Tetric plus Fill: - Missing tooth structure in anterior and posterior teeth Types of restorations: - Restorations of all types of permanent dentition (including the replacement of individual cusps in the posterior region) - Reconstructive build-ups - Restorations of deciduous teeth - Repair of composite / ceramic restorations - Direct veneers - Sealing of implant screw channels Tetric plus Flow: - Missing tooth structure in anterior and posterior teeth Types of restorations: - Initial layer up to 4 mm in Class I and II restorations* - Small direct restorations in permanent teeth (not occlusion-bearing)* - Restorations of deciduous teeth - Liner under direct restorations of all cavity classes - Class V restorations - Extended fissure sealings - Repair of composite / ceramic restorations - Sealing of implant screw channels *In case of occlusion-bearing restorations, apply a final layer using a sculptable composite material.
Device Story
Tetric plus Fill and Tetric plus Flow are light-curing, radiopaque composite resins used by dentists for direct restorative dental procedures. The devices are applied to prepared cavities following standard adhesive techniques, including shade determination, isolation, and conditioning. Tetric plus Fill is a sculptable composite for general restorations; Tetric plus Flow is a flowable composite for liners, small restorations, and fissure sealing. Both utilize an advanced chameleon effect, achieved through an adapted chemical recipe, to match VITA classical shades using only four available shades. The devices are polymerized using dental curing lights. Clinical benefits include restoration of tooth structure and aesthetics. The devices are intended for professional use in a clinical setting.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 4049:2019 and FDA guidance for dental composite resins. Metrics included sensitivity to ambient light, depth of cure (≥ 1.5 mm), Vickers hardness, flexural strength, water sorption, solubility, and radiopacity. Biocompatibility assessed per ISO 10993, ISO 7405, and ISO 21726:2019; no significant cytotoxic, genotoxic, or toxicological risks identified.
Technological Characteristics
Light-curing, radiopaque composite resin. Materials comply with ISO 4049:2019. Available as syringes or cavifills. Polymerization via dental curing light. No sterilization required (consumable).
Indications for Use
Indicated for patients requiring dental restoration of missing tooth structure in anterior and posterior teeth, including permanent and deciduous dentition, reconstructive build-ups, repair of existing composite/ceramic restorations, direct veneers, liner placement, fissure sealing, and sealing of implant screw channels. Contraindicated in patients with known allergies to ingredients or where a dry working field cannot be established.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 1, 2026
Ivoclar Vivadent, Inc.
Anderjeet Gulati
Sr. Manager of QA & Regulatory Affairs
175 Pineview Dr.
Amherst, New York 14228
Re: K261485
Trade/Device Name: Tetric plus Fill;Tetric plus Flow
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth Shade Resin Material
Regulatory Class: Class II
Product Code: EBF
Dated: May 5, 2026
Received: May 5, 2026
Dear Anderjeet Gulati:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261485 - Anderjeet Gulati
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261485 - Anderjeet Gulati
Page 3
Sincerely,
# MICHAEL E. ADJODHA -S
Michael E. Adjodha, M.ChE., RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration **Indications for Use** | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 *See PRA Statement below.* |
| --- | --- |
510(k) Number (if known)
K261485
Device Name
Tetric plus Fill;
Tetric plus Flow
Indications for Use (Describe)
Tetric plus Fill:
- Missing tooth structure in anterior and posterior teeth
Types of restorations:
- Restorations of all types of permanent dentition (including the replacement of individual cusps in the posterior region)
- Reconstructive build-ups
- Restorations of deciduous teeth
- Repair of composite / ceramic restorations
- Direct veneers
- Sealing of implant screw channels
Tetric plus Flow:
- Missing tooth structure in anterior and posterior teeth
Types of restorations:
- Initial layer up to 4 mm in Class I and II restorations*
- Small direct restorations in permanent teeth (not occlusion-bearing)*
- Restorations of deciduous teeth
- Liner under direct restorations of all cavity classes
- Class V restorations
- Extended fissure sealings
- Repair of composite / ceramic restorations
- Sealing of implant screw channels
*In case of occlusion-bearing restorations, apply a final layer using a sculptable composite material.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
K261485
510(K) SUMMARY
ivoclar
1
**Contact:** Anderjeet Gulati, Sr. Manager QA & Regulatory Affairs
Ivoclar Vivadent, Inc.
175 Pineview Drive
Amherst, New York 14228
716-264-2046
Anderjeet.gulati@ivoclar.com
**Company:** Ivoclar Vivadent, AG
Bendererstrasse 2, Schaan, FL-9494, Liechtenstein
+423-235-3535
**Date Prepared:** May 5, 2026
**Proprietary Name:** **Tetric® plus Fill**
**Tetric® plus Flow**
**Classification Name:** Material, Tooth Shade, Resin (872.3790)
(Classification Code EBF)
**Common Name:** Dental Composite
**Tetric plus Fill Predicate Device:** Tetric PowerFill (K183380) by Ivoclar Vivadent AG, Liechtenstein
**Tetric plus Flow Predicate Device:** Tetric PowerFill (K183380) by Ivoclar Vivadent AG,
Liechtenstein
**Device Description:**
**Tetric® plus Fill** is a light-curing, radiopaque composite resin for the direct restorative treatment of anterior and posterior teeth and for restoring occlusal surfaces.
**Tetric® plus Flow** is a flowable, light-curing, radiopaque composite resin for the direct restorative treatment of anterior and posterior teeth and for restoring occlusal surfaces.
In both, the principles of operation are the same. The dental professional starts with determining tooth shade, preparing the cavity, ensuring isolation, placing the matrix, lining the cavity close to the pulp, conditioning the cavity, applying the composite, and finishing/occlusion control/polishing.
510K Summary -1
{5}
510(K) SUMMARY
ivoclar
2
**Indications for Use:**
**Tetric plus Fill**
-Missing tooth structure in anterior and posterior teeth
Types of restorations:
-Restorations of all types of permanent dentition (including the replacement of individual cusps in the posterior region)
-Reconstructive build-ups
-Restorations of deciduous teeth
-Repair of composite / ceramic restorations
-Direct veneers
-Sealing of implant screw channels
**Tetric plus Flow**
-Missing tooth structure in anterior and posterior teeth
Types of restorations:
-Initial layer up to 4mm in Class 1 and II restorations*
-Small direct restorations in permanent teeth (not occlusion-bearing)*
-restorations of deciduous teeth
-Liner under direct restorations of all cavity classes
-Class V restorations
-Extended fissure sealings
-Repair of composite / Ceramic restorations
-Sealing of implant screw channels
*In case of occlusion-bearing restorations, apply a final layer using a sculptable composite material.
**Comparison to Predicate:**
The primary predicate device to which Tetric plus Fill has been compared is Ivoclar Vivadent AG's Tetric PowerFill (K183380). See Comparison Matrix 1.
The primary predicate device to which Tetric plus Flow has been compared is Ivoclar Vivadent AG's Tetric PowerFill (K183380). See Comparison Matrix 2.
510K Summary -2
{6}
510(K) SUMMARY
ivoclar
3
Comparison Matrix 1:
| | Predicate Device: | Proposed Device: | Deviation | |
| --- | --- | --- | --- | --- |
| Device | Ivoclar Vivadent AG: Tetric PowerFill (K183380) | Ivoclar Vivadent AG: Tetric plus Fill | Yes | No |
| Indications for Use | **Conventional application (Light intensity 2,000mW/cm2):** - Restorations in the posterior region (Class I and II, including the replacement of individual cusps) - Class V restorations (cervical caries, root erosion, wedge-shaped defects) - Reconstructive build-ups - Restoration of deciduous teeth **Light-curing using the 3sCure mode of Bluephase PowerCure (Light Intensity 3,050 mW/cm2):** - Restorations in the posterior region of permanent dentition (Class I and II, including the replacement of individual cusps) when light-cured from the occlusal aspect | **Indications** – Missing tooth structure in anterior and posterior teeth Types of restorations: – Restorations of all types of permanent dentition (including the replacement of individual cusps in the posterior region) – Reconstructive build-ups – Restorations of deciduous teeth – Repair of composite / ceramic restorations – Direct veneers – Sealing of implant screw channels | ☑ | ☐ |
| Precaution Measures/ Contraindications / Processing restrictions/ Side effects | **Contraindications:** - If a dry working field cannot be established, or the stipulated working technique cannot be applied - If a patient is known to be allergic to any of the ingredients of Tetric PowerFill | **Contraindications:** The use of this product is contraindicated if the patient is known to be allergic to any of its ingredients | ☑ | ☐ |
| Summary of Indications, Precaution Measures/ Contraindications / Processing restrictions/ Side effects | Tetric PowerFill and the new device Tetric plus Fill are both dental filling materials used by the dentist for the direct filling therapy in dentistry. The new device Tetric plus Fill can also be used for restoration in anterior teeth compared to the predicate. The indications of the new device Tetric plus Fill cover the indications of Tetric PowerFill for posterior teeth. In addition, Tetric plus Fill can also be used for direct veneers, for the repair of composite/ceramic restorations and the closure of implant screw channels. The application of the 3sCure mode of Bluephase PowerCure is described in the instructions for use for the new device under section “Applying the composite resin”. The contraindications are basically the same for both devices. The information regarding the dry working field can be found in section “Limitations of use” for the new device. | | | |
| Working Principle | Tetric PowerFill is a dental composite material for the use of direct filling in teeth. | Tetric plus Fill is a dental composite material for the use of direct filling in teeth. | ☐ | ☑ |
| Summary of Working Principle | No difference | | | |
| Delivery forms/dosage | Tetric PowerFill is available as syringes and cavifills in 3 different shades (IVA, IVB, IVW) | Tetric plus Fill is available as syringes and cavifills in 4 different shades (A2 plus, A3 plus, A3.5 plus, Bleach plus) | ☑ | ☐ |
| Summary of Delivery forms/dosage | Tetric plus Fill is available in 4 different shades with which all classical vita shades can be covered (see table below). This innovation can be achieved due to Tetric plus Fill’s advanced chameleon effect. This chameleon effect is enabled due to an adapted recipe (see section chemical composition) compared to the predicate device. | | | |
| | **Shade correspondence** | | | |
| | Tetric* plus | A2 plus | A3 plus | A3.5 plus Bleach plus |
| | VITA* classical shades | A1, A2, B2, C1, D2 | A3, C2, C3, D3, D4 | A3.5, A4, B3, B4, C4 B1, Bleach |
| | * No registered trademark of Ivoclar. | | | |
| Storage Conditions | 48 months at 2-28 °C / 36-82 °F | | 48 months at 2-28 °C / 36-82 °F | |
510K Summary -3
{7}
510(K) SUMMARY
ivoclar
4
| **Summary of Storage Conditions** | No difference. | | |
| --- | --- | --- | --- |
| **Principles of Operation** | **Step-by-step:** - Shade determination - Isolation - Cavity preparation - Pulp protection/Base (3sCure mode is mentioned) - Apply matrix/interdental wedge - Conditioning/Application of the bonding agent - Application of Tetric PowerFill - Finishing/Checking the occlusion/Finishing | **Step-by-step:** - Determining the tooth shade - Preparing the cavity to the principles of the adhesive technique - Ensuring isolation - Placing the matrix - Lining the cavity close to the pulp - Conditioning the cavity - Applying the composite - Finishing/Occlusion control/Polishing | ☐ |
| **Summary Principles of operation** | No difference – The principle of operation is the same for both devices. | | |
| **Summary of Chemical Composition** | The main ingredients used for the predicate device Tetric PowerFill have also been used for the new device Tetric plus Fill. To achieve the advantage of the chameleon-effect, new ingredients have been used in the recipe of Tetric plus Fill. The biocompatibility has been thoroughly assessed for the new device Tetric plus Fill. This is shown in the Biological Evaluation Report for Tetric plus Fill (OTCS-Doc-ID 36406884). | | |
| **Finished Device Specification** | Applicable Standards: *ISO 4049:2019 – Dentistry – Polymer-based restorative materials* | Applicable Standards: *ISO 4049:2019 – Dentistry – Polymer-based restorative materials* | ☐ |
| | *Applicable FDA Guidance: Dental Composite Resin Devices – Premarket Notification [510(k) Submission]* | *Applicable FDA Guidance: Dental Composite Resin Devices – Premarket Notification [510(k) Submission]* | ☐ |
| **Summary of Finished Device Specification** | Both products fulfill the requirements of ISO 4049:2019 and the FDA Guidance “Composite Resin Devices”. Values for the new device are shown in the documents “Design Verification Matrix” (see OTCS-Doc-ID: 247717871 and LL252535447). The requirement depth of cure out of ISO 4049:2019 is ≥ 1.5 mm. Therefore, the Vickers hardness out of ISO 10477:2020 is considered to be the appropriate testing method to confirm higher claimed values. The differences in sensitivity to ambient light and radiopacity are not considered to be significantly different, as the requirements out of the relevant ISO are met. | | |
| **Sterilization** | Not applicable. No sterilization recommendation. | Not applicable. No sterilization recommendation. | ☐ |
| **Single use** | Consumable material | Consumable material | ☐ |
| **Summary of Performance Specification** | The device performance is tested according to ISO 4049 and the FDA Guidance “Dental Composite Resin Devices” and is substantially equivalent for the new device compared to its predicate. | | |
## Comparison Matrix 2:
| | Predicate Device: | Proposed Device: | Deviation | |
| --- | --- | --- | --- | --- |
| Device | Ivoclar Vivadent AG: Tetric PowerFill Fill (K183380) | Ivoclar Vivadent AG: Tetric plus Flow | Yes | No |
| **Indications for Use** | **Indications** **Conventional application (Light intensity 2,000 mW/cm2):** - Restorations in the posterior region (Class I and II, including the replacement of individual cusps) - Class V restorations (cervical caries, root erosion, wedge-shaped defects) - Reconstructive build-ups | **Indications** – Missing tooth structure in anterior and posterior teeth Types of restorations: – Initial layer up to 4 mm in Class I and II restorations* – Small direct restorations in permanent teeth (not occlusion-bearing)* – Restorations of deciduous teeth | ☑ | ☐ |
510K Summary -4
{8}
510(K) SUMMARY
ivoclar
5
| | - Restoration of deciduous teeth **Light-curing using the 3sCure mode of Bluephase PowerCure (Light Intensity 3,050 mW/cm^{2}) :** - Restorations in the posterior region of permanent dentition (Class I and II, including the replacement of individual cusps) when light-cured from the occlusal aspect | - Liner under direct restorations of all cavity classes - Class V restorations - Extended fissure sealings - Repair of composite / ceramic restorations - Sealing of implant screw channels *In case of occlusion-bearing restorations, apply a final layer using a sculptable composite material. | | |
| --- | --- | --- | --- | --- |
| **Precaution Measures/ Contraindications / Processing restrictions/ Side effects** | **Contraindications** - If a dry working field cannot be established, or the stipulated working technique cannot be applied - If a patient is known to be allergic to any of the ingredients of Tetric PowerFill | **Contraindications:** The use of this product is contraindicated if the patient is known to be allergic to any of its ingredients | ☑ | ☐ |
| **Summary of Indications, Precaution Measures/ Contraindications / Processing restrictions/ Side effects** | Tetric PowerFill and the new device Tetric plus Flow are both dental filling materials used by the dentist for the direct filling therapy in dentistry. The new device Tetric plus Flow can also be used for restoration in anterior teeth compared to the predicate. The indications of the new device Tetric plus Flow cover the indications of Tetric PowerFill. In addition, Tetric plus Flow can be used for extended fissure sealings, repair of composite/ceramic restorations and the closure of implant screw channels.. The contraindications are basically the same for both devices. The information regarding the dry working field can be found in section 'Limitations of use' for the new device. | | | |
| **Working Principle** | Tetric PowerFill is a dental composite material for the use of direct filling in teeth. | Tetric plus Flow is a dental composite material for the use of direct filling in teeth. | ☐ | ☑ |
| **Summary of Working Principle** | No difference | | | |
| **Delivery forms/dosage** | Tetric PowerFill is available as syringes and cavifills in 3 different shades (IVA, IVB, IVW) | Tetric plus Flow is available as syringes and cavifills in 4 different shades (A2 plus, A3 plus, A3.5 plus, Bleach plus) | ☑ | ☐ |
| **Summary of Delivery forms/dosage** | Tetric plus Flow is available in 4 different shades with which all classical vita shades can be covered (see table below). This innovation can be achieved due to Tetric plus Flow's advanced chameleon effect. This chameleon effect is enabled due to an adapted recipe (see section chemical composition) compared to the predicate device. | | | |
| **Shade correspondence** | | | | |
| Tetric® plus | A2 plus | A3 plus | A3.5 plus | Bleach plus |
| VITA® classical shades | A1, A2, B2, C1, D2 | A3, C2, C3, D3, D4 | A3.5, A4, B3, B4, C4 | B1, Bleach |
| * No registered trademark of Ivoclar. | | | | |
| **Storage Conditions** | 48 months at 2-28 °C / 36-82 °F | 48 months at 2-28 °C / 36-82 °F | ☐ | ☑ |
| **Summary of Storage Conditions** | No difference. | | | |
| **Principles of Operation** | **Step-by-step:** - Shade selection - Isolation - Cavity preparation - Pulp protection/Base (3sCure mode is mentioned) - Apply matrix/interdental wedge - Conditioning/Application of the bonding agent - Application of Tetric PowerFill | **Step-by-step:** - Determining the tooth shade - Preparing the cavity to the principles of the adhesive technique - Ensuring isolation - Placing the matrix - Lining the cavity close to the pulp - Conditioning the cavity - Applying the composite - Finishing/Occlusion control/Polishing | ☐ | ☑ |
510K Summary -5
{9}
510(K) SUMMARY
ivoclar
6
| | - Finishing/Checking the occlusion/ Finishing | | | |
| --- | --- | --- | --- | --- |
| **TSummary Principles of operation** | No difference – The principle of operation is the same for both devices. | | | |
| **Summary of Chemical Composition** | The main ingredients used for the predicate device Tetric PowerFill have also been used for the new device Tetric plus Flow. To achieve the advantage of the chameleon-effect, new ingredients have been used in the recipe of Tetric plus Flow. The biocompatibility has been thoroughly assessed for the new device Tetric plus Flow. This is shown in the Biological Evaluation Report for Tetric plus Flow (OTCS-Doc-ID 36395242). | | | |
| **Finished Device Specification** | Applicable Standards: *ISO 4049:2019 – Dentistry – Polymer-based restorative materials* | Applicable Standards: *ISO 4049:2019 – Dentistry – Polymer-based restorative materials* | ☐ | ☑ |
| | Applicable FDA Guidance: *Dental Composite Resin Devices – Premarket Notification [510(k) Submission]* | Applicable FDA Guidance: *Dental Composite Resin Devices – Premarket Notification [510(k) Submission]* | ☐ | ☑ |
| **Summary of Finished Device Specification** | Both products fulfill the requirements of ISO 4049:2019 and the FDA Guidance “Composite Resin Devices”. Values for the new device are shown in the documents “Design Verification Matrix” (see OTCS-Doc-ID 247714197 and LL252535447). The requirement depth of cure out of ISO 4049:2019 is ≥ 1.5 mm. Therefore, the Vickers hardness out of ISO 10477:2020 is considered to be the appropriate testing method to confirm higher claimed values. The differences in the claimed values sensitivity to ambient light are not considered to be significantly different, as the requirements out of the relevant ISO are met. | | | |
| **Sterilization** | Not applicable. No sterilization recommendation. | Not applicable. No sterilization recommendation. | ☐ | ☑ |
| **Single use** | Consumable material | Consumable material | ☐ | ☑ |
| **Summary of Performance Specification** | The device performance is tested according to ISO 4049 and the FDA Guidance “Dental Composite Resin Devices” and is substantially equivalent for the new device compared to its predicate. | | | |
### Substantial Equivalence to the predicate:
The indications of the new device **Tetric plus Fill** cover the indications of Tetric PowerFill, deviations have been explained. The working principle is basically the same for both devices. The majority of the ingredients are the same as used in the recipe of the predicate. The biocompatibility of the new formulation was fully assessed and is equivalent to Tetric PowerFill. The device performance is tested according to ISO 4049 and the FDA Guidance “Dental Composite Resin Devices” and is substantially equivalent for the new device compared to its predicate.
Therefore, **Tetric plus Fill** is substantially equivalent to the predicate device.
The indications of the new device **Tetric plus Flow** cover the indications of Tetric PowerFill, deviations have been explained. The working principle is the same for both devices. The majority of the ingredients are the same as used in the recipe of the predicate. The biocompatibility of the new formulation was fully assessed and is equivalent to Tetric PowerFill. The device performance is tested according to ISO 4049 and the FDA Guidance “Dental Composite Resin Devices” and is substantially equivalent for the new device compared to its predicate.
Therefore, **Tetric plus Flow** is substantially equivalent to the predicate device.
510K Summary -6
{10}
510(K) SUMMARY
ivoclar
7
# **Differences:**
# **Tetric plus Fill**
Indication differs in that Tetric plus Fill can be used for direct veneers, for the repair of composite/ceramic restorations and the closure of implant screw channels. The application of the 3sCure mode of Bluephase PowerCure is described in the instructions for use for the new device under section "Applying the composite resin".
Contraindications differ in that the information regarding the dry working field can be found in section "Limitations of use" for the new device.
Delivery forms/dosage differ in the number of shades and shade names. Tetric plus Fill is available in 4 different shades, which cover all classical vita shades as a result of the adapted recipe compared to the predicate.
The Chemical Composition in Tetric plus Fill uses the same main ingredients as the predicate, but new ingredients have been used in the recipe of Tetric plus Fill. The biocompatibility has been thoroughly assessed for the new device Tetric plus Fill.
There are several performance requirements that differ in the device specification- Sensitivity to ambient light, Depth of cure, Depth of cure by Vickers Hardness, and Radiopacity. The requirement depth of cure per ISO 4049:2019 is $\geq 1.5$ mm. Therefore, the Vickers hardness out of ISO 10477:2020 is considered to be the appropriate testing method to confirm higher claimed values.
The differences in sensitivity to ambient light and radiopacity are not considered to be significantly different, as the relevant ISO requirements are met.
# **Tetric plus Flow**
Indication differs in that Tetric plus Flow can be used for extended fissure sealings, repair of composite/ceramic restorations and the closure of implant screw channels.
Contraindications differ in that the information regarding the dry working field can be found in section "Limitations of use" for the new device.
Delivery forms/dosage differ in the number of shades and shade names. Tetric plus Flow is available in 4 different shades, which cover all classical vita shades as a result of the adapted recipe compared to the predicate.
The Chemical Composition in Tetric plus Flow uses the same main ingredients as the predicate, but new ingredients have been used in the recipe of Tetric plus Flow. The biocompatibility has been thoroughly assessed for the new device Tetric plus Flow.
There are several performance requirements that differ in the device specification- Sensitivity to ambient light, Depth of cure, and Depth of cure by Vickers Hardness. The requirement depth of cure per ISO 4049:2019 is $\geq 1.5$ mm. Therefore, the Vickers hardness out of ISO 10477:2020 is considered to be the appropriate testing method to confirm higher claimed values.
510K Summary -7
{11}
510(K) SUMMARY
ivoclar
8
# **Non-clinical performance testing:**
Bench testing was performed to test the physical properties included in the Finished Device Specification for the subject device including: Sensitivity to ambient light, Depth of cure, Depth of cure by Vickers Hardness, Flexural strength, Water sorption, Solubility, Radiopacity, and Wavelength for Polymerization according to EN ISO 4049:2019 Dentistry- Polymer-based restorative materials and the FDA Guidance "Composite Resin Devices".
# **Biocompatibility:**
The subject devices were evaluated for Biocompatibility according to ISO 10993, ISO 7405, ISO 21726:2019 and ISO 14971:2019. The biological evaluation was performed based on toxicological data on relevant component materials / compounds, information on prior use of relevant component materials / compounds, data from biological tests, data on the history of clinical use or human exposure. Based on the information included in the biological evaluation reports it can be concluded that the products under evaluation are acceptable in relation to its clinical benefit.
The product under assessment **Tetric plus Fill** has no significant cytotoxic potential, holds no potential for material-mediated pyrogenicity, presents no risks for acute, sub-acute, sub-chronic or chronic toxicity, is not an inducer of adverse implantation effects, is not genotoxic, is not carcinogenic, and is not causing reproductive and development toxicity.
The product under assessment **Tetric plus Flow** is not cytotoxic, holds no potential for material-mediated pyrogenicity, presents no risks for acute, sub-acute, sub-chronic toxicity, is not an inducer of adverse implantation effects, is not genotoxic, not carcinogenic, and is not causing reproductive and developmental toxicity.
While Tetric plus Fill and Tetric plus Flow can cause irritation and sensitization mostly likely in its unpolymerized state. If proper safety measures (according to the IFU) during handling and use are employed, health risks can be avoided.
The products do not represent a toxicological risk for the patient and the user.
On the basis of the toxicological evaluation of the products and the longstanding worldwide clinical use of similar materials it can be concluded that the benefits provided by the final product will exceed any potential risks produced by device materials providing that the instructions for use have been followed.
# **Conclusion:**
Tetric PowerFill and the new device **Tetric plus Fill** are both dental filling materials used by the dentist for the direct filling therapy in teeth. The advantage of using the new device Tetric plus Fill lies in the color-feature of the new device: Tetric plus Fill is available in 4 shades. However, all classical vita shades can be covered due to its chameleon effect. This effect is achieved due to an adaptation of the recipe compared to the predicate device. The physical properties of the new device are according to ISO 4049:2019 and the FDA Guidance "Dental Composite Resin" and can be considered as substantially equivalent to Tetric PowerFill.
Therefore, **Tetric plus Fill** is substantially equivalent to the predicate device, Tetric PowerFill.
510K Summary -8
{12}
510(K) SUMMARY
ivoclar
9
Tetric PowerFill and the new device **Tetric plus Flow** are both dental filling materials used by the dentist for the direct filling therapy in teeth. The advantage of using the new device Tetric plus Fill lies in the color-feature of the new device: Tetric plus Flow is available in 4 shades, however, all classical vita shades can be covered due to its chameleon effect. These two effects are achieved due to an adaptation of the recipe compared to the predicate device.
The physical properties of the new device are according to ISO 4049:2019 and the FDA Guidance “Dental Composite Resin” and can be considered as substantially equivalent to Tetric PowerFill.
Therefore, **Tetric plus Flow** is substantially equivalent to the predicate device, Tetric PowerFill.
510K Summary -9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.