← Product Code [QBJ](/productcode/QBJ) · K261402

# Dexcom G7 Continuous Glucose Monitoring System;Dexcom G7 15 Day Continuous Glucose Monitoring System (K261402)

_Dexcom, Inc. · QBJ · Jun 8, 2026 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K261402

## Device Facts

- **Applicant:** Dexcom, Inc.
- **Product Code:** [QBJ](/productcode/QBJ.md)
- **Decision Date:** Jun 8, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1355
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

Dexcom G7 Continuous Glucose Monitoring System: The Dexcom G7 Continuous Glucose Monitoring System (Dexcom G7 CGM System or G7) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 2 years and older. The Dexcom G7 CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. The Dexcom G7 CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes. Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System: The Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System (Dexcom G7 15 Day System or G7 15 Day) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 18 years and older. The Dexcom G7 15 Day CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 15 Day CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 15 Day CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. The Dexcom G7 15 Day CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 15 Day CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes.

## Device Story

Wearable CGM system; sensor/transmitter unit measures interstitial glucose via wire-type sensing mechanism; transmits data wirelessly every 5 minutes to mobile app or receiver. Used at home by patients; provides real-time glucose trends, alerts, and high/low notifications. Enables autonomous communication with AID systems for insulin delivery adjustments. Benefits include reduced fingerstick frequency and improved glycemic management through continuous monitoring.

## Clinical Evidence

No clinical data. Bench testing only; sterilization assessment and mechanical functional testing performed to validate modified sterilization cycle and pallet configuration.

## Technological Characteristics

Wire-type interstitial glucose sensor; radio transmitter; wireless connectivity (Bluetooth/proprietary); 10-day (G7) or 15-day (G7 15 Day) wear life; 12-hour grace period. Sterilization via ethylene oxide (alternate cycle). Interoperable with mobile apps and AID systems.

## Regulatory Identification

An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Robust clinical data demonstrating the accuracy of the device in the intended use population.
(ii) The clinical data must include a comparison between iCGM values and blood glucose values in specimens collected in parallel that are measured on an FDA-accepted laboratory-based glucose measurement method that is precise and accurate, and that is traceable to a higher order (
*e.g.,* an internationally recognized reference material and/or method).(iii) The clinical data must be obtained from a clinical study designed to fully represent the performance of the device throughout the intended use population and throughout the measuring range of the device.
(iv) Clinical study results must demonstrate consistent analytical and clinical performance throughout the sensor wear period.
(v) Clinical study results in the adult population must meet the following performance requirements:
(A) For all iCGM measurements less than 70 milligrams/deciliter (mg/dL), the percentage of iCGM measurements within ±15 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 85 percent.
(B) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 70 percent.
(C) For all iCGM measurements greater than 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 80 percent.
(D) For all iCGM measurements less than 70 mg/dL, the percentage of iCGM measurements within ±40 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 98 percent.
(E) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(F) For all iCGM measurements greater than180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(G) Throughout the device measuring range, the percentage of iCGM measurements within ±20 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 87 percent.
(H) When iCGM values are less than 70 mg/dL, no corresponding blood glucose value shall read above 180 mg/dL.
(I) When iCGM values are greater than 180 mg/dL, no corresponding blood glucose value shall read less than 70 mg/dL.
(J) There shall be no more than 1 percent of iCGM measurements that indicate a positive glucose rate of change greater than 1 mg/dL per minute (/min) when the corresponding true negative glucose rate of change is less than −2 mg/dL/min as determined by the corresponding blood glucose measurements.
(K) There shall be no more than 1 percent of iCGM measurements that indicate a negative glucose rate of change less than −1 mg/dL/min when the corresponding true positive glucose rate of change is greater than 2 mg/dL/min as determined by the corresponding blood glucose measurements.
(vi) Data demonstrating similar accuracy and rate of change performance of the iCGM in the pediatric population as compared to that in the adult population, or alternatively a clinical and/or technical justification for why pediatric data are not needed, must be provided and determined by FDA to be acceptable and appropriate.
(vii) Data must demonstrate that throughout the claimed sensor life, the device does not allow clinically significant gaps in sensor data availability that would prevent any digitally connected devices from achieving their intended use.
(2) Design verification and validation must include a detailed strategy to ensure secure and reliable means of iCGM data transmission to provide real-time glucose readings at clinically meaningful time intervals to devices intended to receive the iCGM glucose data.
(3) Design verification and validation must include adequate controls established during manufacturing and at product release to ensure the released product meets the performance specifications as defined in paragraphs (b)(1) and (b)(2) of this section.
(4) The device must demonstrate clinically acceptable performance in the presence of clinically relevant levels of potential interfering substances that are reasonably present in the intended use population, including but not limited to endogenous substances and metabolites, foods, dietary supplements, and medications.
(5) The device must include appropriate measures to ensure that disposable sensors cannot be used beyond its claimed sensor wear period.
(6) Design verification and validation must include results obtained through a usability study that demonstrates that the intended user can use the device safely and obtain the expected glucose measurement accuracy.
(7) The labeling required under § 809.10(b) of this chapter must include a separate description of the following sensor performance data observed in the clinical study performed in conformance with paragraph (b)(1) of this section for each intended use population, in addition to separate sensor performance data for each different iCGM insertion or use sites (
*e.g.,* abdomen, arm, buttock):(i) A description of the accuracy in the following blood glucose concentration ranges: less than 54 mg/dL, 54 mg/dL to less than 70 mg/dL, 70 to 180 mg/dL, greater than 180 to 250 mg/dL, and greater than 250 mg/dL.
(ii) A description of the accuracy of positive and negative rate of change data.
(iii) A description of the frequency and duration of gaps in sensor data.
(iv) A description of the true, false, missed, and correct alert rates and a description of the available glucose concentration alert settings, if applicable.
(v) A description of the observed duration of iCGM life for the device.

## Predicate Devices

- Dexcom G7 Continuous Glucose Monitoring System ([K253737](/device/K253737.md))
- Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System ([K253737](/device/K253737.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

## I Background Information:

A 510(k) Number

K261402

B Applicant

Dexcom, Inc.

C Proprietary and Established Names

Dexcom G7 Continuous Glucose Monitoring System;

Dexcom G7 15 Day Continuous Glucose Monitoring System

## D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  QBJ | Class II | 21 CFR 862.1355 - Integrated Continuous Glucose Monitoring System | CH – Clinical Chemistry  |

## E Purpose for Submission:

Modification to the sterilization cycle and pallet configuration

## II Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

Dexcom G7 Continuous Glucose Monitoring System:

The Dexcom G7 Continuous Glucose Monitoring System (Dexcom G7 CGM System or G7) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 2 years and older.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

The Dexcom G7 CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments.

The Dexcom G7 CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes.

Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System:

The Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System (Dexcom G7 15 Day System or G7 15 Day) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 18 years and older.

The Dexcom G7 15 Day CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 15 Day CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 15 Day CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments.

The Dexcom G7 15 Day CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 15 Day CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes.

## C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

Dexcom G7 Continuous Glucose Monitoring System:

- Users of the Dexcom G7 Continuous Glucose Monitoring System can take a standard dose of acetaminophen (up to 1 gram every 6 hours in adults) and still use the Dexcom G7 System readings. Sensor glucose readings will be falsely higher if the user is taking more than a standard acetaminophen dose.

- Sensor glucose readings will also be falsely higher if the user is taking hydroxyurea. Users should not use the Dexcom G7 System for diabetes treatment decisions if they are taking hydroxyurea.

- Don't wear any Dexcom G7 CGM System component during magnetic resonance imaging (MRI) or high-frequency electrical heat (diathermy) treatment. However, it's safe to have a CT scan if you keep the sensor out of the scanned area and cover the sensor with a lead apron during the scan.

- Don't ignore low/high symptoms: Use BG meter to make treatment decisions when the sensor readings don't match the user's low/high symptoms. If needed, seek immediate medical attention.

- Use BG meter to make treatment decisions when the Dexcom G7 CGM System doesn't show both a number and trend arrow as well as during the 30-minute sensor warmup period.

- Don't use if on dialysis or critically ill: The Dexcom G7 CGM System performance hasn't been evaluated in these populations and sensor readings may be inaccurate.

K261402 - Page 2 of 10

{2}

- Don't ignore broken or detached sensor wires. If this happens, contact 24/7 technical support.
- Insert only in the arm or buttocks: Don't wear it on other sites as it may not work as expected.
- Store sensors at room temperature or in the refrigerator, between 36° F and 86° F, but not in the freezer.
- Check settings: Make sure the smart device volume is turned up, not muted, and the speaker works. When a headphone is connected, alerts will only sound through the headphone, not on the smart device speaker.
- Make sure Bluetooth is on. Otherwise, the user won't get readings or alerts.
- Make sure the smart device settings follow Dexcom's recommended settings. Certain phone settings such as Android's Digital Wellbeing and Apple's Screen Time may prevent notifications if enabled.
- Allow Dexcom G7 CGM System app notifications to show on Lock screen. This will ensure the user can receive Dexcom notifications and allow the user to see notifications without unlocking the phone.
- Android users must allow Location Permission, Do Not Disturb Access, and Notifications to use the app.
- Apple users must allow Critical Alerts to use the app.
- Battery: Keep the battery charged.
- Quiet Mode (Vibrate): When this setting is enabled all System alerts will vibrate. The Urgent Low Glucose and technical alerts will still escalate to sound if not acknowledged.
- Quiet Mode (Silence All): When this setting is enabled, all System alerts will be silent. The user won't receive sound or vibration for any alerts. The user will still receive visual alerts on the display device. (Exceptions: App Stopped alerts will still sound.) Check the display device frequently to avoid missing a low/high event.
- Before upgrading the smart device or its operating system, check dexcom.com/compatibility. Always update manually and verify correct device settings afterward.
- Let the date and time on the smart device automatically update when traveling across time zones or switch between standard and daylight saving times. Don't manually change the smart device time because the user may not get readings or alerts and it may make the time on the trend screen wrong.

Use electrical equipment as directed:

Portable radio frequency communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 12 inches to any part of the Dexcom G7 CGM System including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

Use of this equipment adjacent to, or stacked with, other equipment should be avoided because it could result in improper operation.

Not using supplied USB charger and cable may cause the receiver battery to not charge. Don't use if the supplied USB charger or cable is damaged. Store supplied USB charger and cable safely. Misuse of the USB cable can be a strangulation risk.

In addition to the special conditions listed above for the Dexcom G7 CGM system, the following special conditions for use were included specifically for the new Direct To Watch feature in K234133:

Watch app settings: The watch app uses settings from your phone app.

K261402 - Page 3 of 10

{3}

Bluetooth wireless technology: Make sure your Bluetooth is on. If not, you won't get readings or alerts.

Make sure your smart device settings follow Dexcom's recommended settings. Certain phone settings such as Android's Digital Wellbeing and Apple's Screen Time may prevent notifications if enabled.

Apple users must allow Location Permission and Critical Alerts to use the phone app.

## Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System:

- Users of the Dexcom G7 15 Day Continuous Glucose Monitoring System can take a standard dose of acetaminophen (up to 1 gram every 6 hours in adults) and still use the Dexcom G7 15 Day CGM System readings. Sensor glucose readings will be falsely higher if the user is taking more than a standard acetaminophen dose.
- Sensor glucose readings will also be falsely higher if the user is taking hydroxyurea. Users should not use the Dexcom G7 15 Day CGM System for diabetes treatment decisions if they are taking hydroxyurea.
- Don't wear any Dexcom G7 15 Day CGM System component during magnetic resonance imaging (MRI) or high-frequency electrical heat (diathermy) treatment. However, it's safe to have a CT scan if you keep the sensor out of the scanned area and cover the sensor with a lead apron during the scan.
- Don't ignore low/high symptoms: Use BG meter to make treatment decisions when the sensor readings don't match the user's low/high symptoms. If needed, seek immediate medical attention.
- Use BG meter to make treatment decisions when the Dexcom G7 15 Day CGM System doesn't show both a number and trend arrow as well as during the sensor warmup period.
- Don't use if on dialysis or critically ill: The Dexcom G7 15 Day CGM System performance hasn't been evaluated in these populations and sensor readings may be inaccurate.
- Don't ignore broken or detached sensor wires. If this happens, contact 24/7 technical support.
- Before upgrading the smart device or its operating system, check dexcom.com/compatibility. Always update manually and verify correct device settings afterward.
- Let the date and time on the smart device automatically update when traveling across time zones or switch between standard and daylight saving times. Don't manually change the smart device time because the user may not get readings or alerts and it may make the time on the trend screen wrong.
- Not using supplied USB charger and cable may cause the receiver battery to not charge. Don't use if the supplied USB charger or cable is damaged.

## III Device Description

The Dexcom G7 Continuous Glucose Monitoring System (G7 System) and the Dexcom G7 15 Day Continuous Glucose Monitoring System (G7 15 Day System) are interoperable continuous glucose monitoring (CGM) systems intended to continuously measure the glucose in the interstitial fluid, calculate the glucose reading and make this value available to the user. The system is intended for single-patient use at home and requires a prescription. The G7 System is intended for persons with diabetes age 2 and older whereas the G7 15 Day System is intended for persons with diabetes age 18 and older.

K261402 - Page 4 of 10

{4}

The G7 System and G7 15 Day System consist of the following primary components: a wearable, consisting of a G7 sensor and transmitter worn on the body and a display device, which can be a G7 Mobile Application1 (Mobile App) on an iOS or Android OS smart device or a G7 Receiver (Receiver). The G7 System and G7 15 Day System use Bluetooth Low Energy (BLE) for wireless communication between these components.

To achieve the intended functions and performance of both the G7 System and the G7 15 Day System, one sensor and at least one display device (app or receiver) must be used together. The receiving device acts as the primary display device and the user may choose to use only the app (with optional watch app) or the receiver as the sole display device; they may also choose to use multiple display devices simultaneously. The user must pair the display device(s) with each unique sensor to enable communication and start a sensor session. During an active session, the sensor reports new glucose data to a receiving device every 5 minutes. The receiving device then displays glucose data and provides alerts and information signals to the user. The reportable glucose range for both the G7 System and the G7 15 Day System is 40 g/dL to 400 mg/dL. Glucose values below this range are reported as 'LOW' and glucose values above this range are reported as 'HIGH'. The sensor has an expected wear period of up to 10 days for the G7 System and up to 15 days for the G7 15 Day System, both with an extended 12-hour grace period after the sensor session. The grace period allows additional time for the user to change the sensor at a convenient time.

The proposed G7 System and G7 15 Day System are based on the same principle of operation and fundamental design as their respective predicate devices (K240902 and K243214) with identical user functionality, intended use, indications for use, display devices, sensor, and alerts/alarms.

The G7 System and G7 15 Day System comprise the following subsystems, which are described in comparison to the respective predicate below:

**Glucose Sensing Subsystem (GSS):** Wearable (Sensor, Transmitter, and Patch) and Applicator. The Sensor is a small and flexible wire inserted by the applicator into subcutaneous tissue where it converts glucose into electric current. The Transmitter is pre-connected to the sensor and is worn on the body by the adhesive patch. The Transmitter measures the electric current produced by the sensor and converts these measurements into estimated glucose values (EGV) using an onboard algorithm. The Transmitter is capable of sending glucose data to a display device (mobile app, receiver, or watch app).

Each GSS box also includes an overpatch, which is a general adhesive tape that helps with the adhesion of the wearable to the user's body.

**Comparison to Predicate Device (K253737)**

There are no changes to the Transmitter, Sensor, Applicator, or adhesive patch (MT-1000356) as a result of this premarket notification.

**Mobile Applications: iOS and Android G7 Mobile Application (iOS App and Android App)**

The G7 Mobile App is available on both iOS and Android platforms for users with a compatible Bluetooth Low Energy (BLE)-enabled smart device. The Mobile App can be used as a primary display to receive and display glucose readings and trend graphs via BLE communication. The Mobile App provides in-app guidance for the user on how to apply and set up the wearable and

K261402 - Page 5 of 10

{5}

how to create a user account. Once the set-up has been completed, the Mobile App receives information from the transmitter and acts as a user interface by indicating the system state (e.g., warm-up period, signal loss, etc.), displaying glucose readings and trend graphs, and alerting the user when glucose levels are outside of a target zone and when specific system states occur. The iOS App (Apple users) can also be used with the Watch Application. When the user downloads the iOS Mobile App from the Apple App Store, the Dexcom Watch App is, by default, automatically downloaded.

## Comparison to Predicate Device (K253737)

The proposed G7 System and the proposed G7 15 Day System utilize the same G7 Mobile App, which is the same app as the predicate device (K253737). There are no changes to the G7 Mobile App as a result of this pre-market notification.

## Watch Application: G7 Watch App (Watch App)

Users with compatible smartwatches can use the G7 Watch App to view glucose information and notifications. The Watch App is bundled as part of the iOS Mobile App software code and is automatically downloaded at the same time the user downloads their Mobile App from the Apple App Store. The Watch App operates in two modes:

- Standard Smartwatch mode – the Watch App connects to the G7 Mobile App and mirrors the G7 Mobile App UI. The user’s smart phone is required to be within close proximity of the watch to utilize the Watch App, as the Watch App receives all information from the Mobile App.
- Direct to Watch (DTW) mode – the Watch App connects directly with the G7 Transmitter after the sensor session has started and the user has enabled the DTW feature on their iOS App. This means that users are not required to keep their smart phones in close proximity to their Watch App in order to receive EGVs and alerts and can instead receive glucose data on their smartwatch directly from the Transmitter.

## Comparison to Predicate Device (K253737)

The proposed G7 System and the proposed G7 15 Day System utilize the same G7 Watch App, which is the same app as the predicate device (K253737). There are no changes to the G7 Watch App as a result of this pre-market notification.

## Receiver Subsystem (RVS): G7 Receiver (Receiver)

The Receiver is a small handheld device that can also be used as a primary display to receive and display glucose readings and trend graphs via BLE communication with the Transmitter. Similar to the Mobile App, the Receiver can alert the user when glucose levels are outside of a target zone and when specific system states occur.

## Comparison to Predicate Device (K253737)

The proposed G7 System and the proposed G7 15 day System use the same Receiver Subsystem (RVS) as the predicate device (K253737). There are no changes to the RVS as a result of this pre-market notification.

## Interoperability

The GSS is designed to communicate to all display devices simultaneously. The proposed G7 System and the proposed G7 15 Day System, same as the respective predicate device (K253737), are also designed to communicate estimated glucose values, trend, and system information to other compatible electronic interfaces via the following secure wireless connections:

K261402 - Page 6 of 10

{6}

- Wireless communication from the transmitter directly to an interoperable device communicating through the same protocol
- The G7 Mobile App communicates to another app on a single mobile platform
- The G7 Mobile App communicates through the cloud to another software device.
- Dexcom Partner Web APIs: The Dexcom Partner Web APIs enable secure and reliable communication of CGM data to authorized client software intended to receive the data through the cloud. The Partner Web APIs is not intended to be used by automated insulin delivery systems (AID).

## IV Substantial Equivalence Information:

### A Predicate Device Name(s):

Dexcom G7 Continuous Glucose Monitoring (CGM) System,
Dexcom G7 15 Day Continuous Glucose Monitoring (CGM) System

### B Predicate 510(k) Number(s):

K253737

### C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K261402 | K253737  |
| --- | --- | --- |
|  Device Trade Name | Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System | Same  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | Dexcom G7 Continuous Glucose Monitoring System: The Dexcom G7 Continuous Glucose Monitoring System (Dexcom G7 CGM System or G7) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 2 years and older. The Dexcom G7 CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. The Dexcom G7 CGM System is also intended to | Same  |

K261402 - Page 7 of 10

{7}

K261402 - Page 8 of 10

|   | autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes. Dexcom G7 15 Day Continuous Glucose Monitoring System: The Dexcom G7 15 Day Continuous Glucose Monitoring System (Dexcom G7 15 Day CGM System or G7 15 Day) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 18 years and older. The Dexcom G7 15 Day CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 15 Day CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 15 Day CGM System also aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. |   |
| --- | --- | --- |
|  Principle of Operation | Amperometric measurement of current proportional to glucose concentration in interstitial fluid via glucose oxidase chemical reaction. | Same  |
|  Compatibility with connected devices | Compatible with digitally connected devices, including automated insulin dosing (AID) systems | Same  |
|  Measurement Range | 40 - 400 mg/dL | Same  |
|  Sterilization Method | Ethylene oxide | Same  |
|  General Device Characteristic Differences |  |   |
|  Alternate Sterilization cycle | Modified cycle parameters and release method | N/A  |
|  Pallet Configuration | Modified pallet configuration for alternate sterilization cycle | N/A  |

V Standards/Guidance Documents Referenced:

- ISO 14971:2019, Medical devices - Application of Risk Management to Medical Devices

{8}

- ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-7:2008, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
- ISO 10993-12:2021, Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
- AAMI TIR 28:2016/(R)2020, Product adoption and process equivalence for ethylene oxide sterilization
- ANSI AAMI ISO 11135:2014/A1:2018, Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices - Amendment 1: Revision of Annex E, Single batch release
- ANSI AAMI ISO 11737-2:2019, Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- ANSI AAMI ISO 11737-1:2018, Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on product [Including Amendment 1 (2021)]
- ISO 11138-7:2019-03, Sterilization of health care products - Biological indicators - Part 7: Guidance for the selection, use and interpretation of results
- ANSI AAMI ISO 11138-1:2017, Sterilization of health care products - Biological indicators - Part 1: General requirements
- ISO 11138-2:2017, Sterilization of health care products - Biological indicators – Part 2: Biological indicators for ethylene oxide sterilization processes
- ISO 10993-2:2022, Biological Evaluation of medical devices - Part 2: Animal welfare requirements
- ASTM D4169-22, Standard Practice for Performance Testing of Shipping Containers and Systems

## VI Performance Characteristics:

### A. Non-Clinical Performance

**Biocompatibility:**
Dexcom conducted biocompatibility testing in accordance with ISO 10993 and followed the same method as the predicate device. Results were consistent with the predicate device.

**Sterilization:**
Dexcom conducted a sterilization assessment in accordance with AAMI TIR 28 and ANSI/AAMI/ISO 11135. Results were consistent with the predicate device.

**Shelf life:**
Dexcom conducted functional testing following accelerated aging conditioning. Results were consistent with the predicate device.

**Packaging**

K261402 - Page 9 of 10

{9}

Dexcom conducted testing per ASTM D4169-23 and results remained consistent with the predicate device.

**Mechanical Functional Testing**
Dexcom executed mechanical functional tests to verify that the proposed alternate EO sterilization cycle and pallet configuration is suitable for use in both the G7 CGM System and the G7 15 Day CGM System. All tests passed and remained consistent with the predicate device.

**B. Clinical Studies:**
N/A. The only update to the subject device is the new sterilization cycle and pallet configuration. The intended users, intended use, and clinical workflow remain unchanged from the predicate device (K253737). Therefore, additional clinical testing was not necessary to demonstrate substantial equivalence.

**C. Other Supportive Device Performance Characteristics Data**
None.

**VII Proposed Labeling:**
The labeling supports the finding of substantial equivalence for this device.

**VIII Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K261402 - Page 10 of 10

---

**Source:** [https://fda-staging.innolitics.com/device/K261402](https://fda-staging.innolitics.com/device/K261402)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
