The BioWave BioWraps are cutaneous electrodes to be used with legally marketed BioWave branded neurostimulators. The knitted garment electrodes are non-sterile reusable prescription-use conductive garments that are intended to deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. These body parts can include hand/wrist, elbow, foot/ankle, knee, thigh/calf, cervical neck and lower back.
Device Story
BioWave BioWraps are washable, wrappable conductive garments; designed for use with BioWavePRO RX and BioWaveHOME RX neurostimulators. Input: electrical stimulation signals from external neurostimulator. Transformation: garments deliver stimulation signals to body surface via silver fiber conductive inner fabric; provide low current density with uniform distribution when used with conductive cream. Output: localized neurostimulation. Usage: clinic or home; operated by patient or clinician. Benefit: provides non-invasive, wearable interface for neurostimulation therapy at various anatomical sites; facilitates treatment of pain or other conditions addressed by BioWave stimulators.
Clinical Evidence
Bench testing only. Previous testing demonstrated conductivity, resistivity, impedance, and uniform current delivery. Biocompatibility and reusability were previously established. No new clinical data required as design changes involve anatomical expansion using previously validated materials and performance characteristics.
Technological Characteristics
Materials: stretchable neoprene-like outer, silver fiber conductive inner fabric. Sensing/Actuation: passive conductive electrode interface. Energy: passive delivery of neurostimulator current. Form factor: wrappable bands for specific anatomical sites. Connectivity: connects to BioWave neurostimulators. Sterilization: non-sterile. Impedance: 1.27 ohms resistance per inch.
Indications for Use
Indicated for adults (22+ years) requiring neurostimulation therapy via BioWave branded neurostimulators on the hand/wrist, elbow, foot/ankle, knee, thigh/calf, cervical neck, or lower back.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 3, 2026
Biowave Corporation
% Emily Andre
Associate Director, Neurology Regulatory Affairs
MCRA, an IQVIA Business
803 7th St. NW
4th Floor
Washington, District of Columbia 20001
Re: K261522
Trade/Device Name: BioWave BioWraps
Regulation Number: 21 CFR 882.1320
Regulation Name: Cutaneous Electrode
Regulatory Class: Class II
Product Code: GXY
Dated: May 7, 2026
Received: May 7, 2026
Dear Emily Andre:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20903
www.fda.gov
{1}
K261522 - Emily Andre
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261522 - Emily Andre
Page 3
Sincerely,
**Tushar Bansal -S**
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K261552 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| BioWave BioWraps | | | |
| Please provide your Indications for Use below. | | | ? |
| The BioWave BioWraps are cutaneous electrodes to be used with legally marketed BioWave branded neurostimulators. The knitted garment electrodes are non-sterile reusable prescription-use conductive garments that are intended to deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. These body parts can include hand/wrist, elbow, foot/ankle, knee, thigh/calf, cervical neck and lower back. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
# 510(k) Summary - K261522
**Device Trade Name:** BioWave BioWraps
**Manufacturer:** BioWave Corporation
19 Newtown Turnpike
Westport, CT 06880
**Contact:** Bradford Siff
Founder & President
BioWave Corporation
19 Newtown Turnpike
Westport, CT 06880
Phone: 203-247-9020
Fax: (475) 444-3400
brad.siff@biowave.com
**Classifications:** 21 CFR §882.1320, Cutaneous Electrodes
**Class:** II
**Product Codes:** GXY
### **Indications for Use:**
The BioWave BioWraps are cutaneous electrodes to be used with legally marketed BioWave branded neurostimulators. The knitted garment electrodes are non-sterile reusable prescription-use conductive garments that are intended to deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. These body parts include hand hand/wrist, elbow, foot/ankle, knee, thigh/calf, cervical neck and lower back.
### **Device Description:**
The BioWave BioWraps are washable, wrappable bands made from a stretchable neoprene-like outer and silver fiber conductive inner fabric. The electrodes are secured to the skin through the wrapping of the bands. The BioWraps are intended for use on the hand/wrist, elbow, foot/ankle, knee, thigh/calf, cervical neck and lower back. When used alongside a conductive cream, the electrodes provide a low current density with uniform current distribution. The BioWraps are to be used alongside cleared BioWave neurostimulator devices.
The current submission adds 2 new BioWraps for the thigh/calf and cervical neck. The addition of these two new BioWraps necessitates an update to the indications for use which explicitly describe the anatomical locations of the cleared BioWraps. These anatomical locations are currently treated by traditional electrodes used with the BioWave family of devices and do not introduce new questions of safety and effectiveness. The generation of a wrappable electrode only provides more options for patients to engage in use of the BioWave family of devices. The BioWave BioWraps are available as
{5}
prescription use only with BioWavePRO® RX and BioWaveHOME® RX neurostimulators.
### **Predicate Device:**
The BioWave Corporation’s BioWave BioWraps are substantially equivalent to the primary predicate (BioWave BioWraps - K203158) previously cleared with respect to indications, design, function, and materials, as outlined below.
### **Performance Testing Summary:**
Bench testing conducted on the BioWave BioWraps has previously demonstrated that the garments meet design controls with regard to conductivity, resistivity, impedance, and uniform delivery of low doses of current consistent with that of the cited predicate. Reusability and biocompatibility have also been previously demonstrated.
### **Substantial Equivalence:**
| Element of Comparison | Primary Predicate K203158 BioWave BioWraps | Subject Device K261522 BioWave BioWraps | Comparison |
| --- | --- | --- | --- |
| **Manufacturer** | BioWave Corporation | BioWave Corporation | Same |
| **Device Name** | BioWave BioWraps | BioWave BioWraps | Same |
| **Regulation Number** | 882.1320 | 882.1320 | Same |
| **Product Code** | GXY | GXY | Same |
| **OTC / Rx** | OTC and Rx | Rx | Similar. The BioWave BioWraps are no longer used with the BioWaveGo device which is available OTC. |
| **Intended Use / Indications for Use** | The BioWave BioWraps are cutaneous electrodes to be used with legally marketed BioWave branded neurostimulators. The knitted garment electrodes are non-sterile reusable prescription-use and OTC conductive garments that are intended to deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. These body parts include hand hands/wrist, elbow, foot/ankle, knee, and lower back. | The BioWave BioWraps are cutaneous electrodes to be used with legally marketed BioWave branded neurostimulators. The knitted garment electrodes are non-sterile reusable prescription-use conductive garments that are intended to deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. These body parts include hand hands/wrist, elbow, foot/ankle, knee, thigh/calf, cervical neck and lower back. | Similar. The current submission includes a minor change to the indications for use to update new anatomical locations of electrodes (thigh/calf and cervical neck). This change does not constitute any new concerns related to safety and effectiveness as they were designed to be used on body locations that have been previously targeted by traditional electrodes. |
{6}
| Element of Comparison | Primary Predicate K203158 BioWave BioWraps | Subject Device K261522 BioWave BioWraps | Comparison |
| --- | --- | --- | --- |
| **Compatible TENS Unit/Neurostimulator** | BioWavePRO® RX (K052289) BioWaveHOME® RX (K152437) BioWaveGO® (K180943) | BioWavePRO® RX (K052289) BioWaveHOME® RX (K152437) | Similar. The BioWave BioWraps are no longer used with the BioWaveGo device. |
| **Design (Shape)** | Wrappable bands for the hand/wrist, elbow, foot/ankle, knee, and lower back | Wrappable bands for the hand/wrist, elbow, foot/ankle, knee, thigh/calf, cervical neck and lower back | Similar. The current submission includes a minor change to the design to add new anatomical locations that are targeted by the BioWraps (thigh/calf and cervical neck). The differences in design do not raise new questions of safety and effectiveness. |
| **Sterile/ Non-Sterile** | Non-Sterile | Non-Sterile | Same |
| **Size** | All BioWraps are available in small/medium (S/M) and large/extra-large (L/XL) sizes: Low Back: - S/M: 28-38' Waist - L/XL: 38-50' Waist Knee: - S/M: 12-15.5' circumference around the kneecap - L/XL: 15.5 - 19' circumference around the kneecap Foot/Ankle: - S/M: Women's shoe size 6-9, Men's shoe size 7-8.5 - L/XL: Women's shoe size 9.5-11, Men's shoe size 9-13 Elbow: - S/M: 8-12' circumference around the elbow joint with arm extended - L/XL: 12-16' circumference around the elbow joint with arm extended Hand/Wrist: - S/M: 6-9' circumference around the dominant hand - L/XL: 9-12' circumference around the dominant hand | Some BioWraps are available in small/medium (S/M) and large/extra-large (L/XL) sizes, and some are one-size fits all: Low Back: - S/M: 28-37' Waist - L/XL: 37-50' Waist Knee: - One size: 12-24' circumference around the kneecap Foot/Ankle: - S/M: Women's shoe size 6-10, Men's shoe size 5-9 - L/XL: Women's shoe size 11-16, Men's shoe size 10-15 Elbow: - S/M: 8-12' circumference around the elbow joint with arm extended - L/XL: 12-16' circumference around the elbow joint with arm extended Hand/Wrist: - S/M: 5.75-7' circumference around the dominant hand - L/XL: 7-8.5' circumference around the dominant hand Thigh/Calf: - One size: 12-30' circumference around the thigh or calf Cervical Neck: - One size: 12-20' circumference around the neck | Similar. The current submission includes minor changes to the sizes of the cleared BioWraps to account for design shape updates. This submission also adds new BioWraps for the thigh/calf and cervical neck. The differences in size do not raise new questions of safety and effectiveness. Previously submitted non-clinical testing information supports safety and effectiveness of the 2 new BioWraps in this submission. |
| | Sizes for the BioWraps are as follows: Low Back: | Sizes for the small/medium and large/extra-large BioWraps are as follows: | Similar. The current submission includes minor changes to the |
{7}
| Element of Comparison | Primary Predicate K203158 BioWave BioWraps | Subject Device K261522 BioWave BioWraps | Comparison |
| --- | --- | --- | --- |
| **Conductive Surface** | - Width: S/M - 4.5'; L/XL - 5.5' - Height: S/M - 3'; L/XL - 3' Knee: - Width in Center: S/M - 3'; L/XL - 3 ½' - Total Height: S/M - 3 ¼'; L/XL - 4' Foot/Ankle: - Ankle Length: S/M - 11 ½'; L/XL - 13' - Ankle Width: S/M - 1 ¾'; L/XL - 2' - Foot Length: S/M - 9 ½'; L/XL - 10 ½' - Foot Width: S/M - 1 ¾'; L/XL - 2' Elbow: - Width in Center: S/M - 2 1/8'; L/XL - 2 5/8' - Total Height: S/M - 3 ¾'; L/XL - 4 ½' Hand/Wrist: - Hand Width in Center: S/M - 8'; L/XL - 11' - Hand Total Height: S/M - 2 ½'; L/XL - 2 ½' - Wrist Width in Center: S/M - 7'; L/XL - 10' - Wrist Total Height: S/M - 2 ½'; L/XL - 2 ½' | Low Back: - Width: S/M - 4.5'; L/XL - 5.5' - Height: S/M - 3'; L/XL - 3' Knee: - Width in Center: S/M - 3'; L/XL - 3 ½' - Total Height: S/M - 3'; L/XL - 4' Foot/Ankle: - Ankle Length: S/M - 7 ¾'; L/XL - 9 ¾' - Ankle Width: S/M - 1 ½'; L/XL - 2' - Foot Length: S/M - 7 ¾'; L/XL - 9 5/8' - Foot Width: S/M - 1 ½'; L/XL - 2' Elbow: - Width in Center: S/M - 2 ½'; L/XL - 3' - Total Height: S/M - 3 ¾'; L/XL - 4' Hand/Wrist: - Hand Width in Center: S/M - 8.5'; L/XL - 9.5' - Hand Total Height: S/M - 2 ½'; L/XL - 2 ½' - Wrist Width in Center: S/M - 7'; L/XL - 8.5' - Wrist Total Height: S/M - 2 ½'; L/XL - 2 ½' Thigh/Calf: - Width in Center: 4 ¼' - Total Height: 3 ¼' Cervical Neck: - Width in Center: 2' - Total Height: 2 ½' | sizes of the cleared BioWraps to account for design shape updates. This submission also adds new BioWraps for the thigh/calf and cervical neck. The differences in conductive surface size do not raise new questions of safety and effectiveness. |
| **Impedance Parameters** | 1.27 ohms resistance per inch | 1.27 ohms resistance per inch | Same |
| **Washable / Not Washable** | Washable | Washable | Same |
| **Reusable** | Single Patient, Reusable | Single Patient, Reusable | Same |
| **Patient Contacting** | External Only, Intact Skin, <24 Hours Silver Fiber | External Only, Intact Skin, <24 Hours Silver Fiber | Same |
| **Performance Qualities** | Bench testing of the conductive fabric confirmed that the garment electrodes do not change their inherent conductivity following washing, indicating that there is no significant adverse effect on conductivity of the device or its inherent ability to deliver low dose current density uniformly to the skin of the wearer. | Bench testing of the conductive fabric submitted under K203158 confirmed that the garment electrodes do not change their inherent conductivity following washing, indicating that there is no significant adverse effect on conductivity of the device or its inherent ability to deliver low dose current density uniformly to the skin of the wearer. This testing applies to the | Same |
{8}
| Element of Comparison | Primary Predicate K203158 BioWave BioWraps | Subject Device K261522 BioWave BioWraps | Comparison |
| --- | --- | --- | --- |
| | | current device since the conductive material has not changed since review of the predicate. | |
# **Conclusion:**
The subject device, which now includes thigh/calf and cervical neck BioWave BioWraps, have the same intended use as the predicate device. The differences between the subject and predicate devices do not raise different questions of safety and effectiveness with respect to indications for use, design, materials, function, availability (i.e., prescription only use), and performance. Therefore, the BioWave BioWraps have been found to be substantially equivalent to the predicate device (BioWave BioWraps, K203158).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.