Profoject™ LOR Syringe

K261372 · CMT Health PTE., Ltd. · CAZ · Sep 25, 2026 · Anesthesiology

Device Facts

Record IDK261372
Device NameProfoject™ LOR Syringe
ApplicantCMT Health PTE., Ltd.
Product CodeCAZ · Anesthesiology
Decision DateSep 25, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5140
Device ClassClass 2
AttributesPediatric

Indications for Use

Profoject™ LOR Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique, it will be filled with air and/or saline during use. The LOR Syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit. The intended patient population includes children, adolescents, and adults.

Device Story

Manual syringe used by clinicians during epidural procedures; verifies needle tip placement in epidural space via Loss of Resistance (LOR) technique. Device filled with air or saline; not for injection or aspiration. Available in 5mL, 7mL, 10mL volumes. Models feature either NRFit (ISO 80369-6) or Luer (ISO 80369-7) connectors. Single-use; supplied sterile or non-sterile for kit integration. Clinician observes tactile resistance feedback during plunger advancement to confirm needle location. Benefits include standardized, reliable verification of epidural space entry.

Clinical Evidence

No clinical data included. Substantial equivalence supported by bench performance testing and biocompatibility evaluation.

Technological Characteristics

Manual syringe; materials include polypropylene, silicone, and polyisoprene rubber. Connectors comply with ISO 80369-6 (NRFit) and ISO 80369-7 (Luer). Sterilized via ethylene oxide (SAL 10^-6). Volume sizes: 5mL, 7mL, 10mL. One-piece or three-piece plunger assemblies.

Indications for Use

Indicated for use in children, adolescents, and adults to verify epidural needle tip placement via Loss of Resistance technique using air or saline.

Regulatory Classification

Identification

An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.

Predicate Devices

Submission Summary (Full Text)

{0} [LOGO] FDA U.S. FOOD & DRUG ADMINISTRATION September 25, 2026 CMT Health PTE., Ltd. Monica Ma Official Correspondent 150 Beach Rd., #28-05, Gateway W., Singapore, 189720 Singapore Re: K261372 Trade/Device Name: Profoject™ LOR Syringe Regulation Number: 21 CFR 868.5140 Regulation Name: Anesthesia conduction kit Regulatory Class: Class II Product Code: CAZ Dated: August 21, 2026 Received: August 21, 2026 Dear Monica Ma: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K261372 - Monica Ma Page 2 may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K261372 - Monica Ma Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JAMES J. LEE -S for Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # **Indications for Use** Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K261372 Device Name Profoject™ LOR Syringe Indications for Use (Describe) Profoject™ LOR Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique, it will be filled with air and/or saline during use. The LOR Syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit. The intended patient population includes children, adolescents, and adults. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. **DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {4} # 510(k) Summary [As required by 21 CFR 807.92] 1. Submission Information | 510(k) Number: | K261372 | | --- | --- | | Date: | April 27, 2026 | | Type of 510(k) Submission: | Traditional | | Submitter: | CMT HEALTH PTE. LTD. 6001 BEACH ROAD, #19-13A, GOLDEN MILE TOWER, SINGAPORE 199589 Contact Person: Monica Ma E-mail: ra@cmthealth.com Tel: +65 6846 1379 | 2. Device Description | Trade Name: | Profoject™ LOR Syringe | | --- | --- | | Common Name: | Loss of Resistance Syringe | | Regulation Name: | Anesthesia Conduction Kit | | Product Code: | CAZ | | Device Class: | II | | Regulation Number: | 21 CFR 868.5140 | | Review Panel: | Anesthesiology | | Indications for use: | Profoject™ LOR Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique, it will be filled with air and/or saline during use. The LOR Syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit. The intended patient population includes children, adolescents, and adults. | | Device Description: | The Profoject™ LOR Syringe is available in two models: Model A and Model B. Model A is a syringe equipped with an NRFit slip or NRFit lock connector that complies with ISO 80369-6, while Model B is a syringe equipped with a Luer slip or Luer lock connector that complies with ISO 80369-7. Depending on the type of plunger assembly, Model B can be further subdivided into Model B1 and Model B2. Model A has only one type of plunger assembly. The product is offered in three volume sizes: 5 mL, 7 mL, and 10 mL. ● Model A: Available with NRFit lock or NRFit slip male connectors (compliant with ISO 80369-6). The type of plunger is one-piece plunger. ● Model B: Model B1: Available with Luer lock or Luer slip male connectors (compliant with ISO 80369-7); and the type of plunger is one-piece plunger. Model B2: Fitted with Luer slip male connector only (compliant with | 1 {5} | | ISO 80369-7). The type of plunger is three-piece plunger.The devices are single-use and are supplied either sterile and individually packaged, or non-sterile and bulk-packed. The sterile devices are sterilized with ethylene oxide gas to achieve an SAL of \( 10^{-6} \) and are provided with a sterility maintenance package that maintains sterility throughout the shelf life. The non-sterile, bulk-packed syringes are supplied to anesthetic conduction kit manufacturers for incorporation into the kit and terminal sterilization. | | --- | --- | ## 3. Predicate Device Identification K061737 - LOSS OF RESISTANCE SYRINGE ## 4. Non-Clinical Test Conclusion Non-clinical verification of the Profoject™ LOR Syringe has been conducted to evaluate their safety, performance, and functionality. The results of these tests have demonstrated the overall safety of the proposed device and its effectiveness in accordance with relevant test methods, and ultimately support a substantial equivalence determination. Particularly, the following was conducted to adequately demonstrate the effectiveness of the proposed device in accordance with relevant test methods cited below: ### 4.1. Performance Testing 1) ISO 7886-1 Second edition 2017-05 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use 2) ISO 80369-1 Second edition 2018-11 Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements 3) ISO 80369-6 First Edition 2016-03-15 Small bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications 4) ISO 80369-7 Second edition 2021-05 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications 5) ISO 80369-20 First edition 2015-05-15 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods 6) ISO 11737-1 Third edition 2018-01 [Including AMD1:2021] Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on product [Including Amendment 1 (2021)] 7) ISO 11135 Second edition 2014-07-15 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)] 8) ASTM F1886/F1886M-16 (2024) Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection 9) ASTM F1140/F1140M-13 (2020) Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages 10) ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials 11) ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration 12) USP - NF <71> Sterility Tests 13) USP - NF <788> Particulate Matter in Injections 2 {6} 14) USP - NF <85> Bacterial Endotoxin Testing 15) ISO 10993-7 Second edition 2008-10-15 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals [Including: Technical Corrigendum 1 (2009), AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019)] 16) ISTA 3A: 2018 Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less 17) USP - NF <788> Particulate Matter in Injections 18) Testing for Scale Marking Stability ### 4.2. Biocompatibility Testing The biocompatibility evaluation for the Profoject™ LOR Syringe was conducted in accordance with ISO 10993-1:2018 Biological Evaluation of Medical Devices - Part 1. Evaluation and Testing within a Risk Management Process, as recognized by FDA. The proposed device is classified as an externally communicating device. 1) ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity 2) ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization 3) ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation 4) USP <151> Rabbit Pyrogen Study 5) ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity 6) ISO 10993-4 Third edition 2017-04 Biological evaluation of medical devices--Part 4: Selection of tests for interactions with blood 7) ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials ## 5. Usability Testing Simulated Clinical Study Report ## 6. Substantial Equivalent Based on Assessment of Clinical Performance Data Clinical data was not included in this submission. ## 7. Substantially Equivalent Comparison Conclusion ### 7.1. Comparison of Technology Characteristics Table 1 Comparison of Technology Characteristics | Parameters | | Proposed Device | Predicate Device | Remark | | --- | --- | --- | --- | --- | | 1 | 510(k) Number | K261372 | K061737 | / | | 2 | 510(k) Holder | CMT HEALTH PTE. LTD. | Robert Busse & Co., Inc. | / | | 3 | Trade Name | ProfojectTM LOR Syringe | LOSS OF RESISTANCE SYRINGE | / | | 4 | Product Code | CAZ | CAZ | Same | | 5 | Regulation Number | 21CFR 868.5140 | 21 CFR 868.5140 | | 3 {7} | 6 | Review Panel | Anesthesiology | Anesthesiology | | | --- | --- | --- | --- | --- | | 7 | Device Class | II | II | Same | | 8 | Indications for use | The Profoject™ LOR Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the loss of Resistance technique, it will be filled with air and/or saline during use. The loss of resistance syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit. | The Busse Loss of Resistance Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the loss of Resistance technique, it will be filled with air and/or saline during use. The loss of Resistance Syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit. | Same | | 9 | Overall design | Barrel; Plunger | Barrel; Plunger | Same | | 10 | Single Use | Single Use | Single Use | Same | | 11 | Priming Volume | 5mL, 7mL, 10mL | 10mL | Different See Note 1 | | 12 | Connectivity | ISO 80369-6 ISO 80369-7 | ISO 80369-7 | Different See Note 2 | | 13 | Materials | Model A Barrel: Polypropylene Plunger: Polypropylene, yellow color additive Plunger stopper: Silicone compound and Light blue color additive | Plastic materials | Different See Note 3 | | | | Model B1 Barrel: Polypropylene Plunger: Polypropylene, blue color additive Plunger stopper: Silicone compound and Light blue color additive | | | | | | Model B2 Barrel: Polypropylene Plunger: Polypropylene, white color additive and blue color additive Plunger stopper: Polyisoprene rubber | | | | 14 | Biocompatibility | Complies with ISO 10993-1 | Complies with ISO | Same | 4 {8} | | | | 10993-1 | | | --- | --- | --- | --- | --- | | 15 | Sterile | Sterile or non-sterile | Sterile or non-sterile | Same | | 16 | Sterile Method | EO Sterilized | EO Sterilized | Same | | 17 | Sterilization | \( SAL=10^{-6} \) | \( SAL=10^{-6} \) | Same | ### 7.2. Substantial Equivalence Discussion Table 2 Substantial Equivalence Discussion | Discussion | Note 1 - Syringe Volume:The syringe volume of the proposed device differs from that of the predicate device. However, this difference is just in capacity. In clinical practice, physicians select the appropriate syringe specification based on individual patient conditions, and this difference does not affect the intended use. Furthermore, bench performance testing was conducted on the proposed device in accordance with ISO 7886-1, ISO 80369-7, and ISO 80369-6, and all results met the requirements of these standards. Therefore, this difference does not raise new questions of safety or effectiveness.Note 2 - Connectivity:The proposed device includes both ISO 80369-7 and ISO 80369-6 connectors, whereas Predicate Device includes only ISO 80369-7 connector. All connectors on the proposed device have been tested and conform to ISO 80369-6 and ISO 80369-7 standards, and labeling addresses misconnection risks. Therefore, this difference does not raise new questions of safety or effectiveness.Note 3 - Materials:The specific materials of the predicate devices are not publicly available. However, the proposed devices were tested, and the results demonstrated that they complied with the ISO 10993 series standards. Predicate devices are legally marketed and have been determined to meet equivalent biocompatibility requirements. Therefore, the lack of publicly available predicate material information does not raise different questions of safety and effectiveness. In addition, bench performance testing per ISO 7886-1, ISO 80369-7, and ISO 80369-6 was conducted on the proposed device, and all results met the standard requirements. Accordingly, the proposed device is substantially equivalent to the predicate devices. | | --- | --- | | Conclusion | The ProfojectTM LOR Syringe is substantially equivalent to its predicate device. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device. | 5
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...