Laser Hair Growth Devices is used to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classification of IIa~V or females with androgenic alopecia who have Ludwig-Savin Classifications of I~II and both with Fitzpatrick Skin Phototypes I to IV.
Device Story
Laser Hair Growth Devices (multiple models) utilize low-level laser therapy (LLLT) to promote hair growth; device consists of cap or helmet housing laser diodes emitting visible red light at 650 ± 10 nm; powered via USB cable; intended for OTC home use by patients; operation involves wearing device for 30-minute sessions; laser diodes deliver 5mW per diode; device promotes hair growth through LLLT mechanism; output is non-invasive light therapy; healthcare providers do not operate device; patient self-administers treatment; device benefits include potential hair growth stimulation in specified alopecia populations.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-11, -1-2, -1-6), laser safety (IEC 60825-1), and functional performance testing (optical output, wavelength accuracy, irradiance, and thermal stability).
Technological Characteristics
LLLT device; 650±10nm wavelength; 5mW per laser diode; Class 3R laser; USB-powered; materials compliant with ISO 10993-5, -10, -23; safety standards IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60601-1-6, IEC 60825-1; form factors include caps and helmets; standalone operation.
Indications for Use
Indicated for males with androgenic alopecia (Norwood-Hamilton IIa-V) and females with androgenic alopecia (Ludwig-Savin I-II) with Fitzpatrick Skin Phototypes I-IV to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 26, 2026
Ziree Co., Ltd.
% Linda Li
Compliance Officer
Huaxiajiamei (Beijing) Information Consulting Co.,Ltd
B2 Floor, Zhongguancun Dongsheng Science And Technology Park
# 66 Xixiaolu Rd., Haidian District
Beijing, 100192
China
Re: K261253
Trade/Device Name: Laser Hair Growth Devices (HR-H1, HR-H2, HR-H3, HR-H4, HR-H5, HR-H6, HR-H7, HR-H8, HR-H9, HR-H10, HR-I1, HR-I2, HR-I3, HR-I4, HR-I5)
Regulation Number: 21 CFR 890.5500
Regulation Name: Infrared Lamp
Regulatory Class: Class II
Product Code: OAP
Dated: April 16, 2026
Received: April 16, 2026
Dear Linda Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}
K261253 - Linda Li
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K261253 - Linda Li
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K261253
Device Name
Laser Hair Growth Devices (HR-H1, HR-H2, HR-H3, HR-H4, HR-H5, HR-H6, HR-H7, HR-H8, HR-H9, HR-H10, HR-I1, HR-I2, HR-I3, HR-I4, HR-I5)
Indications for Use (Describe)
Laser Hair Growth Devices is used to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classification of IIa~V or females with androgenic alopecia who have Ludwig-Savin Classifications of I~II and both with Fitzpatrick Skin Phototypes I to IV.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# 510(K) Summary
## K261253
Prepared in accordance with the requirements of 21 CFR Part 807.92
**Prepared Date: June 18, 2026**
### 1. Submitter's Information
The submitter of this pre-market notification is:
| **Name:** | Ziree Co., Ltd. |
| --- | --- |
| **Address:** | Room 606, Floor 6th, Unit 1, Building 6, Chaoshi Intelligent Industrial Park, No.37 Xinyuan Road, Tianxin District, Changsha City, 410114 Hunan Province , CHINA |
| **Contact person:** | Aijie Peng |
| **Title:** | Manager |
| **E-mail:** | 3527148947@qq.com |
| **Tel:** | 86-13312989605 |
### 2. Device Identification
| **510(K) number:** | K261253 |
| --- | --- |
| **Trade/Device Name:** | Laser Hair Growth Devices (HR-H1、HR-H2、HR-H3、HR-H4、HR-H5、HR-H6、HR-H7、HR-H8、HR-H9、HR-H10、HR-I1、HR-I2、HR-I3、HR-I4、HR-I5) |
| **Models:** | HR-H1、HR-H2、HR-H3、HR-H4、HR-H5、HR-H6、HR-H7、HR-H8、HR-H9、HR-H10、HR-I1、HR-I2、HR-I3、HR-I4、HR-I5 |
| **Common name:** | LASER, COMB, HAIR |
| **Regulation Number:** | 890.5500 |
| **Regulation Description:** | Infrared lamp |
| **Regulation Class:** | Class 2 |
| **Panel:** | General & Plastic Surgery |
| **Product Code:** | OAP |
### 3. Predicate Device
| **510(K) number:** | K192627 |
| --- | --- |
| **Trade/Device Name:** | Laser Hair Growth Cap |
| **Models:** | TW280, TW272, TW147, TW080 |
| **Common name:** | LASER, COMB, HAIR |
| **Regulation Number:** | 890.5500 |
| **Regulation Description:** | Infrared lamp |
| **Regulation Class:** | Class 2 |
1 / 8
{5}
| **Panel:** | General & Plastic Surgery |
| --- | --- |
| **Product Code:** | OAP |
#### 4. Indication for Use
The Laser Hair Growth Devices (HR-H1, HR-H2, HR-H3, HR-H4, HR-H5, HR-H6, HR-H7, HR-H8, HR-H9, HR-H10, HR-I1, HR-I2, HR-I3, HR-I4, HR-I5) is used to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classification of IIa~V or females with androgenic alopecia who have Ludwig-Savin Classifications of I~II and both with Fitzpatrick Skin Phototypes I to IV.
#### 5. Device Description
The Laser Hair Growth Devices consists of Laser hair growth Cap、USB Power Cable、Manual and Warranty Card/Certificate of Conformity.
Laser diodes are the core components of Laser Hair Growth Devices, typically emitting low-energy lasers (LLLT) at a red wavelength of 650 ± 10 nm. The light source of the laser hair Growth Devices is visible red light, it is used for promoting hair growth.
#### 6. Compared to Predicate Device
Compared to the predicate devices, the subject device has the same intended use, similar product design, similar performance, same safety as the predicate device, the summarized comparison information is listed in the following table :
| SE Comparison Elements | Subject Device (Laser Hair Growth Devices) | Predicate Device(Laser Hair Growth Cap) K192627 | Comparison Elements |
| --- | --- | --- | --- |
| Trade name | Laser Hair Growth Devices | Laser Hair Growth Cap | N/A |
| Classification name | Infrared Lamp | Infrared Lamp | |
| Product code | OAP | OAP | |
2 / 8
{6}
| Intended use/Indications for Use | Laser Hair Growth Devices is used to promote hair growth in males with androgenic alopecia who have Norwood-hamilton classification of IIa-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-II and both with Fitzpatrick Skin Prototypes I to IV | Laser Hair Growth Cap is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classification of IIa-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-II and both with Fitzpatrick Skin Prototype I-IV. | Same |
| --- | --- | --- | --- |
| Location for use | OTC application | OTC application | Same |
| Type of Light | Low level laser therapy (LLLT); Laser diodes | Low level laser therapy (LLLT); Laser diodes | Same |
| Wavelength | 650nm±10nm | 650nm | Same |
| Amount of diodes laser | HR-H1: 98 HR-H2: 132 HR-H3: 172 HR-H4: 212 HR-H5: 272 HR-H6: 98 HR-H7: 132 HR-H8: 172 HR-H9: 212 HR-H10:n272 HR-I1: 98 HR-I2: 132 HR-I3: 172 HR-I4: 212 HR-I5: 272 | TW280:280 TW272: 272 TW147: 147 TW080:80 | Note1 |
| Energy of per laser diode | 5mW | 5mW | Same |
| Classification according to IEC60825-1 | Class 3R | Class 3R | Same |
| Treatment time | Each treatment :30 min; 2-3 treatments per weeks; about10-12 weeks. | Each treatment: 30min; 16 weeks, 3 times per week spaced out every other day | Note 2 |
3 / 8
{7}
| Applicable people | Norwood-Hamilton IIa~V (males) Ludwig-Savin I~II (females) | Male: Norwood-Hamilton IIa~V Female: Ludwig-Savin I~II | Same |
| --- | --- | --- | --- |
| Applicable skin | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV | Same |
| Helmet/Cap design | Helmet/Cap | Cap | Note 3 |
| Dimension (L*W*H) | Cap/small size: 18.9*21.7*9.6cm Cap/large size: 19.2*22*11.4mm Helmet: 22*26*13.6mm | 22*18*9cm | |
| Weight | HR-H1: 0.2Kg HR-H2:0.21Kg HR-H3: 0.22Kg HR-H4: 0.23Kg HR-H5: 0.24Kg HR-H6: 0.23Kg HR-H7: 0.24Kg HR-H8:0.25Kg HR-H9: 0.26Kg HR-H10:0.27Kg HR-I1:0.65Kg HR-I2: 0.66Kg HR-I3:0.67Kg HR-I4:0.68Kg HR-I5:0.69Kg | TW280:0.32Kg TW272:0.30Kg TW147: 0.26Kg TW080:0.25Kg | Note 4 |
| Environment for operation | Ambient temperature: 5°C ~ 40 °C (41°F~104°F) Relative humidity: not greater than 80%RH | Temperature:5°C~30°C (41°F~8°F) Humidity: 15%~90% | Note 5 |
| Environment for storage | Temperature: -20°C~50°C (-4 ~122) relative humidity: not exceeding 93% | Temperature: -40°C ~70°C (-40°F~158°F) Humidity: 10%~100% | |
4 / 8
{8}
| Safety feature | Complied with IEC60601-1, IEC60601-1-11, IEC60601-1-2, IEC 60601-1-6 and IEC60825-1 | Complied with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60825-1 Complied with IEC 62133 (Battery pack) Complied with IEC 60950-1 (Adapter) | Note 6 |
| --- | --- | --- | --- |
| Biocompatibility feature | All body-contacting materials are complied with ISO10993- 5,ISO 10993-10 and ISO 10993-23. | All body-contacting materials are complied with ISO10993- 5 and ISO 10993-10 | Note 7 |
| Power Density (mW/cm2) | HR-H11.06 HR-H2: 1.42 HR-H3: 1.85 HR-H4: 2.28 HR-H5: 2.93 HR-H6: 0.88 HR-H7: 1.18 HR-H8: 1.54 HR-H9: 1.89 HR-H10: 2.43 HR-I1: 0.99 HR-I2:1.34 HR-I3: 1.74 HR-I4: 2.15 HR-I5:2.76 | Not Publicly Available | Note 8 |
# **Note 1:**
The subject device utilizes a different number of laser diodes compared to the predicate device. However, both devices use the same wavelength, operating mode (continuous low-level laser therapy), and same energy of per laser diode.
Performance testing demonstrates that the variation in the number of diodes does not alter the device's mechanism of action, energy delivery per treatment area, or safety profile.
Therefore, this difference represents an engineering design variation rather than a new technological characteristic. It does not raise new questions of safety or effectiveness, supporting a conclusion of substantial equivalence between the subject and predicate devices.
# **Note 2:**
The subjects and predicate devices have the same intended use: to promote hair growth through low-intensity laser therapy (LLLT).
The test equipment was programmed with different recommended treatment cycles and
5 / 8
{9}
frequencies (each treatment: 30 minutes; 16 weeks, 3 times per week spaced out of every other day) compared to the equivalent device.
This difference represents a change in user instructions, rather than a change in device technology.
The energy delivered in each treatment course is maintained within the same clinically established safe and effective range as the equivalent device.
Performance tests have confirmed that differences in treatment regimens do not lead to new safety or efficacy issues. Therefore, the main device is considered to be essentially equivalent to the equivalent device.
# **Note 3:**
The differences in appearance and structure do not introduce new safety or effectiveness issues, and the expected use is consistent; The difference between Helmet and Cap belongs to the design/configuration difference in appearance and wearing structure, rather than the difference in technical principles or energy transfer methods;
The wavelength and energy of per laser diode are consistent, and structural differences do not pose any new risks.
# **Note 4:**
The technical features remain unchanged, and the laser wavelength, power output, electrical parameters, and thermal performance are consistent. The weight difference comes from the shell material, structural design, or heat dissipation method, which does not affect the laser performance or user safety.
It can be proven through performance testing, electrical safety testing (IEC 60601 series), ISO 10993-23, and risk analysis that existing differences in weight has not caused new safety issues.
# **Note 5:**
The above items, 'Environment for Operation' and 'Environment for Storage' of subject device are slightly different from the predicate devices, but it will not affect the main function and the intended use of the device as they all also comply with IEC 60601-1 requirements. **Note 6:**
The tested equipment does not have a power adapter as a factory component, but users are required to use a medical grade adapter that complies with IEC 60601-1 (as reflected in the instruction manual). This is a system configuration variation that does not change the electrical principles or affect safety. As long as the manufacturer's instructions and control measures are sufficient, it is considered an acceptable variation.
# **Note 7:**
The subject device underwent stimulus assessment based on the latest ISO 10993-23 standard, which provides a more scientific and ethical evaluation method. Compared to traditional animal testing based on ISO 10993-10 for predicate devices, the testing coverage of Subject devices is wider and meets current regulatory requirements, further verifying the safety equivalence or superiority of the device in terms of biocompatibility. **Note 8:**
Although quantitative irradiance and energy density data are not available for the predicate device (K192627), a comparative engineering assessment was performed based on device
6 / 8
{10}
design specifications, including wavelength (650 nm), number of laser diodes, and optical output power per diode (5 mW).
The calculated irradiance and energy density of the subject device fall within expected ranges for LLLT-based devices and are consistent with the intended therapeutic mechanism.
Differences in the number of laser diodes result in variations in treatment coverage area rather than changes to the fundamental technological characteristics or mode of action of the device.
## 7. Non-Clinical Performance Testing
The following non-clinical performance testing was conducted to support the substantial equivalence determination for the subject device.
### 1) Biocompatibility Testing
The biocompatibility evaluation of the body-contacting materials of the Laser Hair Growth Device was conducted in accordance with ISO 10993-1. Testing included:
- ISO 10993-5:2009 Tests for In Vitro Cytotoxicity
- ISO 10993-10:2021 Tests for Skin Sensitization
- ISO 10993-23:2021 Tests for Skin irritation
The results demonstrated that the device is biocompatible for its intended use.
### 2) Electrical and EMC Safety
Electrical safety and EMC safety testing was performed to, and passed, the following standards:
- IEC 60601-1:2005+A1:2012+A2:2020+ES Medical electrical equipment –Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11:2015+A1:2020 Medical electrical equipment –Part 1-11: General requirements for basic safety and essential performance –Collateral standard:
Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-1-2:2014+A1:2020 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility – Requirements and tests
- IEC 60601-1-6:2010+A1:2013+A2:2020 Medical electrical equipment Part 1-6 General requirements for safety - Collateral Standard: Usability
- IEC TR 60601-4-2: 2016 Medical electrical equipment. Part 4-2: Guidance and interpretation- Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- IEC 60825-1:2014 Safety of laser products - Part 1: Equipment classification and requirements certifies the laser system to classification 3R, which is the same as the predicate devices.
### 3) Software Verification and Validation Testing
Software verification and validation testing were conducted and basic level of documentation
7 / 8
{11}
was provided as recommended by FDA'S Guidance for Industry and FDA
Staff, 'Content of Premarket Submissions for Device Software Functions (2023).'
#### 4) Performance Testing
A dedicated performance testing program was conducted to evaluate the functional performance of the device under normal operating conditions.
The performance testing included the following evaluations:
- \* Optical output power verification per laser diode
- \* Laser wavelength accuracy verification
- \* Irradiance (power density) measurement across the treatment area
- \* Spatial uniformity of optical output
- \* Output stability under continuous operation
- \* Operating time / treatment cycle verification
- \* Surface temperature evaluation during operation
- \* Acoustic noise measurement during normal use conditions
All performance testing was conducted using calibrated test equipment as described in the System Test Report and associated validation documentation.
The results of these evaluations confirm that the device operates within its specified performance specifications and consistently delivers the intended low-level laser therapy (LLLT) optical output under normal operating conditions for its intended use.
#### 8. Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
#### 9. Conclusion
The subject device and predicate device have similar technical features in terms of basic design, characteristics, operating methods, working modes, applications, and expected uses. The subject device will not pose any new potential safety risks and its performance is substantially equivalent to the predicate device.
8 / 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.