K261179 · Gimmi GmbH · GEI · Jul 22, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261179
Device Name
AlphaTriPart - Monopolar MIC Instruments
Applicant
Gimmi GmbH
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 22, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Instruments for laparoscopic surgery are intended for cutting, resecting, gripping and coagulation of body tissue.
Device Story
AlphaTriPart is a 3-part monopolar laparoscopic instrument system comprising a handle, sheath, and instrument insert. Used in medical facilities by trained personnel; device connects to an external HF generator via a cable. Electrical current flows from the plug to the distal instrument tip upon activation of a foot pedal. Mechanical jaw movement is controlled by a pull/push wire connected to the jaw joint. Features include rotatable sheath/insert, Luer Lock cleaning port, and optional ratchet or non-locking handles. Designed for use through a trocar to perform tissue cutting, resection, gripping, and coagulation. Benefits include modular design for reprocessing and surgical versatility.
Clinical Evidence
No clinical data is required for this submission. Evidence consists of bench testing, including performance verification (impact, drop, and molding stress tests) and compliance with IEC 60601-2-2, IEC 60601-1-6, IEC 62366-1, and ISO 10993 series standards.
Technological Characteristics
Monopolar electrosurgical instrument; 3-part modular design (handle, sheath, insert). Materials include PEEK (handle). Sensing/actuation: mechanical pull/push wire for jaw movement. Energy source: external HF generator. Dimensions designed for trocar access. Reprocessing: Luer Lock cleaning port. Sterilization: not specified. Connectivity: manual connection to HF generator.
Indications for Use
Indicated for cutting, resecting, gripping, and coagulating body tissue in adult patients undergoing laparoscopic surgery. Contraindicated if a qualified physician determines the patient's general condition is inadequate or if endoscopic methods are contraindicated.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 22, 2026
Gimmi GmbH
Nicolas Geisheimer
Regulatory Affairs Representative
Carl-Zeiss-Str. 6 -
Tuttlingen, BW 78532
Germany
Re: K261179
Trade/Device Name: AlphaTriPart - Monopolar MIC Instruments
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: June 25, 2026
Received: June 25, 2026
Dear Nicolas Geisheimer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261179 - Nicolas Geisheimer
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261179 - Nicolas Geisheimer
Page 3
Sincerely,
**Colin K.
Chen -S**
Digitally signed by Colin K. Chen -
S
Date: 2026.07.22 15:27:04 -04'00'
Colin K. Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
K261179 Page 1 of 1
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261179 |
| --- | --- |
| Please provide the device trade name(s). | |
| AlphaTriPart - Monopolar MIC Instruments | |
| Please provide your Indications for Use below. | |
| Instruments for laparoscopic surgery are intended for cutting, resecting, gripping and coagulation of body tissue. | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) |
{4}
K261179
Page 1 of 3
Surgical technology
adapted to life.
Gimmi
®
### 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. All data included in this document are accurate and complete to the best of Gimmi's knowledge.
**Applicant:**
Gimmi GmbH
Carl-Zeiss-Str. 6
78532 Tuttlingen
Germany
Tel: +49 (7461) 96590 c
Fax: +49 (7461) 96590 33
**Contact:**
Nicolas Geisheimer / Anagha Balaji
Regulatory Affairs
RA@gimmi.de
**Date of Preparation:**
July 22nd, 2026
**Device Identification:**
Trade name:
AlphaTriPart – Monopolar MIC Instruments
Common name:
Monopolar MIC Instruments
Classification name:
Electrosurgical, cutting & coagulation & accessories
**Product Code:**
GEI
**Regulation:**
21 CFR 878.4400
**510(k) Type:**
Traditional
**Predicate Device:**
CLICKLINE; K954122, K935071; KARL STORZ SE & Co. KG
Laparoscopic Instruments, 3parts; K060232; Trokamed GmbH
**Device Description:**
The Gimmi AlphaTriPart system is a monopolar laparoscopic system consisting of 3-parts. One part is the handle, second the sheath and the third part is the instrument insert which is inserted through the sheath into the handle. Handle and sheath are isolated. Therefore, the electrical monopolar current (only when pressing the foot pedal of the HF unit) is flowing from the plug, where the HF cable is fixed, to the distal end of the instrument tip.
The Luer Lock cleaning port allows flushing of the inside of the sheath during reprocessing. Sheath and insert are rotatable. For the handle there are different types available – with or without ratchet. For AlphaTriPart there is in addition a variant with non-locking handle available.
Page 2 of 4
{5}
K261179
Page 2 of 3
Surgical technology adapted to life.
G
Gimmi
®
Principle of operation / mode of action:
The monopolar products are monopolar forceps and scissors electrodes, which consist of two movable jaw parts. The movement is carried out by a pull/ push wire, which is connected to the corresponding mouth part via a joint. Shaft and handles are isolated. The instrument jaw parts are available in different shapes and figures. The dimensions of the mouth parts as well as of the electrodes were designed for use through a trocar.
### Indication:
Instruments for laparoscopic surgery are intended for cutting, resecting, gripping and coagulating of body tissue.
Patient population
| Patients | Human medicine |
| --- | --- |
| Age | Use on Adults only |
| Health condition | Relevant, because patient is treated under anesthesia. |
Note: The products must be used only in medical facilities by trained and skilled medical personnel. The products must not be used if, according to a qualified physician, the general condition of the patient is not adequate or if the endoscopic methods are contraindicated.
### Technological Characteristics:
The products under consideration & the predicate device(s) are similar in all application related points and technical characteristics, including being monopolar and connecting to an HF generator. The labeling is consistent with the valid requirements. Both are made of the same materials, have identical indications and contraindications, and are used at the same site in the body. Some minor differences exist, such as the color of the PEEK handle and the connection type between parts. However, these variation's do not impact safety or effectiveness. Biocompatibility, mechanical, and electrical tests were successfully completed for all products. Consequently, the product under consideration and the predicate devices are considered to be substantially equivalent.
### Non- Clinical Performance Data:
The AlphaTriPart Instruments are tested according to the following standards:
IEC 60601-2-2
IEC 60601-1-6
IEC 62366-1
ISO 10993-1
ISO 10993-5
ISO 10993-10
ISO 10993-11
Additional bench testing for performance verification and validation purposes:
- Impact test
- Drop test
• Reduction of molding stresses
### Biological characteristics:
The predicate and the subject devices uses the same materials in contact with the same human tissues or body fluids for a similar kind and duration of contact and similar release characteristics of substances, including degradation products and leachables.
Page 3 of 4
{6}
K261179
Page 3 of 3
Surgical technology
adapted to life.
Gimmi®
# Clinical characteristics:
The predicate and the subject devices are used for the same clinical condition or purpose, including similar severity and stage of disease, at the same site in the body, in a similar population, including as regards age, anatomy and physiology; has the same kind of user; has similar relevant critical performance in view of the expected clinical effect for a specific intended purpose.
# Clinical Performance Data:
No Clinical data is required for this submission
# Conclusion:
The AlphaTriPart Instruments and the predicate device are similar in all application related points and technical characteristics, including being monopolar and connecting to an HF generator. The labeling is consistent with the valid requirements. They are made of the same materials, have identical indications and contraindications, and are used at the same site in the body. Some minor differences exist, such as the color of the PEEK handle and the connection type between parts. However, these variations do not impact safety or effectiveness. Biocompatibility, mechanical, and electrical tests were successfully completed for all products. Consequently, they are considered to be substantially equivalent.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.