids. The Vertex Thrombectomy System is intended for the treatment of pulmonary embolism.
Device Story
Vertex Thrombectomy System is a catheter-based device for non-surgical removal of pulmonary emboli/thrombi. System includes a Fixation Catheter (26F) with variable stiffness shaft, Aspiration Catheter (20F), and Extraction Set. Operated by physicians in healthcare facilities; catheters introduced over guidewires into pulmonary arteries. Aspiration is achieved via connection to a commercially available vacuum pump. Device enables mechanical thrombus removal and fluid management. Clinical benefit includes reduction of RV/LV ratio in patients with intermediate-risk pulmonary embolism.
Clinical Evidence
Prospective, multicenter, single-arm pivotal study (SPIRARE II) of 123 subjects. Primary safety endpoint: composite of device-related death, major bleeding, and adverse events at 48 hours; rate was 2.4% (p<0.0001 vs 25% goal). Primary effectiveness endpoint: reduction in RV/LV ratio from baseline to 48 hours; mean reduction was 0.39 (95% CI: 0.332, 0.452), exceeding the 0.20 performance goal (p<0.0001).
Technological Characteristics
System comprises 26F Fixation Catheter, 20F Aspiration Catheter, and Extraction Set. Materials include medical-grade plastics and metals. Features variable stiffness shaft, hemostatic valves, and radiopaque markers. Compatible with 0.035" guidewires. Sterilized via Ethylene Oxide (EO). Single-use. No integrated software or electronic sensing.
Indications for Use
Indicated for adults (≥22 years) with pulmonary embolism requiring non-surgical removal of emboli/thrombi from pulmonary arteries or injection/infusion/aspiration of fluids.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 4, 2026
Jupiter Endovascular
Nicole Barber
Sr. Director, Regulatory Affairs
155 Jefferson Dr.
Menlo Park, California 94025
Re: K261523
Trade/Device Name: Vertex™ Thrombectomy System
Regulation Number: 21 CFR 870.5150
Regulation Name: Embolectomy catheter
Regulatory Class: Class II
Product Code: QEW, KRA
Dated: July 6, 2026
Received: July 6, 2026
Dear Nicole Barber:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261523 - Nicole Barber
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261523 - Nicole Barber
Page 3
Sincerely,
KRINA M.
PATEL -S
Digitally signed by
KRINA M. PATEL -S
Date: 2026.08.04
15:13:03 -04'00'
For,
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261523 | ? |
| Please provide the device trade name(s). | | ? |
| Vertex™ Thrombectomy System | | |
| Please provide your Indications for Use below. | | ? |
| The Vertex Thrombectomy System is indicated for use in the pulmonary arteries for: •The non-surgical removal of emboli and thrombi. •Injection, infusion, and/or aspiration of contrast media and other fluids. The Vertex Thrombectomy System is intended for the treatment of pulmonary embolism. | | |
| Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
510(k) #: K261523
Jupiter Endovascular - Vertex™ Thrombectomy System
Prepared on: 2026-08-03
# 510(k) Summary
This 510(k) summary is submitted in accordance with the requirements of 21 CFR 807.92
## 1 Contact Details: 21 CFR 807.92(a)(1)
| Applicant Name | Jupiter Endovascular |
| --- | --- |
| Applicant Address | 155 Jefferson Dr., Menlo Park, CA 94025 |
| Applicant Contact Telephone | (888) 271-9499 |
| Applicant Contact | Nicole Barber, Senior Director Regulatory Affairs |
| Applicant Contact Email | nicole@jupiterendo.com |
## 2 Device Name: 21 CFR 807.92(a)(2)
| Trade Name | Vertex™ Thrombectomy System |
| --- | --- |
| Common Name | Embolectomy Catheter |
| Classification Name | Embolectomy Catheter |
| Regulation Number | 870.5150, 870.1210 |
| Product Code(s) | QEW, KRA |
## 3 Legally Marketed Predicate Device: 21 CFR 807.92(a)(3)
Predicate Device: K252057, Imperative Care / Truvic Medical, Inc. Symphony Thrombectomy System
Reference Device: K191710, Inari Medical FlowTriever 24 System
## 4 Device Description Summary: 21 CFR 807.92(a)(4)
The Vertex™ Thrombectomy System is comprised of:
- Vertex Fixation Catheter
- Vertex Aspiration Catheter
- Vertex Extraction Set
The Vertex Fixation Catheter is a single-use catheter having a multilayer shaft with stiffness that is selectively variable between flexible and fixed, a hemostatic valve, and an included obturator (“dilator”). The working length of the catheter is coated to improve lubricity. The handle of the Vertex Fixation Catheter includes a flush line, a fixation line for fixation control, and a hemostatic valve. The catheter shaft and dilator bodies are radiopaque.
The Vertex Aspiration Catheter is a single-use catheter with a hemostatic valve and an included dilator. The handle of the Vertex Aspiration Catheter includes a flush line and a hemostatic valve. The catheter shaft and dilator bodies are radiopaque. The Vertex Aspiration Catheter may be introduced through the Vertex Fixation Catheter.
Page 1 of 7
{5}
510(k) #: K261523
Jupiter Endovascular - Vertex™ Thrombectomy System
Prepared on: 2026-08-03
Either catheter may be used with a commercially available vacuum pump connected via the Vertex Extraction Set to aspirate thrombi. The Vertex Extraction Set includes a distal connector, proximal connector, extraction handle, and clot capture.
## 5 Intended Use/Indications for Use: 21 CFR 807.92(a)(5)
The Vertex Thrombectomy System is indicated for use in the pulmonary arteries for:
•The non-surgical removal of emboli and thrombi.
•Injection, infusion, and/or aspiration of contrast media and other fluids.
The Vertex Thrombectomy System is intended for the treatment of pulmonary embolism.
### Indications for Use Comparison: 21 CFR 807.92(a)(5)
The indications for use for the proposed device are substantially equivalent to the predicate device. Both the subject and the predicate devices are indicated for use in adults undergoing thrombectomy procedures in a healthcare facility; they are intended for use in the treatment of pulmonary embolism.
6 Technological Comparison: 21 CFR 807.92(a)(6)
| Attribute | Subject Device Vertex Thrombectomy System K261523 | Predicate Device Symphony Thrombectomy System K252057 | Reference Device FlowTriever 24 System K191710 |
| --- | --- | --- | --- |
| Product Code | QEW, KRA | QEW, KRA | QEW, KRA |
| Regulation | 21 CFR 870.5150 Embolectomy Catheters | 21 CFR 870.5150 Embolectomy Catheters | 21 CFR 870.5150 Embolectomy Catheters |
| Prescription Use Only | Yes | Yes | Yes |
| Indications for Use | The Vertex Thrombectomy System is indicated for use in the pulmonary arteries for:•The non-surgical removal of emboli and thrombi.•Injection, infusion, and/or aspiration of contrast media and other fluids. The Vertex Thrombectomy System is intended for the treatment of pulmonary embolism. | The Symphony Thrombectomy System is intended for:• The non-surgical removal of fresh, soft emboli and thrombi from blood vessels.• Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The Symphony Thrombectomy System is intended for use in the peripheral vasculature and for the treatment of pulmonary embolism. | The FlowTriever Retrieval/Aspiration System is indicated for:• The non-surgical removal of emboli and thrombi from blood vessels.• Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The FlowTriever Retrieval/Aspiration System is intended for use in the peripheral vasculature and for the treatment of pulmonary embolism. |
Page 2 of 7
{6}
510(k) #: K261523
Jupiter Endovascular - Vertex™ Thrombectomy System
Prepared on: 2026-08-03
| Attribute | Subject Device Vertex Thrombectomy System K261523 | Predicate Device Symphony Thrombectomy System K252057 | Reference Device FlowTriever 24 System K191710 |
| --- | --- | --- | --- |
| Main Components | The Vertex Thrombectomy System is comprised of: - Vertex Fixation Catheter (26F) with dilator - Vertex Aspiration Catheter (20F) with dilator - Vertex Extraction Set | The system is comprised of: - 24F Symphony Catheter - 24F Symphony Standard Dilator - 24F Symphony Advance™ Long Dilator - 24F Symphony ProHelix™ - 16F Symphony Catheter - 16F Symphony Dilator - 16F Symphony ProHelix™ - Symphony Clot Container - TRUVIC Generator - TRUVIC Canister - TRUVIC Tubeset | The system is comprised of: - Triever Catheters (16, 20, and 24 Fr) with dilator - FlowTriever Catheters (6-10 mm, 11-14 mm, 15-18 mm, and 19-25 mm) with dilator |
| Principle of Operation | The Vertex Fixation Catheter is intended to be introduced over a guidewire through compatible introducer sheaths. The Vertex Aspiration Catheter may be introduced through the Vertex Fixation Catheter and may be advanced over a guidewire. As needed, the Fixation and Aspiration Catheters may be used with a commercially available vacuum pump connected via the Vertex Extraction Set to aspirate thrombi. | The Symphony Thrombectomy System is designed to remove thrombus/embolus using controlled aspiration. The Symphony Catheter targets aspiration from the TRUVIC Generator directly to the thrombus. The Symphony ProHelix may be used to facilitate aspiration and removal of the thrombus through the Symphony Catheter. | The FlowTriever Catheter is inserted through the Triever Catheter and advanced to the thrombus. Self-expanding wireform disks are deployed to engage thrombus by retracting the outer delivery catheter. The FlowTriever Catheter is retracted into the Triever Catheter to capture the targeted thrombus. Additional clot may be removed by aspiration with the provided 60 cc VacLok Vacuum syringe. |
| Dimensions | Vertex Fixation Catheter - 26F x 85cm | - 16F x 121 cm - 24F x 85 cm | Triever Catheters: - 16, 20, and 24F x 107, 90, 90 cm |
Page 3 of 7
{7}
510(k) #: K261523
Jupiter Endovascular - Vertex™ Thrombectomy System
Prepared on: 2026-08-03
| Attribute | Subject Device Vertex Thrombectomy System K261523 | Predicate Device Symphony Thrombectomy System K252057 | Reference Device FlowTriever 24 System K191710 |
| --- | --- | --- | --- |
| | Vertex Aspiration Catheter • 20F x 105cm | | FlowTriever Catheters: • 6-10, 11-14, 15-18, and 19-25 mm x 115 cm |
| Guidewire Compatibility | 0.035” | 0.035” | 0.035” |
| Vacuum Source | Aspiration Pump | Aspiration Pump | 60 cc Syringe |
| Sterilization Method | Ethylene Oxide (EO) Sterilization | Ethylene Oxide (EO) Sterilization | Ethylene Oxide (EO) Sterilization |
| Single-Use Disposition | Single-use device | Single-use device | Single-use device |
| Materials | Commonly used medical grade plastics and metals | Commonly used medical grade plastics and metals | Commonly used medical grade plastics and metals |
| Non-Clinical Testing | • Visual and Dimensional Verification • Kink Resistance • Actuation Force • Tensile and Torque Strength • System Freedom from Leakage (positive and negative pressure) • Burst Pressure • Radiodetectability • Simulated Use • Coating Integrity Testing (lubricity) • Particulate Testing • Air Leak • Sheath Stiffness (relaxed state and fixed state) • Obturator Removal Force • Simulated Shipping • Climate Conditioning • Sterile Barrier Integrity | • Visual and Dimensional Verification • Kink / Bend Verification • Actuation Force Verification • Tensile and Torque Strength Verification • Positive Pressure / Fluid Leak Verification • Negative Pressure / Air Leak Verification • Lumen Integrity Verification • Burst Pressure Verification • Fluoroscopy Validation (Visibility test) • Simulated Use Performance Validation • Corrosion Resistance Testing • Coating Integrity Testing • Acute Particulate Testing | • Visual & Dimensional Inspection • Guidewire and Sheath Compatibility Verification • Snap Fit, Dilator Luer to Guide Catheter Hemostasis Valve • Leakage Testing • Vacuum • Air leakage During Aspiration • Retraction Force Testing • Kink Radius Testing • Determination of Flowrate • Burst Testing • Clot Burden Removal Validation • Push Button Force • Simulated Use and Tensile Testing • Simulated Use and Torque |
Page 4 of 7
{8}
510(k) #: K261523
Jupiter Endovascular - Vertex™ Thrombectomy System
Prepared on: 2026-08-03
| Attribute | Subject Device Vertex Thrombectomy System K261523 | Predicate Device Symphony Thrombectomy System K252057 | Reference Device FlowTriever 24 System K191710 |
| --- | --- | --- | --- |
| | | Drop Testing VerificationComponent Fatigue Testing Verification | |
| Biocompatibility | Biocompatibility testing in accordance with ISO 10993-1:CytotoxicitySensitizationIrritation/Intracutaneous ReactivityAcute Systemic ToxicityMaterial Mediated PyrogenicitySterilization residualsHemocompatibility ISO 10993-4:Direct and Indirect HemolysisPartial Thromboplastin TimeComplement ActivationIn vitro Hemocompatibility | Biocompatibility testing in accordance with ISO 10993-1:CytotoxicitySensitizationIntracutaneous ReactivityAcute Systemic ToxicityMaterial Mediated PyrogenicityHemocompatibilityDirect and Indirect HemolysisComplement ActivationIn-Vivo Thrombogenicity (from GLP animal study) | Biocompatibility testing in accordance with ISO 10993-1:MEM elutionISO Guinea Pig Maximization SensitizationISO Intracutaneous IrritationAcute systemic toxicityHemolysis, direct contact and extract methodComplement Activation, C3a and SC5b-9 AssayUnactivated Partial Thromboplastin Time (UPTT) Test (from K180466) |
| Animal Study | A GLP animal study was conducted to confirm the safety and performance of the Vertex Thrombectomy System. A reference device was used for comparative testing. In-vivo thrombogenicity was also assessed | A GLP animal study was performed to assess the acute and chronic safety and performance of the Symphony Thrombectomy System. In-vivo thrombogenicity was also assessed. | Animal testing confirmed the safety and performance of the Triever24 Catheter. |
Page 5 of 7
{9}
510(k) #: K261523
Jupiter Endovascular - Vertex™ Thrombectomy System
Prepared on: 2026-08-03
## 7 Clinical Study
### Introduction
The SPIRARE II study, a prospective, multicenter, single-arm, pivotal clinical investigation, was conducted to evaluate the Vertex Thrombectomy System for the treatment of pulmonary embolism. A total of 123 subjects were enrolled at 19 sites in the United States, Austria, Poland, and Switzerland. The primary effectiveness endpoint was the reduction in RV/LV ratio from baseline to 48 hours or discharge by core lab assessed computed tomography angiography. The primary safety endpoint was the composite of device related major adverse events (MAEs) within 48 hours of the index procedure. Major adverse events were adjudicated for clinical endpoint analysis by an independent clinical events committee (CEC).
### Demographics
Of the 123 enrolled and treated subjects, mean age was \(61 \pm 14\) years; 63 (51%) were male and 60 (49%) were female. Mean body mass index was \(34 \pm 9 \mathrm{~kg} / \mathrm{m}^2\). Cardiovascular comorbidities were reported in 72% of subjects, including hypertension in 59% and hyperlipidemia in 41%. Pulmonary history was reported in 44%, including prior deep vein thrombosis (DVT) in 19%, concomitant DVT at enrollment in 20%, and prior PE in 7%. Baseline disease characteristics reflected intermediate-risk PE severity: mean RV/LV ratio \(1.44 \pm 0.33\), mean Refined Modified Miller Score \(21.79 \pm 3.54\), mean pulmonary artery pressure \(29.6 \pm 6.9 \mathrm{mmHg}\), mean total pulmonary vascular resistance \(6.6 \pm 2.5\) Wood units, mean cardiac index \(2.3 \pm 0.6 \mathrm{~L} / \mathrm{min} / \mathrm{m}^2\), and mean mMRC dyspnea score \(2.72 \pm 1.24\) (with 58% reporting Grade 3 or 4 dyspnea).
### Primary Safety Endpoint
The primary safety endpoint was met. The composite endpoint comprises device-related death; major bleeding (BARC Type 3b, 3c, 5a, 5b, or bleeding requiring blood transfusion); and device-related adverse events including clinical deterioration, pulmonary vascular injury, or cardiac injury. The endpoint was formally tested against a pre-specified performance goal of \(< 25\%\). The 48-hour MAE rate was \(2.4\%\) (3/123); one-sided exact binomial \(\mathrm{p} < 0.0001\).
Composite Safety Endpoint - ITT (N=123)
| Major Adverse Events | 0 to 48 Hours | Comparator | p-value |
| --- | --- | --- | --- |
| Subjects with ≥1 MAE | 3/123 (2.4%) | 25% | <0.0001 |
| Device-related death within 48 hours of the procedure | 0/123 (0%) | | |
| Major bleeding within 48 hours of the procedure | 2/123 (1.6%) | | |
| Device-related adverse events within 48 hours of the procedure | 1/123 (0.8%) | | |
| Clinical deterioration | 1/123 (0.8%) | | |
| Pulmonary vascular injury | 0/123 (0%) | | |
| Cardiac injury | 0/123 (0%) | | |
Results are presented as subjects with events/total subjects (%).
Page 6 of 7
{10}
510(k) #: K261523
Jupiter Endovascular - Vertex™ Thrombectomy System
Prepared on: 2026-08-03
### Primary Effectiveness Endpoint
The primary effectiveness endpoint was met. The reduction in RV/LV ratio from baseline to 48 hours was assessed by CTA and evaluated by an independent core laboratory. The mean RV/LV ratio decreased from \(1.44 \pm 0.33\) at baseline to \(1.05 \pm 0.21\) at 48 hours, with a mean reduction of \(0.39 \pm 0.33\) (95% CI: 0.332, 0.452), evaluated by a one-sided paired t-test against \(\mathrm{H}_0\): mean reduction \(\leq 0.20\) (one-sided \(p < 0.0001\)). The lower bound of the \(95\%\) confidence interval substantially exceeded the 0.20 performance goal.
Reduction in RV/LV Ratio from Baseline to 48 Hours – ITT (N = 123)
| | Baseline | 48 Hours | Change from Baseline to 48 Hours |
| --- | --- | --- | --- |
| RV/LV ratio (CT-based) | \( 1.44 \pm 0.33 \) | \( 1.05 \pm 0.21 \) | \( -0.39 \pm 0.33 \) |
| N | 121 | 122 | 120 |
## Conclusion
All pre-specified performance goals were met, demonstrating substantial equivalence of the Vertex Thrombectomy System for the treatment of acute pulmonary embolism.
## 8 Non-clinical and/or Clinical Tests Summary & Conclusions: 21 CFR 807.92(b)
The Vertex Thrombectomy System has the same intended use as the predicate device. After analyzing technology, materials, bench tests, biocompatibility, animal data, and clinical data, it can be concluded that the Vertex Thrombectomy System is substantially equivalent to the predicate device.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.