K261265 · Orthocon, Inc. · GXP · Jun 15, 2026 · Neurology
Device Facts
Record ID
K261265
Device Name
Permatage XT Settable Bone Putty
Applicant
Orthocon, Inc.
Product Code
GXP · Neurology
Decision Date
Jun 15, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Permatage XT Settable Bone Putty is a self-setting cement indicated for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects. Permatage XT Settable Bone Putty should be used only in skeletally mature individuals.
Device Story
Permatage XT is a sterile, nonabsorbable, two-component putty-like material used for cranial defect repair. Components consist of granular calcium phosphate, paraffin oil, vitamin E acetate, a triglyceride, and polyether-based polymers. Components are mixed manually by a clinician immediately prior to surgery to form a cohesive, spreadable material that adheres to bone and hardens in place. The device is applied manually to the bone surface in a clinical setting. It provides a 4-minute working time and cures without an appreciable exothermic reaction. The resulting hardened material is primarily calcium phosphate and nonabsorbable polymers. The device is intended for single-patient use.
Clinical Evidence
No clinical data. Bench testing only. A single benchtop study was conducted to verify the extended working time of 4 minutes compared to the predicate device.
Technological Characteristics
Two-part putty system; components: granular calcium phosphate, paraffin oil, vitamin E acetate, triglyceride, polyether-based polymers. Radiopaque (contains hydroxyapatite and β-tricalcium phosphate). Nonabsorbable. Manual mixing (45 seconds). Working time: 4 minutes. Sterilization: Gamma irradiation. Packaging: Foil packages in Tyvek/LDPE pouches with desiccant.
Indications for Use
Indicated for repair of neurosurgical burr holes, contiguous craniotomy cuts, and other cranial defects in skeletally mature individuals (adolescents and adults).
Regulatory Classification
Identification
Methyl methacrylate for cranioplasty (skull repair) is a self-curing acrylic that a surgeon uses to repair a skull defect in a patient. At the time of surgery, the surgeon initiates polymerization of the material and forms it into a plate or other appropriate shape to repair the defect.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 15, 2026
Orthocon, Inc.
Alyssa Mcdermott
Sr. Director, Quality and Regulatory Affairs
700 Fairfield Ave., Suite 1
Stamford, Connecticut 06902
Re: K261265
Trade/Device Name: Permatage XT Settable Bone Putty
Regulation Number: 21 CFR 882.5300
Regulation Name: Methyl Methacrylate For Cranioplasty
Regulatory Class: Class II
Product Code: GXP
Dated: April 16, 2026
Received: April 16, 2026
Dear Alyssa Mcdermott:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261265 - Alyssa Mcdermott
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261265 - Alyssa Mcdermott
Page 3
Sincerely,
YEN-CHIH LIN -S
Digitally signed by YEN-CHIH LIN
-S
Date: 2026.06.15 14:40:07 -04'00'
for Jaime Raben, Ph.D.
Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K261265 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| Permatage XT Settable Bone Putty | | | |
| Please provide your Indications for Use below. | | | ? |
| Permatage XT Settable Bone Putty is a self-setting cement indicated for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects. Permatage XT Settable Bone Putty should be used only in skeletally mature individuals. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
K261265 Page 1 of 3
# 510(k) SUMMARY
## General Company Information:
Name: Orthocon, Inc.
Contact: Alyssa McDermott
Senior Director of Quality and Regulatory Affairs
Address: 700 Fairfield Avenue, Suite 1
Stamford, CT 06902
Telephone: (855) 475 - 9175
## Date Prepared:
June 9th, 2026
## General Device Information:
Product Name: Permatage XT Settable Bone Putty
Common Name: Methyl Methacrylate for Cranioplasty
Classification: Class II
Product codes: GXP
## Predicate Device:
Orthocon, Inc. Permatage Settable Bone Putty
## Device Description
Permatage XT Settable Bone Putty is a sterile, biocompatible, nonabsorbable material of putty-like consistency for in the repair of cranial defects. The single use Permatage XT device contains two separate components of putty-like consistency comprised of granular calcium phosphate, paraffin oil, vitamin E acetate, a triglyceride, and a mixture of nonabsorbable, polyether-based polymers. When mixed together, the components of the Permatage XT device form a nonabsorbable cohesive, putty-like material that adheres to the bone surface and remains in place following application. The resulting hardening material is primarily calcium phosphate and nonabsorbable polymer materials. Permatage XT components must be mixed immediately prior to use.
## Indications for Use
Permatage XT Settable Bone Putty is a self-setting cement indicated for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects. Permatage XT Settable Bone Putty should be used only in skeletally mature individuals.
The following table shows comparisons of characteristics of Permatage XT Settable Bone Putty and the predicate device.
{5}
K261265 Page 2 of 3
| | Permatage Settable Bone Putty (K241027, Primary Predicate) | Permatage XT Settable Bone Putty (K261265, Subject Device) |
| --- | --- | --- |
| Indications | Permatage Settable Bone Putty is intended for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects. | Permatage XT Settable Bone Putty is intended for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects. |
| Product Code | Product Code GXP | Product Code GXP |
| Product Consistency | Permatage Settable Bone Putty is formulated as a two-part putty/putty device that forms a "settable" (hardening) material when mixed at the time of surgery. At the time of application, device is in the form of a spreadable material. | Permatage XT Settable Bone Putty is formulated as a two-part putty/putty device that forms a "settable" (hardening) material when mixed at the time of surgery. At the time of application, device is in the form of a spreadable material. |
| Application | Device is designed to be manually applied to the bone surface. | Device is designed to be manually applied to the bone surface. |
| Components | Permatage Settable Bone Putty is a sterile, biocompatible, nonabsorbable material of putty-like consistency for use in repair of cranial defects. The single use Permatage device contains two separate components of putty-like consistency comprised of granular calcium phosphate, paraffin oil, vitamin E acetate, a triglyceride, and a mixture of nonabsorbable, polyether-based polymers. When mixed together, the components of the Permatage device form a nonabsorbable cohesive putty-like material that adheres to the bone surface and remains in place following application. The resulting hardening material is primarily comprised of calcium phosphate and nonabsorbable polymer materials. | Permatage XT Settable Bone Putty is a sterile, biocompatible, nonabsorbable material of putty-like consistency for use in repair of cranial defects. The single use Permatage device contains two separate components of putty-like consistency comprised of granular calcium phosphate, paraffin oil, vitamin E acetate, a triglyceride, and a mixture of nonabsorbable, polyether-based polymers. When mixed together, the components of the Permatage XT device form a nonabsorbable cohesive putty-like material that adheres to the bone surface and remains in place following application. The resulting hardening material is primarily comprised of calcium phosphate and nonabsorbable polymer materials. |
| Radiopacity | Radiopaque – Contains hydroxyapatite and β-tricalcium phosphate. | Radiopaque – Contains hydroxyapatite and β-tricalcium phosphate. |
| Implant Life | Implanted device is nonabsorbable. | Implanted device is nonabsorbable. |
| Sterilization | Single-patient-use device is provided sterile by gamma irradiation. | Single-patient-use device is provided sterile by gamma irradiation. |
| Unit Sizes | The bone putty is available in individual sizes of up to 10cc. | The bone putty is available in individual sizes of up to 10cc. |
| Packaging | The putty is provided in two foil packages each within a Tyvek/LDPE pouch, within a single outer foil pouch. The outer foil pouch contains a desiccant. The pouch is heat sealed and terminally sterilized via gamma irradiation. | The putty is provided in two foil packages each within a Tyvek/LDPE pouch, within a single outer foil pouch. The outer foil pouch contains a desiccant. The pouch is heat sealed and terminally sterilized via gamma irradiation. |
| Mixing | Mixing for homogeneity takes 45 seconds. | Mixing for homogeneity takes 45 seconds. |
| Setting | Material sets within minutes following application. | Material sets within minutes following application. |
| Working Time | Material provides a working time of 2 minutes. | Material provides a working time of 4 minutes. |
| Exotherm | Device cures with no appreciable exothermic reaction. | Device cures with no appreciable exothermic reaction. |
{6}
K261265 Page 3 of 3
## Testing Completed
### Biocompatibility Testing
Biocompatibility testing was previously conducted on the primary predicate device, Permatage, to evaluate biocompatibility in accordance with the recommendations of ISO 10993. The following biocompatibility studies were conducted on the final, finished, gamma-irradiation sterilized device and in accordance with the GLP requirements: cytotoxicity, irritation, sensitization, systemic toxicity, genotoxicity, local tissue toxicity, hemolysis, pyrogenicity and neurotoxicity. Because the minute adjustments made to the polymer content of the device decrease the overall polymer content and increase the calcium phosphate content, the conclusions of the predicate biocompatibility are still applicable and no additional testing is needed to support the biocompatibility of the Permatage XT Settable Bone Putty.
### Bench testing
The only difference between the predicate device in K241027 and the subject device is a minute adjustment in the polymer content to extend the working time of the subject device. All other studies submitted for the predicate device in K241027 are directly applicable to the subject device, Permatage XT. Only one benchtop study was conducted to compare the extended working time via physical handling of the subject device to the predicate device, Permatage. Results confirm the extended working time of 4 minutes presented in the labeling.
### Conclusions
This submission supports the position that Orthocon Permatage XT Settable Bone Putty is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.