Retrospective clinical PET/CT and PET/MR imaging datasets
A retrospective multi-rater study was conducted to evaluate the quantitative and qualitative performance of the MICSI-PET software by comparing its output against standard-of-care (SOC) clinical imaging data across various amyloid and tau tracers.
Retrospective study; Clinical imaging data; Standard of Care (SOC); PET imaging
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective Multi-Rater Study; Retrospective, multi-rater study comparing device output to standard-of-care (SOC) clinical imaging
Patients undergoing amyloid and tau PET imaging; Sample Size: N=664 paired reads
MICSI-PET is an image processing software that can be used for image enhancement in PET images. It computes quantitative metrics including regional SUVr (Standardized Uptake Value ratio), Centiloid scores for amyloid and tau PET, and z-scores compared to normative controls.
Device Story
MICSI-PET is SaMD for PET brain image enhancement and quantitative analysis. Inputs: DICOM PET images (amyloid, tau, FDG tracers) and T1-weighted MR images for anatomical reference. Operation: Software resides on a DICOM router/processor; performs physics-based super-resolution enhancement and atlas-based brain segmentation using co-registered MRI. Outputs: Enhanced PET images, quantitative reports (SUVr, Centiloid, z-scores) sent to PACS. Used in clinics/hospitals by radiologists/nuclear medicine physicians. Benefits: Improved image quality, reduced artifacts, and standardized quantitative metrics to assist clinical decision-making.
Clinical Evidence
Retrospective multi-rater study (N=664 paired reads). Metrics: Image quality (3.04 to 4.59 on 5-point scale), artifact reduction (+0.32 improvement), tracer positivity agreement (0.87, p=0.203), and concordance (AUC) between quantitative metrics and visual reads. Amyloid AUCs consistently ≥0.80. Tau AUCs ≥0.80 for universal, mesial, meta-temporal, and temporoparietal ROIs. Bench testing confirmed SUVr agreement (R² 0.979–0.998).
Technological Characteristics
Software-based image processing; physics-based super-resolution algorithm; requires T1-weighted MRI for anatomical registration/segmentation; DICOM 3.0 compatible; standalone deployment on DICOM router/processor; no direct user interface.
Indications for Use
Indicated for adults (22+ years) undergoing PET brain imaging (amyloid, tau, or FDG tracers) to provide image enhancement and quantitative metrics (SUVr, Centiloid, z-scores) for clinical assessment.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 11, 2026
Microstructure Imaging, Inc.
% Adam Heroux
Consultant
Highland Biomedical, Inc.
4190 Grove St.
Denver, Colorado 80211
Re: K261305
Trade/Device Name: MICSI-PET
Regulation Number: 21 CFR 892.1200
Regulation Name: Emission computed tomography system
Regulatory Class: Class II
Product Code: KPS, LLZ
Dated: August 10, 2026
Received: August 11, 2026
Dear Adam Heroux:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261305 - Adam Heroux
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K261305 - Adam Heroux
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261305 | ? |
| Please provide the device trade name(s). | | ? |
| MICSI-PET | | |
| Please provide your Indications for Use below. | | ? |
| MICSI-PET is an image processing software that can be used for image enhancement in PET images. It computes quantitative metrics including regional SUVr (Standardized Uptake Value ratio), Centiloid scores for amyloid and tau PET, and z-scores compared to normative controls. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
MiCSI
510(k)# K261305
510(k) Summary
11SEP2026
# Contact Details
21 CFR 807.92(a)(1)
Applicant Information:
Microstructure Imaging, Inc.
370 Jay St, FL 7
Brooklyn, NY 11201 United States
718-801-2324
Dr. Gregory Lemberskiy
gregory.lemberskiy@micsi.com
Correspondent:
Highland Biomedical, Inc.
4190 Grove St
Denver, CO 80211 United States
617-823-0515
Adam Heroux
adam.heroux@highlandbiomed.com
# Device Name
21 CFR 807.92(a)(2)
Device Trade Name: MICSI-PET
Classification/Common Name: Emission computed tomography system
Regulation Number: 21 CFR 892.1200
Product Code(s): KPS, LLZ
Device Class: Class II
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate Device # K211964
Predicate Trade Name: SubtlePET
Regulation Number: 21 CFR 892.1200
Product Code: KPS, LLZ
Reference Device:
Reference Device # K233620
Reference Trade Name: MIM - Centiloid Scaling
Regulation Number: 21 CFR 892.2050
Reference Product Code: LLZ
# Device Description Summary
21 CFR 807.92(a)(4)
MICSI-PET Super-Resolution (MICSI-PET) is a Software as a Medical Device (SaMD) intended to enhance positron emission tomography (PET) brain images using super-resolution techniques and to compute standardized quantitative metrics for amyloid and tau PET imaging.
The software supports three PET tracer categories: amyloid (Florbetapir, Florbetaben,
{5}
MiCSI
510(k)# K261305
510(k) Summary
11SEP2026
Flutemetamol, PiB), tau (Flortaucipir), and FDG. The device is intended to be used by radiologists and nuclear medicine physicians in an imaging center, clinic, or hospital. The device is compatible with the DICOM 3.0 standard.
The subject device has no direct user interface. The DICOM images obtained from a compatible PET/CT and/or PET/MR scanner are streamed from the scanner to the designated DICOM destination, e.g., picture archiving and communication system (PACS), then to a DICOM router and processor. It is within this router and processor where the subject device performs super-resolution enhancement and quantitative analysis. The software requires MR images (T1-weighted) for anatomical reference which are co-registered to PET images during processing; the MR images are used both to guide the super-resolution algorithm and to perform atlas-based brain segmentation for quantitative analysis. Once the processing is complete, the device outputs are then routed to the target PACS (which can be the same one as before or a new DICOM destination) where the intended user can view the enhanced images, quantitative reports, and the original images. Image processing time varies as it is dependent on the central processing unit (CPU), input image size, and the selected processing configuration. When multiple jobs have been submitted to MICSI-PET, they are processed sequentially based on the order received.
### Intended Use/Indications for Use
21 CFR 807.92(a)(5)
MICSI-PET is an image processing software that can be used for image enhancement in PET images. It computes quantitative metrics including regional SUVr (Standardized Uptake Value ratio), Centiloid scores for amyloid and tau PET, and z-scores compared to normative controls.
### Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use for the subject device are substantially similar to those for the predicate device. Both the proposed device and the predicate are intended to process and enhance PET images. MICSI-PET compared to the reference predicate has a narrower scope but is substantially equivalent to the indications associated with Centiloid scoring, and the computation of z-scores.
### Technological Comparison
21 CFR 807.92(a)(6)
The following information provides a summary of how the technological characteristics of the devices compare. MICSI-PET is similar to the predicate with the following characteristics:
- Same intended use environment and user
- Similar output of enhanced PET imaging
- Both software process FDG and Amyloid Tracer cases
{6}
MICSI
510(k)# K261305
510(k) Summary
11SEP2026
MICSI-PET has the following differences from the predicate and reference devices:
- Different tracers- the predicate device also includes the following tracers: 18F-DOPA, 18F-DCFPyL, Ga-68 Dotatate, and Ga-68 PSMA. MICSI-PET also includes the processing of Flortaucipir.
- Different imaging anatomy - In addition to brain imaging, the reference device can process alternate anatomical locations while MICSI-PET is indicated for brain imaging.
- Different software architecture - MICSI-PET utilizes physics-based software and MRI anatomical registration, whereas the predicate utilizes an AI/ML model to provide the enhanced imaging output.
- Different functionality - while the predicate and reference have features not included in MICSI-PET, the functionality of image enhancement, centiloid display and z-scores are included in the predicate and reference devices.
MICSI-PET has undergone software and performance testing to ensure that any differences in the technological characteristics do not raise different questions of safety and effectiveness
### Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Summary of Non-clinical Testing:
Non-clinical testing consisted of Software Verification and Validation testing at the unit, integration, and system level to demonstrate that software requirements were implemented.
### Rater Study:
A retrospective, multi-rater study was conducted to Evaluate Quantitative and Qualitative performance of the device including Image Quality Enhancement, Reduction in Artifacts, Tracer Positivity, SNR, Median SUVr (mSUVr), STD SUVr (sSUVr), Median Z-Score, STD Z-Score, and SUVr Correlation. Qualitative-Quantitative Concordance testing was also conducted to demonstrate Visual vs Centiloid Positivity, and Visual vs Z-Score Positivity.
### Rater Study Results:
- MICSI-PET improved reader-rated Image Quality relative to SOC across all tracers. The pooled mean Image Quality increased from 3.04 (SOC) to 4.59 (MICSI-PET) on a 5 point scale, corresponding to a paired mean improvement (MICSI-PET – SOC) of +1.55;
- MICSI-PET reduced reader-rated Artifact Severity relative to SOC across all tracers. The pooled mean Artifact Severity decreased from 1.43 (SOC) to 1.11 (MICSI-PET) on a 3
{7}
MICSI
510(k)# K261305
510(k) Summary
11SEP2026
point scale, corresponding to a paired mean improvement (SOC – MICSI-PET; positive indicates fewer artifacts) of +0.32 across N=664 paired reads
- MICSI-PET maintained high overall agreement in reader binary tracer positivity classifications relative to SOC across amyloid and tau tracers. The pooled Tracer Positivity Agreement between SOC and MICSI-PET was 0.87, and the pooled McNemar test indicated no significant directional difference in positivity classification, p=0.203. Agreement was consistently high across individual tracers, FLP 0.87, FBB 0.86, FMT 0.88, and TAU 0.88. These results show that MICSI-PET generally preserved binary tracer positivity determinations relative to SOC.
- MICSI-PET demonstrated statistically significant improvement in SNR relative to SOC, with statistically significant processing differences in mSUVr and sSUVr.
- SUVr agreement met the acceptance criterion under both ROI frameworks (GAAIN reference ROIs and MICSI-derived ROIs), with R² values of [GAAIN: 0.997] and [MICSI: 0.995] across amyloid tracers. SUVr agreement met the acceptance criterion for the Universal ROI under both ROI frameworks. (GAAIN reference ROIs and MICSI-derived ROIs), with R² values of [GAAIN: 0.998] and [MICSI: 0.998]. SUVr agreement met the acceptance criterion using MICSI-derived cortical ROIs with R² values of [MICSI: 0.979].
- Z-scores met pre-defined acceptance criteria: For the amyloid aggregate, the SOC/GAAIN vs MICSI-PET/MICSI combined comparison yielded mZ 1.739 to 1.839 and sZ 1.080 to 0.975. The reciprocal combined comparison (SOC/MICSI vs MICSI-PET/GAAIN) yielded mZ 1.555 to 1.863 and sZ 0.985 to 1.038. For the tau Universal ROI, SOC/GAAIN vs MICSI-PET/MICSI: mZ changed from 1.973 to 1.100 and sZ changed from 1.079 to 1.092. SOC/MICSI vs MICSI-PET/GAAIN): mZ changed from 1.016 to 2.006 and sZ changed from 1.118 to 1.052. The opposing directions across the two combined comparisons are consistent with ROI framework differences shifting the absolute z-score scale; therefore, combined contrasts should be interpreted in the context of the ROI-only comparisons rather than as processing-only effects.
- Across pooled amyloid and individual amyloid tracers, concordance between visual amyloid tracer positivity classification and centiloid values consistently exceeded the pre-specified acceptance threshold of AUC ≥ 0.80 under SOC and MICSI-PET processing and under both MICSI and GAAIN ROI definitions.
- Amyloid (pooled): AUCs were 0.909 (SOC+GAAIN) and 0.914 (SOC+MICSI), and 0.959 (MICSI-PET+GAAIN) and 0.959 (MICSI-PET+MICSI).
- Florbetapir (FLP): AUCs were 0.934 (SOC+GAAIN) and 0.940 (SOC+MICSI), and 0.949(MICSI-PET+GAAIN) and 0.958 (MICSI-PET+MICSI).
{8}
MiCSI
510(k)# K261305
510(k) Summary
11SEP2026
○ Florbetaben (FBB): AUCs were 0.870 (SOC+GAAIN) and 0.875 (SOC+MICSI), and 0.951 (MICSI-PET+GAAIN) and 0.953 (MICSI-PET+MICSI).
○ Flutemetamol (FMT): AUCs were 0.929 (SOC+GAAIN) and 0.933 (SOC+MICSI), and 0.969 (MICSI-PET+GAAIN) and 0.965 (MICSI-PET+MICSI).
- Across pooled amyloid and individual amyloid tracers, concordance between visual amyloid tracer positivity classification and z-score values consistently exceeded the pre-specified acceptance threshold of AUC ≥ 0.80 under SOC and MICSI-PET processing and under both MICSI and GAAIN ROI definitions.
○ Amyloid (pooled): AUCs were 0.904 (SOC+GAAIN) and 0.911 (SOC+MICSI), and 0.951 (MICSI-PET+GAAIN) and 0.958 (MICSI-PET+MICSI).
○ Florbetapir (FLP): AUCs were 0.922 (SOC+GAAIN) and 0.940 (SOC+MICSI), and 0.930 (MICSI-PET+GAAIN) and 0.959 (MICSI-PET+MICSI).
○ Florbetaben (FBB): AUCs were 0.855 (SOC+GAAIN) and 0.868 (SOC+MICSI), and 0.939 (MICSI-PET+GAAIN) and 0.952 (MICSI-PET+MICSI).
○ Flutemetamol (FMT): AUCs were 0.927 (SOC+GAAIN) and 0.928 (SOC+MICSI), and 0.962 (MICSI-PET+GAAIN) and 0.963 (MICSI-PET+MICSI)
- Across tau ROIs, concordance between visual tau tracer positivity classification and centiloid values generally met the pre-specified acceptance threshold of AUC ≥ 0.80, with the strongest performance in the meta-temporal and temporoparietal ROIs.
○ Universal: AUCs were 0.901 (SOC+GAAIN) and 0.840 (SOC+MICSI), and 0.872 (MICSI-PET+GAAIN) and 0.817 (MICSI-PET+MICSI).
○ Mesial: AUCs were 0.880 (SOC+GAAIN) and 0.878 (SOC+MICSI), and 0.879 (MICSI-PET+GAAIN) and 0.834 (MICSI-PET+MICSI).
○ Meta: AUCs were 0.902 (SOC+GAAIN) and 0.910 (SOC+MICSI), and 0.901 (MICSI-PET+GAAIN) and 0.907 (MICSI-PET+MICSI).
○ TemporoParietal: AUCs were 0.903 (SOC+GAAIN) and 0.864 (SOC+MICSI), and 0.872 (MICSI-PET+GAAIN) and 0.833 (MICSI-PET+MICSI).
○ Frontal: AUCs were 0.809 (SOC+GAAIN), with 0.759 (SOC+MICSI), 0.795 (MICSI-PET+GAAIN), and 0.770 (MICSI-PET+MICSI), indicating that frontal ROI concordance did not consistently meet the AUC ≥ 0.80 acceptance threshold across configurations.
- Across tau ROIs, concordance between visual tau tracer positivity classification and z-score values similarly met the pre-specified acceptance threshold of AUC ≥ 0.80 for most ROIs, with strongest performance again in the meta-temporal and temporoparietal ROIs.
○ Universal: AUCs were 0.901 (SOC+GAAIN) and 0.864 (SOC+MICSI), and 0.869 (MICSI-PET+GAAIN) and 0.861 (MICSI-PET+MICSI).
{9}
MICSI
510(k)# K261305
510(k) Summary
11SEP2026
- ○ **Mesial:** AUCs were 0.899 (SOC+GAAIN) and 0.884 (SOC+MICSI), and 0.892 (MICSI-PET+GAAIN) and 0.860 (MICSI-PET+MICSI).
- ○ **Meta:** AUCs were 0.925 (SOC+GAAIN) and 0.904 (SOC+MICSI), and 0.912 (MICSI-PET+GAAIN) and 0.907 (MICSI-PET+MICSI).
- ○ **TemporoParietal:** AUCs were 0.902 (SOC+GAAIN) and 0.878 (SOC+MICSI), and 0.869 (MICSI-PET+GAAIN) and 0.858 (MICSI-PET+MICSI).
- ○ **Frontal:** AUCs were 0.820 (SOC+GAAIN), with 0.726 (SOC+MICSI), 0.794 (MICSI-PET+GAAIN), and 0.754 (MICSI-PET+MICSI), indicating that frontal ROI concordance did not consistently meet the AUC ≥ 0.80 acceptance threshold across configurations.
### Rater Study Conclusion:
The non-clinical bench performance testing demonstrated that MICSI-PET processing improved qualitative image quality and reduced artifact severity across all evaluated tracers while maintaining high agreement with standard-of-care (SOC) expert visual tracer positivity classifications. Quantitatively, MICSI-PET preserved expected SUVr relationships and showed strong concordance between continuous Centiloid and z-score measures and expert visual reads. Amyloid concordance met the pre-specified acceptance threshold across all evaluated configurations, while tau concordance generally met the acceptance threshold across the universal, mesial, meta-temporal, and temporoparietal ROI definitions. Taken together, these findings support that MICSI-PET is substantially equivalent to SOC processing, while providing enhanced visual interpretability and precise quantitative outputs.
### Overall Conclusions:
The Non-Clinical and Rater Study testing have demonstrated that MICSI-PET shows substantially equivalent performance to the predicate device and is as safe and as effective for the proposed intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.