The Indira Cloud Viewer is a visualization software intended to process and display image and associated data in a clinical setting. The software displays images of different modalities and timepoints, and performs digital image processing, measurement, manipulation, quantification and communication. The Indira Cloud viewer is a web-based application and does not perform automated diagnosis and is not intended to replace clinical judgment.
Device Story
Indira Cloud Viewer is a web-based, zero-footprint SaMD for visualization, processing, and manipulation of DICOM medical images. It enables authorized healthcare professionals (radiologists, technicians) to access and review studies stored in secure cloud infrastructure. Input consists of multi-modality DICOM images; output is displayed on a web interface for clinical review. The device provides tools for 2D/3D viewing, image manipulation, measurement, segmentation, and cine playback. It does not perform automated diagnosis, CAD, or algorithmic interpretation. It is used in clinical environments to support physician decision-making by facilitating image analysis and communication. The cloud-based architecture eliminates the need for local software installation, streamlining access to imaging data across clinical workflows.
Clinical Evidence
Bench testing only. Verification and validation activities were conducted per IEC 62304, ISO 14971, and IEC 62366-1. Testing included unit, integration, and system-level verification, alongside performance and regression testing to ensure software functionality and usability in clinical scenarios.
Technological Characteristics
Web-based SaMD; zero-footprint architecture. Supports DICOM standard. Features 2D/3D visualization, measurement (line, ROI, angle), segmentation (CT/MR), and cine viewing. Cloud-hosted deployment. Cybersecurity controls per ISO/IEC 27001, AAMI TIR57, and AAMI TIR97. Software lifecycle per IEC 62304; usability per IEC 62366-1.
Indications for Use
Indicated for use by trained healthcare professionals, including radiologists and radiology technicians, for the processing, display, and analysis of multi-modality medical images (CT, CBCT, Spectral CT, MR, EMR, NM, PET, SPECT, US, XA, DX, CR, RF, ECG) in clinical settings. Not for use in mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 6, 2026
Indigo Technologies
Andres Felipe Sanabria Diaz
Head of Product
Carrera 7 155 C 30 To E Of 3205 ED North Point
Bogota,
Colombia
Re: K261465
Trade/Device Name: Indira Cloud Viewer
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: May 2, 2026
Received: May 4, 2026
Dear Andres Felipe Sanabria Diaz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261465 - Andres Felipe Sanabria Diaz
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261465 - Andres Felipe Sanabria Diaz
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Assistant Director, Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261465
Device Name
Indira Cloud Viewer
Indications for Use (Describe)
The Indira Cloud Viewer is a visualization software intended to process and display image and associated data in a clinical setting.
The software displays images of different modalities and timepoints, and performs digital image processing, measurement, manipulation, quantification and communication. The Indira Cloud viewer is a web-based application and does not perform automated diagnosis and is not intended to replace clinical judgment.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# K261465
# 510(K) SUMMARY
# INDIRA CLOUD VIEWER
| Company's Name and Address | INDIGO TECHNOLOGIES SAS CARRERA 7 155 C 30 TO E OF 3205 ED NORTH POINT BOGOTA COLOMBIA |
| --- | --- |
| Contact Person | Andres Felipe Sanabria Diaz Head of Product +57 6013849900 calidad@indigo.tech |
| Date | 02 May 2026 |
| Device Trade name | INDIRA CLOUD VIEWER |
| Classification Name | System, image processing, radiological |
| Product Code | LLZ |
| Classification | 21 CFR 892.2050 |
| Predicate Device | Philips Medical Systems' AV Viewer (K250181) |
# Device description
Indira Cloud Viewer is a Software as a Medical Device (SaMD) developed for the visualization, processing, and manipulation of medical images in Digital Imaging and Communications in Medicine (DICOM) format.
The software enables authorized healthcare professionals to securely access and review medical imaging studies through a web-based, zero-footprint viewer without requiring local software installation. The system allows users to display and interact with DICOM images stored in secure cloud infrastructure.
Indira Cloud Viewer provides visualization and image manipulation tools that support clinical review of medical images by qualified healthcare professionals. The software does not perform automated diagnosis, computer-aided detection (CAD), or algorithmic image interpretation.
# Indications for use:
The Indira Cloud Viewer is a visualization software intended to process and display image and associated data in a clinical setting.
The software displays images of different modalities and timepoints, and performs digital image processing, measurement, manipulation, quantification and communication. The Indira Cloud viewer is a web-based application and does not perform automated diagnosis and is not intended to replace clinical judgment
{5}
## Verification and validation:
The verification and validation activities for the Indira Cloud Viewer were conducted in accordance with a comprehensive quality and regulatory framework aligned with internationally recognized standards.
Software lifecycle processes followed IEC 62304:2006 + AMD1:2015, including software development planning, requirements definition, architectural design, implementation, and systematic verification and validation. Traceability between requirements, risk controls, and test cases was maintained throughout the lifecycle.
Risk management activities were performed in accordance with ISO 14971:2019. Hazards, including those related to software functionality, usability, and cybersecurity, were identified and evaluated. Risk control measures were implemented and verified to ensure that residual risks are acceptable and that the benefit-risk profile supports safe and effective use.
Information security considerations were addressed in alignment with ISO/IEC 27001:2022, ensuring appropriate controls for data protection, access management, and system integrity within the cloud-based environment.
Cybersecurity principles were applied following AAMI TIR57 and AAMI TIR97, covering both premarket secure design and postmarket monitoring, vulnerability management, and incident response processes.
Usability engineering activities were conducted in accordance with IEC 62366-1:2015 + AMD1:2020 to minimize use-related risks. The user interface and workflows were evaluated to ensure safe and effective interaction by intended users in the intended use environment.
Verification activities included unit, integration, and system-level testing, as well as performance and regression testing. Validation activities confirmed that the software meets user needs and intended use in representative clinical scenarios.
{6}
| Comparison feature | Subject Device Indira Cloud Viewer | Predicate device Philips Medical Systems' AV Viewer K250181 | Comparison |
| --- | --- | --- | --- |
| Device Class | II | II | Equivalent |
| Classification panel | Radiology | Radiology | Equivalent |
| Product code | LLZ | LLZ | Equivalent |
| Regulatory Description | Medical image management and processing system | Medical image management and processing system | Equivalent |
| Regulation Number | 892.2050 | 892.2050 | Equivalent |
| Indications for Use | The Indira Cloud Viewer is a visualization software intended to process and display image and associated data in a clinical setting. The software displays images of different modalities and timepoints, and performs digital image processing, measurement, manipulation, quantification and communication. The Indira Cloud Viewer is not to be used for mammography. The Indira Cloud viewer is a web-based application and does not perform automated diagnosis and is not intended to replace clinical judgment' | The AV Viewer is an advanced visualization software intended to process and display image and associated data in a clinical setting. The software displays images of different modalities and timepoints, and performs digital image processing, measurement, manipulation, quantification and communication. The AV Viewer is not to be used for mammography. | Equivalent |
| Clinical characteristics | | | |
| Intended users | Trained professionals: radiologists and radiology technicians | Trained professionals, including but not limited to, physicians and medical technicians | Similar |
| Supported imaging modalities | CT, CBCT –CT format, Spectral CT, MR, EMR, NM, PET, SPECT, US, XA, DX, CR, RF, ECG. | CT, CBCT –CT format, Spectral CT, MR, EMR, NM, PET, SPECT, US, XA, DX, CR, RF, ECG. | Equivalent |
| Technological features | | | |
| 2D and 3D viewing capabilities | Yes | Yes | Equivalent |
{7}
| 2D measurements (line, ROI, angle etc.) | Yes | Yes | Equivalent |
| --- | --- | --- | --- |
| 3D Segmentation on CT and MR datasets | Yes | Yes | Equivalent |
| Cine View – sequence of images as a movie | Yes | Yes | Equivalent |
| Finding dashboard | Yes | Yes | Equivalent |
| Reporting | No | Yes | Different |
| Exporting reports and derived data | Yes | Yes | Equivalent |
| Cloud based application | Yes | It has the possibility of a cloud base application | Similar |
## Substantial Equivalence
The Indira Cloud Viewer has the same intended use as the predicate device from Philips Medical Systems (AV Viewer, K250181). Both are medical image visualization software intended for use by trained healthcare professionals to process, display, and analyze multi-modality medical images in clinical environments, without performing automated diagnosis or replacing clinical judgment.
Technologically, both devices share the same core functionalities, including 2D/3D visualization, image manipulation, measurement, segmentation, and communication tools. The subject device is cloud-based, while the predicate can also be deployed in a cloud environment; this difference relates only to deployment and does not impact performance or safety. The absence of reporting functionality in the subject device does not affect the intended use and does not introduce new risks. All identified differences have been assessed through risk management and supported by verification and validation (V&V) activities.
Clinically, both devices are used by similar user groups (e.g., radiologists and physicians), support the same imaging modalities (CT, MR, PET, SPECT, ultrasound, and X-ray), and are integrated into comparable clinical workflows. There are no differences that impact clinical performance, safety, or effectiveness.
## Conclusion on Substantial Equivalence:
The Indira Cloud Viewer is substantially equivalent to the predicate device from Philips Medical Systems, as it has the same intended use, similar technological characteristics, and no differences that raise new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.