The GI Genius™ system is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius™ computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.
Device Story
AI-based computer-assisted detection (CADe) system; processes live white-light endoscopic video streams; identifies regions consistent with colorectal lesions (polyps/adenomas); generates real-time visual overlays (green square markers) and audible alerts on endoscopic display when lesions detected; used in clinical endoscopy suites by endoscopists; device sits between video processor and display monitor; provides real-time decision support to improve lesion detection rates; subject device adds multicontainer management functionality allowing concurrent execution of non-medical software (e.g., clinical management, reporting) in background containers; connectivity features enabled for user interaction with background software via external devices; medical software container (ColonPRO™ US) remains isolated with prioritized system resources; no changes to core lesion detection algorithm or clinical performance.
Clinical Evidence
No new clinical data provided. Baseline clinical validation conducted under DEN200055 remains applicable. Performance metrics for the medical software (ColonPRO™ US) are identical to the predicate: 88.07% lesion-based sensitivity, 60.14% mean true positive rate per frame, 1.79% mean false positive rate per frame, and 0.826 AUC.
Technological Characteristics
Hardware: GI Genius™ Module 300; interfaces with standard video processors/monitors via SDI/HD-SDI/3G-SDI/12G-SDI. Software: ColonPRO™ US (v4.0.3-S) and OS (v1.0.3). Architecture: Multicontainer management system isolating medical software (foreground) from non-medical software (background). Connectivity: Local/remote data exchange enabled for non-medical functions. Safety: Compliant with IEC/EN 60601-1 and 60601-1-2. Cybersecurity: Software architecture controls, access limitations, and security monitoring for container isolation.
Indications for Use
Indicated for adult patients (22 years and older) undergoing screening and surveillance endoscopic mucosal evaluations to aid endoscopists in real-time detection of colonic mucosal lesions (polyps and adenomas) during standard white-light endoscopy.
Regulatory Classification
Identification
A gastrointestinal lesion software detection system is a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device may contain hardware to support interfacing with an endoscope.
Special Controls
In combination with the general controls of the FD&C Act, the gastrointestinal lesion software detection system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including detection of gastrointestinal lesions and evaluation of all adverse events.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
(i) Standalone algorithm performance testing;
(ii) Pixel-level comparison of degradation of image quality due to the device;
(iii) Assessment of video delay due to marker annotation; and
(iv) Assessment of real-time endoscopic video delay due to the device.
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Performance data must demonstrate electromagnetic compatibility and electrical safety, mechanical safety, and thermal safety testing for any hardware components of the device.
(5) Software verification, validation, and hazard analysis must be provided. Software description must include a detailed, technical description including the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, the associated inputs and outputs, mapping of the software architecture, and a description of the video signal pipeline.
(6) Labeling must include:
(i) Instructions for use, including a detailed description of the device and compatibility information;
(ii) Warnings to avoid overreliance on the device, that the device is not intended to be used for diagnosis or characterization of lesions, and that the device does not replace clinical decision making;
(iii) A summary of the clinical performance testing conducted with the device, including detailed definitions of the study endpoints and statistical confidence intervals; and
(iv) A summary of the standalone performance testing and associated statistical analysis.
Predicate Devices
GI Genius™ Module 300 with ColonPRO™ 4.0 (K241887)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2026
Cosmo Artificial Intelligence - Ai, Ltd.
% Roger Gray
VP Quality and Regulatory
Donawa Lifescience Consulting Srl
Piazza Albania 10
Rome, 00153
Italy
Re: K261369
Trade/Device Name: GI Genius™ Module 300 (GGM300-US); ColonPRO™ US (CPRO403S-US)
Regulation Number: 21 CFR 876.1520
Regulation Name: Gastrointestinal Lesion Software Detection System
Regulatory Class: Class II
Product Code: QNP
Dated: April 26, 2026
Received: April 27, 2026
Dear Roger Gray:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K261369 - Roger Gray
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261369 - Roger Gray
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
SIVAKAMI VENKATACHALAM -S
for
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261369 | ? |
| Please provide the device trade name(s). | | ? |
| GI Genius™ Module 300 (GGM300-US); ColonPRO™ US (CPRO403S-US) | | |
| Please provide your Indications for Use below. | | ? |
| The GI Genius™ system is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius™ computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
K261369
Page 1 of 6
# 510(k) Summary
510(k) Reference: K261369
Device Name: GI Genius™ Module 300 (GGM300-US); ColonPRO™ US (CPRO403S-US).
Type of 510(k) submission: Special
Date of submission: 24 April 2026
510(k) Owner and Submitter: Cosmo Artificial Intelligence - AI Ltd
Riverside II, Sir John Rogerson’s Quay
Dublin D02 KV60
Ireland
FDA Establishment Reg. Number: 3018899987
Specification Developer: Linkverse Srl
via Ostiense 131/L
00154 Rome, Italy
Owner/Operator Reg. Number: 3018901422
510(k) Application Correspondent: Roger Gray
VP Quality and Regulatory
Donawa Lifescience Consulting
Piazza Albania 10
00153 Rome, Italy
Phone: +39 06 578 2665
Email: rgray@donawa.com
FDA Product Code: QNP
FDA Regulation Number: 21 CFR 876.1520
FDA Classification Name: Gastrointestinal lesion software detection system
Classification Panel: Gastroenterology and Urology
FDA Classification: Class II
# Indications for Use:
The GI Genius™ system is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius™ computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.
Special 510(k) for GI Genius
Page 1 of 6
{5}
K261369
Page 2 of 6
Cosmo
The indications for use statement is the same as that of the original unmodified (predicate) device, as cleared under K241887.
## Device Description:
GI Genius™ is an artificial intelligence-based device that has been trained to process colonoscopy images containing regions consistent with colorectal lesions like polyps, including those with flat (non-polypoid) morphology.
The GI Genius™ system is composed of software (namely, ColonPRO™ US) and hardware (namely, GI Genius™ Module 300).
GI Genius™ Module 300 is compatible with Video Processors featuring SDI (SMPTE 259M), HD-SDI (SMPTE 292M), 3G/SDI (SMPTE 424 M) output ports and endoscopic display monitors featuring SDI (SMPTE 259M), HD-SDI (SMPTE 292M), 3G/SDI (SMPTE 424 M) input ports. GI Genius™ Module 300 is also compatible with Video Processors featuring the 4K UHD standard (12G-SDI SMPTE ST 2082).
The GI Genius™ system is connected between the video processor and the endoscopic display monitor. When first switched on, the endoscopic field of view is clearly identified by four corner markers, and a blinking green square indicator appears on the connected endoscopic display monitor to state that the system is ready to function.
During live video streaming of the endoscopic video image, GI Genius™ system generates a video output on the endoscopic display monitor that contains the original live video together with superimposed green square markers that will appear when a polyp or other lesion of interest is detected, accompanied by a short sound. These markers will not be visible when no lesion detection occurs.
The operating principle of the subject device is identical to that of the predicate device, this being a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device includes hardware to support interfacing with video endoscopy systems.
The baseline clinical validation for the subject device was conducted and reviewed in DEN200055 and is still applicable to the version of the device that is the subject of this submission.
## Design changes:
This Special 510(k) submission describes the design changes incorporated into GI Genius™ system following FDA clearance under K241887. The device software version number as cleared under K241887 was 4.0.0-S and Operating System version was 1.0.0; the device software version that is subject to this Special 510(k) is 4.0.3-S and Operating System version is 1.0.3.
Two design changes are detailed in this submission that fall within the FDA guideline for submittal of a new 510(k). The first is to allow the use of background non-medical software in separate 'software containers' at the same time as the medical lesion detection software ColonPRO™ US in its separate container. The second is to allow the existing connectivity features of the device to be used for user-interaction with the background software containers via external communication devices. Both design changes introduce 'other functions' in the meaning of the FDA guideline on multiple function device products, given that they are unrelated to the medical application of the subject device, which is the same as in the predicate device.
The proposed first change is based on the device's internal software features that allow more than one software container to run at the same time. In the unmodified version of the device there is only one software
Special 510(k) for GI Genius
Page 2 of 6
{6}
C Cosmo
K261369
Page 3 of 6
container being executed, that is ColonPRO™ US. Nonetheless, the device is also capable of managing the concurrent execution of more software containers as an additional non-medical function, thus qualifying the proposed subject device as a multiple function device product with a multicontainer management functionality that is available to users.
The multicontainer management functionality is intended to allow the user to operate non-medical device software, in real time, that can support daily manual operations commonly executed by the user, such as clinical management, reporting, and health IT functions, thus easing user workload. The multicontainer management functionality allows only the medical software container to be executed in foreground and thus the subject device still provides the same onscreen overlay as the predicate device, while any other software container can be executed in background only, with no output on the main endoscopy display, which has no means of interaction with any background software containers through the subject device user interface. Availability of any software container remains under the exclusive control of the manufacturer.
The proposed second change is to allow use of the current device connectivity features, both local and remote, to allow the user to interact with the background software containers on external devices and exchange and transfer data in real time. The unmodified device already features connectivity functionalities, but they have previously been used only by the manufacturer for service activities. In the proposed subject device, connections will be made available to users and to authorized third parties for exchange of information and data, under conditions and limitations defined exclusively by the manufacturer.
The two changes will result in an update of the labelling of the predicate device to inform the user of the multicontainer management functionality and of its connectivity features, as newly available functions of the device.
These design changes do not affect the intended use of the GI Genius™ system, nor do they modify the algorithmic function, performance, or clinical output of the medical software container ColonPRO™ US. The fundamental scientific technology remains unchanged compared to the predicate device cleared under K241887.
## Non-clinical testing:
The following non-clinical verification/validation activities have been completed:
- Risk mitigation measures identified during Risk Management have been successfully verified or validated, as applicable, with specific reference to:
- Verification and validation that resources management controls ensure that execution of background software containers does not impact the performance, timing, image processing, or output generation of the medical software container ColonPRO™ US. The foreground medical software container is always executed with prioritized system resources and with exclusive control of the foreground execution environment.
- Verification and validation that a failure or malfunction of a background software container or associated connectivity feature cannot directly or indirectly alter the medical function and output provided by ColonPRO™ US.
- Verification and validation that built-in design features prevent presence of critical tasks that can lead the user to use errors when the device is used for its established medical purpose.
- Verification that labeling updates inform the users of the availability, purpose, and limitations of the multicontainer management functionality and associated connectivity features. Clear distinction between the medical and non-medical functions and statement that background software containers do not provide medical information, clinical decision support, or user interaction through the GI Genius™ system interface.
Special 510(k) for GI Genius
Page 3 of 6
{7}
K261369
Page 4 of 6
Cosmo
- Cybersecurity of activated communication and data transmission has been included and risk mitigation measures introduced and successfully verified or validated, as applicable, with specific reference to:
- Verification and validation that external connections do not permit control of, modification to, or interaction with the foreground medical software container.
- Verification and validation that the potential attack surface is addressed through software architecture controls, access limitations, and security monitoring measures.
- Verification and validation that container isolation failure, resource contention, and container coexistence is addressed through software architecture controls, access limitations, and security monitoring measures.
- Verification of the software at the system level. Each element of the SRS was tested and found to meet specified requirements, testing 37 Units and 25 Items, encompassing 206 software requirements.
- Validation of the software at the user level has been carried out, incorporating tests sufficient for the validation of the SRS.
The results of the above testing aid demonstration of substantial equivalence of the subject device with the predicate device, as the same methodology and test protocols have been used where applicable.
## Substantial equivalence:
The predicate device for the subject device is the pre-modification version of the same device, GI Genius™ Module 300 with software ColonPRO™ 4.0, FDA-cleared under K241887 on 25 July 2024:
| Predicate Device: | GI Genius™ Module 300 with ColonPRO™ 4.0 |
| --- | --- |
| Sponsor: | Cosmo Artificial Intelligence - AI Ltd |
| 510(k) Number: | K241887 |
| Clearance Date: | 25 July 2024 |
| FDA Product Code: | QNP |
| Classification Name: | Gastrointestinal lesion software detection system |
| Regulation No: | 21 CFR 876.1520 |
| Class: | II |
## Predicate device comparison table:
The two proposed changes relate to the update of labeling to inform the user of the available other functions given by the multicontainer management functionality and associated connectivity features, whereas any other technical feature, use, and medical claim between the subject and predicate device remain the same.
Table 1 provides evidence of substantial equivalence of the subject device with the predicate device.
| Table 1: Predicate device comparison table | | | |
| --- | --- | --- | --- |
| Characteristic | Subject device | Predicate device | Comparison |
| Device name | GI Genius™ Module 300 (GGM300-US, operating system v.1.0.3) ColonPRO™ US (CPRO403S-US) | GI Genius™ Module 300 (GGM300-US, operating system v.1.0.0) ColonPRO™ 4.0 (CPRO40S-US) | Minor software updates to internal software functionalities |
| Manufacturer | Linkverse S.r.l., Italy | Linkverse S.r.l., Italy | Same |
| FDA clearance | This submission | K241887 | N/A |
| FDA Reg name | Gastrointestinal lesion software detection system | Gastrointestinal lesion software detection system | Same |
| FDA Reg # | 21 CFR 876.1520 | 21 CFR 876.1520 | Same |
Special 510(k) for GI Genius
Page 4 of 6
{8}
K261369
Page 5 of 6
Cosmo
| Table 1: Predicate device comparison table | | | |
| --- | --- | --- | --- |
| Characteristic | Subject device | Predicate device | Comparison |
| FDA Product Code | QNP | QNP | Same |
| Indications for Use | The GI Genius™ system is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius™ computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making. | The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making. | Same |
| Video delay, signal in to signal out | 29 μs (< 5.75 ms limit) | 29 μs (< 5.75 ms limit) | Same |
| Lesion-based sensitivity | 88.07% | 88.07% | Same |
| Frame level performance (150 videos / 338 polyps) | True positive: 277,738 True negative: 5,248,406 False positive: 95,391 False negative: 184,052 | True positive: 277,738 True negative: 5,248,406 False positive: 95,391 False negative: 184,052 | Same |
| True positive rate per frame | Mean: 60.14 % % of polyps: 100 % | Mean: 60.14 % % of polyps: 100 % | Same |
| False positive rate per frame | Mean: 1.79% | Mean: 1.79% | Same |
| Frame-Based TPr/FPr ROC curve, AOC | 0.826 | 0.826 | Same |
| False positive clusters per patient | < 500 ms: 97 > 500 ms: 10 | < 500 ms: 97 > 500 ms: 10 | Same |
| Additional video processor | none | N/A | Same |
| Software other function | Procedure Highlights function Multicontainer management | Procedure Highlights function | Background software containers available to users through an internal non-medical function of the device |
| Electrical safety | IEC/EN 60601-1 | IEC/EN 60601-1 | Same |
| Electromagnetic compatibility | IEC/EN 60601-1-2 | IEC/EN 60601-1-2 | Same |
| LAN port USB port QSFP port | Yes, functional to user | Yes, non-functional to user | Connectivity available to users as non-medical function |
## Conclusion
The subject and predicate devices have identical indications for use and fundamental technological characteristics. The subject device and predicate device share the same medical performance as no change has been made to the software functions related to detection of suspected lesions. Verification and validation tests performed on risk mitigation measures for both safety of use and cybersecurity demonstrate that the
Special 510(k) for GI Genius
Page 5 of 6
{9}
K261369
Page 6 of 6
Cosmo
subject device is substantially equivalent to the predicate device, which is already in interstate commerce within the USA. Therefore, the subject device is as safe and effective as the predicate device.
Special 510(k) for GI Genius
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.