# Temporary Keratoprosthesis (QuickSet TKP™) (K261312)

_Eyekon Medical, Inc. · MLP · Jul 24, 2026 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K261312

## Device Facts

- **Applicant:** Eyekon Medical, Inc.
- **Product Code:** [MLP](/productcode/MLP.md)
- **Decision Date:** Jul 24, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.3400
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The QuickSet Temporary Keratoprosthesis (QuickSet TKP™) is a surgical aid to provide a transparent optical pathway that permits closed pars plana vitrectomy in eyes with abnormal corneas.

## Device Story

QuickSet TKP™ is a sterile, single-use temporary keratoprosthesis; replaces pathologic host cornea during combined corneal and vitreoretinal surgery. Device provides transparent optical pathway for posterior segment visualization; maintains intraocular pressure (IOP) during surgery. Used in clinical settings by ophthalmologists; device features peripheral grooves for fixation via horizontal mattress sutures. Input is surgical site; output is clear optical view of intraocular structures. Device facilitates procedures like pars plana vitrectomy, retinal reattachment, and foreign body removal. Post-surgery, device is removed for corneal management. Benefits include preventing treatment delays in ocular trauma cases by enabling visualization through compromised corneas.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), EO sterilization validation (ISO 11135), optical performance (MTF, SEP), and simulated clinical wet lab use. Wet lab testing confirmed watertight sealing at pressures up to 120 mmHg and visualization during complex maneuvers.

## Technological Characteristics

Material: PMMA with proprietary UV absorber. Refractive index: 1.49. Light transmittance: >90%. Dimensions: 12.0 mm collar diameter, 8.0 mm optical cylinder diameter, 0.82 mm thickness. Fixation: 4 grooves for horizontal mattress sutures. Sterilization: Ethylene oxide. Packaging: Tyvek/Mylar pouch. Standards: ISO 10993, ISO 11135, ISO 11737-1, ANSI/AAMI ST72, ASTM D4332/D4169/F1886/F2096/F88/F1980.

## Regulatory Identification

A keratoprosthesis is a device intended to provide a transparent optical pathway through an opacified cornea, either intraoperatively or permanently, in an eye that is not a reasonable candidate for a corneal transplant.

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(2) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” and
(3) “Guidance on 510(k) Submissions for Keratoprostheses.”

## Predicate Devices

- Landers Wide Field Temporary Keratoprosthesis ([K910828](/device/K910828.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD & DRUG ADMINISTRATION

July 24, 2026

Eyekon Medical, Inc.
% Robert Lundberg
President
OneRA, LLC
2043 Evanston Cir.
Corona, California 92881

Re: K261312

Trade/Device Name: Temporary Keratoprosthesis (QuickSet TKP™)

Regulation Number: 21 CFR 886.3400

Regulation Name: Keratoprosthesis

Regulatory Class: Class II

Product Code: MLP

Dated: April 21, 2026

Received: April 21, 2026

Dear Robert Lundberg:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K261312 - Robert Lundberg

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K261312 - Robert Lundberg

Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CLAUDINE H. KRAWCZYK -S

Claudine Krawczyk

Assistant Director

DHT1A: Division of Ophthalmic Devices

OHT1: Office of Ophthalmic, Anesthesia,

Respiratory, ENT, and Dental Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261312 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Temporary Keratoprosthesis (QuickSet TKP™)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The QuickSet Temporary Keratoprosthesis (QuickSet TKP™) is a surgical aid to provide a transparent optical pathway that permits closed pars plana vitrectomy in eyes with abnormal corneas.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |
|  Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ?  |

{4}

EyeKon

510(k) SUMMARY

K261312

Page 1 of 4

## Submitter Information

|  Company: | EyeKon Medical, Inc. 2451 Enterprise Road Clearwater, FL 33763 Phone: 727-793-0170 Fax: 727-799-2212  |
| --- | --- |
|  Contact Person: | Robert Lundberg OneRA, LLC 2043 Evanston Circle Corona, CA 92881 Tel: (951) 741-2776 Email: bob@oneregulatory.com  |
|  Date Prepared: | 23 July 2026  |

## Device Information

|  Device Trade Name: | QuickSet Temporary Keratoprosthesis (QuickSet TKP™)  |
| --- | --- |
|  Device Common Name: | Keratoprosthesis  |
|  Classification Regulation: | 21 CFR 886.3400  |
|  Classification Name: | Keratoprosthesis, Temporary Implant, Surgical Use  |
|  Device Class: | Class II  |
|  Device Regulation Panel: | Ophthalmic  |
|  Product Code: | MLP  |

## Predicate/Reference Devices

|  Device Trade Name: | Landers Wide Field Temporary Keratoprosthesis  |
| --- | --- |
|  Device Common Name: | Keratoprosthesis  |
|  Classification Regulation: | 21 CFR 886.3400  |
|  Classification Name: | Keratoprosthesis, Temporary Implant, Surgical Use  |
|  Device Class: | Class II  |
|  Device Regulation Panel: | Ophthalmic  |
|  Product Code: | MLP  |

## Device Description

The QuickSet TKP™ allows for visualization of intraocular structures during combined corneal and vitreoretinal surgery. This indication is useful in preventing the delay of posterior segment treatment, especially in the setting of ocular trauma where corneal visibility is often limited from corneal blood staining, large edematous corneal wounds, or corneal explosive injury with corneal foreign bodies. Common concurrent vitreoretinal procedures include pars plana vitrectomy with scleral buckling, retinal reattachment, and foreign body removal.

The QuickSet TKP™ is a sterile, single-use, temporary keratoprosthesis intended to replace the pathologic host cornea during surgery to facilitate posterior segment visualization during vitreoretinal surgery and allow intraocular pressure (IOP) to be maintained. Once the posterior segment procedure is completed, the QuickSet TKP™ is removed and penetrating keratoplasty or other appropriate corneal management may be performed.

{5}

EyeKon

510(k) SUMMARY

K261312

Page 2 of 4

The QuickSet TKP™ is manufactured from clinical quality polymethyl methacrylate (PMMA) with a proprietary ultraviolet absorber. The device incorporates a transparent optical element and peripheral features that allow fixation using horizontal mattress sutures placed into corresponding device grooves. The design is intended to provide optical clarity, mechanical stability, ease of insertion and removal, a watertight temporary seal, and posterior segment visualization during the intended intraoperative use.

Material and physical properties include PMMA with proprietary UV absorber, refractive index of 1.49, light transmittance greater than 90%, tensile strength of 73, flexural strength of 118, and Rockwell hardness of 105.

## Indications for Use

The QuickSet Temporary Keratoprosthesis (QuickSet TKP™) is a surgical aid to provide a transparent optical pathway that permits closed pars plana vitrectomy in eyes with abnormal corneas.

## Comparison of Technological Characteristics

Table 1: Substantial Equivalence

|  Product Characteristic | Landers Wide Field TKP (K910828) | QuickSet TKP™ (K261312) | Comparison / Rationale  |
| --- | --- | --- | --- |
|  Indications for Use | Transparent optical pathway that permits closed pars plana vitrectomy in eyes with abnormal corneas. | The QuickSet Temporary Keratoprosthesis (QuickSet TKP™) is a surgical aid to provide a transparent optical pathway that permits closed pars plana vitrectomy in eyes with abnormal corneas. | Same intended use and similar indications; supported by simulated use.  |
|  Material | PMMA | PMMA | Same ophthalmic material family.  |
|  Refractive Index | 1.49 | 1.49 | Same.  |
|  Optical Function | Transparent optical pathway for posterior segment visualization. | Transparent optical pathway for posterior segment visualization. | Similar; supported by optical testing.  |
|  Collar Diameter | 15.5 mm | 12.0 mm | Different; supported by design and testing.  |
|  Optical Cylinder Diameter | 8.2 mm / 6.2 mm | 8.0 mm | Similar.  |
|  Optical Cylinder Thickness | 1.0 mm | 0.82 mm | Similar; supported by dimensional and optical testing.  |
|  Attachment Mechanism | 6 holes for limbal sutures | 4 grooves for horizontal mattress sutures | Different; simulated use demonstrated securement and watertight sealing.  |
|  Use | Multiple use | Single use | Different; single-use sterile presentation reduces reprocessing-related risks.  |
|  Sterilization | Non-sterile | Ethylene oxide sterilized | Different; EO validation and EO residuals support sterile single-use presentation.  |
|  Packaging | Medical pouch | Tyvek/Mylar pouch and box configuration | Different; sterile barrier, distribution, and aging testing support packaging integrity.  |

## Performance Testing - Nonclinical Tests

The following nonclinical performance testing and assessments were conducted for the QuickSet TKP™ to support substantial equivalence to the identified predicate device. The testing evaluated biological safety, sterilization, residuals, microbial cleanliness, endotoxin, packaging and sterile barrier integrity, shelf-life, dimensional and visual characteristics, optical performance, and simulated use sealing and visualization performance. The results support the claimed performance characteristics of the QuickSet TKP™ as a sterile, single-use temporary surgical aid that provides a transparent optical pathway and maintains a watertight seal during representative vitreoretinal surgical use.

|  Test / Assessment | Purpose | Summary of Results and Support for Substantial Equivalence  |
| --- | --- | --- |
|  Biocompatibility evaluation and testing | Evaluate biological safety for intended temporary ophthalmic surgical use. | ISO 10993-1 evaluation included cytotoxicity, sensitization, and ocular irritation. Cytotoxicity testing resulted in Grade 0 responses at 24 and 48 hours and the device was non-cytotoxic. Sensitization testing found the device was not a contact skin sensitizer. Primary eye irritation testing found no ocular irritation and classified the device as a non-irritant.  |

{6}

EyeKon

510(k) SUMMARY

K261312

Page 3 of 4

|  Test / Assessment | Purpose | Summary of Results and Support for Substantial Equivalence  |
| --- | --- | --- |
|  Ethylene oxide sterilization validation | Validate EO sterilization for sterile, single-use presentation. | EO sterilization validation was performed in accordance with ISO 11135 and supports the sterile, single-use presentation of the finished device.  |
|  Ethylene oxide residuals testing | Confirm residual EO and related residuals are within applicable limits. | EO residuals testing was performed in accordance with ISO 10993-7. Results supported that residual sterilant levels were within applicable limits for the device and intended patient contact.  |
|  Bioburden testing | Characterize and control pre-sterilization microbial load. | Bioburden testing was performed in accordance with ISO 11737-1 and USP <61>. Results supported the manufacturing and sterilization controls for the finished device.  |
|  Bacterial endotoxins testing | Confirm acceptable endotoxin levels for ophthalmic surgical use. | Bacterial endotoxins testing was performed in accordance with ANSI/AAMI ST72 and USP <85>. Results met the applicable acceptance criterion for the finished sterile device.  |
|  Packaging and sterile barrier integrity testing | Verify package integrity after sterilization, handling, transport simulation, and aging. | Packaging validation included environmental conditioning and distribution simulation, visual inspection, bubble leak testing, and seal strength testing using ASTM D4332, ASTM D4169, ASTM F1886/F1886M, ASTM F2096, ASTM F88/F88M, and ASTM F1980, as applicable. The Tyvek/Mylar pouch met predefined sterile barrier acceptance criteria.  |
|  Shelf-life / aging support | Support maintenance of package integrity and device performance through labeled shelf life. | Accelerated and real-time aging supported package integrity and continued conformance to critical performance expectations through the claimed shelf life. One-year and two-year aging reports concluded that tested samples met requirements for visual inspection, bubble leak testing, and seal strength testing.  |
|  Dimensional and visual inspection | Verify conformance to critical dimensions and workmanship requirements. | Dimensional and visual inspection confirmed conformance to critical device specifications, including optical geometry and absence of burrs, cracks, sharp edges, and visually significant defects.  |
|  Optical performance testing | Verify optical characteristics needed for posterior segment visualization. | Optical testing evaluated higher-order aberration, spherical equivalent power (SEP), optical diameter, MTF at 50 lp/mm, and sagittal/thickness using 30 returned samples with repeated measurements. Results supported design targets: low HOA, mean SEP approximately +30.0 D, optical diameter approximately 7.60 mm, MTF at 50 lp/mm greater than 0.40, and sagittal/thickness approximately 2.23 mm. Finished device inspection and rejection controls prevent release of nonconforming units.  |
|  Simulated use sealing and visualization validation | Evaluate placement, securement, pressure sealing, and posterior segment visualization under representative surgical conditions. | Simulated clinical wet lab use demonstrated successful device placement and suture engagement, watertight sealing after securement, no leakage during pressurization from 35 mmHg to 120 mmHg, and excellent visualization during representative complex vitreoretinal maneuvers. Additional details are provided below.  |

## Simulated Use / Sealing and Visualization Validation

Simulated use testing was performed in a clinical wet lab by qualified ophthalmologists trained and experienced in the use of TKPs. The procedure included vitrectomy trocar placement, limbal marking, pre-placement of two 7-0 vicryl horizontal mattress sutures, 8 mm corneal trephination, removal of the central corneal button, placement of the QuickSet TKP™, suture engagement into the device grooves, and securement of the device.

After the second suture was tied, the globe was secure and the QuickSet TKP™ was watertight. The eye was pressurized using an Alcon Constellation system from 35 mmHg to 120 mmHg in 10 mmHg increments, and no TKP wound leak was observed throughout the tested pressure range.

Visualization was excellent under representative challenging conditions and remained excellent after additional simulated complex maneuvers, including complete vitrectomy, induction of retinal detachment, peripheral retinectomy, crystalline lens manipulation, and extensive scleral depression. These results support the claimed performance characteristics of maintaining intraocular pressure without leakage and providing posterior segment visualization during representative vitreoretinal surgical maneuvers.

{7}

EyeKon

510(k) SUMMARY

K261312

Page 4 of 4

## Clinical Performance Testing

Clinical performance testing was not required to support substantial equivalence. The determination of substantial equivalence is based on comparison to the predicate device, similarities in intended use and technological characteristics, risk-based assessments, and the results of nonclinical performance testing.

## Substantial Equivalence Discussion

The QuickSet TKP™ and the predicate device have the same intended use and similar indications for use: both provide a transparent optical pathway to permit closed pars plana vitrectomy in eyes with abnormal corneas. Both devices are temporary ophthalmic surgical aids made from PMMA and are intended to facilitate posterior segment visualization during combined corneal and vitreoretinal surgery.

The QuickSet TKP™ differs from the predicate in certain technological characteristics, including dimensions, attachment mechanism, single-use sterile presentation, EO sterilization, and packaging configuration. The differences are addressed by nonclinical testing and risk-based assessments, including simulated use testing demonstrating secure placement and watertight sealing under simulated intraocular pressure conditions, optical testing demonstrating that the device provides an adequate transparent optical pathway, biocompatibility testing demonstrating biological acceptability, and sterilization/packaging testing supporting sterile single-use presentation.

## Conclusion

EyeKon Medical, Inc. considers the QuickSet TKP™ to be substantially equivalent to the predicate Landers Wide Field Temporary Keratoprosthesis cleared under K910828. This conclusion is based on the same intended use, similar indications for use, comparable PMMA material and optical function, and supporting nonclinical performance data. The nonclinical test results demonstrate that the QuickSet TKP™ meets its predefined acceptance criteria and support the claimed performance characteristics of providing a transparent optical pathway, maintaining intraocular pressure during simulated use without leakage, and being provided sterile for single use. Therefore, the QuickSet TKP™ nonclinical test results demonstrate substantial equivalence to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K261312](https://fda-staging.innolitics.com/device/K261312)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).