K261054 · Stryker Leibinger GmbH & Co KG · HRS · Jul 1, 2026 · Orthopedic
Device Facts
Record ID
K261054
Device Name
RibPlate System
Applicant
Stryker Leibinger GmbH & Co KG
Product Code
HRS · Orthopedic
Decision Date
Jul 1, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RibPlate System is intended for use in the stabilization and fixation of fractures and reconstructions in the chest wall, including sternal reconstructive surgical procedures to promote bony fusion. The RibPlate system can be used in normal and poor bone.
Device Story
RibPlate System consists of titanium plates and screws for chest wall fracture stabilization and reconstruction; operates via mechanical fixation where screws lock into plates to provide structural support; used in surgical settings by surgeons; output is physical stabilization of bone segments; benefits include promotion of bony fusion in trauma or reconstructive cases.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including static and dynamic 4-point bending, pull-out, insertion/removal, shear-off, and locking torque tests. MR compatibility evaluated via computational modeling and comparison to reference devices per ASTM F2182, F2052, F2213, and F2119.
Technological Characteristics
Commercially pure titanium or titanium alloy plates and screws. Mechanical locking interface. Non-sterile, intended for end-user steam sterilization (ISO 17665, ANSI/AAMI ST79). MR Conditional. Evaluated per ISO 10993-1 for biocompatibility.
Indications for Use
Indicated for stabilization and fixation of chest wall fractures and reconstructions, including sternal reconstructive procedures to promote bony fusion, in adult patients with normal or poor bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Biomet Microfixation RibFix Blu Thoracic Fixation System (K212608)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 1, 2026
Stryker Leibinger GmbH & Co KG
Amelia Kesti
Senior Staff Regulatory Affairs Specialist
Boetzinger Strasse 41
Freiburg, D-79111
Germany
Re: K261054
Trade/Device Name: RibPlate System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: HRS, HWC
Dated: June 3, 2026
Received: June 4, 2026
Dear Amelia Kesti:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261054 - Amelia Kesti
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261054 - Amelia Kesti
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261054 | ? |
| Please provide the device trade name(s). | | ? |
| RibPlate System | | |
| Please provide your Indications for Use below. | | ? |
| The RibPlate System is intended for use in the stabilization and fixation of fractures and reconstructions in the chest wall, including sternal reconstructive surgical procedures to promote bony fusion. The RibPlate system can be used in normal and poor bone. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
Page 1 of 1
{4}
# K261054
# 510(k) Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
# I. SUBMITTER [§807.92(a)(1)]
510(k) Owner: Stryker Leibinger GmbH & Co. KG Boetzinger Strasse 41 D-79111 Freiburg, Germany
Submitter/Contact Person: Amelia Kesti Senior Staff Regulatory Affairs Specialist Stryker Craniomaxillofacial (CMF) 1941 Stryker Way Portage, MI 49002
Phone: 269-330-5919
Date prepared: 07/01/2026
# II. DEVICE [§807.92(a)(2)]
| Trade Name: | RibPlate System |
| --- | --- |
| Abbreviated Name: | RibPlate System |
| Common or Usual Name: | Plate, Fixation, Bone Screw, Fixation, Bone |
| Device: | RibPlate System |
| Classification Name & Regulation Description: | Plate, Fixation, Bone; per 21 CFR §888.3030 Single/multiple component metallic bone fixation appliances and accessories Screw, Fixation, Bone; per 21 CFR §888.3040 Smooth or threaded metallic bone fixation fastener |
| Regulation Medical Specialty & Review Panel: | Office of Orthopedic Devices, Office of Product Evaluation and Quality, Division of Stereotaxic, Trauma, and Restorative Devices, DHT6C |
| Product Code: | HRS (Primary), HWC |
| Regulatory Device Class: | Class II |
| *Note the company Stryker or legacy name Stryker Leibinger precedes the product/trade name and predicate device in some documentation. | |
K261054 - Page 1 of 5
{5}
### III. PREDICATE DEVICE [§807.92(a)(3)]
A. Predicate Devices: The predicate devices for this Traditional 510(k) are:
1. Primary Predicate: Biomet Microfixation RibFix Blu Thoracic Fixation System – K212608
i. Submission Branch of Predicate Device- Division of Orthopedic Devices, Office of Device Evaluation, Center for Devices and Radiological Health
2. Additional Predicate: KLS Martin Thoracic Plating System – K153482
i. Submission Branch of Predicate Device- Division of Orthopedic Devices, Office of Device Evaluation, Center for Devices and Radiological Health
3. Additional Predicate: DePuy Synthes MatrixRIB Fixation System – K161590
i. Submission Branch of Predicate Device- Division of Orthopedic Devices, Office of Device Evaluation, Center for Devices and Radiological Health
Biomet Microfixation RibFix Blu Thoracic Fixation System (K212608) was selected as the Primary Predicate for substantial equivalence in this submission. Additional FDA cleared predicate devices were referenced to support evaluation of performance characteristics.
We are only utilizing the Primary Predicate, Biomet Microfixation RibFix Blu Thoracic Fixation System (K212608), for our substantial equivalence comparison.
### IV. DEVICE DESCRIPTION [§807.92(a)(4)]
A. Submission Branch of Subject Device: Office of Orthopedic Devices, Office of Product Evaluation and Quality, Division of Stereotaxic, Trauma, and Restorative Devices, DHT6C
B. Subject Device: RibPlate System
This Traditional 510(k) is submitted for the Stryker RibPlate System. The Subject Device contains various commercially pure titanium or titanium alloy plates and screws for use in the chest wall.
### V. INDICATIONS FOR USE [§807.92(a)(5)]
The wording of the Indications for Use statements for the Predicate device in comparison to the Subject is fundamentally equivalent. See Table 5-1 below.
K261054 - Page 2 of 5
{6}
TABLE 5-1: COMPARISON OF INDICATIONS FOR USE
| Aspect | Subject Device | Primary Predicate Device (K212608) |
| --- | --- | --- |
| Intended Use/Indications for Use | The RibPlate system is indicated for use in the stabilization and fixation of fractures and reconstructions in the chest wall, including sternal reconstructive surgical procedures to promote bony fusion. The RibPlate system can be used in normal and poor bone. | The Biomet Microfixation RibFix Blu Thoracic Fixation System is indicated for use in the stabilization and fixation of fractures in the chest wall including sternal reconstructive surgical procedures, trauma, rib fractures associated with flail chest, or planned osteotomies. The system may be used in normal and poor bone to promote union. |
Please note that this 510(k) lists several additional Predicate devices, however, these additional Predicate devices were used for testing and not overall substantial equivalence evaluation.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE [§807.92(a)(6)]
The Subject Device is compared to its Predicate Devices for substantial equivalence of technological characteristics based on the following criteria:
- A. Principles of Operation
- B. Technological Characteristics
#### A. Principles of Operation / Operating Principle
The basic operating principle of the Subject Device is the same as the Predicate Device. The basic operating principle of the Subject Device, as well as the Predicate Device is the stabilization and fixation of fractures and reconstruction in the chest wall using titanium plates and screws that lock into each other.
#### B. Technological Characteristics
The operating principle, fixation method, material, sterility and sterilization method, and patient population are identical between Subject Device and Predicate Device Systems. The dimensions of the Subject Device plates and screws are similar to the Predicate Device in size and screw/plate interface.
### VI. PERFORMANCE DATA [§807.92(b)(1 and 2)]
#### Biocompatibility
The Subject Devices have been evaluated per the requirements of ISO 10993-1:2018 Biological evaluation of medical devices – Part1: Evaluation and testing within a risk management process and in alignment with FDA guidance “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – part 1: Evaluation and testing with a risk management process Issued September 4, 2020. This evaluation has concluded that no new or increased biocompatibility risks are associated with the Subject Device.
K261054 - Page 3 of 5
{7}
## Reprocessing and Sterility
The reprocessing instructions (cleaning, disinfection, and sterilization) for the Stryker RibPlate system (Subject Device) were validated to ensure the device conforms to the defined user needs and Intended Use/Indication for Use. The sterility testing was conducted according to BS ISO 17664-1, ISO 17665, AAMI TIR12, AAMI TIR39, ANSI/AAMI ST77, ANSI/AAMI ST79 and FDA guidance using a steam autoclave.
The Subject Device is provided non-sterile and is intended to be sterilized by the end-user via steam sterilization methods.
## Performance Bench Testing
All Verification and Validation (V&V) tests, which were identified in the risk analysis, met their respective acceptance criteria as compared to the Predicate Devices. These V&V tests included static 4-point-bending, dynamic 4-point-bending, pull-out, insertion and removal, shear-off, profile height, locking torque to failure, locking removal torque, self-retention, and end-user tests.
The results of these tests support the fact that the Subject Device fulfils the design inputs claimed in the Design Input / Output Verification and Validation (DIOVV) document. The results of this testing show that Subject Device is substantially equivalent to the Predicate Devices.
An assessment regarding magnetic resonance (MR) compatibility for the Subject Device was conducted. Magnetically induced displacement force, magnetically induced torque, and MR image artifact were evaluated through comparison to previously tested reference devices, while RF-induced heating was evaluated using computational modeling and simulation. Based on a worst-case analysis using computational modeling and scientific rationales, along with previously conducted testing on a Reference Device, the Subject Device was determined to be MR Conditional. The assessment was conducted per FDA Guidance, “Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, FDA, 2023” and utilized methods described in ASTM F2182, ASTM F2052, ASTM F2213, and ASTM F2119.
## Animal Testing
Animal testing was not required as a basis for substantial equivalence.
## Clinical Testing
Clinical testing was not required as a basis for substantial equivalence.
## VII. CONCLUSIONS [§807.92(b)(3)]
The Intended Use and Technological Characteristics comparisons provide evidence to prove substantial equivalence between the Subject and Predicate Device. There are no substantial differences between the devices. Therefore, Subject Device is substantially
K261054 - Page 4 of 5
{8}
equivalent to the Predicate Device with respect to Intended Use, General Characteristics, and Technological Characteristics.
K261054 - Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.