CVAC® Aspiration System

K261447 · Calyxo, Inc. · FGB · Sep 15, 2026 · Gastroenterology, Urology

Device Facts

Record IDK261447
Device NameCVAC® Aspiration System
ApplicantCalyxo, Inc.
Product CodeFGB · Gastroenterology, Urology
Decision DateSep 15, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The CVAC Aspiration System and CVAC Image Processor consists of a sterile single use, steerable ureteral catheter and a reusable software-controlled image processor. It is intended to establish a conduit during endoscopic urological procedures for the treatment and removal of urinary stones (e.g., kidney stones, fragments, and dust). It employs flexible ureteroscopy within the urinary tract for endoscopic examination of the urinary tract and the interior of the kidney. The steerable ureteral catheter is used for irrigation and aspiration of kidney stones and stone dust with dedicated irrigation and vacuum channels under ureteroscopy. The CVAC Aspiration System and CVAC Image Processor can be used with additional accessories to perform various diagnostic and therapeutic procedures.

Device Story

System comprises sterile single-use steerable ureteral catheter and reusable software-controlled image processor; used in surgical suites by physicians. Catheter features integrated irrigation and vacuum channels; distal tip includes CMOS chip and dual LEDs for visualization. Device establishes conduit for stone removal; physician navigates ureteroscope to kidney; stone fragments/dust aspirated via vacuum channel. System connects to standard hospital monitors, vacuum, and irrigation sources. Modifications in subject device (CVAC 2.1) include re-routed outflow pathway to improve passive aspiration rates. Output is real-time endoscopic video displayed on monitor; aids physician in stone localization and removal. Benefits include improved stone clearance efficiency during urological procedures.

Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included design verification (dimensional, flow, and intra-renal pressure tests) and design validation to confirm that the modified outflow path and handle geometry meet customer requirements and maintain safety/effectiveness compared to the predicate.

Technological Characteristics

Single-use steerable ureteral catheter; reusable software-controlled image processor. Features: recessed CMOS chip, dual LEDs, 180-degree tip angulation. Materials: sterile, EtO sterilized (SAL 10^-6). Connectivity: video cable to image processor. Dimensions: 70 cm length, <12 Fr shaft OD, 7 Fr working channel ID. Software-controlled image processing. Passive aspiration via irrigation-driven pressure, enhanced by re-routed outflow path.

Indications for Use

Indicated for patients undergoing endoscopic urological procedures for the treatment and removal of urinary stones (e.g., kidney stones, fragments, and dust) and endoscopic examination of the urinary tract and kidney interior. Applicable to all age groups including neonates, infants, children, adolescents, and adults.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} ![LOGO](U.S. Food & Drug Administration) September 15, 2026 Calyxo, Inc. Nitya Narayanan Director, Regulatory Affairs 4473 Willow Rd., Suite 100 Pleasanton, California 94588 Re: K261447 Trade/Device Name: CVAC® Aspiration System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FGB Dated: August 21, 2026 Received: August 21, 2026 Dear Nitya Narayanan: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K261447 - Nitya Narayanan Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2} K261447 - Nitya Narayanan Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark R. Kreitz -S for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261447 | ? | | Please provide the device trade name(s). | | ? | | CVAC® Aspiration System | | | | Please provide your Indications for Use below. | | ? | | The CVAC Aspiration System and CVAC Image Processor consists of a sterile single use, steerable ureteral catheter and a reusable software-controlled image processor. It is intended to establish a conduit during endoscopic urological procedures for the treatment and removal of urinary stones (e.g., kidney stones, fragments, and dust). It employs flexible ureteroscopy within the urinary tract for endoscopic examination of the urinary tract and the interior of the kidney.The steerable ureteral catheter is used for irrigation and aspiration of kidney stones and stone dust with dedicated irrigation and vacuum channels under ureteroscopy. The CVAC Aspiration System and CVAC Image Processor can be used with additional accessories to perform various diagnostic and therapeutic procedures. | | | | Please select the types of uses (select one or both, as applicable). | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? | | Please select the age group(s) for which the device(s) is to be used. | Neonates/Newborns (Birth to < 29 days old)Infants (29 days old to < 2 years old)Children (2 years old to < 12 years old)Adolescents (12 years old to < 22 years old)Adults (22 years old and greater) | ? | {4} [LOGO] Calyxo. K261447 Page 1 of 4 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS # 1. GENERAL INFORMATION # 1.1 Submitter and 510(k) Owner Calyxo, Inc. 4473 Willow Road, Suite 100 Pleasanton, CA 94588 # 1.2 Official Correspondent Ms. Nitya Narayanan 4473 Willow Road, Suite 100 Pleasanton CA 94588 Telephone: (925) 526-5900 Email: nitya.narayanan@calyxoinc.com # 1.3 Date of Preparation August 21, 2026 # 2. NAME OF THE DEVICE # 2.1.1 Trade/Proprietary Name CVAC® Aspiration System # 2.1.2 Common/Usual Name Ureteroscope and Accessories, Flexible/Rigid # 2.1.3 Classification Information Trade/Proprietary Name: CVAC® Aspiration System Classification Name: Endoscope and Accessories Classification Regulation: 21 CFR 876.1500 Class: II Product Codes: FGB, Ureteroscope and Accessories, Flexible/Rigid Panel: Gastroenterology / Urology # 3. PREDICATE DEVICE The predicate device is the CVAC Aspiration System, which is part of the CVAC Set cleared under K233472. The reference device is the CVAC Image Processor, cleared under K260013. # 4. DESCRIPTION OF THE DEVICE The single use CVAC Aspiration System and the reusable CVAC Image Processor are designed to work together. The devices allow physicians to locate kidney stones and to establish a conduit during endoscopic urological procedures for the treatment and removal of urinary stones (e.g., kidney stones, 1 {5} K261447 Page 2 of 4 fragments, and dust). The devices are used in a surgical suite and are compatible with standard hospital monitors, vacuum sources, and irrigation sources. The CVAC Aspiration System is a sterile single-use steerable ureteral catheter employing ureteroscopy with built-in irrigation and vacuum functionality. Accessories provided with the CVAC Aspiration System include an introducer, a laser bridge, a trigger tie, a 3-way stopcock, and a vacuum port cap. The CVAC Aspiration System catheter is compatible with commercially available accessories for stone removal such as appropriately sized single-use laser fibers and stone baskets, which are inserted and used in the vacuum channel. The CVAC Aspiration System connects to the CVAC Image Processor via an attached video cable. There are no changes to the CVAC Image Processor in this 510(k). ## **5. INTENDED USE** The intended use / indications for use for the CVAC Aspiration System and CVAC Image Processor is as follows: The CVAC Aspiration System and CVAC Image Processor consists of a sterile single use, steerable ureteral catheter and a reusable software-controlled image processor. It is intended to establish a conduit during endoscopic urological procedures for the treatment and removal of urinary stones (e.g., kidney stones, fragments, and dust). It employs flexible ureteroscopy within the urinary tract for endoscopic examination of the urinary tract and the interior of the kidney. The steerable ureteral catheter is used for irrigation and aspiration of kidney stones and stone dust with dedicated irrigation and vacuum channels under ureteroscopy. The CVAC Aspiration System and CVAC Image Processor can be used with additional accessories to perform various diagnostic and therapeutic procedures. ## **6. INTENDED USE COMPARED TO THE PREDICATE** The intended use and indications for use statement is unchanged from the predicate device cleared under K233472 except for the addition of “e.g.” for clarification purpose. ## **7. TECHNOLOGY CHARACTERISTICS COMPARED TO THE PREDICATE** The outflow pathway of the CVAC Aspiration System was re-routed to improve passive aspiration resulting in a higher passive outflow rate. The different technological characteristics between the subject device and the predicate has been evaluated and was demonstrated not to raise new questions of safety and effectiveness. ---2 {6} K261447 Page 3 of 4 **Table 1. Overview of Technology Comparison of Subject and Predicate Devices** | | CVAC Aspiration System (CVAC 2.0) Predicate Device (K233472) | CVAC Aspiration System (CVAC 2.1) Subject Device | Same or Different? | | --- | --- | --- | --- | | Dimensions | Dimensions: 70 cm Total Length: 100 cm Shaft OD <12 Fr (4 mm) Working Channel ID 7 Fr (2.3 mm) UAS compatibility 12 Fr Inner Diameter or larger | | Same | | Navigation | Ureteroscopic navigation | | Same | | Distal Tip Configuration (Camera) | Recessed CMOS chip and dual LEDs for direct vision capability | | Same | | Steerable technology Angulation range | Yes; Distal tip (7 cm) moves up and down, reaches all calyces 180° up and down | | Same Same | | Stone Collector | Integrated into handle | | Same | | Accessories | Five accessories provided in the CVAC Aspiration System box: trigger tie, a three-way stopcock, working channel cap, laser bridge and introducer | | Same | | Image System | The CVAC Image Processor is connected to the CVAC Aspiration System and provides power to the ureteroscope | | Same | | Single or multiple use | Single patient use | | Same | | Sterilization | EtO, Sterilization Assurance Level 10^{-6} | | Same | | Shelf Life | 1 year | | Same | | Outflow Path – Passive Mode | Passive aspiration relies primarily on irrigation-driven pressure. | Passive aspiration is enhanced by modification of outflow path. | Different. CVAC 2.1 improves passive aspiration while maintaining the same operating principle; difference does not raise new questions of safety or effectiveness. (Performance Testing – Bench) | | Handle Components | Original handle components configuration supporting outflow path. | Updated handle component geometry and interfaces to support revised outflow path routing. | Different. Component changes support internal flow improvement and do not change user interaction or clinical workflow. Design Verification and Design Validation (Performance Testing – Bench) | 3 {7} K261447 Page 4 of 4 # 8. PERFORMANCE TESTING The 510(k) submission provides performance data to establish the substantial equivalence of the CVAC 2.1 Aspiration System (the subject device) to CVAC 2.0 Aspiration System (predicate device, K233472). A summary of these performance tests is provided below. Design Verification Testing: Verification (dimensional, flow and intra renal pressure tests) and validation activities demonstrated that all customer requirements passed validation and all impacted product requirements passed verification. Supplemental Bench Testing: Supplemental tests were performed to further characterize the intrarenal pressure performance of the CVAC 2.1 Aspiration System under various operating conditions. Design Validation: Design validation activities demonstrated that all customer requirements passed validation and all impacted product requirements passed verification. Leveraged Data: The following data were leveraged from the predicate or reference device submission as the device modifications were demonstrated not to impact the results of this testing: - Software and Cybersecurity - Sterilization and Shelf Life - Electrical Safety and EMC # 9. CONCLUSIONS This 510(k) submission demonstrates that the CVAC Aspiration System is substantially equivalent to the predicate device cleared under K233472. 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...