The LigaSure™ Robotically Assisted Surgery (RAS) Maryland Jaw Vessel Sealer/Divider is a bipolar electrosurgical instrument intended for use with the Hugo™ RAS System where ligation, and division of vessels, thick tissue (tissue bundles), and lymphatics is desired. The LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use with the Hugo™ RAS System during the representative surgical procedures set forth in the Hugo™ RAS System Indications Document and the Hugo™ RAS System User's Guide Instrument Addendum. The LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider for these procedures.
Device Story
Sterile, single-use, bipolar electrosurgical instrument; designed for use with Hugo™ RAS System and Valleylab™ FT10 Energy Platform. Surgeon controls instrument movements via Hugo™ RAS console; device performs blunt dissection, vessel sealing, and tissue division. Distal jaw/shaft design equivalent to predicate; proximal end modified for robotic interface. Benefits include minimally invasive vessel ligation and division in robotic-assisted procedures. Safety and performance verified through bench testing, animal studies (porcine models), and human factors/usability engineering (cadaver models).
Clinical Evidence
No clinical performance data relied upon. Evidence includes bench testing (mechanical/functional performance, ex vivo tissue effects), animal studies (porcine models for seal quality), and human factors/usability engineering studies (cadaver models).
Technological Characteristics
Bipolar RF electrosurgical instrument. Materials: Stainless steel (seal plate, insert, blade, outer tube), PPA (insulation), Hexamethyldisiloxane (coating), Encapsulant paste (gap spacer). Sterilization: EO. Connectivity: Robotic interface for Hugo™ RAS System. Standards: ISO 10993-1, ISO 11135, ISO 11737-1, ISO 11607, IEC 60601-1, IEC 60601-2-2.
Indications for Use
Indicated for ligation and division of vessels (arteries/veins up to 7mm), thick tissue bundles, and lymphatics during surgical procedures using the Hugo™ RAS System. Contraindicated for tubal sterilization or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 15, 2026
Covidien, LLC
Marissa Miller
Senior Regulatory Affairs Specialist
Contact Address
Re: K261070
Trade/Device Name: LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider, Nano-coated
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI, SCV
Dated: August 6, 2026
Received: August 7, 2026
Dear Marissa Miller:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K261070 - Marissa Miller
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261070 - Marissa Miller
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARK
Digitally signed by MARK
TRUMBORE -S
TRUMBORE -S
Date: 2026.09.15 11:36:56
-04'00'
Mark Trumbore, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K261070
Device Name
LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider, Nano-coated
Indications for Use (Describe)
The LigaSure™ Robotically Assisted Surgery (RAS) Maryland Jaw Vessel Sealer/Divider is a bipolar electrosurgical instrument intended for use with the Hugo™ RAS System where ligation, and division of vessels, thick tissue (tissue bundles), and lymphatics is desired. The LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use with the Hugo™ RAS System during the representative surgical procedures set forth in the Hugo™ RAS System Indications Document and the Hugo™ RAS System User's Guide Instrument Addendum.
The LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider for these procedures.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# K261070
510(k) Summary (21 CFR 807.92)
Date: September 14, 2026
510(k) Submitter
Covidien llc
200 Medtronic Dr.
Lafayette, CO 80026
Contact
Marissa Miller, Sr Regulatory Affairs Specialist
Telephone: 601-790-9820
Email: Marissa.miller@medtronic.com
Subject Device
Trade Name: LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider Sealing, Nano-coated
Common Name: Bipolar Vessel Sealing Device
Classification Name: Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400)
Regulatory Class: II
Product Code: GEI
Associated Product Code: SCV
Predicate Device
Trade Name: LigaSure™ Maryland Jaw Sealer/Divider One-step Sealing, Nano-coated
Common Name: Bipolar Vessel Sealing Device
Classification Name: Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400)
Regulatory Class: II
Product Code: GEI
510(k): K170869 (cleared: April 21, 2017)
Manufacturer: Covidien llc
Reference Device
Trade Name: Hugo™ RAS Bipolar Maryland Forceps (cleared under Hugo Robotically Assisted Surgery (RAS) System)
Common Name: System, surgical, computer-controlled instrument
Classification Name: Modular Electromechanical Surgical System (21 CFR 878.4964)
Regulatory Class: II
Product Code: SCV
510(k): K250725 (cleared: December 03, 2025)
Manufacturer: Covidien llc
Page 1
{5}
## Device Description
The LigaSure™ Robotically Assisted Surgery (RAS) Maryland Jaw Vessel Sealer/Divider device is a sterile, single-use, minimally invasive, radiofrequency (RF) advanced bipolar vessel sealer/divider, designed for use with the Hugo™ RAS System and powered by an electrical surgical unit (ESU), the Valleylab™ FT10 Energy Platform.
The LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider, henceforth referred to as the LigaSure™ RAS Maryland device, is designed to perform blunt dissection, vessel sealing (ligation), and division (cutting) of vessels, thick tissue bundles, and lymphatic tissue. All instrument movements are controlled by the surgeon via the surgeon console on the Hugo™ RAS System.
## Indications for Use
The LigaSure™ Robotically Assisted Surgery (RAS) Maryland Jaw Vessel Sealer/Divider is a bipolar electrosurgical instrument intended for use with the Hugo™ RAS System where ligation, and division of vessels, thick tissue (tissue bundles), and lymphatics is desired. The LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use with the Hugo™ RAS System during the representative surgical procedures set forth in the Hugo™ RAS System Indications Document and the Hugo™ RAS System User's Guide Instrument Addendum.
The LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider for these procedures.
## Comparison of Technological Characteristics with the Predicate
The LigaSure™ RAS Maryland device is equivalent to the predicate device (LigaSure™ Maryland Jaw Sealer/Divider One-step Sealing, henceforth referred to as LigaSure™ Maryland device) in terms of in terms of intended use, fundamental performance and distal-end design (jaw and shaft), electrosurgical unit (ESU) connector, materials, energy type, sterilization, packaging, shelf life, and risk profile.. While the distal portion of the subject device is equivalent to the predicate device, the proximal end has been modified to enable compatibility with the Hugo™ robotic-assisted surgery (RAS) System. A reference device is necessary and appropriate to support verification and validation methods of the proximal end of the subject device. The reference device is utilized exclusively to demonstrate the functionality of the proximal interface with the robotic system. The intended use and performance characteristics remain equivalent to those of the predicate device, and the proximal modifications for Hugo™ RAS System compatibility do not adversely affect the subject device's safety or effectiveness.
In conclusion, the differences between the predicate and subject device are not critical to the intended surgical use of the device, and do not affect the safety and effectiveness of the device when used as labeled.
Page 2
{6}
Table 1: Substantial Equivalence Comparison Table
| Characteristic | Predicate Device LigaSure™ Maryland Jaw Sealer/Divider One-step Sealing (K170869) | Subject Device LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider | Comparison |
| --- | --- | --- | --- |
| Intended Use | The intended purpose of the devices is to ligate and divide vessels, tissue bundles, and lymphatics. | The intended purpose of the devices is to ligate and divide vessels, tissue bundles, and lymphatics. | Same |
| Indications for Use | The LigaSure™ Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure™ Sealer/ Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and such surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc. The LigaSure™ system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure™ system for these procedures. | The LigaSure™ Robotically Assisted Surgery (RAS) Maryland Jaw Vessel Sealer/Divider is a bipolar electrosurgical instrument intended for use with the Hugo™ RAS System where ligation, and division of vessels, thick tissue (tissue bundles), and lymphatics is desired. The LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use with the Hugo™ RAS System during the representative surgical procedures set forth in the Hugo™ RAS System Indications Document and the Hugo™ RAS System User's Guide Instrument Addendum. The LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider for these procedures. | Different Both devices have the same intended use however the indications for LigaSure™ RAS Maryland are limited to the indications for the Hugo™ RAS System. The reduced scope of procedure indications compared to the predicate does not affect the safety and effectiveness of the subject device. |
| Contraindications | None | None | Same |
| Maximum Vessel Diameter | 7 millimeters (mm) | 7 mm | Same |
| Single Use | Yes | Yes | Same |
| Sterile | Yes | Yes | Same |
| Sterilization Method | EO | EO | Same |
| Electrical Characteristics | | | |
| Energy Type | Electrical (RF) bipolar energy | Electrical (RF) bipolar energy | Same |
Page 3
{7}
| Characteristic | Predicate Device LigaSure™ Maryland Jaw Sealer/Divider One-step Sealing (K170869) | Subject Device LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider | Comparison |
| --- | --- | --- | --- |
| **Mechanical Characteristics** | | | |
| **Proximal end Design** | Handheld Pistol Grip | RAS Compatible | **Different** The subject device is designed to be used with the Hugo RAS system. This difference is covered by the reference device. |
| **Jaw Characteristics** | | | |
| **Functional Materials** | | | |
| **Jaw Seal Plate** | Stainless Steel | Stainless Steel | **Same** |
| **Jaw Insert** | Stainless Steel | Stainless Steel | **Same** |
| **Jaw Gap Spacer** | Encapsulant Paste | Encapsulant Paste | **Same** |
| **Jaw Insulation Material** | Polypththalamide (PPA) | Polypththalamide (PPA) | **Same** |
| **Blade Material** | Stainless Steel | Stainless Steel | **Same** |
| **Outer Tube Material** | Stainless Steel | Stainless Steel | **Same** |
| **Jaw Coating Material** | Hexamethyldisiloxane | Hexamethyldisiloxane | **Same** |
| **Patient Contacting Materials Biocompatible per ISO 10993-1** | Yes | Yes | **Same** |
## Verification and Validation Data
Sterilization and shelf-life testing of the subject device confirmed that residual levels remained below allowable limits while maintaining the required sterilization assurance level, in accordance with applicable validation standards ISO 11135, ISO 11737-1 and ISO 10993-7. Additionally, the packaging and product integrity of the subject device were found to be acceptable for a five-year shelf life, in accordance with the applicable standards ISO 11607-1, ISO 11607-2, ASTM D4169 and ASTM F1980.
Page 4
{8}
Biocompatibility requirements of ISO 10993-1 were met by the subject device. The following endpoints were evaluated to ensure the device is biocompatible:
- Cytotoxicity
- Sensitization
- Irritation
- Acute Systemic Toxicity
- Material Mediated Pyrogenicity
- Physical/Chemical Information
- Genotoxicity
- Systemic toxicity
- Carcinogenicity
- Reproductive/Developmental Toxicity
Electromagnetic compatibility (EMC), electrical and mechanical safety, as well as thermal safety testing was conducted. This testing showed the subject device, and compatible system, demonstrates conformance to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-2-18, IEC TS 60601-4-2, and IEC 80601-2-77.
Performance data included bench, animal and human factors/usability engineering studies. These studies demonstrated that the subject device performs as intended and is substantially equivalent to the predicate device, with modifications required for robotic-assisted surgery substantiated by the reference device.
- Performance bench testing met clinically relevant, predetermined device acceptance criteria (requirements) for mechanical and functional performance, visual inspection, device reliability and ex vivo tissue effects and device performance.
- Performance animal studies were conducted to evaluate the short- and long-term seal quality and overall performance of the subject device using clinically relevant tissue and vessels. The studies utilized porcine models because of their anatomical and physiological similarities to humans.
- Human Factors and Usability Engineering studies were performed in accordance with FDA guidance Applying Human Factors and Usability Engineering to Optimize Medical Device Design (Feb. 3, 2016). These studies utilized cadaver models, focused on identifying the subject device's critical tasks and ensuring the safety and effectiveness of the system with the introduction of the subject device. Documentation is provided based on FDA draft guidance Content of Human Factors Information in Medical Device Marketing Submission (December 9, 2022).
## Clinical Studies
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
Page 5
{9}
# Conclusions
The intended use, technological characteristics (supported by the reference device for proximal end design characteristics) and non-clinical bench performance data demonstrate that the LigaSure™ RAS Maryland Jaw Sealer/Divider, Nano-coated is substantially equivalent to the predicate LigaSure™ Maryland Jaw Sealer/Divider One-step Sealing, Nano-coated cleared in K170869. The verification and validation studies support the safety of the device and demonstrate that the subject LigaSure™ RAS Maryland device performs as intended when used with the compatible robotic assisted surgery system.
Page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.