← Product Code [NUC](/productcode/NUC) · K261177

# PINKSLIP Water-Based Personal Lubricant and Moisturizer (K261177)

_PINKSLIP, LLC · NUC · Jul 9, 2026 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K261177

## Device Facts

- **Applicant:** PINKSLIP, LLC
- **Product Code:** [NUC](/productcode/NUC.md)
- **Decision Date:** Jul 9, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

PINKSLIP Water-Based Personal Lubricant and Moisturizer is a water-based personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication. PINKSLIP Water-Based Personal Lubricant and Moisturizer is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

## Device Story

Water-based personal lubricant; non-sterile; formulated with water, sodium benzoate, potassium sorbate, organic aloe barbadensis leaf juice, sorbitol, xanthan gum, hyaluronic acid, and citric acid. Packaged in 74ml PET bottle. Intended for OTC use to supplement natural lubrication and enhance comfort during sexual activity. Compatible with natural rubber latex and polyisoprene condoms; incompatible with polyurethane condoms. Applied topically by user. Benefits include reduced friction and increased comfort during sexual activity.

## Clinical Evidence

Bench testing only. Biocompatibility testing per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization, vaginal irritation), and ISO 10993-11 (acute systemic toxicity) confirmed the device is non-cytotoxic, non-sensitizing, non-irritating, and non-systemically toxic. Condom compatibility tested per ASTM D7661-18. Shelf-life validated via accelerated aging study.

## Technological Characteristics

Water-based lubricant; pH 4.0-4.5; osmolality 220-290 mOSm/kg; viscosity 5,000-17,800 cPs. Ingredients: water, sodium benzoate, potassium sorbate, organic aloe barbadensis leaf juice, sorbitol, xanthan gum, hyaluronic acid, citric acid. Non-sterile. Packaged in PET bottle. Microbial testing per USP <61>, <62>, <1111>, and <51>.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- Desnuda Reflect ([K211998](/device/K211998.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**U.S. FOOD & DRUG**
ADMINISTRATION

July 9, 2026

Pinkslip, LLC
Louie Goryoka
Sr. VP. Regulatory and Quality Consultant
267 Flathead Ave.
Whitefish, Montana 59937

Re: K261177
Trade/Device Name: PINKSLIP Water-Based Personal Lubricant and Moisturizer
Regulation Number: 21 CFR 884.5300
Regulation Name: Condom
Regulatory Class: II
Product Code: NUC
Dated: April 10, 2026
Received: April 10, 2026

Dear Louie Goryoka:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K261177 - Louie Goryoka

Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}

K261177 - Louie Goryoka

Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Monica D. Garcia -S

Monica D. Garcia, Ph.D.

Assistant Director

DHT3B: Division of Reproductive,

Gynecology, and Urology Devices

OHT3: Office of Gastrorenal, ObGyn,

General Hospital, and Urology Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2023

See PRA Statement below.

510(k) Number (if known)

K261177

Device Name

PINKSLIP Water-Based Personal Lubricant and Moisturizer

Indications for Use (Describe)

PINKSLIP Water-Based Personal Lubricant and Moisturizer is a water-based personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication.

PINKSLIP Water-Based Personal Lubricant and Moisturizer is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

Type of Use (Select one or both, as applicable)

☐

Prescription Use (Part 21 CFR 801 Subpart D)

☑

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (301) 443-6740

EF

{4}

510(k) Summary

K261177

Page 1 of 5

K261177

PINKSLIP Water-Based Personal Lubricant and Moisturizer

# 1. Submitter Information:

PINKSLIP LLC

267 Flathead Ave,

Whitefish, MT. 59937, USA

(406) 314-2865

# 2. Submission Correspondent:

Louie Goryoka

Sr. VP. Regulatory and Quality Consultant

Med-Device Consulting, Inc.

Email: mdci@m-dci.us

(818) 585-7488

# 3. Date of Summary:

July 9, 2026

# 4. Device Identification:

Device Trade Name: PINKSLIP Water-Based Personal Lubricant and Moisturizer

Common Name: Personal Lubricant

Regulation Number: 21 CFR §884.5300

Regulation Name: Condom

Device Class: Class II

Product Code: NUC (lubricant, personal)

# 5. Predicate Device

Product Name: Desnuda Reflect

510(k) Number: K211998

The predicate device has not been subject to a design-related recall.

# 6. Device Description

PINKSLIP Water-Based Personal Lubricant and Moisturizer is a non-sterile personal lubricant. It is water-based, available over the counter, and formulated to be slightly yellow-colored and odorless. The subject device is compatible with natural rubber latex and lubricated polyisoprene male condoms. It is not compatible with polyurethane male condoms.

The formulation consists of water, sodium benzoate, potassium sorbate, organic aloe barbadensis leaf juice, sorbitol, xanthan gum, hyaluronic acid, and citric acid.

The device is neither a contraceptive nor a spermicide. The device is packaged in a pre-filled polyethylene terephthalate (PET) bottle (74ml).

{5}

510(k) Summary

K261177

Page 2 of 5

K261177

PINKSLIP Water-Based Personal Lubricant and Moisturizer

The table below lists the specifications of the subject device:

Table 1: Device Specifications for PINKSLIP Water-Based Personal Lubricant and Moisturizer

|  Property | Specification  |
| --- | --- |
|  Appearance | Liquid  |
|  Color | Clear, slightly yellow  |
|  Odor | Fragrance Free  |
|  Viscosity@25°C ; Spindle #5@10rpm (cps) | 5,000 – 17,800 cPs  |
|  pH @ 25° | 4.0 - 4.5  |
|  Osmolality (mOSm/kg) | 220 – 290 mOSm/kg  |
|  Total aerobic microbial count (TAMC) per USP <61> and <1111> | <10 cfu/g  |
|  Total yeast and mold count (TYMC) per USP <61> and <1111> | <10 cfu/g  |
|  Antimicrobial effectiveness per USP <51> | Meets USP <51> acceptance criteria for Category 2 products. Category 2 bacteria should show a log reduction of less than 2.0 at 14 days and no increase from the 14-day count to the 28-day count. Yeast and molds should show no increase from the initial calculated count at 14 and 28 days.  |
|  Presence of Pathogens per USP <62> | Specification  |
|  Absence of pathogenic organisms (*Staphylococcus aureus*, *Pseudomonas aeruginosa*, *Escherichia coli*, and *Candida albicans*) | Absent  |

### 7. Indication for Use

PINKSLIP Water-Based Personal Lubricant and Moisturizer is a water-based personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication. PINKSLIP Water-Based Personal Lubricant and Moisturizer is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

{6}

510(k) Summary

K261177

K261177

Page 3 of 5

PINKSLIP Water-Based Personal Lubricant and Moisturizer

### 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

The table below compares the intended use and technological characteristics of the subject and predicate device.

Table 2: Intended Use and Technological Characteristics Comparison

|  Characteristic/Feature | PINKSLIP Water-Based Personal Lubricant and Moisturizer K261177 | Desnuda Reflect K211998  |
| --- | --- | --- |
|  Device Name | PINKSLIP Water-Based Personal Lubricant and Moisturizer | Desnuda Reflect  |
|  Indications for Use | PINKSLIP Water-Based Personal Lubricant and Moisturizer is a water-based personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication. PINKSLIP Water-Based Personal Lubricant and Moisturizer is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. | Desnuda Reflect is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.  |
|  Rx/OTC | OTC | OTC  |
|  Sterile | No | No  |
|  Base type | Water-based | Water-based  |
|  Formulation | Water, sodium benzoate, potassium sorbate, organic aloe barbadensis leaf juice, sorbitol, xanthan gum, hyaluronic acid, and citric acid. | Water, hyaluronic acid, hydroxyethylcellulose, xanthan gum, zemea propanediol, sodium benzoate, potassium sorbate, dl lactic acid  |

{7}

510(k) Summary

K261177

K261177

Page 4 of 5

PINKSLIP Water-Based Personal Lubricant and Moisturizer

|  Appearance | Clear, water-based liquid | Gel  |
| --- | --- | --- |
|  Color | Slight Yellow | Clear to Cloudy  |
|  Odor | Odorless | Odorless  |
|  Viscosity | 5,000 – 17,800 cPs | 12,500 – 25,000 cPs  |
|  pH@25°C | 4.0 -4.5 | 3.5 – 4.0  |
|  Osmolality | 220-290 mOSm/kg | 250 – 400 mOsm/kg  |
|  Antimicrobial Effectiveness Tested per USP <62> | Absent | Absent  |
|  Total mold/yeast count (TYMC) per USP <61> | <10 cfu/g | <10 cfu/g  |
|  Total aerobic microbial count (TAMC) per USP <61> | <10 cfu/g | <100 cfu/g  |
|  Antimicrobial Effectiveness Testing (USP<51> Category 2) | Yes | Yes  |
|  Condom Compatibility | Compatible with Natural Rubber Latex and Polyisoprene Condoms. It is incompatible with lubricated polyurethane male condoms. | Compatible with Natural Rubber Latex and Polyisoprene Condoms. It is incompatible with lubricated polyurethane male condoms.  |
|  Shelf life | 8.5 month | 1 Year  |

The subject device and predicate device have the same indications for use and intended use. The subject device and predicated devices differ in physical features, total aerobic microbial count (TAMC), formulation, and shelf-life. These differences in technological characteristics do not raise different questions of safety and effectiveness.

{8}

510(k) Summary

K261177

K261177

Page 5 of 5

PINKSLIP Water-Based Personal Lubricant and Moisturizer

### 9. Summary of Non-Clinical Performance Testing

# **Biocompatibility**

Biocompatibility studies were performed in accordance with the 2020 FDA Guidance Document, Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices-Part 1: Evaluation and testing within a risk management process."

The following testing was conducted:

a. Cytotoxicity (ISO 10993-5:2009)
b. Sensitization (ISO 10993-10:2010)
c. Vaginal Irritation (ISO 10993-10:2010)
d. Acute Systemic Toxicity (ISO 10993-11:2017)

The test results demonstrated that the subject device is non-cytotoxic, non-sensitizing, non-irritating, and non-systemically toxic.

# **Condom Compatibility:**

The subject device was tested for compatibility with natural rubber latex, polyisoprene, and polyurethane condoms in accordance with the ASTM D7661-18 Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms. The results show that PINKSLIP Water-Based Personal Lubricant and Moisturizer is compatible with natural rubber latex and polyisoprene male condoms and is not compatible with lubricated polyurethane male condoms.

# **Shelf Life:**

The subject device has a shelf-life of 8.5 months based on the results of an accelerated aging study. The shelf-life study evaluated all device specifications listed in Table 1 and met all device specifications throughout the stated shelf-life.

### 10. Conclusions

The results of performance testing described above demonstrate that the PINKSLIP Water-Based Personal Lubricant and Moisturizer is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K261177](https://fda-staging.innolitics.com/device/K261177)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
