K261415 · Guilin Kevin Peter Technology Co., Ltd. · KIF · Jun 29, 2026 · Dental
Device Facts
Record ID
K261415
Device Name
Injectable Root Canal Sealer (KP-Root SP)
Applicant
Guilin Kevin Peter Technology Co., Ltd.
Product Code
KIF · Dental
Decision Date
Jun 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Permanent obturation of the root canal following vital pulp-extirpation. Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings. KP-Root SP is suitable for use in the single cone and lateral condensation technique.
Device Story
KP-Root SP is a premixed, ready-to-use, injectable bioceramic paste for root canal sealing and permanent filling. Material requires water to set and harden; radiopaque. Supplied in preloaded syringes (0.5g, 1g, 2g, 3g). Used by dental professionals in clinical settings for endodontic procedures. Applied via syringe into the root canal; provides a tight seal. Benefits include convenience of premixed delivery and biocompatible sealing properties for permanent obturation.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including ISO 6876:2012 performance standards (flow, setting time, film thickness, solubility, radiopacity, pH, strength) and comprehensive biocompatibility testing (ISO 10993 series: cytotoxicity, sensitization, systemic toxicity, pyrogenicity, genotoxicity, and implantation effects).
Technological Characteristics
Bioceramic paste containing zirconium oxide, strontium-doped calcium silicates, calcium hydroxide, and calcium phosphate monobasic. Premixed, ready-to-use, injectable form factor. Radiopaque. Sets via hydration. Conforms to ISO 6876:2012. Non-sterile.
Indications for Use
Indicated for permanent root canal obturation following vital pulp-extirpation or removal of infected/necrotic pulp with intracanal dressings. Suitable for single cone and lateral condensation techniques.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 29, 2026
Guilin Kevin Peter Technology Co., Ltd.
% Boyle Wang
General Manager
ABMED Medical Technology PTY, Ltd.
48/6 Herbert St.
St Leonards, NSW 2065
Australia
Re: K261415
Trade/Device Name: Injectable Root Canal Sealer (KP-Root SP)
Regulation Number: 21 CFR 872.3820
Regulation Name: Root Canal Filling Resin
Regulatory Class: Class II
Product Code: KIF
Dated: April 30, 2026
Received: April 30, 2026
Dear Boyle Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261415 - Boyle Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261415 - Boyle Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FDA Bobak
Shirmohammadi -
S
For Michael E. Adjodha, M.ChE., RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261415 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Injectable Root Canal Sealer (KP-Root SP) | | |
| Please provide your Indications for Use below. | | ? |
| • Permanent obturation of the root canal following vital pulp-extirpation. • Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings. KP-Root SP is suitable for use in the single cone and lateral condensation technique. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary---
# 510(k) Summary
## K261415
This summary of 510(k) substantial equivalence is being submitted in accordance with requirements of 21 CFR 807.92.
### 1.0 Submitter’s Information
Name: Guilin Kevin Peter Technology Co., Ltd.
Address: North Side of Linsu Rd, Lingui Town, Lingui Dist., Guilin, Guangxi, 541100, P. R. China
Tel: 86- 773-2351806
Contact: Zhang Qian
**Registration Number:** 3023816720
### Designated Submission Correspondent
Contact: Mr. Boyle Wang
Company Name: ABMED MEDICAL TECHNOLOGY PTY LTD.
Address: 48/6 Herbert Street, St Leonards NSW 2065, Australia
Telephone: + 61 432 758 222
Email: info@abmed.com.au
Date of Preparation: June 27, 2026
### 2.0 Device Information
Trade Name: KP-Root SP
Device Name: Injectable Root Canal Sealer
Common Name: Root Canal Sealer
Classification Name: Resin, Root Canal Filling
Model: KP-Root SP
### 3.0 Classification
Production code: KIF
Regulation number: 21 CFR 872.3820
Classification: Class II
Panel: Dental
### 4.0 Identification of Predicate Device and Reference Device
1 of page 6
{5}
510(k) Summary
Predicate Device:
510(k) Number: K080917
Product Name: iRoot SP
Manufacturer: Innovative BioCeramix, Inc.
This predicate device has not been subject to a design-related recall.
### 5.0 Indication for Use Statement
- Permanent obturation of the root canal following vital pulp-extirpation.
- Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings.
KP-Root SP is suitable for use in the single cone and lateral condensation technique.
### 6.0 Device Description
KP-Root SP is a premixed ready-to-use, injectable bioceramic paste, which is used for tight sealing and permanent filling of root canal. KP-Root SP is a radiopaque material, which requires the presence of water to set and harden. KP-Root SP is packaged in a preloaded syringe and is very convenient to use.
KP-Root SP is available in two preloaded syringe mode that provide different in specification in 0.5g,1g, 2g,3g; The only difference between the types is the net weight.
### 7.0 Summary of Non-Clinical Testing
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the Technical Requirements of Injectable Root Canal Sealer issued by Guilin Kevin Peter Technology Co., Ltd. and the standards ISO 6876:2012 Dentistry - Root canal sealing materials, including:
- Composition and appearance
- Packing volume
- Flow
- Setting time
- Film thickness
- Solubility and disintegration
- Dimensional change after setting
- Radiopacity
- pH value
2 of page 6
{6}
510(k) Summary
All test results demonstrated that the subject device meets the requirements of ISO 6876 and the product technical specifications.
# **Biocompatibility testing**
- Ames Test - ISO 10993-3:2014 Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity.
- In vitro mammalian chromosomal aberration test - ISO 10993-3:2014 Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity.
- In Vitro Cytotoxicity Test - ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- Sensitization - ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- Acute systemic Toxicity Test - ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- Pyrogenicity Test- ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- Subchronic systemic Toxicity Test - ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- Implantation effects - ISO 10993-6:2016 Biological evaluation of the medical devices – Part 6: Tests for Local Effects after Implantation.
# **8.0 Summary of Clinical Testing**
Clinical testing was not required for this submission.
# **9.0 Technological Characteristics and Substantial Equivalence**
The following table shows similarities and differences of use, design, and material between our device and the predicate devices.
Table 1 General Device Characteristics Comparison Table
3 of page 6
{7}
510(k) Summary
| Item | Subject device K261415 | Predicate device K080917 | Remark |
| --- | --- | --- | --- |
| Product Name | KP-Root SP | iRoot SP | -- |
| Product Code | KIF | KIF | Same |
| Regulation No. | 21 CFR 872.3820 | 21 CFR 872.3820 | Same |
| Class | II | II | Same |
| Intended Use | - Permanent obturation of the root canal following vital pulp-extirpation. - Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings. KP-Root SP is suitable for use in the single cone and lateral condensation technique. | - * Permanent obturation of the root canal following vital pulp-extirpation. - * Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings. iRoot SP is suitable for use in the single cone and lateral condensation technique. | Same |
| Prescription Use | Yes | Yes | Same |
| Basic Chemical Composition | Zirconium oxide, Strontium doped calcium silicates, calcium hydroxide, calcium phosphate monobasic and Filler agents. | Zirconium Oxide, Calcium Silicates, Calcium Hydroxide, Calcium Phosphate Monobasic and Filler Agents | Analysis 1 |
| Performance Standard Conformance | Conformed to ISO 6876 and Technical Requirements of Injectable Root Canal Sealer | Conformed to ISO 6876 | Same |
| Flow | Average Min.value. 26.66 | Average Min.value 26.89 | Analysis 2 |
| Setting time | 9h | 10.5h | Analysis 2 |
| Dimensional change following setting | 0.04 in shrinkage | 0.06 in shrinkage | Analysis 2 |
| Film thickness | Average 21.3 | Average 19.3 | Analysis 2 |
| Solubility and disintegration | Solubility: Average 0.67% Disintegration: Compliance requirement | Solubility: Average 0.93% Disintegration: Compliance requirement | |
4 of page 6
{8}
510(k) Summary
| Radio-opacity | Compliance requirement. Equivalent to the thickness of 6 mm aluminum plate | Compliance requirement. Equivalent to the thickness of 6 mm aluminum plate (maybe affected by bubbles) | Analysis 2 |
| --- | --- | --- | --- |
| pH | Average 12.66 | Average 12.62 |
| Microbial Limit | Aerobic Plate Count: < 10 cfu/g Yeast and Mold: <10 cfu/g Escherichia ciol:ND Staphylococcus aureus:ND Pseudomonas Aeruginosa:ND | Aerobic Plate Count: <10 cfu/g Yeast and Mold: <10 cfu/g Escherichia ciol:ND Staphylococcus aureus:ND Pseudomonas Aeruginosa:ND |
| Acid Soluble Arsenic Content | Acid soluble arsenic < 2mg/kg Acid <100mg/kg | Acid soluble arsenic < 2mg/kg Acid <100mg/kg |
| Flexural strength | (6.24 ± 0.13) MPa | (6.46 ± 0.13) MPa |
| Compressive strength | (49.8 ± 0.6)MPa | (49.2 ± 1.3) MPa |
| Root Canal Sealing Ability | There was no significant difference in root canal sealing between the subject device and the predicate device. |
| Treatment Site | Root canal following vital pulp-extirpation | Root canal following vital pulp-extirpation | Same |
| Sterile | Non-sterile | Non-sterile | Same |
| Biocompatibility | Comply with ISO 10993-1:2018, FDA Guidance, tests included cytotoxicity, sensitization, pyrogenicity, acute systemic toxicity, subchronic systemic toxicity test, implantation effect , Ames test and In vitro mammalian chromosomal aberration test. | Comply with ISO 10993-1, FDA Guidance | Same |
| Label and Labeling | Conforms to FDA Regulatory Requirements | Conforms to FDA Regulatory Requirements | Same |
# Analysis 1:
iRoot SP Root Canal Sealer was identified as the predicate device due to the subject device having similar materials and delivery form (i.e. premixed ready-to-use injectable paste) to the predicate. Analysis and Research Report on the chemical
5 of page 6
{9}
510(k) Summary---
composition of the predicate device iRoot SP Root Canal Sealer to provide reference for the formulation design of KP-Root SP.
Also, the biocompatibility testing completed in alignment ISO 10993-1 and ISO 7405:2018 respectively demonstrate that any material differences between the subject device and predicate device do not raise any new questions as to safety and effectiveness.
# **Analysis 2:**
The performance technological parameters are little different with those of the predicate. Performance comparison testing was conducted to the predicate device and current subject device to compare their performance according to the Technical Requirements of “Injectable Root Canal Sealer”, which is identical to ISO 6876:2012, and the comparison testing results shown both the subject device and the predicate device are all complied with Technical Requirements of “Injectable Root Canal Sealer” issued by Guilin Kevin Peter Technology Co., Ltd.
Therefore, it is concluded that the subject device is substantially equivalent to the predicate device.
## 10.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the differences between the proposed device and the predicated device are insignificant in terms of substantial equivalence. The subject device is substantially equivalent to the predicate device in K080917.
6 of page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.