K261399 · Nipro Medical Corporation · QAX · Jul 28, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K261399
Device Name
ELISIO™-HX
Applicant
Nipro Medical Corporation
Product Code
QAX · Gastroenterology, Urology
Decision Date
Jul 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5862
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ELISIO™-HX dialyzer is indicated for patients with chronic kidney failure who are prescribed intermittent hemodialysis. It provides an expanded solute removal profile with increased removal of various middle molecules (up to 45 KDa) that may play a pathologic role in the uremic clinical syndrome. This dialyzer is not intended for hemofiltration or hemodiafiltration therapy. The total extracorporeal blood volume for this dialyzer and the set should represent less than 10% of the patient's blood volume.
Device Story
Extracorporeal hemodialyzer for end-stage renal disease (ESRD); utilizes semi-permeable Polyethersulfone (PES) hollow-fiber membrane; blood and dialysate flow in countercurrent directions; selective diffusion of uremic toxins, electrolytes, and excess fluid driven by concentration gradients and transmembrane pressure; connects to dialysis machine via ISO 8637-2 compliant tubing; requires machine with ultrafiltration controller; enlarged pore structure enhances clearance of middle-molecule uremic toxins (up to 45 kDa) while retaining albumin; used by clinicians in clinical settings; provides improved removal of middle molecules (e.g., λ-FLC) to manage uremic syndrome; benefits patient through effective solute clearance and fluid management.
Clinical Evidence
Prospective clinical study (n=18) in adult chronic kidney disease patients. Primary safety endpoint: change in serum albumin (66 kDa) from baseline to 12 weeks (mean change -0.04±0.419 g/dL). Primary efficacy endpoint: reduction ratio of λ-FLC (-11.20% improvement vs. baseline). Secondary endpoints showed increased clearance of TNFα, Complement Factor D, κ-FLC, and β2-Microglobulin compared to ELISIO™-H. No device-related AEs or UADEs reported. Bench testing per ISO 8637-1 and FDA guidance confirmed structural integrity, solute clearance, and sieving coefficients.
Indicated for patients with chronic kidney failure prescribed intermittent hemodialysis. Not for hemofiltration or hemodiafiltration. Total extracorporeal blood volume must be <10% of patient blood volume.
Regulatory Classification
Identification
A hemodialyzer with expanded solute removal profile is a device intended for use a part of an artificial kidney system for the treatment of patients with renal failure by performing such therapies as hemodialysis, hemofiltration, and hemodiafiltration. A hemodialyzer with expanded solute removal profile includes modifications to the semipermeable membrane that allows for increased removal of uremic retention solutes compared with standard high-flux hemodialyzers of the high permeability hemodialysis system classification (§ 876.5860), including solutes at the upper end of the “middle” molecular weight range (0.5 kDa to 60 kDa). This device is intended to be used with the extracorporeal hemodialysis delivery systems, blood tubing sets, blood access devices, and accessories regulated under §§ 876.5820, 876.5860, 876.5540, and/or 876.5600.
Special Controls
In combination with the general controls of the FD&C Act, the hemodialyzer with expanded solute removal profile is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing under anticipated conditions of use must evaluate the solute removal profile and document all adverse events.
(2) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Ultrafiltration;
(ii) Blood and dialysate pressure drop;
(iii) Clearance rates;
(iv) Sieving coefficients;
(v) Mechanical hemolysis;
(vi) Structural integrity;
(vii) Blood compartment integrity;
(viii) Volume of the blood compartment; and
(ix) Endotoxin retention of the dialyzer membrane.
(3) The tissue-contacting components of the device must be demonstrated to be biocompatible. Biocompatibility evaluation must include a chemical analysis of the dialyzer membrane.
(4) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(5) The patient-contacting components of the device must be demonstrated to be non-pyrogenic.
(6) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(7) Device labeling must include:
(i) Shelf life;
(ii) Storage conditions;
(iii) Instructions for the preparation of the hemodialyzer, initiation of dialysis, troubleshooting, and discontinuance of dialysis;
(iv) Membrane surface area, priming (blood) volume, maximum transmembrane pressure, maximum blood flow and maximum dialysate rate for each model;
(v) A non-pyrogenic statement;
(vi) A summary of the in vitro performance data, provided in tabular form; and
(vii) A summary of the clinical performance data.
{0}

July 28, 2026
Nipro Medical Corporation
Jessica Oswald-Mcleod
Director, Regulatory Affairs
3150 NW 107th Ave.
Miami, Florida 33172
Re: K261399
Trade/Device Name: Elisio™-HX
Regulation Number: 21 CFR 876.5862
Regulation Name: Hemodialyzer With Expanded Solute Removal Profile
Regulatory Class: Class II
Product Code: QAX
Dated: April 29, 2026
Received: April 29, 2026
Dear Jessica Oswald-Mcleod:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261399 - Jessica Oswald-Mcleod
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261399 - Jessica Oswald-Mcleod
Page 3
Sincerely,
MAURA ROONEY -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261399 | ? |
| Please provide the device trade name(s). | | ? |
| ELISIO™-HX | | |
| Please provide your Indications for Use below. | | ? |
| The ELISIO™-HX dialyzer is indicated for patients with chronic kidney failure who are prescribed intermittent hemodialysis. It provides an expanded solute removal profile with increased removal of various middle molecules (up to 45 KDa) that may play a pathologic role in the uremic clinical syndrome. This dialyzer is not intended for hemofiltration or hemodiafiltration therapy. The total extracorporeal blood volume for this dialyzer and the set should represent less than 10% of the patient's blood volume. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K261399
Page 1 of 3

510(k) Summary: ELISIO™-HX | Page 1
## 510(k) Summary: ELISIO™-HX
Date Prepared: April 28, 2026
### Contact Details (21 CFR 807.92(a)(1))
| Applicant Name | Nipro Medical Corporation |
| --- | --- |
| Applicant Address | 3150 NW 107th Ave. Miami FL 33172 United States |
| Applicant Contact Telephone | 305-432-6699 |
| Applicant Contact | Mrs. Jessica Oswald-McLeod, Director, Regulatory Affairs |
| Applicant Contact Email | Jessicao@nipromed.com |
### Device Name (21 CFR 807.92(a)(2))
| Device Trade Name | ELISIO™-HX |
| --- | --- |
| Common Name | Hemodialyzer |
| Classification Name | Hemodialyzer with expanded solute removal profile |
| Regulation Number | 876.5862 |
| Product Code | QAX |
### Legally marketed predicate device (21 CFR 807.92(a)(3))
| Predicate # DEN190042 | Predicate Trade Name: Theranova | Product Code: QAX |
| --- | --- | --- |
| Predicate # K260533 | Predicate Trade Name: ELISIO™-H | Product Code: KDI |
### Device Description Summary (21 CFR 807.92(a)(4))
The ELISIO™-HX hemodialyzer is an extracorporeal medical device designed to perform the function of an artificial kidney in patients with end-stage renal disease (ESRD). It is comprised of two fluid pathways, blood and dialysate, separated by a semi-permeable membrane composed of Polyethersulfone (PES). During hemodialysis, the patient's blood circulates through the blood compartment of the dialyzer, while dialysate flows in a countercurrent direction through the adjacent dialysate compartment. The semipermeable membrane facilitates selective diffusion of uremic toxins, electrolytes, and excess fluid from the blood into the dialysate, driven by concentration gradients and transmembrane pressure. The purified blood is then returned to the patient, while the spent dialysate containing the removed solutes is discarded.
The ELISIO™-HX connects to the patient through an ISO 8637-2 compliant blood tubing set and is intended for use exclusively with dialysis machines equipped with an ultrafiltration controller or accurate fluid-balancing system. The POLYNEPHRON™ (Polyethersulfone) membrane incorporates an enlarged pore structure and optimized membrane geometry to enhance the clearance of middle-molecule uremic toxins (up to 45 kDa) while maintaining minimal albumin loss. The polypropylene housing is not made with BPA or DEHP.
The ELISIO™-HX is available in four sizes distinguished by membrane surface area: 1.5, 1.7, 1.9, and 2.1 m². It is sterile (gamma radiation) and non-pyrogenic with a three-year shelf life. No integrated components or accessories are included.
NIPRO MEDICAL CORPORATION
3150 NW 107th Ave. | Miami, FL | 33172 USA
T: +1 305.599.7174
info@nipromed.com | www.nipro.com
{5}
K261399
Page 2 of 3
510(k) Summary: ELISIO™-HX | Page 2
### **Intended Use/Indications for Use (21 CFR 807.92(a)(5))**
The ELISIO™-HX dialyzer is indicated for patients with chronic kidney failure who are prescribed intermittent hemodialysis. It provides an expanded solute removal profile with increased removal of various middle molecules (up to 45 KDa) that may play a pathologic role in the uremic clinical syndrome. This dialyzer is not intended for hemofiltration or hemodiafiltration therapy. The total extracorporeal blood volume for this dialyzer and the set should represent less than 10% of the patient's blood volume.
### **Indications for Use Comparison (21 CFR 807.92(a)(5))**
The subject device Indications for Use are identical to the predicate, Theranova (DEN190042)
### **Technological Comparison (21 CFR 807.92(a)(6))**
The ELISIO™-HX and Theranova dialyzers share the same technological characteristics of intended use, indications for use, operating principles, and key design features. Both devices are sterile, single-use, medium cutoff (MCO) hemodialyzers intended for the extracorporeal treatment of patients with renal failure and are designed with an expanded solute removal profile that targets middle molecules. They function through the same mechanism of diffusion across a semi-permeable hollow-fiber membrane. Each is intended for use by qualified clinicians within an in-patient clinical setting and must be used with dialysis machines equipped with ultrafiltration control.
The subject device is identical in materials, manufacturing processes, sterilization method, shelf life, and packaging to the reference device ELISIO™-H (K260533).
### **Non-Clinical and/or Clinical Tests Summary & Conclusions (21 CFR 807.92(b))**
Performance testing was conducted according to ISO 8637-1: 2017 - Extracorporeal systems for blood purification - Part 1: Haemodialysers, haemodiafilters, haemofilters and haemoconcentrators and the FDA Guidance for the Content of Premarket Notifications for Conventional and High Permeability Hemodialyzers. Testing included Mechanical Characteristics (Structural integrity: positive and negative pressure, Blood compartment integrity, Connectors verification) and Performance Characteristics (Solute clearance: Urea, Creatinine, Phosphate, Vitamin B12, Inulin, Myoglobin, Potassium (K), TNF-α and IL-6 and Cytochrome C; Sieving coefficients: Urea, Creatinine, Inulin, Vitamin B12, β2-Microglobulin, Myoglobin and Albumin; Ultrafiltration rate; Ultrafiltration coefficient; Blood compartment volume; Pressure drop; Endotoxin testing (ANSI/AAMI ST72), Mechanical Hemolysis (ASTM F756-17) and expiration date (ASTM F1980)).
The ELISIO™-HX clinical study evaluated the safety and effectiveness of the ELISIO™-HX in adult patients with chronic kidney disease who are suitable candidates for high-flux hemodialysis (HD). The primary safety endpoint was the change in serum albumin level (66 kDa) from baseline to 12 weeks post-treatment. The primary efficacy endpoint was the change in reduction ratio of post-dialysis lambda free light chains (λ-FLC) from baseline to 12 weeks post-treatment. Secondary efficacy endpoints included the reduction ratio of other middle molecules and Kt/V. Secondary safety endpoints included changes in Factor VII, Protein C, Vitamin A and nPNA (nPCR), and laboratory assessments (Comprehensive Metabolic Panel, Hematology Panel). All adverse events, regardless of their relationship to the study device or procedure, were recorded during the study.
NIPRO MEDICAL CORPORATION
3150 NW 107th Ave. | Miami, FL | 33172 USA
T: +1 305.599.7174
info@nipromed.com | www.nipro.com
{6}
K261399
Page 3 of 3

510(k) Summary: ELISIO™-HX | Page 3
The study screened 22 participants at a single site in the United States between September and December 2025. A total of 18 participants underwent a baseline HD treatment with their regular ELISIO™-H dialyzer followed by treatment with the ELISIO™-HX three times a week over a period of 12 weeks. Of the 18 participants, the mean age was 55.5±13.01 years, 55.6% (10/18) were men, 72.2% (13/18) were Black or African American, with mean time on renal replacement therapy of 5.7±5.52 years. Thirteen participants completed all 36 study treatment visits.
Serum albumin levels remained stable from baseline to 12 weeks. The mean level of serum albumin was 4.44±0.355 g/dL at baseline and 4.37±0.534 g/dL at 12 weeks, representing a mean change from baseline of -0.04±0.419 g/dL (95% CI: -0.29, 0.21). Clinical laboratory assessments of coagulation factors, Vitamin A levels, comprehensive metabolic panel and hematology results, as well as nPNA (nPCR) evaluations, revealed no safety concerns associated with ELISIO™-HX. Any changes observed were within expected levels for this patient population and not deemed clinically significant.
The reduction ratio of λ-FLC from baseline to 12 weeks was -11.20% (95% CI: -20.41%, -1.99%) indicating improved removal of λ-FLC after treatment with ELISIO™-HX over baseline treatment with ELISIO™-H. After 12 weeks of treatment with ELISIO™-HX, secondary endpoints showed trends toward increased clearance of TNFα, Complement Factor D, κ-FLC, and β2-Microglobulin compared to ELISIO™-H, but not IL-6 or myoglobin. Single-pool Kt/V was assessed every 4 weeks and remained consistent throughout the study treatment period.
A total of 13 participants experienced 38 Adverse Events (AEs). There were four participants who experienced at least one Serious Adverse Event (SAE), however none of the events were deemed related to the study device or procedure. Eight participants reported experiencing at least one procedure-related AE, most commonly muscle spasms and hypotension events. There were no device-related AEs and no Unanticipated Adverse Device Effects (UADEs). Most AEs were mild or moderate in severity. All adverse events were resolved. Treatment-related adverse events were as expected and consistent with those observed in patients with chronic kidney disease undergoing maintenance hemodialysis. Monthly water and dialysate quality monitoring showed acceptable levels of endotoxin and microbial counts for all samples tested.
Non-clinical performance testing demonstrated that the ELISIO™-HX consistently met its intended performance characteristics across the full range of evaluated membrane surface areas and operating conditions. The device exhibited robust structural integrity, effective and predictable clearance of small and middle molecular weight solutes, controlled solute permeability with strong albumin retention, stable ultrafiltration behavior, and acceptable flow-dependent pressure performance. Clearance and sieving results showed reproducibility and repeatability, and all measurements met predefined acceptance criteria. Clinical data shows hemodialysis treatment with ELISIO™-HX provides improved removal of middle molecules as represented by λ-FLC while maintaining serum albumin levels. These results suggest that hemodialysis using ELISIO™-HX dialyzers can have a persistent effect (i.e. net removal) on the serum levels of some middle molecules that can accumulate in ESRD. Collectively, the performance data confirm that ELISIO™-HX performs as intended without identifying new safety or effectiveness concerns.
NIPRO MEDICAL CORPORATION
3150 NW 107th Ave. | Miami, FL | 33172 USA
T: +1 305.599.7174
info@nipromed.com | www.nipro.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.