GuangDong Yong Yi Rehabilitation Equipment Technology Co., Ltd.
Product Code
IOR · Physical Medicine
Decision Date
Jun 22, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
Caregiver-assisted wheelchair (model:MS02,MA02) Caregiver-assisted wheelchair is intended for medical purpose to provide mobility to persons limited to a sitting position.
Device Story
Manual wheelchair (models MS02, MA02) provides mobility to persons limited to sitting position; operated entirely by trained caregiver. Features steel or aluminum frame, 12-inch rear wheels, 8-inch front swiveling castors, flip-back armrests, and adjustable footrests. Includes fireproof nylon/foam backrest and seat cushion for postural support. Caregiver controls steering and movement; manual brakes immobilize rear wheels for transfers. Used in clinical or home settings. Benefits include improved patient mobility and postural stability. Device is foldable for storage.
Clinical Evidence
No clinical data. Bench testing only. Device complies with ISO 7176 series (stability, brakes, dimensions, fatigue strength) and ISO 16840-10 for ignition resistance. Biocompatibility testing performed per ISO 10993-5, -10, and -23.
Technological Characteristics
Manual, foldable wheelchair. Frame: steel or aluminum. Upholstery: fireproof nylon fabric and foam. Dimensions: 916mm (L) x 574mm (W) x 896mm (H). Features: 12" rear wheels, 8" front castors, manual brakes, flip-back armrests, swing-away footrests. Standards: ISO 7176-1, -3, -5, -7, -8, -11, -13, -15; ISO 16840-10; ISO 10993-5, -10, -23.
Indications for Use
Indicated for persons limited to a sitting position requiring mobility assistance. Contraindicated for users exceeding 100 kg, use as a motor vehicle seat, use on gradients >10°, reverse travel on gradients, use on escalators/stairs, or multi-person use.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 22, 2026
GuangDong Yong Yi Rehabilitation Equipment Technology Co., Ltd.
Sally Peng
Primary Correspondent
#39, Baozhu E. Rd., Huangpu Town
Zhongshan, Guangdong 528400
China
Re: K261076
Trade/Device Name: Caregiver-assisted wheelchair (MS02); Caregiver-assisted wheelchair (MA02)
Regulation Number: 21 CFR 890.3850
Regulation Name: Mechanical Wheelchair
Regulatory Class: Class I, reserved
Product Code: IOR
Dated: March 31, 2026
Received: April 1, 2026
Dear Sally Peng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261076 - Sally Peng
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261076 - Sally Peng
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

for Tushar Bansal, PhD
Acting Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261076 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Caregiver-assisted wheelchair (MS02); Caregiver-assisted wheelchair (MA02) | | |
| Please provide your Indications for Use below. | | ? |
| Caregiver-assisted wheelchair (model:MS02,MA02) Caregiver-assisted wheelchair is intended for medical purpose to provide mobility to persons limited to a sitting position. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Traditional 510(k) Submission of Caregiver-assisted wheelchair
# 510(K) Summary - K261076
Prepared in accordance with the requirements of 21 CFR Part 807.92
**Prepared Date:** March 20,2026
## 1. Submitter's Information
The submitter of this pre-market notification is:
| Name: | GuangDong Yong Yi Rehabilitation Equipment Technology Co., Ltd. |
| --- | --- |
| Address: | No.39, Baozhu East Road, Huangpu Town, Zhongshan City, Guangdong, China. |
| Contact person: | Sally Peng |
| Title: | Manager |
| E-mail: | sally@yongyicn.com |
| Tel: | 0086-760-23235999 |
## 2. Device Identification
| 510(K) number: | |
| --- | --- |
| Trade/Device Name: | Caregiver-assisted wheelchair |
| Models: | MS02,MA02 |
| Common name: | Manual Wheelchair |
| Regulation Number: | 21CFR 890.3850 |
| Regulation Name: | Mechanical wheelchair |
| Regulation Class: | Class 1 |
| Panel: | Physical Medicine |
| Product Code: | IOR |
## 3. Predicate Device
| 510(K) number: | K231750 |
| --- | --- |
| Trade/Device Name: | MA012 Aluminum wheelchair, MS019 Steel wheelchair |
| Models: | MA012,MS019 |
| Common name: | Manual Wheelchair |
| Regulation Number: | 21CFR 890.3850 |
| Regulation Name: | Mechanical wheelchair |
| Regulation Class: | Class 1 |
| Panel: | Physical Medicine |
| Product Code: | IOR |
510(k) Summary - K261076
Page 1 of Page 5
{5}
Traditional 510(k) Submission of Caregiver-assisted wheelchair
#### **4. Indication for Use**
Caregiver-assisted wheelchair (model:MS02,MA02) Caregiver-assisted wheelchair is intended for medical purpose to provide mobility to persons limited to a sitting position.
#### **5. Device Description**
Caregiver-assisted wheelchair (model:MS02,MA02) are intended for medical purpose to provide mobility to persons limited to a sitting position. The Caregiver-assisted wheelchair (model:MS02,MA02) is controlled, steered and operated completely by a trained caregiver.
This caregiver-assisted wheelchair (model: MS02,MA02) is contraindicated for overloading beyond the specified maximum occupant mass of 100 kg; the wheelchair is strictly prohibited from being used as a seat in a motor vehicle. It is forbidden to use the wheelchair on gradients exceeding 10°, reverse down a gradient, or on escalators, and the wheelchair is only suitable for single occupancy with no multi-person use allowed. Standing on the footrests and overloading the footrests are prohibited; leaning or pushing down on the push handles is not permitted to avoid structural damage to the wheelchair. The wheelchair is not designed for ascending or descending staircases, and if no elevators or ramps are available, manual carrying is required with the strict prohibition of carrying the wheelchair by its push handles or any removable parts. The storage bag on the back upholstery shall not hold items exceeding 5kg to prevent the wheelchair from tipping backwards. During use and assembly, fingers and foreign objects are forbidden to be caught in wheel spokes and other moving parts; users shall not move their center of gravity out of the seating area or reach for items beyond arm's length. The wheelchair is prohibited from being used when damaged or severely worn, and long-term storage near high-temperature sources, in direct sunlight, in non-ventilated environments with corrosive harmful gases, as well as stacking and pressing the wheelchair during storage are all forbidden.
The wheelchair is designed with two 12' rear wheels, and can be moved forward and in reverse.
Wheelchair frames are made of either steel or aluminum. Frame serves as the primary structural chassis of the device, providing the foundational support, mechanical integrity, and overall dimensional form to the wheelchair assembly. It dictates the essential geometry and foldability of the device.
The back upholstery consists of a backrest made of fireproof nylon fabric and foam. Its function is to provide postural support to the user's back and pelvis, distribute pressure.
The seat upholstery consists of a cushion made of fireproof nylon fabric and foam. It is to provide seating interface for the user, contributing to pressure distribution and postural stability.
Armrest is a horizontally oriented support pad mounted laterally to the frame. It functions to support the user's forearm and elbow, aid in upper body stability and posture.
Footrests and foot plate are adjustable, cantilevered support structure extending from the lower front frame. The function is to provide a stable, ergonomic platform.
510(k) Summary - K261076
Page 2 of Page 5
{6}
Traditional 510(k) Submission of Caregiver-assisted wheelchair
Brake is to securely immobilize the wheelchair by preventing rotation of the rear wheel, thereby stabilizing the device during stationary periods such as user transfers from wheelchair.
Castors are 8 inches in size, swiveling wheel assembly mounted at the front of the frame. The functions are to provide directional stability during straight-line travel, facilitate tight-radius turns by pivoting.
The wheelchair can support users of up to 100 kg.
## 6. Compared to Predicate Device
Compared to the predicate devices, the subject device has the same intended use, similar product design, similar performance, same safety as the predicate device, the summarized comparison information is listed in the following table
| SE Comparisons | Subject Device Manual Wheelchair (Model: MS02,MA02) | Predicate Device MA012 Aluminum wheelchair, MS019 Steel wheelchair (Model: MA012,MS019:) | Similarities/ Differences |
| --- | --- | --- | --- |
| 510(K) number | / | K231750 | / |
| Indication for Use | MS02/MA02 Caregiver-assisted wheelchair is intended for medical purpose to provide mobility to persons limited to a sitting position. | MA012 Aluminum wheelchair and MS019 steel wheelchair are intended for medical purpose to provide mobility to persons limited to a sitting position. | Same |
| Product code | IOR | IOR | Same |
| Class | I | I | Same |
| Regulation Number | 21 CFR 890. 3850 | 21 CFR 890.3850 | Same |
| Common name | Mechanical Wheelchair | Mechanical Wheelchair | Same |
| Type of Use | Over the Counter (OTC Only) | Over the Counter (OTC Only) | Same |
| Device Length | 916mm | MA012: 950mm MS019: 916mm | Different See Note 1 |
| Device Width | 574mm | MA012:610mm MS019:574mm | Different See Note 1 |
| Device Height | 896mm | MA012:970mm MS019:896mm | Different See Note 1 |
| Total weight | MS02: 18.9kg/42 lbs MA02: 14kg/ 31lbs | MA012: 11.5kg/25 lbs MS019: 18.7kg/ 41 lbs | Different See Note 1 |
| Weight Capacity | 100kg/220lbs | 115 kg/250lbs | Different See Note 1 |
| Seat depth | 420 mm | MA012: 390mm MS019: 420mm | Different See Note 2 |
| Seat width | 450 mm | MA012: 420mm MS019: 450mm | Different See Note 2 |
510(k) Summary
Page 3 of Page 5
{7}
Traditional 510(k) Submission of Caregiver-assisted wheelchair
| Seat surface height at front edge | 596mm | MA012: 520mm MS019: 596mm | Different See Note 2 |
| --- | --- | --- | --- |
| Frame type | Foldable | Foldable | Same |
| Frame material | Model MA02: Aluminum Model MS02: Steel | Model MA012: Aluminum Model MS019: Steel | Same |
| Armrest | Flip back armrest | Flip back armrest | Same |
| Footrest | Optional/ swing away | Optional/ swing away | Same |
| Tires | Front: 200mm (8') Rear: 300mm (12') | Front: 200mm (8') Rear: 315mm (12.5') | Different See Note 2 |
| Non clinical testing | | | |
| Performance | wheelchair conforms to the ISO 7176 standards | wheelchair conforms to the ISO 7176 standards | Same |
| Flammability Testing | wheelchair conforms to the ISO 16840-10 standards | wheelchair conforms to the ISO 7176-16 standards | Different See Note 3 |
| Biocompatibility | All user directly contacting materials are compliance with ISO10993-5:2009, ISO10993-10:2021, ISO 10993-23:2021 requirements. | All user directly contacting materials are compliance with ISO10993-5, ISO10993-10, ISO 10993-23 requirements. | Different See Note 4 |
#### Substantial equivalence Analysis:
Note 1: The predicate device and subject device have different dimension. Both of them comply with ISO 7176-5:2008 Wheelchairs – Part 5: Determination of dimensions, mass, and maneuverings space, these differences do not affect safety and effectiveness.
Note 2: Both of the subject device and the predicate device comply with ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions, these differences do not affect safety and effectiveness.
Note 3: The FDA no longer recognizes ISO 7176-16; its recognition has been superseded by ISO 16840-10, and the relevant transition period has long since ended. Both the subject device and the predicate device passed the FDA-approved ignition resistance test at that time.
Note 4: Both of the subject device and the predicate device evaluated biocompatibility according to ISO 10993 series standards. Although standard version updated, test methods for the subject device and the predicate device are same.
The subject device and predicate device are substantially equivalent in the areas of technological characteristics such as basic design, features, method of operation, general function and intended use. The subject device device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices.
510(k) Summary - K261076
Page 4 of Page 5
{8}
Traditional 510(k) Submission of Caregiver-assisted wheelchair
## 7. Performance Testing Summary
The subject device Caregiver-assisted wheelchair(Model:MS02,MA02) comply with:
### Clinical test:
Clinical testing is not required.
### Non-clinical data
#### *Safety and performance*
#### *Safety and performance*
ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
ISO 7176-5:2008 Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space
ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
ISO 7176-11:2012 Wheelchairs - Part 11: Test dummies
ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces
ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
ISO 16840-10:2021+A1:2024 Wheelchair seating - Part 10: Resistance
to ignition of postural support devices - Requirements and test method, heat source: heating cartridge.
#### *Biocompatibility*
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ISO 10993-10:2021 Biological Evaluation of Medical Devices - Part 10: Tests for Skin Sensitization
ISO 10993-23:2021 Biological Evaluation of Medical Devices - Part 23: Tests for Irritation
## 8. Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device Caregiver-assisted wheelchair(Model:MS02,MA02) is as safe, as effective, and performs as well the legally marketed predicate device K231750.
510(k) Summary - K261076
Page 5 of Page 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.