The SimplGuide guiding sheath is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The SimplGuide guiding sheath is intended to be used in the coronary or peripheral vascular system.
Device Story
SimplGuide is a guiding sheath (GS) and vessel dilator (VD) system used by clinicians to provide a pathway for therapeutic/diagnostic devices into coronary or peripheral vasculature. The device is advanced over a 0.035" guidewire. It features a reinforced body, distal radiopaque marker, and hydrophilic coating on the distal 75cm. The device is operated manually. It facilitates interventional procedures by maintaining access and guiding instruments to target sites. It is a single-use, sterile device.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-4, -5, -7, -10, -11, -18, -23) and performance testing (inner/outer diameter, kink resistance, radial stiffness, coating integrity, pull force, radiopacity, and USP <788> particulate testing).
Technological Characteristics
Reinforced sheath with vessel dilator. Materials: Inner layer (PolyAmide/PolyEthylene blend), outer layer (Nylon/polyether-blok-amide), stainless steel reinforcement, radiopaque marker (Nylon with Tungsten). Hydrophilic coating (Polyvinylpyrrolidone). Sterilization: Ethylene Oxide (ISO 11135-1, SAL 10^-6). Dimensions: 100cm length, 2.10mm ID, 2.48mm OD. Straight tip configuration. No software or electronics.
Indications for Use
Indicated for the introduction of therapeutic and diagnostic devices into the peripheral and coronary vasculature through an access site.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 25, 2026
Pendracare
Erendira Rodriguez
Director of Quality Assurance and Regulatory Affairs
Van Der Waalspark 20-22
Leek, 9351VC
Netherlands
Re: K261242
Trade/Device Name: SimplGuide
Regulation Number: 21 CFR 870.1340
Regulation Name: Catheter Introducer
Regulatory Class: Class II
Product Code: DYB
Dated: April 15, 2026
Received: April 15, 2026
Dear Erendira Rodriguez:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
{1}
K261242 - Erendira Rodriguez
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261242 - Erendira Rodriguez
Page 3
Sincerely, **FINN E. DONALDSON - S** Digitally signed by FINN E. DONALDSON -S
Date: 2026.06.25
07:59:09 -04'00'
Finn Donaldson
Acting Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K261242
Device Name
SimplGuide
Indications for Use (Describe)
The SimplGuide guiding sheath is indicated for the introduction of therapeutic and diagnostic devices into the peripheral and coronary vasculature through an access site.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Traditional 510(k) Notification
PendraCare SimplGuide Guiding Sheath
K261242
Page 1 of 7
# **510(k) Summary**
Per 21 CFR 807.92
**A. Submitter information (807.92 (a) (1))**
**Prepared by:**
Eréndira Rodríguez
Quality Assurance and Regulatory Affairs Director
PendraCare International B.V.
Tel +31 85 0073 000
**Owner/Operator:**
***Manufacturer (510(k) Applicant)***
PendraCare International B.V.
Van der Waalspark 22
9351VC, Leek
The Netherlands
Registration: 3005635992
FEI: 3005635992
***Sterilization Facility***
Sterigenics Belgium Petit Rechain S.A.
Zoning Industriel De Petit-Rechain
Avenue Andre Ernst 21
Verviers Liege, BE B-4800
Registration Number: 3005941719
FEI Number: 3002807111
**Contact Person:**
Eréndira Rodríguez
Quality Assurance and Regulatory Affairs Director
PendraCare International B.V.
Van der Waalspark 22
9351VC, Leek
The Netherlands
Tel +31 85 0073 000
Email: erendira.rodriguez@welling.com
**Date Prepared:**
April 14, 2026
**B. Device Name**
| Trade or proprietary name: | SimplGuide |
| --- | --- |
| Common or usual name: | Guiding Sheath |
| Classification name: | Class II, 21 CFR 870.1340 |
| Review Panel: | Cardiovascular |
| Product Code: | DYB |
**C. Predicate Device information**
The legally marketed devices which the 510(k) used for claiming equivalence are
- R2P Destination Slender Guiding Sheath (K193125) (Predicate device)
- Convey 7F Guiding Catheters (K132197, July 11, 2013) (Reference device)
- Primum Hydrophilic Guiding Catheter ( K250972, June 29, 2025) (Reference device)
{5}
Traditional 510(k) Notification
PendraCare SimplGuide Guiding Sheath
K261242
Page 2 of 7
# D. Device Description
Component 1: The SimplGuide guiding sheath (GS) consists of a reinforced body with a hub and strain relief at the proximal end and a radiopaque marker band and distal tip at the distal end. The distal part of the GS is straight without a shape. A part of the GS body features a hydrophilic coating. The GS is developed without side holes.
Component 2: The vessel dilator (VD) is supplied together with the GS. The combination GS and VD is introduced through the skin tissue into the vascular system of the patient, and advanced through the vasculature to the target location over a routinely used 0.035" guidewire.
The Design Inputs and Design Outputs are gathered in attachment 1 (Design Input – Design Output Table) per SOP R&D-006 (Design Input – Design Output).
# E. Intended use of Device
The SimplGuide guiding sheath is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The SimplGuide guiding sheath is intended to be used in the coronary or peripheral vascular system.
# F. Indications for Use
The SimplGuide guiding sheath is indicated for the introduction of therapeutic and diagnostic devices into the peripheral and coronary vasculature through an access site.
# G. Technological Characteristics / Substantial Equivalence
See Comparison tables (Table 1 and Table 2) below.
{6}
Traditional 510(k) Notification
PendraCare SimplGuide Guiding Sheath
K261242
Page 3 of 7
**Table 1: Comparison Predicate Device and Proposed Device**
| | Predicate Device | New Proposed Device |
| --- | --- | --- |
| | **R2P Destination Slender guiding sheath (6F) (K193125)** | **SimplGuide guiding sheath (6F) (K261242)** |
| **Product Code** | **DYB (Introducer, catheter)** | **Same** |
| **Trade Name** | R2P™ Destination Slender™ Guiding Sheath | SimplGuide |
| **Indications for use** | R2P™ Destination Slender™ Guiding Sheath is indicated for the introduction of interventional and diagnostic devices into the human vasculature through an access site, including but not limited to the radial artery. | The SimplGuide guiding sheath is indicated for the introduction of therapeutic and diagnostic devices into the peripheral and coronary vasculature through an access site. |
| **Operating Principle** | Operated manually or by a manual process | Same |
| **Design** | Length: 75cm, 85cm, **95cm, 105cm**, 119cm and 149cm | Length: 100 cm |
| | ID: 2.2mm | ID: 2.10mm |
| | OD: 2.5mm | OD: 2.48mm |
| *- design / construction* | Sheath, Dilator, Hemostatic Valve with side tube and three-way stopcock | Sheath, Vessel Dilator |
| *- outer hydrophilic coating* | Yes | Yes |
| *- hydrophilic coating length* | Hydrophilic Coating: full usable length | Hydrophilic Coating: 75cm distal length |
| *- tip configuration* | Straight tip | Same |
| *- tip visibility material* | Radiopaque Tip: Nylon with Tungsten | Same |
| *- body reinforcement config* | Coiled reinforcement | Braided reinforcement |
| *- hub* | Nylon | Nylon Colorant |
| **Vessel dilator** | Vessel dilator included | Same |
| **Vessel dilator extended length** | 5cm | Same |
| **Hemostasis device** | Hemostasis device: cross cut hemostatic valve, three way stopcock | No hemostasis device included |
| **Materials** | Inner layer: PTFE | Inner layer: PolyAmide/PolyEthylene blend |
| | Outer Layer: Nylon / polyether-blok-amide | Same |
{7}
Traditional 510(k) Notification
PendraCare SimplGuide Guiding Sheath
K261242
Page 4 of 7
| | Predicate Device | New Proposed Device |
| --- | --- | --- |
| | **R2P Destination Slender guiding sheath (6F) (K193125)** | **SimplGuide guiding sheath (6F) (K261242)** |
| | Stainless steel reinforcement | Same |
| | Markerband Nylon with Tungsten | Same |
| **- hydrophilic coating composition** | Polyvinylpyrrolidone-based coating | Polyvinylpyrrolidone water-soluble polymer (hydrophilic coating) & primer |
| **DILATOR** | **Dilator Assembly** Tubing: Polypropylene. Hub: Polypropylene/Thermoplastic Elastomer. Blend Coating: Silicone Caulking Pin: Stainless steel. | **Dilator Assembly** Tubing: Pebax 7033 / Bismuth subcarbonate / Zinc stearate. Glue: Acrylate glue. Hub: Polycarbonate |
| **Shelf-life** | 30 months | 36 months |
| **Sterility** | Ethylene Oxide (validated in accordance with ANSI / AAMI / ISO 11135-1 to achieve SAL 10^{-6}) | Same |
| **Packaging** | Unit Pouch; Shelf Box; Shipping Carton | Same |
**Table 2: Comparison Reference Device and Proposed Device**
| | Reference Device | New Proposed Device |
| --- | --- | --- |
| | **Convey Guiding Catheter (7F) (K132197)** ***Primum Hydrophilic Guiding Catheter (K250972)** | **SimplGuide guiding sheath (6F) K261242** |
| **Product Code** | **DQY (Catheter percutaneous)** | **DYB (Introducer, catheter)** |
| **Trade Name** | Convey™ Guiding Catheter *Primum Hydrophilic Guiding Catheter | SimplGuide |
| **Indications for use** | The Convey Guiding/*Primum Guiding Catheter is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The Convey Guiding Catheter is intended to be used in the coronary or peripheral vascular system. | The SimplGuide guiding sheath is indicated for the introduction of therapeutic and diagnostic devices into the peripheral and coronary vasculature through an access site. |
| **Operating Principle** | Operated manually or by a manual process | Same |
| **Design** | Length: 100 cm | Same |
| | ID: 2.08mm | ID: 2.10mm |
| | OD: 2.37mm | OD: 2.48mm |
{8}
Traditional 510(k) Notification
PendraCare SimplGuide Guiding Sheath
K261242
Page 5 of 7
| | Reference Device | New Proposed Device |
| --- | --- | --- |
| | **Convey Guiding Catheter (7F) (K132197)** ***Primum Hydrophilic Guiding Catheter (K250972)** | **SimplGuide guiding sheath (6F) K261242** |
| - *design / construction* | Catheter | Sheath, Vessel Dilator |
| - *outer hydrophilic coating* | Yes | Same |
| - *hydrophilic coating length* | Hydrophilic Coating Length: 75cm proximal of the intermediate tip | Hydrophilic Coating: 75cm distal length |
| - *tip configuration* | Only shaped tips | Straight tip |
| - *tip visibility material* | Radiopaque soft tip | Radiopaque Tip: Nylon with Tungsten |
| - *body reinforcement config* | Braided reinforcement | Same |
| - *hub* | Nylon Colorant | Same |
| **Vessel dilator** | No vessel dilator included | Vessel dilator included |
| **Vessel dilator extended length** | N.A. | 5cm |
| **Hemostasis device** | No hemostasis device included | Same |
| **Materials** | Inner layer: PolyAmide/PolyEthylene blend | Same |
| | Outer Layer: Nylon / polyether-blok-amide | Same |
| | Stainless steel reinforcement | Same |
| | Bismuth oxychloride filled marker material | Markerband Nylon with Tungsten |
| - *hydrophilic coating composition* | Polyvinylpyrrolidone PVP water-soluble polymer (hydrophilic coating) & primer | Same |
| **DILATOR** | N/A | **Dilator Assembly** Tubing: Pebax 7033 / Bismuth subcarbonate / Zinc stearate. Glue: Acrylate glue. Hub: Polycarbonate |
| **Shelf-life** | 36 months | Same |
| **Sterility** | Ethylene Oxide (validated in accordance with ANSI / AAMI / ISO 11135-1 to achieve SAL 10^{-6}) | Same |
| **Packaging** | Unit Pouch; Shelf Box; Shipping Carton | Same |
{9}
Traditional 510(k) Notification
PendraCare SimplGuide Guiding Sheath
K261242
Page 6 of 7
### H. Summary biocompatibility and bench testing
The following biocompatibility tests were completed on the SimplGuide Guiding Sheath considering the categorization of the Material Characterization for medical application per ISO 10993-1 and USP.
#### ISO 10993-4: Hemocompatibility
- Hemolysis Test (Complete - Direct and Indirect Contact)
- Partial Thromboplastin Time (UPTT) Test – Direct Contact Method
- Complement Activation (C3 Complement Protein) Test – Indirect Contact Method
- Complement Activation (SC5b-9 (C5) Complement Protein) Test – Indirect Contact Method
- Platelet and Leukocyte Count (Direct Contact)
#### ISO 10993-5: Cytotoxicity
- L929 Minimal Essential Medium (MEM) Elution Test
#### ISO 10993-7: Ethylene Oxide Sterilization Residuals
- Simulated use: residue value after 24 hours at 37°C by Gas chromatography.
#### ISO 10993-10: Sensitization
- Kligman Maximization Test – Normal Saline and Vegetable (Cottonseed) Oil Extracts
#### ISO 10993-10: Irritation/ Intracutaneous Reactivity
- Intracutaneous Injection Test – Normal Saline and Vegetable (Cottonseed) Oil Extracts
#### ISO 10993-11: Systemic toxicity
- Systemic Injection Test
#### ISO 10993-18: 2020 /Amd: 2023 Biological evaluation of medical devices - Part 18: Chemical characterization of materials
- Chemical characterization of the materials is performed.
#### ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Test for irritation
- Intracutaneous Injection Test
#### USP <85> Endotoxin-Mediated Pyrogenicity
- LAL test.
The following in-vitro performances tests (following simulated use, if applicable) were completed of the SimplGuide Guiding Sheath:
- Inner Diameter
- Outer Diameter
- Catheter Usable Length
- Coating Length
- C-Kink (Bending Kink Diameter)
- Euler Kink (axial Kink Displacement)
- Radial Stiffness (Collapse)
- Coating Integrity (Visual Inspection)
- Outer Friction & Wear (Coating Integrity – functional tests)
- Three Point Bending Test (ending Stiffness Body)
- Pull Force (after simulated use)
- Radiopacity
- USP <788> Particulate testing (Light Obscuration after simulated use). Particle Count Test
{10}
Traditional 510(k) Notification
PendraCare SimplGuide Guiding Sheath
K261242
Page 7 of 7
### I. Clinical and Nonclinical Performance Data and Conclusions
This 510(k) does not include data from clinical tests.
The User requirements were validated and concluded that it meets the Product Description and design for its intended purpose.
Performance testing was conducted to ensure that the SimplGuide Guiding Sheath met the applicable design and performance requirements during the shelf life, verify conformity to the applicable external and internal standards, and demonstrate substantial equivalence to the predicate device.
### J. Conclusion
In summary, the PendraCare SimplGuide Guiding Sheath is substantially equivalent to its legally marketed predicate device, the R2P Destination Slender Guiding Sheath (K193125) with respect to the intended use, operating principles, fundamental design, catheter dimension, materials, technology, packaging, labeling and sterility.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.