Automatic Biopsy Needles are intended in obtaining biopsy samples from soft tissues such as liver, kidney, prostate, breast, lymph nodes and various soft tissue tumors. It is not intended for use in bone.
Device Story
Hand-operated, non-electronic, sterile, single-use automatic biopsy needle; used by healthcare professionals in professional healthcare facilities. Device utilizes a spring-loaded guillotine mechanism to extract soft tissue specimens. Operator pushes outer needle backward to compress a spring; spring resilience triggers outer needle movement to capture tissue in the inner needle sampling slot. Used under imaging guidance (ultrasound, X-ray, CT); not MRI compatible. Output is a tissue specimen for pathological examination. Benefits include automated sampling with minimal surrounding tissue damage.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including cleanliness, puncture force, mechanical power unit coordination, bond strength, and resistance to breakage/corrosion.
Technological Characteristics
Disposable, sterile, single-use, spring-loaded biopsy system. Materials: 304 stainless steel (X5CrNi18-10) and X12CrNi17-7 stainless steel. Sizes: 12G-20G; lengths: 90mm-250mm. Sterilization: EO (SAL 10^-6). Standards: ISO 9626:2016, ISO 10993 series. Non-electronic, no energy source.
Indications for Use
Indicated for adults requiring biopsy for sampling of soft tissue abnormalities, including liver, kidney, prostate, breast, lymph nodes, and soft tissue tumors. Not for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 12, 2026
Promisemed Hangzhou Meditech Co., Ltd.
Zearou Yang
Regulatory affairs manager
#1388 Cangxing St., Cangqian Community, Yuhang District
Hangzhou City, 311121
China
Re: K261196
Trade/Device Name: Promisemed VeriEcto Automatic Biopsy Needles
Regulation Number: 21 CFR 876.1075
Regulation Name: Gastroenterology-Urology Biopsy Instrument
Regulatory Class: Class II
Product Code: KNW
Dated: April 11, 2026
Received: April 13, 2026
Dear Zearou Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261196 - Zearou Yang
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261196 - Zearou Yang
Page 3
Sincerely,
JESSICA CARR -S
Jessica Carr, PhD
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261196 | ? |
| Please provide the device trade name(s). | | ? |
| Promisemed VeriEcto Automatic Biopsy Needles | | |
| Please provide your Indications for Use below. | | ? |
| Automatic Biopsy Needles are intended in obtaining biopsy samples from soft tissues such as liver, kidney, prostate, breast, lymph nodes and various soft tissue tumors. It is not intended for use in bone. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
Promisemed
Promisemed Hangzhou Meditech Co., Ltd.
No. 1388 Cangxing Street, Cangqian Community, Yuhang District,
Hangzhou City, 311121 Zhejiang, P. R. China.
# K261196-510(k) Summary
Date prepared: 2026-05-12
| 1. Manufacturer [21 CFR 807.92 (a) (1)] | |
| --- | --- |
| Name | Promisemed Hangzhou Meditech Co., Ltd. |
| Address | No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou City, 311121 Zhejiang, China. |
| Contact Person | Zearou Yang |
| Phone | +86 571 88772985 |
| Email | zearou.yang@promisemed.ca |
| 2. Device [21 CFR 807.92 (a) (2)] | |
| Name | Promisemed VeriEcto Automatic Biopsy Needles |
| Common Name | Automatic Biopsy Needles |
| Classification Name | Gastroenterology-urology biopsy instrument |
| Regulation Number | 21 CFR 876.1075 |
| Class | Class II |
| Product Code | KNW, FCG |
| 3. Legally Marketed Predicate Device [21 CFR 807.92 (a) (3)] | |
| Predicate Name | Promisemed Automatic Biopsy Needles |
| 510(k) Number | K210946 |
| Product Code | KNW, FCG |
| Reference Devices | No reference devices were used in this submission. |
| 4. Device Description [21 CFR 807.92 (a) (4)] | |
| Promisemed VeriEcto Automatic Biopsy Needles are hand-operated, non-electronic, surgical instruments. Promisemed VeriEcto Automatic Biopsy Needle is designed for the automatic extraction of a specimen from soft tissues, while causing minimal surrounding tissue damage, for tissue pathological examination/ testing. Push the outer needle backward to expose the sampling slot of the inner needle by compression spring, then the spring's resilience realizes the movement of the outer needle and the excitation function. It is a sterile with a Sterility Assurance Level (SAL) of 10-6, non-pyrogenic and single-use device. The ultrasound, X-ray, CT and other equipment are used to guide the puncture and sampling. This product can't be used under MRI. | |
| 5. Indication for use [21 CFR 807.92 (a) (5)] | |
| The automatic biopsy needles are intended in obtaining biopsy samples from soft tissues such as liver, kidney, prostate, breast, lymph nodes and various soft tissue tumors. It is not intended for use in bone. | |
Page 1 of 4
{5}
Promisemed
Promisemed Hangzhou Meditech Co., Ltd.
No. 1388 Cangxing Street, Cangqian Community, Yuhang District,
Hangzhou City, 311121 Zhejiang, P. R. China.
# 6. Comparison of Technological Characteristic [21 CFR 807.92 (a) (6)]
The Promisemed VeriEcto Automatic Biopsy Needle is substantially equivalent to the predicate device, K210946. Both devices have the same intended use. The basic technological and operating principles are the same for both devices. Both the subject and predicate devices are disposable, sterile, single-patient-use devices.
While certain technological differences exist between the subject and predicate devices, these differences do not raise different questions of safety and effectiveness and do not affect the fundamental design principles or safe usage of the device.
| Item of description | Predicate device (K210946) | Subject device (K261196) | | | Similar/ Differences | Safety / Effectiveness Statement |
| --- | --- | --- | --- | --- | --- | --- |
| Common name | Promisemed Automatic Biopsy Needles | Promisemed VeriEcto Automatic Biopsy Needles | | | N/A | / |
| Manufacturer | Promisemed Hangzhou Meditech Co., Ltd. | Promisemed Hangzhou Meditech Co., Ltd. | | | Same | All produced by the same manufacturer. |
| Device class | Class II | Class II | | | Same | No new concerns. |
| Product Code | KNW, FCG | KNW, FCG | | | Same | No new concerns. |
| Intended use/ Indications for use | Automatic biopsy needles are intended in obtaining biopsy samples from soft tissues such as liver, kidney, prostate, breast, lymph nodes and various soft tissue tumors. It is not intended for use in bone. | The automatic biopsy needles are intended in obtaining biopsy samples from soft tissues such as liver, kidney, prostate, breast, lymph nodes and various soft tissue tumors. It is not intended for use in bone. | | | Same | No new concerns. More details about warning was added for safely usage. |
| Target population | Individuals requiring biopsy for sampling of soft tissue abnormalities | Individuals requiring biopsy for sampling of soft tissue abnormalities | | | Same | No new concerns. |
| Anatomical site | Soft tissues such as liver, kidney, prostate, breast, lymph nodes and various soft tissue tumors | Soft tissues such as liver, kidney, prostate, breast, lymph nodes and various soft tissue tumors | | | Same | No new concerns. |
| Where used | Professional Healthcare Facility | Professional Healthcare Facility | | | Same | No new concerns. |
| Human factors | Healthcare professional users | Healthcare professional users | | | Same | No new concerns. |
| Design | Disposable programmable automatic spring-loaded guillotine style biopsy system for histological biopsy on soft tissue. | Disposable programmable automatic spring-loaded guillotine style biopsy system for histological biopsy on soft tissue. | | | Same | No new concerns. The fundamental technology and operation remain the same. |
| Structure | The disposable automatic biopsy needle is composed of biopsy needle and coaxial needle (optional). The device is composed of the inner needle, outer needle, mechanical power unit and sheath. The Co-Axial Biopsy Device is composed of the inner needle, outer needle, sheath, casing seat, base and screw cap for the models with the Co-Axial Biopsy Device. | The disposable automatic biopsy needle is composed of biopsy needle and coaxial needle (optional). The device is composed of the inner needle, outer needle, mechanical power unit and sheath. The Co-Axial Biopsy Device is composed of the inner needle, outer needle, sheath, casing seat, base and screw cap for the models with the Co-Axial Biopsy Device. | | | Similar | No new concerns. The fundamental technology and operation remain the same.. |
| Materials (tissue-contacting) | Inner needle and outer needle are made out of 304 stainless steel(X5CrNi18-10). Only Stainless steel is in direct surgical contact with all soft tissues | Component | Material (ISO number) | Contact methods | Difference | Subject device inner needle of X12CrNi17-7 is contacting the tissue with Less than one hour. It does not introduce new risks through biocompatibility testing. |
| | | Inner needle (Biopsy Needle) | Stainless steel (X5CrNi17-7) | Direct contact Internal tissue | | |
| | | Outer needle (Biopsy Needle, Co-axial needle) | Stainless steel (X5CrNi18-10) | Externally communicating, direct contact tissue. | | |
{6}
Promisemed
Promisemed Hangzhou Meditech Co., Ltd.
No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou City, 311121 Zhejiang, P. R. China.
| | | inner needle (Co-axial needle) | Stainless steel (XSC/Ni:1 B-10) | Externally communicating, direct contact tissue | | |
| --- | --- | --- | --- | --- | --- | --- |
| Needle size (Biopsy Needle) | 12G- 20G | 12G, 14G, 16G, 18G, 20G | | | Same | No new concerns. |
| Needle length (Biopsy Needle) | 90mm- 200mm. | 90mm, 100mm, 130mm, 150mm, 160mm, 200mm, 220mm, 250mm | | | Difference | 220mm and 250mm were added. Broader needle length for more selection during clinical procedure. Difference of needle length does not introduce new risks as the fundamental technology and operation remain the same. |
| Needle size (Co-axial needle) | 11G-19G. | 11G, 13G, 15G, 17G, 19G. | | | Same | No new concerns. |
| Needle length (Co-axial needle) | 60mm-190mm. | 60mm, 70mm, 100mm, 120mm, 130mm, 170mm, 190mm, 220mm. | | | Difference | 220mm was added. Broader needle length for more selection during clinical procedure. Difference of needle length does not introduce new risks as the fundamental technology and operation remain the same. |
| Sterilization method | EO sterilization | EO sterilization | | | Same | No new concerns. |
| Sterility | SAL of 10^{-6} | SAL of 10^{-6} | | | Same | No new concerns. |
| Single use | Yes | Yes | | | Same | No new concerns. |
| Biocompatibility | Biocompatibility established | Biocompatibility established | | | Same | No new concerns. |
| Standards met | - ISO 9626:2016; - ISO 10993 series; | - ISO 9626:2016; - ISO 10993 series; | | | Same | No new concerns. |
| Compatibility with the environment and other devices | The introduction of the needle into the body should be carried out under imaging guidance (ultrasound, X-Ray, CT, etc.) | The introduction of the needle into the body should be carried out under imaging guidance (ultrasound, X-Ray, CT, etc.) | | | Same | No new concerns. |
| Energy used and/or delivered | Not applicable | Not applicable | | | / | / |
| Electrical safety | Not applicable | Not applicable | | | / | / |
| Mechanical safety | Not applicable | Not applicable | | | / | / |
| Chemical safety | Not applicable | Not applicable | | | / | / |
| Thermal safety | Not applicable | Not applicable | | | / | / |
| Radiation safety | Not applicable | Not applicable | | | / | / |
The following changes were identified between the subject and predicate devices:
- Adding of needle lengths: The addition of new needle length expands options and new type without altering the device's safety or effectiveness when compared to the predicate device.
- Difference of inner needle material: biocompatibility of subject device material has been tested in accordance with ISO 10993 series that indicated is safely.
Page 3 of 4
{7}
Promisemed
Promisemed Hangzhou Meditech Co., Ltd.
No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou City, 311121 Zhejiang, P. R. China.
The subject device has the same intended use and technological characteristics as the predicate device. The basic technological and operating principles are the same for both devices. Both the subject and predicate devices are disposable, sterile, single patient use devices. As evidenced by comparison table above, the subject device is substantially equivalent to the predicate device.
## 7. Nonclinical test [21 CFR 807.92 (b) (1)]
Performance testing confirms that the output meets the design inputs and specifications. Internal verification and validation testing confirms that product specifications are met, which support the intended use and technological characteristics as compared to the predicate devices. The information supports the substantial equivalence to the referenced predicate device:
- Cleanliness;
- Size of outer needle;
- Needle points;
- Puncture force of needle point;
- Coordination of biopsy needle;
- Mechanical power unit;
- Bond between outer needle and upper slider;
- Bond between Inner needle and lower slider;
- Sheath;
- Resistance to breakage;
- Resistance to corrosion;
- Scale identification;
- Sampling type;
The subject device is meeting all the requirements for overall design and performance.
## 8. Clinical test [21 CFR 807.92 (b) (2)]
No clinical testing was conducted for this submission.
## 9. Conclusion [21 CFR 807.92 (b) (3)]
Based on the nonclinical performance testing described above, the Promisemed VeriEcto Automatic Biopsy Needle (subject device) has been demonstrated to be substantially equivalent to the predicate device (K210946). The subject device has the same intended use as the predicate device. While certain technological differences exist (needle length and material), these differences do not raise different questions of safety and effectiveness. Therefore, the subject device is substantially equivalent to the predicate device and is appropriate for clearance under the 510(k) pathway.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.