The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
Device Story
IOPS is an electromagnetic tracking system for endovascular interventions; utilizes previously acquired CT scan data to create 3D vascular maps; tracks sensor-equipped catheters and guidewires in real time; superimposes device position onto vascular map on an onboard monitor; used in clinical settings by physicians; acts as an adjunct to fluoroscopy to assist navigation; does not provide diagnosis; system includes cart, tracking unit (SIU/SCU), field generator, and accessories (catheters, guidewires, fiducial tracking pad).
Clinical Evidence
No clinical or animal testing conducted. Substantial equivalence supported by system/software verification testing, packaging integrity, catheter dimensional measurements, sensor functionality, tensile strength, and simulated use testing.
Technological Characteristics
Electromagnetic tracking system; components include cart, Sensor Interface Unit (SIU), System Control Unit (SCU), and Field Generator. Accessories: 6 Fr sensorized catheters, 0.035" sensorized guidewires, fiducial tracking pad. Sterilization: Ethylene Oxide (EtO). Connectivity: USB ports for image import. Software: v1.6.
Indications for Use
Indicated for patients undergoing endovascular interventions in peripheral, aortic, and aortic side branch vasculature. Used for real-time tip positioning and navigation of sensor-equipped catheters and guidewires as an adjunct to fluoroscopy. No specific age or gender restrictions.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Intra-Operative Positioning System (IOPS®) (K243842)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 10, 2026
Centerline Biomedical
Carroll Martin
Sr. Regulatory Affairs Manager
4535 Renaissance Pkwy.
Warrensville Heights, Ohio 44128
Re: K261329
Trade/Device Name: IOPS® (Intra-Operative Positioning System)
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable diagnostic computer
Regulatory Class: Class II
Product Code: DQK, DQY, DQX
Dated: April 22, 2026
Received: April 22, 2026
Dear Carroll Martin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261329 - Carroll Martin
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261329 - Carroll Martin
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for: **MARCO CANNELLA -S**
Aneesh Deoras
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261329 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| IOPS® (Intra-Operative Positioning System) | | |
| Please provide your Indications for Use below. | | ? |
| The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Centerline Biomedical
510(k) PREMARKET NOTIFICATION Intra-
Operative Positioning System (IOPS®)
Centerline
Biomedical
# **510(k) Summary for the IOPS® (Intra-Operative Positioning System)**
# **I. Device Name**
| Trade Name | IOPS® (Intra-Operative Positioning System) |
| --- | --- |
| Device Class | Class II |
| Common/Usual Name | Programmable Diagnostic Computer |
| Classification Name | Computer, diagnostic, programmable |
| Product Codes | DQK, DQY, DQX |
| Regulation Number | 870.1425 |
# **II. Submitter Information**
| Applicant Name | Centerline Biomedical, Inc. |
| --- | --- |
| Applicant Address | 4535 Renaissance Parkway, Warrenville Heights, Ohio 44120 |
| Telephone Number | 216-200-6366 |
| Date of Preparation | 22APR2026 |
| Contact Person | Carroll L. Martin, Sr. Regulatory Affairs Manager |
# **III. Predicate Device**
The predicate device to which substantial equivalence is claimed is the Intra-Operative Positioning System (IOPS®), cleared under K243842.
# **IV. Device Description**
The IOPS® (Intra-Operative Positioning System) displays the position and orientation of sensor-equipped IOPS® guidewires and IOPS® catheters utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously-acquired scan data (CT). The Intra-Operative Positioning System (IOPS) tracks the location and orientation of the sensors in real time, superimposing navigation of the IOPS Catheter and IOPS Guidewire to the patient's vascular map. The system consists of a surgical navigation technology and a number of associated accessories. The navigation technology is a non-contact, reusable, multi-patient use device. The associated accessories are single use devices provided sterile (EtO). The Intra-Operative Positioning System (IOPS) is indicated for the evaluation of vascular anatomy as captured via 3D modeling from previously-acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic, and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The Intra-Operative Positioning System (IOPS) does not make a diagnosis.
The associated accessories include:
- Guidewire
510(k) K261329
Page 1 of 8
{5}
Centerline Biomedical
510(k) PREMARKET NOTIFICATION Intra-
Operative Positioning System (IOPS®)
Centerline
Biomedical
- Fiducial Tracking Pad
The system consists of three major sections: the cart, the tracking system (Sensor Interface Unit (SIU), System Control Unit (SCU) and Field Generator) and the accessories.
# **V. Indications for Use**
The IOPS (Intra-Operative Position System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
# **VI. Substantial Equivalence**
| Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion |
| --- | --- | --- | --- |
| Intended Use | The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis. | The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis. | Identical |
| Onboard Monitor | Displays patient maps and superimposed images | Displays patient maps and superimposed images | Identical |
| Computer | Provides an operating system | Provides an operating system | Identical |
| Keyboard | Allows data input | Allows data input | Identical |
| Mouse | Allows data input | Allows data input | Identical |
510(k) K261329
Page 2 of 8
{6}
Centerline Biomedical
510(k) PREMARKET NOTIFICATION Intra-
Operative Positioning System (IOPS®)
| Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion |
| --- | --- | --- | --- |
| Uninterruptible power supply (UPS) | Provides power when electricity is not available | Provides power when electricity is not available | Identical |
| Power Requirements | 120V, 60HZ | 120V, 60HZ | Identical |
| Voltage Transformer | Not present | Not present | Identical |
| Video Splitter | The video out enables the Mobile Cart to simultaneously display the user interface on the onboard monitor and provide a video signal for an external display. | The video out enables the Mobile Cart to simultaneously display the user interface on the onboard monitor and provide a video signal for an external display. | Identical |
| Cables | Allows data transfer | Allows data transfer | Identical |
| IOPS Software Application | Software which creates vascular map used to show location of sensorized catheter and guidewires in real time. Vascular map is generated from a preoperative CT scan which is processed to segment the vasculature and perform centerline and surface analysis. Version 1.5 | Software which creates vascular map used to show location of sensorized catheter and guidewires in real time. Vascular map is generated from a preoperative CT scan which is processed to segment the vasculature and perform centerline and surface analysis. Version 1.6 | Different The new software version addresses anomalies discovered through the use of device and also provides new features to the user. These changes do not affect the safety and effectiveness of the device. They only address inconvenience to the user and provide for desired functionality |
510(k) K261329
Page 3 of 8
{7}
Centerline Biomedical
510(k) PREMARKET NOTIFICATION Intra-
Operative Positioning System (IOPS®)
| Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion |
| --- | --- | --- | --- |
| System Control Unit (SCU) | Collects information from the SIUs, calculates the position and orientation of each sensor and interfaces with the host computer. | Collects information from the SIUs, calculates the position and orientation of each sensor and interfaces with the host computer. | Identical |
| Sensor Interface Unit (SIU) | Amplifies and digitizes the electrical signals from the sensors and provides an increased distance between the SCU and sensors, while minimizing the potential for data noise | Amplifies and digitizes the electrical signals from the sensors and provides an increased distance between the SCU and sensors, while minimizing the potential for data noise | Identical |
| Electromagnetic Field Generator | Emits a low-intensity, varying electromagnetic field and establishes the position of the tracking volume | Emits a low-intensity, varying electromagnetic field and establishes the position of the tracking volume | Identical |
| Mounting brackets | Affixes field generator to operating table | Affixes field generator to operating table | Identical |
| USB Ports | Present. Allows the user to copy fluoroscan images from a flash drive that was obtained from the user’s fluoroscan. These images are used with IOPS to create vessel and bones models for use during procedures. | Present. Allows the user to copy fluoroscan images from a flash drive that was obtained from the user’s fluoroscan. These images are used with IOPS to create vessel and bones models for use during procedures. | Identical |
510(k) K261329
Page 4 of 8
{8}
Centerline Biomedical
510(k) PREMARKET NOTIFICATION Intra-
Operative Positioning System (IOPS®)
[LOGO]
Centerline
Biomedical
| Device Features | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion |
| --- | --- | --- | --- |
| Functionality | The IOPS (MC-1) displays the position and orientation of sensor equipped catheters, guidewires, and tracking pad utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously acquired CT scan data. IOPS registers the location and orientation of the sensors in real time, superimposing navigation of the catheters and guidewires to the patient's vascular map displayed on a monitor. The system is for use as an adjunct to fluoroscopy and does not make a diagnosis. | The IOPS (MC-1) displays the position and orientation of sensor equipped catheters, guidewires, and tracking pad utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously acquired CT scan data. IOPS registers the location and orientation of the sensors in real time, superimposing navigation of the catheters and guidewires to the patient's vascular map displayed on a monitor. The system is for use as an adjunct to fluoroscopy and does not make a diagnosis. | Identical |
510(k) K261329
Page 5 of 8
{9}
Centerline Biomedical
510(k) PREMARKET NOTIFICATION Intra-
Operative Positioning System (IOPS®)
| Device Accessories | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion |
| --- | --- | --- | --- |
| IOPS Viewpoint Simple Curve Catheters (C00751; C01251) and Double Curve Catheters (C00752; C01252) | 6 Fr catheters equipped with multiple tracking sensors allowing the IOPS to detect and visualize the catheter tip position and shape, in real time, on a 3D rendering of the patient’s vascular map. They are available in two tip shape configurations: simple curve and double curve. Each curve is offered in two different working lengths (75cm and 125cm). | 6 Fr catheters equipped with multiple tracking sensors allowing the IOPS to detect and visualize the catheter tip position and shape, in real time, on a 3D rendering of the patient’s vascular map. They are available in two tip shape configurations: simple curve and double curve. Each curve is offered in two different working lengths (75cm and 125cm). | Identical |
| | Shelf Life: 1 year | Shelf life: 2 years | Different Testing showed that extension of shelf life has no effect on safety or effectiveness. |
| | Sterilization: EO in a rigid chamber. | Sterilization: EO in a flexible bag system. Data provided in section 002_Device Description and Principles of Operation in this submission. | Different The change has no impact on the safety or performance of the device. Requirements for EO and ECH residuals, and device functionality were met. |
510(k) K261329
Page 6 of 8
{10}
Centerline Biomedical
510(k) PREMARKET NOTIFICATION Intra-
Operative Positioning System (IOPS®)
[LOGO]
Centerline
Biomedical
| Device Accessories | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion |
| --- | --- | --- | --- |
| IOPS Guidewire (ATW-2) | Guidewire is a 0.035" diameter wire with a single sensor at the distal end used to navigate through vasculature to facilitate placement of a catheter. | Guidewire is a 0.035" diameter wire with a single sensor at the distal end used to navigate through vasculature to facilitate placement of a catheter. | Identical |
| | Shelf life: 2 years | Shelf life: 2 years | Identical |
| | Sterilization: EO in a rigid chamber. | Sterilization: EO in a rigid chamber. | Identical |
| IOPS Fiducial Tracking Pad (T02111) | Fiducial Tracking Pad is equipped with a single tracking sensor allowing IOPS to track gross patient motion to allow maintenance of patient registration during endovascular procedure. It will not track minor patient motion such as breathing or cardiac movement. The tracking pad is equipped with radiopaque beads which allow registration of a conebeam CT scan of the patient in their current position to a previously acquired CT scan. | Fiducial Tracking Pad is equipped with a single tracking sensor allowing IOPS to track gross patient motion to allow maintenance of patient registration during endovascular procedure. It will not track minor patient motion such as breathing or cardiac movement. The tracking pad is equipped with radiopaque beads which allow registration of a conebeam CT scan of the patient in their current position to a previously acquired CT scan. | Identical |
| | Shelf life: 3 years | Shelf life: 3 years | Identical |
| | Sterilization: EO in a flexible bag: (K242133) | Sterilization: EO in a flexible bag: (K242133) | Identical |
510(k) K261329
Page 7 of 8
{11}
Centerline Biomedical
510(k) PREMARKET NOTIFICATION Intra-
Operative Positioning System (IOPS®)
Centerline
Biomedical
| Device Accessories | Predicate Device Intra-Operative Positioning System, K243842 | Subject Device Modified Intra-Operative Positioning System | Equivalence Discussion |
| --- | --- | --- | --- |
| IOPS Guidewire Handle (H01035) | Guidewire Handle is a non-sensorized handle that connects to an IOPS sensorized guidewire to allow detection and visualization of the guidewire tip position, in real time, on a 3D rendering of the patient’s vascular map. It consists of a handle housing which encases a cable assembly that connects to a sensorized IOPS Guidewire. | Guidewire Handle is a non-sensorized handle that connects to an IOPS sensorized guidewire to allow detection and visualization of the guidewire tip position, in real time, on a 3D rendering of the patient’s vascular map. It consists of a handle housing which encases a cable assembly that connects to a sensorized IOPS Guidewire. | Identical |
| | Shelf-Life: 3 years (K242133) | Shelf-Life: 3 years (K242133) | Identical |
| | Sterilization: EO in a flexible bag (K242133) | Sterilization: EO in a flexible bag (K242133) | Identical |
#### VII. Performance Testing
System and software verification testing was conducted to establish equivalency to the predicate device in safety and effectiveness and to ensure the Intra-operative Positioning System met established requirements.
Also, packaging and catheter visual inspection, package integrity (bubble leak, seal strength) catheter dimensional measurement, sensor functionality, sensor placement, tensile strength, and simulated use testing were conducted to verify the extension of catheter shelf life.
#### VIII. Animal and Clinical Testing
No animal or clinical testing was conducted.
#### IX. Conclusions
The modifications made to the Intra-Operative Positioning System do not raise new or different questions of safety or effectiveness. The successful completion of non-clinical testing demonstrates that the Intra-Operative Positioning System performs as intended and is substantially equivalent to the predicate device.
510(k) K261329
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.