← Product Code [DQK](/productcode/DQK) · K261329

# Intra-Operative Positioning System (IOPS) (MC-1);IOPS Viewpoint Simple Curve Catheter, 75cm (C00751);IOPS Viewpoint Simple Curve Catheter, 125cm (C01251);IOPS Viewpoint Double Curve Catheter, 75cm (C00752);IOPS Viewpoint Double Curve Catheter, 125cm (C02152);IOPS Guidewire 2 (ATW-2);IOPS Fiducial Tracking Pad (T02111);IOPS Guidewire Handle (H01035) (K261329)

_Centerline Biomedical · DQK · Jun 10, 2026 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K261329

## Device Facts

- **Applicant:** Centerline Biomedical
- **Product Code:** [DQK](/productcode/DQK.md)
- **Decision Date:** Jun 10, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1425
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.

## Device Story

IOPS is an electromagnetic tracking system for endovascular interventions; utilizes previously acquired CT scan data to create 3D vascular maps; tracks sensor-equipped catheters and guidewires in real time; superimposes device position onto vascular map on an onboard monitor; used in clinical settings by physicians; acts as an adjunct to fluoroscopy to assist navigation; does not provide diagnosis; system includes cart, tracking unit (SIU/SCU), field generator, and accessories (catheters, guidewires, fiducial tracking pad).

## Clinical Evidence

No clinical or animal testing conducted. Substantial equivalence supported by system/software verification testing, packaging integrity, catheter dimensional measurements, sensor functionality, tensile strength, and simulated use testing.

## Technological Characteristics

Electromagnetic tracking system; components include cart, Sensor Interface Unit (SIU), System Control Unit (SCU), and Field Generator. Accessories: 6 Fr sensorized catheters, 0.035" sensorized guidewires, fiducial tracking pad. Sterilization: Ethylene Oxide (EtO). Connectivity: USB ports for image import. Software: v1.6.

## Regulatory Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

## Predicate Devices

- Intra-Operative Positioning System (IOPS®) ([K243842](/device/K243842.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD & DRUG ADMINISTRATION

June 10, 2026

Centerline Biomedical
Carroll Martin
Sr. Regulatory Affairs Manager
4535 Renaissance Pkwy.
Warrensville Heights, Ohio 44128

Re: K261329

Trade/Device Name: IOPS® (Intra-Operative Positioning System)
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable diagnostic computer
Regulatory Class: Class II
Product Code: DQK, DQY, DQX
Dated: April 22, 2026
Received: April 22, 2026

Dear Carroll Martin:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K261329 - Carroll Martin

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K261329 - Carroll Martin

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for: **MARCO CANNELLA -S**

Aneesh Deoras  
Assistant Director  
Division of Cardiac Electrophysiology,  
Diagnostics, and Monitoring Devices  
Office of Cardiovascular Devices  
Office of Product Evaluation and Quality  
Center for Devices and Radiological Health

Enclosure

{3}

# Indications for Use

|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261329 | ?  |
| --- | --- | --- |
|  Please provide the device trade name(s). |   | ?  |
|  IOPS® (Intra-Operative Positioning System)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

Centerline Biomedical  
510(k) PREMARKET NOTIFICATION Intra-  
Operative Positioning System (IOPS®)

Centerline  
Biomedical

# **510(k) Summary for the IOPS® (Intra-Operative Positioning System)**

# **I. Device Name**

|  Trade Name | IOPS® (Intra-Operative Positioning System)  |
| --- | --- |
|  Device Class | Class II  |
|  Common/Usual Name | Programmable Diagnostic Computer  |
|  Classification Name | Computer, diagnostic, programmable  |
|  Product Codes | DQK, DQY, DQX  |
|  Regulation Number | 870.1425  |

# **II. Submitter Information**

|  Applicant Name | Centerline Biomedical, Inc.  |
| --- | --- |
|  Applicant Address | 4535 Renaissance Parkway, Warrenville Heights, Ohio 44120  |
|  Telephone Number | 216-200-6366  |
|  Date of Preparation | 22APR2026  |
|  Contact Person | Carroll L. Martin, Sr. Regulatory Affairs Manager  |

# **III. Predicate Device**

The predicate device to which substantial equivalence is claimed is the Intra-Operative Positioning System (IOPS®), cleared under K243842.

# **IV. Device Description**

The IOPS® (Intra-Operative Positioning System) displays the position and orientation of sensor-equipped IOPS® guidewires and IOPS® catheters utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously-acquired scan data (CT). The Intra-Operative Positioning System (IOPS) tracks the location and orientation of the sensors in real time, superimposing navigation of the IOPS Catheter and IOPS Guidewire to the patient's vascular map. The system consists of a surgical navigation technology and a number of associated accessories. The navigation technology is a non-contact, reusable, multi-patient use device. The associated accessories are single use devices provided sterile (EtO). The Intra-Operative Positioning System (IOPS) is indicated for the evaluation of vascular anatomy as captured via 3D modeling from previously-acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic, and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The Intra-Operative Positioning System (IOPS) does not make a diagnosis.

The associated accessories include:

- Guidewire

510(k) K261329

Page 1 of 8

{5}

Centerline Biomedical  
 510(k) PREMARKET NOTIFICATION Intra-  
 Operative Positioning System (IOPS®)

Centerline  
 Biomedical

- Fiducial Tracking Pad

The system consists of three major sections: the cart, the tracking system (Sensor Interface Unit (SIU), System Control Unit (SCU) and Field Generator) and the accessories.

# **V. Indications for Use**

The IOPS (Intra-Operative Position System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.

# **VI. Substantial Equivalence**

|  Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion  |
| --- | --- | --- | --- |
|  Intended Use | The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis. | The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis. | Identical  |
|  Onboard Monitor | Displays patient maps and superimposed images | Displays patient maps and superimposed images | Identical  |
|  Computer | Provides an operating system | Provides an operating system | Identical  |
|  Keyboard | Allows data input | Allows data input | Identical  |
|  Mouse | Allows data input | Allows data input | Identical  |

510(k) K261329

Page 2 of 8

{6}

Centerline Biomedical  
 510(k) PREMARKET NOTIFICATION Intra-  
 Operative Positioning System (IOPS®)

|  Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion  |
| --- | --- | --- | --- |
|  Uninterruptible power supply (UPS) | Provides power when electricity is not available | Provides power when electricity is not available | Identical  |
|  Power Requirements | 120V, 60HZ | 120V, 60HZ | Identical  |
|  Voltage Transformer | Not present | Not present | Identical  |
|  Video Splitter | The video out enables the Mobile Cart to simultaneously display the user interface on the onboard monitor and provide a video signal for an external display. | The video out enables the Mobile Cart to simultaneously display the user interface on the onboard monitor and provide a video signal for an external display. | Identical  |
|  Cables | Allows data transfer | Allows data transfer | Identical  |
|  IOPS Software Application | Software which creates vascular map used to show location of sensorized catheter and guidewires in real time. Vascular map is generated from a preoperative CT scan which is processed to segment the vasculature and perform centerline and surface analysis. Version 1.5 | Software which creates vascular map used to show location of sensorized catheter and guidewires in real time. Vascular map is generated from a preoperative CT scan which is processed to segment the vasculature and perform centerline and surface analysis. Version 1.6 | Different The new software version addresses anomalies discovered through the use of device and also provides new features to the user. These changes do not affect the safety and effectiveness of the device. They only address inconvenience to the user and provide for desired functionality  |

510(k) K261329

Page 3 of 8

{7}

Centerline Biomedical  
 510(k) PREMARKET NOTIFICATION Intra-  
 Operative Positioning System (IOPS®)

|  Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion  |
| --- | --- | --- | --- |
|  System Control Unit (SCU) | Collects information from the SIUs, calculates the position and orientation of each sensor and interfaces with the host computer. | Collects information from the SIUs, calculates the position and orientation of each sensor and interfaces with the host computer. | Identical  |
|  Sensor Interface Unit (SIU) | Amplifies and digitizes the electrical signals from the sensors and provides an increased distance between the SCU and sensors, while minimizing the potential for data noise | Amplifies and digitizes the electrical signals from the sensors and provides an increased distance between the SCU and sensors, while minimizing the potential for data noise | Identical  |
|  Electromagnetic Field Generator | Emits a low-intensity, varying electromagnetic field and establishes the position of the tracking volume | Emits a low-intensity, varying electromagnetic field and establishes the position of the tracking volume | Identical  |
|  Mounting brackets | Affixes field generator to operating table | Affixes field generator to operating table | Identical  |
|  USB Ports | Present. Allows the user to copy fluoroscan images from a flash drive that was obtained from the user’s fluoroscan. These images are used with IOPS to create vessel and bones models for use during procedures. | Present. Allows the user to copy fluoroscan images from a flash drive that was obtained from the user’s fluoroscan. These images are used with IOPS to create vessel and bones models for use during procedures. | Identical  |

510(k) K261329

Page 4 of 8

{8}

Centerline Biomedical

510(k) PREMARKET NOTIFICATION Intra-

Operative Positioning System (IOPS®)

[LOGO]

Centerline

Biomedical

|  Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion  |
| --- | --- | --- | --- |
|  Functionality | The IOPS (MC-1) displays the position and orientation of sensor equipped catheters, guidewires, and tracking pad utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously acquired CT scan data. IOPS registers the location and orientation of the sensors in real time, superimposing navigation of the catheters and guidewires to the patient's vascular map displayed on a monitor. The system is for use as an adjunct to fluoroscopy and does not make a diagnosis. | The IOPS (MC-1) displays the position and orientation of sensor equipped catheters, guidewires, and tracking pad utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously acquired CT scan data. IOPS registers the location and orientation of the sensors in real time, superimposing navigation of the catheters and guidewires to the patient's vascular map displayed on a monitor. The system is for use as an adjunct to fluoroscopy and does not make a diagnosis. | Identical  |

510(k) K261329

Page 5 of 8

{9}

Centerline Biomedical  
 510(k) PREMARKET NOTIFICATION Intra-  
 Operative Positioning System (IOPS®)

|  Device Accessories | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion  |
| --- | --- | --- | --- |
|  IOPS Viewpoint Simple Curve Catheters (C00751; C01251) and Double Curve Catheters (C00752; C01252) | 6 Fr catheters equipped with multiple tracking sensors allowing the IOPS to detect and visualize the catheter tip position and shape, in real time, on a 3D rendering of the patient’s vascular map. They are available in two tip shape configurations: simple curve and double curve. Each curve is offered in two different working lengths (75cm and 125cm). | 6 Fr catheters equipped with multiple tracking sensors allowing the IOPS to detect and visualize the catheter tip position and shape, in real time, on a 3D rendering of the patient’s vascular map. They are available in two tip shape configurations: simple curve and double curve. Each curve is offered in two different working lengths (75cm and 125cm). | Identical  |
|   |  Shelf Life: 1 year | Shelf life: 2 years | Different Testing showed that extension of shelf life has no effect on safety or effectiveness.  |
|   |  Sterilization: EO in a rigid chamber. | Sterilization: EO in a flexible bag system. Data provided in section 002_Device Description and Principles of Operation in this submission. | Different The change has no impact on the safety or performance of the device. Requirements for EO and ECH residuals, and device functionality were met.  |

510(k) K261329

Page 6 of 8

{10}

Centerline Biomedical

510(k) PREMARKET NOTIFICATION Intra-

Operative Positioning System (IOPS®)

[LOGO]

Centerline

Biomedical

|  Device Accessories | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion  |
| --- | --- | --- | --- |
|  IOPS Guidewire (ATW-2) | Guidewire is a 0.035" diameter wire with a single sensor at the distal end used to navigate through vasculature to facilitate placement of a catheter. | Guidewire is a 0.035" diameter wire with a single sensor at the distal end used to navigate through vasculature to facilitate placement of a catheter. | Identical  |
|   |  Shelf life: 2 years | Shelf life: 2 years | Identical  |
|   |  Sterilization: EO in a rigid chamber. | Sterilization: EO in a rigid chamber. | Identical  |
|  IOPS Fiducial Tracking Pad (T02111) | Fiducial Tracking Pad is equipped with a single tracking sensor allowing IOPS to track gross patient motion to allow maintenance of patient registration during endovascular procedure. It will not track minor patient motion such as breathing or cardiac movement. The tracking pad is equipped with radiopaque beads which allow registration of a conebeam CT scan of the patient in their current position to a previously acquired CT scan. | Fiducial Tracking Pad is equipped with a single tracking sensor allowing IOPS to track gross patient motion to allow maintenance of patient registration during endovascular procedure. It will not track minor patient motion such as breathing or cardiac movement. The tracking pad is equipped with radiopaque beads which allow registration of a conebeam CT scan of the patient in their current position to a previously acquired CT scan. | Identical  |
|   |  Shelf life: 3 years | Shelf life: 3 years | Identical  |
|   |  Sterilization: EO in a flexible bag: (K242133) | Sterilization: EO in a flexible bag: (K242133) | Identical  |

510(k) K261329

Page 7 of 8

{11}

Centerline Biomedical  
 510(k) PREMARKET NOTIFICATION Intra-  
 Operative Positioning System (IOPS®)

Centerline  
 Biomedical

|  Device Accessories | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion  |
| --- | --- | --- | --- |
|  IOPS Guidewire Handle (H01035) | Guidewire Handle is a non-sensorized handle that connects to an IOPS sensorized guidewire to allow detection and visualization of the guidewire tip position, in real time, on a 3D rendering of the patient’s vascular map. It consists of a handle housing which encases a cable assembly that connects to a sensorized IOPS Guidewire. | Guidewire Handle is a non-sensorized handle that connects to an IOPS sensorized guidewire to allow detection and visualization of the guidewire tip position, in real time, on a 3D rendering of the patient’s vascular map. It consists of a handle housing which encases a cable assembly that connects to a sensorized IOPS Guidewire. | Identical  |
|   |  Shelf-Life: 3 years (K242133) | Shelf-Life: 3 years (K242133) | Identical  |
|   |  Sterilization: EO in a flexible bag (K242133) | Sterilization: EO in a flexible bag (K242133) | Identical  |

#### VII. Performance Testing

System and software verification testing was conducted to establish equivalency to the predicate device in safety and effectiveness and to ensure the Intra-operative Positioning System met established requirements.

Also, packaging and catheter visual inspection, package integrity (bubble leak, seal strength) catheter dimensional measurement, sensor functionality, sensor placement, tensile strength, and simulated use testing were conducted to verify the extension of catheter shelf life.

#### VIII. Animal and Clinical Testing

No animal or clinical testing was conducted.

#### IX. Conclusions

The modifications made to the Intra-Operative Positioning System do not raise new or different questions of safety or effectiveness. The successful completion of non-clinical testing demonstrates that the Intra-Operative Positioning System performs as intended and is substantially equivalent to the predicate device.

510(k) K261329

Page 8 of 8

---

**Source:** [https://fda-staging.innolitics.com/device/K261329](https://fda-staging.innolitics.com/device/K261329)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
