K261175 · Whill, Inc. · ITI · Jun 12, 2026 · Physical Medicine
Device Facts
Record ID
K261175
Device Name
WHILL (WHILL Model C2)
Applicant
Whill, Inc.
Product Code
ITI · Physical Medicine
Decision Date
Jun 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the WHILL Model C2 powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.
Device Story
WHILL Model C2 is a battery-operated, rear-wheel drive powered wheelchair; intended for indoor/outdoor mobility. Input: user commands via joystick on armrest-mounted control pad or smartphone app (Bluetooth Low Energy). Transformation: control system processes inputs to drive motors and electromagnetic braking system. Output: physical movement of wheelchair; status information (battery, speed) on app. Used by patients in various environments; operated by patient. Smartphone app allows remote driving, speed/acceleration adjustment, and device locking. Benefits: provides mobility to individuals with limited walking ability. Modifications from predicate include upgraded BLE security (Mode 1/Level 3), new lithium-ion battery options, new USB Type-C charger, and adjustable backrest mounting system.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 7176 series (stability, braking, energy consumption, dimensions, speed, fatigue, climatic, obstacle climbing), electrical safety (IEC 60335-2-29, EN 60601-1-2), battery safety (IEC 62133-2), biocompatibility (ISO 10993-5, -10, -23), and wireless coexistence (ANSI C63.27).
Technological Characteristics
Solid aluminum frame; 2-wheel rear-drive; electromagnetic braking. Powered by 25.3V/25.6V/25.83V DC rechargeable Li-ion batteries. Connectivity: BLE (IEEE 802.15.4) for smartphone app/key fob control. Standards: ISO 7176 series, IEC 62133-2, IEC 60335-2-29, EN 60601-1-2, ISO 10993. Software: IEC 62304 compliant.
Indications for Use
Indicated for persons limited to a seated position capable of operating a powered wheelchair for indoor and outdoor mobility.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 12, 2026
Whill, Inc.
Tsuyoshi Iriyama
2-1-11, Higashi-Shinagawa
Shinagawa, Tokyo 1400002
Japan
Re: K261175
Trade/Device Name: WHILL (WHILL Model C2)
Regulation Number: 21 CFR 890.3860
Regulation Name: Powered Wheelchair
Regulatory Class: Class II
Product Code: ITI
Dated: April 10, 2026
Received: April 10, 2026
Dear Tsuyoshi Iriyama:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20903
www.fda.gov
{1}
K261175 - Tsuyoshi Iriyama
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261175 - Tsuyoshi Iriyama
Page 3
Sincerely,
**Tushar Bansal -S**
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K2611175
?
Please provide the device trade name(s).
?
WHILL (WHILL Model C2)
Please provide your Indications for Use below.
?
The intended use of the WHILL Model C2 powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.
Please select the types of uses (select one or both, as applicable).
☐ Prescription Use (21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
?
{4}
| 510(k) #: | 510(k) Summary | Prepared on: |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | WHILL.Inc | |
| Applicant Address | 2-1-11 Higashi-Shinagawa, Shinagawa-Ku Tokyo, 140-0002, Japan | |
| Applicant Contact Telephone | +81-70-4414-098 | |
| Applicant Contact | Mr. TSUYOSHI IRIYAMA | |
| Applicant Contact Email | tsuyoshi.iriyama@whill.inc | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | WHILL (WHILL Model C2) | |
| Common Name | Powered wheelchair | |
| Classification Name | Wheelchair, Powered | |
| Regulation Number | 890.3860 | |
| Product Code(s) | ITI | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K213383 | WHILL Model C2 | ITI |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The subject device, the WHILL Model C2 series, incorporates the following modifications to the previously cleared WHILL Model C2 (K213383). Major Design Change (Risk Level Impacting) from K213383 1. BLE Security Level Upgrade (Encryption Algorithm) to Mode:1/Level:3. The Bluetooth Low Energy (BLE) security level has been upgraded, including implementation of an encryption algorithm, in accordance with IEC 81001-5-1. For additional details, please refer to the Software/Firmware and Cybersecurity/Interoperability section. Minor Design Changes (Addition of Accessories) from K213383 1. New Off-Board Lithium-Ion Battery charger Model: ADP-69CR A (Color change of status indication LED from RED to Yellow) 2. New Rechargeable Lithium-Ion Battery Lineup Standard Type: 23-15000-0 (272.4 Wh) H098-8S4P Note: Distance range: 16km High-Capacity Type: 23-15002-0 (496 Wh) M043-7S4P Note: Distance range: 27.9km | | |
{5}
3. New USB Type-C Off-Board Lithium-Ion Battery Charger
Model: A100d
4. Add accessories
24-10065-1 Model: Model C2 Backrest Mounting System
For further information, please refer to the EMC, Wireless, Electrical, Mechanical, and Thermal Safety sections, which include testing conducted with the Model C2 in combination with the above accessories. No additional risk regarding safety and effectiveness of the device.
Except for the design changes described above, the subject device is identical to the previously cleared WHILL Model C2 (K213383) in terms of intended use, fundamental scientific technology, and overall device design.
The WHILL Model C2 is an indoor/outdoor battery-operated 2-wheel drive (rear-wheel drive) powered wheelchair. It consists of four parts: seat system, control system, braking system, and drive system. It consists of two motors drive systems, an electromagnetic braking system, an electric motor controller, and a lithium-ion battery with an off-board battery charger. The wheelchair is powered by a 25.3V DC 10.6A rechargeable lithium-ion battery charged by an off-board lithium-ion battery charger.
The control system, including the directional controller (joystick), is equipped on the control pad that attaches to the armrest. When the joystick is released, the electromagnetic brakes will be actuated, and the powered wheelchair is slowed to a stop.
As with all commercially available powered wheelchairs, the user sits in the wheelchair seat and uses the control system such as the control pad positioned on either of the two arms to turn the chair on, control the speed, and direct the movement. Adjustments can be made to the seating to fit the user's body. Like the predicate device WHILL Model C2, the two side-arms can be rotated out of the way to make it easier for the user to get into and out of the device.
Model C2 also contains Bluetooth-based RF wireless technology. The device can be controlled by the directional controller or remote control by a smartphone app via Bluetooth Low Energy (BLE) wireless communication interface. The smartphone app is used to drive the chair remotely (For safety reasons, Joystick control takes priority over remote control by design). The smartphone app can also view the battery's status, adjust the speed and acceleration setting and lock the unattended device. The user can lock and unlock the device remotely via the BLE interface using the smartphone app or using a smart key fob.
The device supports a maximum weight of 136Kg (300lbs.), including the weight of the occupant and any carried items. It has a maximum driving range of 12miles (20km) with a maximum speed limit of up to 5mph (8km/h).
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The intended use of the WHILL Model C2 powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use are the same.
## Technological Comparison
21 CFR 807.92(a)(6)
{6}
## Substantial Equivalence Analysis
Predicate Device
| 510(K) Number | Manufacturer | Device Name | Product Code |
| --- | --- | --- | --- |
| K213383 | WHILL, Inc | WHILL Model C2 | ITI |
## Comparison with Predicate Device
Table 1 Comparison of subject device with predicate device (K213383)
| Element of Comparison | Subject Device WHILL Model C2 (K261175) | Predicate Device WHILL Model C2 (K213383) | Remark |
| --- | --- | --- | --- |
| Manufacturer | WHILL, Inc. | WHILL, Inc. | - |
| Common or Usual Name | Powered Wheelchair | Powered Wheelchair | Same |
| Product Code | ITI | ITI | Same |
| Product Classification | Class II | Class II | Same |
| Device Classification Name | Powered Wheelchair | Powered Wheelchair | Same |
| Regulation Number | 21 CFR 890.3860 | 21 CFR 890.3860 | Same |
| **Indications for Use** | | | |
| Indications for Use | The intended use of the Model C2 powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair. | The intended use of the Model C2 powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair. | Same |
| Type of Use | Over the counter (OTC) | Over the counter (OTC) | Same |
| **Physical Characteristics** | | | |
| Device weight including battery | 116.4 lbs (52.8 kg) | 116.4 lbs (52.8 kg) | Same |
| Device length | 38.8" | 38.8" | Same |
| Device width | 21.8" to 25.6" | 21.8" to 25.6" | Same |
| Device Height | 26.4" to 28.7" | 26.4" to 28.7" | Same |
| Device Construction | Solid aluminum frame | Steel frame | Same |
| Number of front wheels | 2 | 2 | Same |
| Number of rear wheels | 2 | 2 | Same |
| Diameter of front wheel(s) | 10.11" | 10.11" | Same |
{7}
| Diameter of rear wheels | 10.4" | 10.4" | Same |
| --- | --- | --- | --- |
| Ground clearance | 3" | 3" | Same |
| Battery pack | 1 rechargeable lithium-ion battery 1.23-00006-0, 7S4PMH1: Ratings: 25.3 V, 10.5Ah 2.23-15000-0, H098-8S4P Ratings: 25.6 V, 10.6Ah 3.23-15002-0, M043-7S4P Ratings: 25.83 V, 19.2Ah | 1 rechargeable lithium-ion battery 1.23-00006-0, 7S4PMH1: Ratings: 25.3 V, 10.5Ah | SE-Note 1 |
| Battery weight | 1. 23-00006-0, 7S4PMH1: 6.0 lbs. 2. 23-15000-0, H098-8S4P: 6.6 lbs 3. 23-15002-0, M043-7S4P: 6.9 lbs | 23-00006-0: 6.0 lbs. | SE-Note 1 |
| Charger type | Off-board | Off-board | SE-Note 1 |
| Charger ratings | 1. ADP-69BR/CR Series: 28.49 V_{DC} / 2.4 A 2. A100d: 5V/3.0A 9V/3.0A 12V/3.0A 15V/3.0A 20V/5.0A | 1. ADP-69BR Series: 28.49 V_{DC} / 2.4 A | SE-Note 2 |
| Charger Port Type | 1. ADP-69BR/CR Series: XLR Type Connector 2. A100d: USB Type-C | 1. ADP-69BR Series: XLR Type Connector | SE-Note 2 |
| Status Indication LED Color | 1. ADP-BR A 2. ADP-CR A 3. A100d Charging: 1.2.3: Green blinking Full charge: 1.2.3: Solid Green Power On/Standby: 1.: Solid RED 2.3: Solid Yellow Safety Protection: 1.: RED blinking 2.: Yellow blinking 3.: Solid RED | 1. ADP-BR A Charging: 1: Green blinking Full charge: 1: Solid Green Power On/Standby: 1: Solid RED Safety Protection: 1: RED blinking | SE-Note 2 |
| Battery charging time | 1. 23-00006-0, 7S4PMH1: ~5hours. 2. 23-15000-0, H098-8S4P: ~4.5hours 3. 23-15002-0, M043-7S4P: ~ 9hours | 1. 23-00006-0, 7S4PMH1: ~5hours. | SE-Note 1 |
| **Operating Characteristics** | | | |
| Operating environments | Indoor and outdoor use | Indoor and outdoor use | Same |
{8}
| Maximum weight capacity | 300lb(136kg) | 300lb(136kg) | Same |
| --- | --- | --- | --- |
| Maximum forward speed | 5 mph (8 km/h) | 5 mph (8 km/h) | Same |
| Braking system | Electromagnetic | Electromagnetic | Same |
| Turning radius | 30" | 30" | Same |
| Obstacle climbing height | 2" | 2" | Same |
| Drive system | 2 Wheel Drive (Rear Wheel Drive) | 2 Wheel Drive (Rear Wheel Drive) | Same |
| Dynamic stability on incline (Maximum allowable inclination) | 10° | 10° | Same |
| Driving range on full battery charge | 1. 23-00006-0, 7S4PMH1: 19.2km 2. 23-15000-0, H098-8S4P: 16km 3. 23-15002-0, M043-7S4P: 27.9km | 1. 23-00006-0, 7S4PMH1: 19.2km | Same |
| **Design Features** | | | |
| Motor rating | 150 W x 2 pcs | 150 W x 2 pcs | Same |
| Motor controller | Manufacturer: WHILL Model: 21-00011-0 | Manufacturer: WHILL Model: 21-00011-0 | Same |
| Portability | Disassemble for transport | Disassemble for transport | Same |
| Joystick Location | Left or right arm | Left or right arm | Same |
| Joystick | Users can select the WHILL mouse controller, WHILL easy-grip controller, or a Body point controller. | Users can select the WHILL mouse controller, WHILL easy-grip controller, or a Body point controller. | Same |
| User control interface | User controls are housed in a single component—the control pad subassembly. The control pad subassembly may be placed on either the right- or left-hand side of the device to match the user's preference. | User controls are housed in a single component—the control pad subassembly. The control pad subassembly may be placed on either the right- or left-hand side of the device to match the user's preference. | Same |
| Seat Width | 16", 18" and 20" | 16", 18" and 20" | Same |
| Back support Height | 13.4 – 18.1" | 13.4 – 18.1" | Same |
| Adjustable Backrest Mounting System | Prepared | None | SE-Note 3 |
| Disassembly | Users can disassemble model C2 without using tools into four components: Seat, Front Drive Base, Rear Drive Base, and Battery. | Users can disassemble model C2 without using tools into four components: Seat, Front Drive Base, Rear Drive Base, and Battery. | Same |
| **Wireless Communication Characteristics** | | | |
| Remote control | Bluetooth key fob: lock and unlock device | Bluetooth key fob: lock and unlock device | Same |
{9}
| | Mobile app: -Display general device information (e.g. driving distance, battery capacity) -Power on/off control -Speed profile setting -Firmware update/patch -Remote control | Mobile app: -Display general device information (e.g. driving distance, battery capacity) -Power on/off control -Speed profile setting -Firmware update/patch -Remote control | |
| --- | --- | --- | --- |
| Type of wireless technology | IEEE 802.15.4 (Bluetooth Low Energy) | IEEE 802.15.4 (Bluetooth Low Energy) | Same |
| FCC compliance | 47 CFR, Part 15 | 47 CFR, Part 15 | Same |
| EMC compliance | ISO 7176-21:2009 | ISO 7176-21:2009 | Same |
| Wireless coexistence compliance | ANSI C63.27:2017 (EN 12184) | ANSI C63.27:2017 (EN 12184) | Same |
| Wireless functions | adjust speed, acceleration, turning settings, and lock the device when it is unattended | adjust speed, acceleration, turning settings, and lock the device when it is unattended | Same |
| Mobile app | iOS and Android | iOS and Android | Same |
| Wireless RF frequency range | 2.402 GHz to 2.480 GHz | 2.402 GHz to 2.480 GHz | Same |
| Wireless RF maximum output power | 6 dBm | 6 dBm | Same |
| Wireless operating range | 10 m | 10 m | Same |
| BLE Security Mode Level | Mode: 1, Level: 3 | Mode: 0, Level:1 | SE-Note 4 |
Notes to analyze the differences in technological characteristics between the subject device and the predicate device are summarized below.
The subject device is identical technological characteristics as the predicate device in terms of principle of operation, portability design and construction and the drive and braking systems. The differences in technological characteristics may be summarized in 2 major categories
1. Ad-hoc Accessories of Li-ion Batteries, Charger, Adjustable backrest mounting system
2. BLE security level
[Note1] Ad-hoc accessories of Li-ion Battery
The subject device added two types of new lithium-ion battery. The lithium-ion battery passed the safety testing per IEC 62133-2:2017 Secondary Cells and Batteries containing Alkaline or other Non-Acid Electrolytes – Safety Requirements for Portable Sealed Secondary Cells, and Batteries made from them, for use in Portable Applications – Part 2: Lithium systems.
{10}
The subject device with ad-hoc accessories passed the safety and performance testing per ISO 7176 standards (Table 3). Lithium-ion battery is a proven technology that is widely used in medical devices and does not introduce new and intolerable safety risks. These test results demonstrate that the subject device has the same level of safety and effectiveness as the predicate device.
[Note2] Ad-hoc accessories of Charger
Although the chargers of the subject and predicate devices have different specifications, indication LED color, and connector type, the WHILL-proprietary lithium-ion battery charger manufactured was tested to rigorous safety standards—
○ IEC 60335-2-29:2016 Household and Similar Electrical Appliances - Safety - Part 2-29: Particular Requirements For Battery Chargers
○ EN 60601-1-2: 2015+A1:2021 Medical electrical equipment — Part 1-2: General requirements for basic safety and essential performance — Collateral Standard: Electromagnetic disturbances — Requirements and tests
[Note 3] Ad-hoc accessories of adjustable backrest mounting system
The subject device with ad-hoc accessories passed the safety and performance testing per ISO 7176 standards (Table 3).
[Note 4] BLE security level
-Predicate device: BLE security Level: Mode 0 / Level 1
-Subject device: BLE security Level: Mode 1 / Level 3
BLE Security level is upgraded from the predicate device. For this change, the subject device of security level is much improved than the predicate device. However, regarding the level of safety and effectiveness are substantially equivalent as the predicate device.
## Summary of Technical Characteristics of Device Compared to Predicate Device
The subject device is designed to fulfill the requirements of the following recognized consensus standards:
### Non-clinical performance testing
Table 2 Comparison of non-clinical performance testing standards
| Standard | Subject Device | Predicate Device | Result |
| --- | --- | --- | --- |
| ISO 7176-1:2014 Wheelchairs – Part 1: Determination of static stability | Yes | Yes | Same |
{11}
| ISO 7176-2:2017 Wheelchairs – Part 2: Determination of dynamic stability of electrically powered wheelchairs | Yes | Yes | Same |
| --- | --- | --- | --- |
| ISO 7176-3:2012 Wheelchairs – Part 3: Determination of effectiveness of brakes | Yes | Yes | Same |
| ISO 7176-4:2008 Wheelchairs – Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range | Yes | Yes | Same |
| ISO 7176-5:2008 Wheelchairs – Part 5: Determination of dimensions, mass and manoeuvring space | Yes | Yes | Same |
| ISO 7176-6:2018 Wheelchairs – Part 6: Determination of maximum speed of electrically powered wheelchairs | Yes | Yes | Same |
| ISO 7176-7:1998 Wheelchairs – Part 7: Measurement of seating and wheel dimensions | Yes | Yes | Same |
| ISO 7176-8:2014 Wheelchairs – Part 8: Requirements and test methods for static, impact and fatigue strengths | Yes | Yes | Same |
| ISO 7176-9:2009 Wheelchairs – Part 9: Climatic tests for electric wheelchairs | Yes | Yes | Same |
| ISO 7176-10:2008 Wheelchairs – Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs | Yes | Yes | Same |
| ISO 7176-11:2012 Wheelchairs – Part 11: Test dummies | Yes | Yes | Same |
| ISO 7176-13:1989 Wheelchairs – Part 13: Determination of coefficient of friction of test surfaces | Yes | Yes | Same |
| ISO 7176-14: 2022 Wheelchairs – Part 14: Power and control systems for electrically powered wheelchairs and scooters — Requirements and test methods | Yes | Yes | Same |
| ISO 7176-15:1996 Wheelchairs – Part 15: Requirements for information disclosure, documentation and labelling | Yes, alternative state-of-the-art standards. See SE-Note 12. | Yes | Same |
| ISO 7176-16:2012 Wheelchairs – Part 16: Resistance to ignition of postural support devices | Yes | Yes | Same SE-Note 13. |
{12}
Notes to analyze the differences in non-clinical performance testing among the subject device, the predicate device are summarized below.
- Note 12: In place of the wheelchair and scooter-specific standard ISO 7176-15:1996, the labeling of the subject device conforms to
- ISO 20417:2021 Medical devices — Information to be supplied by the manufacturer
- ISO 15223-1:2021 Medical devices — Symbols to be used with information to be supplied by the manufacturer – Part 1: General requirements, and
- EN 12184:2022 Electrically powered wheelchairs, scooters and their chargers – Requirements and test methods, Section 13 Information Supplied by the manufacturer
in alignment with the state-of-the-art in wheelchair and scooter development. Conformance to these standards ensures that the labeling of the subject device provides both the generic information concerning medical devices and the specific information relevant to powered scooters to users in an effective manner. The labeling of the subject device is considered substantially equivalent to that of the predicate device and does not raise concerns about safety or effectiveness.
- Note 13: WHILL-manufactured specialty cushion that tested to ISO 8191-1/8191-2 that is equivalent to ISO 7176-16.
## Electrical Safety and Electromagnetic Compatibility Testing
Table 3 Comparison of electrical safety and electromagnetic compatibility testing standards
| Standard | Subject Device | Predicate Device | Result |
| --- | --- | --- | --- |
| ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers | Yes | Yes | Same |
| ISO 7176-25:2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs | N/A | N/A | Same |
| IEC 62133-2:2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems | Yes | Yes | Same |
{13}
| IEC 60335-2-29:2016+A1:2019 Safety of household and similar electrical appliances Part 2-29: Particular requirements for battery chargers | Yes | Yes | Same |
| --- | --- | --- | --- |
| EN 60601-1-2:2015+A1:2021 Medical electrical equipment — Part 1-2: General requirements for basic safety and essential performance — Collateral Standard: Electromagnetic disturbances — Requirements and tests | Yes | Yes | Same |
## Biocompatibility Testing
The biocompatibility testing for the subject device was conducted according to EN ISO 10993-1:2020 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
Table 4 Comparison of biocompatibility testing standards
| Standard | Subject Device | Predicate Device | Result |
| --- | --- | --- | --- |
| ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity | Yes | Yes | Same |
| ISO 10993-10:2021 Biological evaluation of medical devices – Part 10: Tests for skin sensitization | Yes | Yes | Same |
| ISO 10993-23:2021 Biological evaluation of medical devices – Part 23: Tests for irritation | Yes | Yes | Same |
## Software Verification Testing
Software verification testing was performed to evaluate the functionality of the subject device's design and operational principles. Software verification testing was conducted, and the documentation level was determined on the subject device as recommended by the FDA's guidance document "FDA Guidance: Content of Premarket Submissions for Device Software Functions." Software verification testing conforms to IEC 62304:2006/Amd1:2015 Medical Device Software – Software Life Cycle.
## FCC Radio Frequency Testing
The radiofrequency wireless function of the subject device was tested to FCC requirements and found to comply with 47 CFR 15.249.
{14}
# Wireless Coexistence Testing
The performance of WHILL Model C2 was evaluated in an environment with other WHILL Model C2 devices and with different types of 2.4 GHz wireless devices. The device met all specified requirements listed in ANSI C63.27 2017 American National Standard for Evaluation of Wireless Coexistence.
# Conclusion on Substantial Equivalence
The WHILL Model C2 (subject device) described herein has an equivalent intended use and the same fundamental technology as the cleared primary predicate device, WHILL Model C2 (K213383). Based on the performance data presented for the design differences between the subject device and the predicate device, it can be concluded that the WHILL Model C2 is as safe and effective as, and substantially equivalent to, the predicate device. Hence the subject and the predicate device are considered substantially equivalent concerning the aspects of wireless technology.
The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed devices identified in this submission. Thus, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.