Surgical Gown Level2 (S, M, L, XL, XXL, XXXL)

K261028 · Yadu Medical (Henan) Co., Ltd. · FYA · Aug 7, 2026 · General, Plastic Surgery

Device Facts

Record IDK261028
Device NameSurgical Gown Level2 (S, M, L, XL, XXL, XXXL)
ApplicantYadu Medical (Henan) Co., Ltd.
Product CodeFYA · General, Plastic Surgery
Decision DateAug 7, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

Level2 Surgical Gown is sterile, single-use surgical apparel intended to be worn by healthcare professionals during surgical procedures to provide barrier protection against the transfer of microorganisms, body fluids, and particulate material for both patients and healthcare personnel. The product meets the Level 2 liquid barrier performance requirements of ANSI/AAMI PB70:2022 "Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities".

Device Story

Disposable, sterile surgical gown; constructed from 35g blue anti-SMMS non-woven fabric; features collar, body, sleeves, and lanyard/belt ties. Used in surgical procedures and invasive examinations to provide Level 2 liquid barrier protection per ANSI/AAMI PB70:2022. Operated by healthcare professionals in clinical settings. Protects both patient and provider from cross-contamination of fluids and particulates. Subject device is non-reinforced; provides barrier protection equivalent to predicate.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via AATCC TM42 (impact penetration), AATCC 127 (hydrostatic pressure), ISO 9073-10 (linting), ASTM D5034 (tensile), ASTM D5587 (tear), ASTM D1683 (seam strength), and 16 CFR 1610 (flammability). Biocompatibility confirmed per ISO 10993-5, -10, and -23. Sterility and packaging integrity verified per ISO 11737-2, ASTM F88, ASTM F1929, and ASTM D3078.

Technological Characteristics

Material: 35g blue anti-SMMS non-woven fabric. Design: Non-reinforced, raglan/set-in sleeves, hook-and-loop neck closure, belt ties. Barrier: Level 2 per ANSI/AAMI PB70:2022. Sterilization: Ethylene Oxide (EtO) per ISO 10993-7. Standards: ASTM D5034, ASTM D5587, ASTM D1683, ISO 9073-10, 16 CFR 1610.

Indications for Use

Indicated for healthcare professionals during surgical procedures to provide barrier protection against microorganisms, body fluids, and particulate material for patients and personnel.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0} **FDA** U.S. FOOD & DRUG ADMINISTRATION August 7, 2026 Yadu Medical (Henan) Co., Ltd. Winnie Li Regulatory Affairs Manager 66 Yangze Rd., Nanpu District, Changyuan, Xinxiang He Nan, Henan 453400 China Re: K261028 Trade/Device Name: Surgical Gown Level2 (S, M, L, XL, XXL, XXXL) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: July 8, 2026 Received: July 8, 2026 Dear Winnie Li: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K261028 – Winnie Li Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K261028 – Winnie Li Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Bifeng Qian, M.D., Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # **Indications for Use** Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K261028 Device Name Surgical Gown Level2 (S, M, L, XL, XXL, XXXL) Indications for Use (Describe) Level2 Surgical Gown is sterile, single-use surgical apparel intended to be worn by healthcare professionals during surgical procedures to provide barrier protection against the transfer of microorganisms, body fluids, and particulate material for both patients and healthcare personnel. The product meets the Level 2 liquid barrier performance requirements of ANSI/AAMI PB70:2022 "Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities". Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. **\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {4} # 510(k) Summary # K261028 # I 510(k) Submitter | Device Submitter: | YADU Medical (Henan) Co., Ltd. 66 Yangze Road, Nanpu District, 453400 Changyuan, Xinxiang, Henan Province PEOPLE'S REPUBLIC OF CHINA | | --- | --- | | Contact Person: | Winnie Li Regulatory Affairs Manager Phone: +86 18637387159 E-mail: winnie.li@yadugroup.com | | Summary | 2026.08.06 | | Preparation Date | | # II Device | Trade Name of Device: | Surgical Gown Level2 (S, M, L, XL, XXL, XXXL) | | --- | --- | | Regulation Number: | 21 CFR 878.4040 | | Classification Name: | Surgical gown | | Review Panel | General Hospital | | Regulatory Class | 2 | | Product Code: | FYA | | Common Device name | Surgical gown | | Size | S,M,L,XL,XXL,XXXL | # III Predicate Devices | 510k Number | K190950 | | --- | --- | | Trade Name of Device: | Medline Level 2 Surgical Gown (Eclipse Non-Reinforced) | | Regulation Number: | 21 CFR 878.4040 | | Classification Name: | Surgical Gown | | Review Panel | General Hospital | | Regulatory Class | 2 | | Product Code: | FYA | | Common Device name | Surgical gown | | Size | Small/Medium, Large, X-Large, XX-Large, XXX-Large, XXXX-Large | # IV Device Description Level 2 Surgical Gowns is made of non-woven fabric. This product is a disposable sterilized product. The product is composed of collar, body, sleeve and lanyard. It is Page 1 of 7 {5} used to prevent transmission of infectious materials between patients and health care workers during surgical procedures and other invasive examinations. Gowns have six different sizes: Small (S), Medium (M), Large (L), Extra Large (XL), Extra Extra Large (XXL), Extra extra extra Large (XXXL). These products meet the Level 2 liquid barrier performance requirements of ANSI/AAMI PB70:2022 “Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities”. #### **V Indication For Use** Level2 Surgical Gown is sterile, single-use surgical apparel intended to be worn by healthcare professionals during surgical procedures to provide barrier protection against the transfer of microorganisms, body fluids, and particulate material for both patients and healthcare personnel. The product meets the Level 2 liquid barrier performance requirements of ANSI/AAMI PB70:2022 "Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities". #### **VI Technological Characteristics Comparison** | Device Characteristic | Subject Device | Predicate Device (K190950) | Discussion | | --- | --- | --- | --- | | Trade name | Surgical Gown Level2 (S, M, L, XL, XXL, XXXL) | Medline Level 2 Surgical Gown (Eclipse Non-Reinforced) | N/A | | Product code | FYA | FYA | Same | | Regulation Number | 21 CFR 878.4040 | 21 CFR878.4040 | Same | | Manufacturer | YADU Medical (Henan) Co., Ltd. | Medline Industries, Inc. | N/A | | Device Classification | 2 | 2 | Same | Page 2 of 7 {6} | Device Characteristic | Subject Device | Predicate Device (K190950) | Discussion | | --- | --- | --- | --- | | Intended Use / Indications for Use | Level2 Surgical Gown is sterile, single-use surgical apparel intended to be worn by healthcare professionals during surgical procedures to provide barrier protection against the transfer of microorganisms, body fluids, and particulate material for both patients and healthcare personnel. The product meets the Level 2 liquid barrier performance requirements of ANSI/AAMI PB70:2022 "Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities". | The Medline Level 2 Surgical Gowns are sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Medline Level 2 Surgical Gowns meets the respective level requirements of ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities. | Different Comment 1 | | Color | Blue | Blue | Same | | Design Features | A. Collar ; B. Collar Closure ; C. Sleeve Opening ; D. Belt ; E. Body ; F. Sleeve | Available in Fabric non-Reinforced Hook and Loop Closure at neck Belt Ties Knit Cuffs Transfer Tab Raglan or Set-in/Standard Sleeves | Different Comment4 | | Sizes | Small to XXX-Large | Large, X-Large, XX-Large, XXX-Large, Small/Medium, XXXX-Large | Different Comment 2 | | Materials | 35g blue anti-SMMS | Nonwoven SMS polypropylene/Polyolefin | Different Comment 3 | Page 3 of 7 {7} | Device Characteristic | Subject Device | Predicate Device (K190950) | Discussion | | --- | --- | --- | --- | | Performance Specifications | Level 2 PB70 Barrier Protection | Level 2 PB70 Barrier Protection | Same | | Tensile strength | Meet the ASTM D5034-2021 Standard Test Method for Breaking Strength and Elongation of Textile Fabrics (Grab Test) | Meet the ASTM D5034-09 (2013) Breaking Strength and Elongation of Textile Fabrics (Grab Test) | Same | | Tear Strength | Meet the ASTM D5587-2015(2024) Standard Test Method for Tearing Strength of Fabrics by Trapezoid Procedure | Meet the ASTM D5587-15 Tearing Strength of Fabrics by Trapezoid Procedure | Same | | Seam Strength | Meet the ASTM D1683/D1683M | None | Different Comment 5 | | Lint Generation | Meet the ISO 9073-10:2003 | None | Different Comment 5 | | Package Test | Under the test conditions, the subject device was shown to be the seal strength compliant, No dye penetration, Microbial Barrier Properties meet the requirement, No bacteria growth and No leak per ASTM F88/F88M, ASTM F1929, DIN 58953-6, ISO11737-2& ASTM D3078-02. | Meet the ASTM F1929-15 Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration | Different Comment 6 | | Flammability | Meets requirements of Flame Resistant CFR Part 1610, Class 1 | Meets requirements of Flame Resistant CFR Part 1610, Class 1 | Same | | Prescription vs. OTC | OTC | OTC | Same | | Contact type and Durations | Surface, breached or compromised , < 24 ours | Surface, breached or compromised , < 24 hours | Same | | Single Use | Yes | Yes | Same | Page 4 of 7 {8} | Device Characteristic | Subject Device | Predicate Device (K190950) | Discussion | | --- | --- | --- | --- | | Biocompatibility | Under the test conditions, the subject device was shown to be non-cytotoxic, non-irritating and non-sensitizing per ISO 10993-5, ISO 10993-23 & ISO 10993-10. | Under the test conditions, the subject device was shown to be non-cytotoxic, non-irritating and non-sensitizing per ISO 10993-5 & ISO 10993-10. | Same | | Disposable vs Non Disposable | Disposable | Disposable | Same | | Sterile vs Non Sterile | Sterile | Sterile | Same | | Sterilization Method | Ethylene Oxide (EtO) | Ethylene Oxide (EtO) | Same | | Ethylene Oxide (EtO) Residuals | The residual values are within the limits specified by ISO 10993-7 standard. EO < 4mg/device ECH < 9 mg/device | The residual values are within the limits specified by ISO 10993-7 standard. | Same | # **Comment 1** The indications for use statement is worded slightly differently and both have the same intended use when used. All achieved the effect of the intended protection # **Comment 2** The only difference is that the subject gown does not have XXXX-Large size. # **Comment 3** Although the product materials are different, biocompatibility and specification performance testing data demonstrate that the subject device is as safe, as effective and can perform as well as the predicate device. # **Comment 4** The subject Surgical Gown adopts non-reinforced fabric construction. While the predicate device supplies optional reinforced zone material, the tested SMMS non-woven fabric of our device meets all ANSI/AAMI PB70-specified Level 2 liquid barrier, tensile strength and tear resistance performance requirements. The neck adopts hook and loop fastener closure; waist belt ties, knit cuffs, raglan cut sleeves and the overall structural layout are consistent with the predicate. The above mentioned structural difference will not affect the safety, barrier protection function and intended use performance of the device. # **Comment 5** The predicate device 510(k) Summary does not include seam strength and Lint Page 5 of 7 {9} Generation tests. The subject device conducts the above mentioned tests following ASTM D1683/D1683M and ISO 9073-10:2003. Test results meet standard specifications. The extra characteristic verification does not change core function, intended use or safety profile, and substantial equivalence is preserved. # **Comment 6** Compared with the predicate device which carries out ASTM F1929-15 dye-leak test, our device implements multiple sterile-package tests under multiple standards. All indicators are qualified. The extra-test items only verify packaging reliability and will-not affect substantial equivalence. # **VII Summary of Non-clinical Testing** | Test Item | | Test Standard | Level 2 Result | Acceptance Criteria | Pass/Fail | | --- | --- | --- | --- | --- | --- | | Surgical Gown | Impact Penetration | AATCC TM42-2017 | 0.8 g | ≤ 1.0 g | Pass | | | Hydrostatic Pressure | AATCC 127-2018e | 25 cm H₂O | Level 2: ≥ 20 cm H₂O | Pass | | | Lint Generation | IS09073-10:2003 | Minimal linting | Minimal linting allowed | Pass | | | Tensile Strength | ASTM D5034-2021 | MD, CD > 95N | ≥30N | Pass | | | Tear Strength | ASTM D5587-2015(2024) | MD, CD > 30N | ≥10N | Pass | | | Seam Strength | ASTM D1683/D1683M-2022 | > 70N | ≥30N | Pass | | | Flammability | US 16 CFR Part 1610 | IBE | Class 1 | Pass | | | Skin Irritation Test | ISO 10993-23 | Under the test conditions, the subject device was non-irritating | The erythema and edema scores are ≤ 0.4 | Pass | | | Test for In vitro cytotoxicity | ISO 10993-5 | Cell viability is 86.57%. There was no significant cytotoxicity. | Cell viability≥70% | Pass | | | Skin Sensitization | ISO 10993-10 | Sensitization rate is 0. No Sensitization reactions. | GPMT sensitization rate<30% | Pass | | | EtO residual testing | ISO 10993-7:2008 | EO < 4mg/device ECH < 9 mg/device | EO < 4mg/device ECH < 9 mg/device | Pass | | Package | Seal Strength | ASTM F88/F88M-2023 | ≥292.67N | ≥80N/m | Pass | Page 6 of 7 {10} | | Dye Penetration | ASTM F1929-2023 | No penetration | No penetration was found | Pass | | --- | --- | --- | --- | --- | --- | | | Microbial Barrier Properties | DIN 58953-6:2023 | Viable Count = 0 | No penetration | Pass | | | Sterility Test | ISO11737-2-2019 | No bacteria growth | Should be sterile. | Pass | | | Vacuum leak | ASTM D3078-02 | No leak | No leak was found. | Pass | ### VIII Clinical Test Conclusion No clinical study is included in this submission. ### IX Conclusion The conclusions drawn from the nonclinical tests demonstrate that subject Surgical Gown is as safe, as effective, and performs as well as or better than the legally marketed predicate device. Page 7 of 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...