NOVEOS Specific IgE (sIgE); Capture Reagent F014, Soybean (Glycine max); Capture Reagent F017, Hazelnut (Corylus avellana); Capture Reagent F352, Ara h 8 PR-10, Peanut; Capture Reagent F423, Ara h 2, Peanut; Capture Reagent F447, Ara h 6, Peanut
Applicant
Hycor Biomedical
Product Code
DHB · Immunology
Decision Date
Jul 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5750
Device Class
Class 2
Indications for Use
The NOVEOS™ Specific IgE assay is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum. NOVEOS Specific IgE assay is to be used with the NOVEOS Immunoassay Analyzer. It is intended for use as an in vitro diagnostic aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings and is to be used in clinical laboratories.
Device Story
The NOVEOS Specific IgE assay is an in vitro diagnostic test performed on the NOVEOS Immunoassay Analyzer in clinical laboratories. It uses patient serum samples to detect allergen-specific IgE. The assay employs fluorescent-labeled magnetic streptavidin-coated microparticles incubated with biotinylated allergenic capture reagents, patient sample, and HRP-labeled anti-human IgE monoclonal antibody. A magnetic field collects beads for washing to remove unbound components. Chemiluminescent substrate is added; light generation is proportional to the amount of bound IgE. Fluorescence corrects for bead loss. The instrument compares corrected chemiluminescence to a calibration curve traceable to WHO standard 11/234 to quantify sIgE. Results aid clinicians in diagnosing IgE-mediated allergic disorders when combined with other clinical findings.
Clinical Evidence
Clinical performance was evaluated by comparing test results from atopic (diagnosed via oral food challenge or physician) and non-atopic (ImmunoCAP negative) cohorts. Sensitivity ranged from 33.3% to 94.0% and specificity from 98.4% to 100.0% across the five allergens. Literature was provided to support the observed sensitivity for peanut components.
Technological Characteristics
Chemiluminescent immunometric assay using magnetic streptavidin-coated microparticles. Reagents include HRP-labeled mouse monoclonal anti-IgE. Operates at 37°C. Traceable to WHO 11/234. Quantitative measurement range 0.10–100 kU/L. Analyzed on the NOVEOS Immunoassay Analyzer.
Indications for Use
Indicated for the quantitative measurement of allergen-specific IgE in human serum to aid in the clinical diagnosis of IgE-mediated allergic disorders in patients suspected of having allergies to soybean, hazelnut, or peanut (Ara h 8, Ara h 2, or Ara h 6 components).
Regulatory Classification
Identification
A radioallergosorbent immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the allergen antibodies (antibodies which cause an allergic reaction) specific for a given allergen. Measurement of specific allergen antibodies may aid in the diagnosis of asthma, allergies, and other pulmonary disorders.
Special Controls
*Classification.* Class II (special controls). The device, when intended to detect any of the allergens included in Table 1 in this paragraph, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9. Table 1—Class II Exempt Allergens Under § 866.5750—Radioallergosorbent (RAST) Immunological Test Systems
Allergen code Allergen product Source (taxonomical name) Grass Pollens g1 Sweet vernal grass Anthoxanthum odoratum. g3 Cocksfoot grass, Orchard grass Dactylis glomerata. g4 Meadow fescue Festuca elatior. g5 Rye-grass (perennial rye grass) Lolium perenne. g7 Common reed (common reed grass) Phragmites communis. g8 Meadow grass, Kentucky blue (June grass) Poa pratensis. g9 Redtop, Bentgrass Agrostis stolonifera ,Agrostis gigantea (Agrostis alba ). g11 Brome grass Bromus inermis. g12 Cultivated rye (cultivated rye grass) Secale cereale. g13 Velvet grass Holcus lanatus. g14 Cultivated oat (cultivated oat grass) Avena sativa. g15 Cultivated wheat (cultivated wheat grass) Triticum aestivum (Triticum spp.). g16 Meadow foxtail (meadow foxtail grass) Alopecurus pratensis. g17 Bahia grass Paspalum notatum. g24 Wheat grass, Western Agropyron smithii (Elymus smithii ). g30 Bluegrass, annual Poa annua. g70 Wild rye grass Elymus triticoides Elymus condensatus. g71 Canary grass Phalaris arundinacea. g201 Barley, cultivated Hordeum vulgare. g202 Maize, corn (cultivated corn) Zea mays. g203 Salt grass Distichlis spicata. g204 False oat-grass Arrhenatherum elatius. g216 Cyn d 1 Cynodon dactylon. g701 Phl p 1.0102, Phl p 5.0101 Phleum pratense. g702 Phl p 7.0101 Phleum pratense. g703 Phl p 12.0101 Phleum pratense. Weed Pollens w2 Western ragweed Ambrosia psilostachya. w4 False ragweed Ambrosia acanthicarpa (Franseria acanthicarpa ). w5 Wormwood Artemisia absinthium Artemisia annua. w6 Mugwort Artemisia vulgaris. w7 Marguerite, ox-eye daisy Chrysanthemum leucanthemum. w8 Dandelion Taraxacum vulgare ,Taraxacum officinale. w9 Plantain (English), Ribwort Plantago lanceolata. w10 Goosefoot, lamb's quarters Chenopodium album. w11 Saltwort (prickly), Russian thistle Salsola kali (Salsola pestifer ). w12 Goldenrod Solidago virgaurea (Solidago spp.). w13 Cocklebur, common Xanthium commune. w14 Common pigweed (rough pigweed) Amaranthus retroflexus. w15 Scale, Lenscale Atriplex lentiformis. w16 Rough marsh elder Iva ciliate ,Iva annua. w17 Firebush (Kochia) Kochia scoparia. w18 Sheep sorrel Rumex acetosella. w19 Wall pellitory Parietaria officinalis. w20 Nettle (Common stinging nettle) Urtica dioica. w21 Wall pellitory Parietaria judaica. w22 Japanese hop (careless weed) Humulus japonicas (Humulus scandens ). w23 Yellow dock, Yellow dockweed Rumex crispus. w24 Spiny pigweed Amaranthus spinosus. w27 Carnation Dianthus spp. w28 Rose Rosa rugosa. w33 Clover Trifolium pratense. w35 Mexican tea Chenopodium ambrosioides. w36 Rabbit bush Ambrosia deltoidea (Franseria deltoides ). w37 Salt bush, annual Atriplex wrightii. w39 Water hemp, Western Amaranthus rudis (Acnida tamariscina ). w41 Burrobrush Hymenoclea salsola. w42 Poverty weed Baccharis neglecta. w43 Common sagebrush Artemisia tridentata. w45 Alfalfa Medicago sativa. w46 Dog fennel Eupatorium capillifolium. w53 Geranium Geranium spp. w67 Groundsel bush Baccharis halimifolia. w69 Iodine bush Allenrolfea occidentalis. w70 Ragweed, slender Ambrosia confertiflora. w75 Wing scale (wingscale) Atriplex canescens. w82 Careless weed Amaranthus palmeri ,Amaranthus hybridus. w90 Japanese hop Humulus japonicas (Humulus scandens ). w203 Rape (rape pollen) Brassica napus. w204 Sunflower Helianthus annuus. w206 Camomile Matricaria chamomilla. w207 Lupin Lupinus spp. w210 Sugar-beet Beta vulgaris. w211 Par j 2.0101 Parietaria judaica. w231 Art v 1 Artemisia vulgaris (Mugwort). w232 Sal k 1 Salsola kali. w233 Art v 3 Artemisa vulgaris (LTP, Mugwort). w234 Pla l 1 Plantago lanceolata. w235 Che a 1.0101 Chenopodium album. w236 Mer a 1.0101 Mercurialis annua. a753 Art v 1 Artemisia vulgaris (Mugwort weed). Tree Pollens t1 Box-elder (Maple) Acer negundo ,Acer saccharum. t2 Gray alder, speckled alder (alder) Alnus incana. t4 Hazel, hazelnut Corylus avellana ,Corylus americana. t5 American beech (beech) Fagus grandifolia (Fagus americana ). t6 Mountain juniper, Mountain cedar Juniperus ashei (Juniperus sabinoides ). t8 Elm Ulmus americana. t9 Olive Olea europaea. t10 Walnut Juglans californica ,Juglans nigra. t11 Maple leaf sycamore, London plane, Plane tree Platanus acerifolia. t61 Sycamore Platanus occidentalis. t12 Willow Salix caprea ,Salix nigra. t14 Cottonwood (Eastern Cottonwood/Black Cottonwood) Populus deltoides. t15 White ash Fraxinus americana. t16 White pine Pinus strobus. t18 Eucalyptus, gum-tree Eucalyptus globulus (Eucalyptus spp.). t19/t26 Acacia Acacia longifolia (Acacia spp.). t20 Mesquite Prosopis glandulosa/ Prosopis juliflora. t21 Melaleuca, cajeput tree Melaleuca quinquenervia (Melaleuca leucadendron ). t22 Pecan, hickory Carya illinoinensis (Carya pecan ). t23 Italian/Mediterranean/funeral cypress Cupressus sempervirens. t24 Japanese cypress Chamaecyparis obtusa (Chamaecyparis spp.). t25 Ash Fraxinus excelsior. t27 Maple, red Acer rubrum. t29 Acacia Acacia spp. t30 Birch, white Betula populifolia. t32 Willow, black Salix nigra. t33 Ash, Arizona Fraxinus velutina. t35 Cedar, salt Tamarix gallica. t37 Bald cypress (white bald cypress) Taxodium distichum. t38 Elm, Chinese/Siberian Ulmus pumila. t40 Hazelnut tree Corylus americana. t41 White hickory Carya alba (Carya tomentosa ). t42 Oak, red Quercus rubra. t43 Loblolly pine Pinus taeda. t44 Hackberry Celtis occidentalis. t45 Cedar elm Ulmus crassifolia. t47 Juniper, one seed Juniperus monosperma. t48 Pine, lodgepole Pinus contorta. t49 Pine, ponderosa Pinus ponderosa. t50 Beech, European Fagus sylvatica. t51 Tree of Heaven Ailanthus altissima. t52 Western white pine Pinus monticola. t54 Russian olive Elaeagnus angustifolia. t55 Scotch broom Cytisus scoparius. t56 Bayberry Myrica cerifera. t57 Red cedar Juniperus virginiana. t60 Western juniper Juniperus occidentalis. t61 Sycamore Platanus occidentalis. t70 Mulberry (white mulberry) Morus alba. t71 Red mulberry Morus rubra. t72 Queen palm Arecastrum romanzoffiamon. t73 Australian pine Casuarina equisetifolia. t77 Oak mix (red, white, black) Quercus spp. t80 Japanese cypress Chamaecyparis obtusa. t81 Japanese alder Alnus japonica. t83 Mango tree Mangifera indica. t90 Walnut, black Juglans nigra. t96 Poplar, white (poplar) Populus alba. t103/t218 Virginia live oak (live oak) Quercus virginiana. t105 Pepper tree Schinus molle. t110 Orange tree Citrus sinensis. t201 Spruce, Norway spruce Picea abies (Picea excelsa ). t202 Alder, smooth Alnus incana spp.Rugosa (Alnus rugosa ). t203 Horse chestnut Aesculus hippocastanum. t205 Elder Sambucus nigra. t206 Chestnut Castanea sativa. t207 Douglas fir Pseudotsuga menziesii (Pseudotsuga taxifolia ). t208 Linden Tilia cordata. t209 Horn beam Carpinus betulus. t210 Privet Ligustrum vulgare. t211 Sweet gum Liquidambar styraciflua. t212 Cedar Libocedrus decurrens. t213 Pine Pinus radiata. t214 Date palm Phoenix canariensis. t215 Lilac Syringa vulgaris. t217 Pepper tree Schinus molle. t217 Red alder Alnus rubra. t218 Virginia live oak Quercus virginiana. t218 Bayberry (bayberry/sweet gale) Myrica gale. t219 Palo verde Cercidium floridum. t219 Red cedar Juniperus virginiana. t220 Bet v 4 Betula verrucosa (Birch). t221 Bet v 2.0101, Bet v 4 Betula verrucosa (Birch). t222 Cypress (Arizona cypress) Cupressus arizonica. t223 Oil palm Elaeis guineensis. t224 Ole e 1 Olea europaea. t225 Bet v 6 Betula verrucosa (Birch). t226 Cup a 1 Cupressus arizonica. t227 Ole e 7 Olea Europaea. t228 Aspen, quaking Populus tremuloides. t229 Eastern hemlock Tsuga canadensis. t230 Redwood (sequoia) Sequoia sempervirens. t232 Pussy willow Salix discolor. t240 Ole e 9.0101 Olea Europaea. t241 Pla a 1.0101 Platanus acerifolia. t242 Pla a 2 Platanus acerifolia. t243 Pla a 3.0101 Platanus acerifolia. t244 Cor a 1.0103 Corylus avellana. t245 Aln g 1.0101 Alnus glutinosa. t246 Cry j 1 Cryptomeria japonica. t280 Locust tree Robinia pseudoacacia. t401 Brazilian peppertree Schinus terebinthifolius. t402 Mastic tree Pistacia lentiscus. t404 Tree of heaven Ailanthus altissima. t406 Date palm Phoenix dactylifera. a482 Ole e 1 Olea europaea (Olive Oil). Mites d207 Blo t 5.0101 Blomia tropicalis. d208 Lep d 2.0101 Lepidoglyphus destructor. Microorganisms, Molds, Yeast m1 Penicillium chrysogenum (Penicillium notatum ) Penicillium chrysogenum (Penicillium notatum ). m2 Cladosporium herbarum (Hormodendrum ) Cladosporium herbarum (Hormodendrum ). m3 Aspergillus fumigatus Aspergillus fumigatus. m4 Mucor racemosus Mucor racemosus. m5 Candida albicans Candida albicans. m7 Botrytis cinerea Botrytis cinerea. m8 Drechslera halodes (Setomelanomma rostrata, Helminthosporium halodes, Helminthosporium interseminatum ) Drechslera halodes (Setomelanomma rostrata, Helminthosporium halodes. m9 Fusarium moniliforme (Fusarium proliferatum ) Fusarium moniliforme (Fusarium proliferatum ). m10 Stemphylium botryosum Stemphylium herbarum (Stemphylium botryosum ). m11 Rhizopus nigricans Rhizopus nigricans. m12 Aureobasidium pullulans Aureobasidium pullulans. m13 Phoma betae Phoma betae. m14 Epicoccum purpurascens Epicoccum purpurascens (Epicoccum nigrum ). m15 Trichoderma viride Trichoderma viride. m16 Curvularia lunata Curvularia lunata ,Curvularia specifera (K923044 ). m17 Cladosporium fulvum Cladosporium fulvum. m18 Fusarium culmorum Fusarium culmorum. m19 Aspergillus versicolor Aspergillus versicolor. m20 Mucor mucedo Mucor mucedo. m21 Aspergillus clavatus Aspergillus clavatus. m22 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m23 Thermoactinomyces vulgaris Thermoactinomyces vulgaris. m24 Stachybotrys atra Stachybotrys chartarum (Stachybotrys atra ). m24 Paecilomyces spp Paecilomyces spp. m25 Aspergillus versicolor Aspergillus versicolor. m25 Penicillium brevicompactum Penicillium brevicompactum. m26 Cladosporium cladosporioides Cladosporium cladosporioides. m26 Penicillium citrinum Penicillium citrinum. m27 Penicillium spp Penicillium spp. m29 Aspergillus repens Aspergillus repens. m30 Penicillium roqueforti Penicillium roqueforti. m32 Cladosporium cladosporioides Cladosporium cladosporioides. m34 Serpula lacrymans Serpula lacrymans. m36 Aspergillus terreus Aspergillus terreus. m37 Trichophyton mentagrophytes Trichophyton mentagrophytes. m40 Aspergillus amstelodami Aspergillus amstelodami. m43 Saccharomyces Carlsberg Saccharomyces carlsbergensis. m44 Saccharomyces cerevisiae Saccharomyces cerevisiae. m45 Hormodendrum hordei Hormodendrum hordei. m46 Bipolaris spicifera Bipolaris spicifera. m47 Aspergillus nidulans Aspergillus nidulans. m48 Aspergillus oryzae Aspergillus oryzae. m49 Fusarium oxysporum Fusarium oxysporum. m50 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m51 Thermoactinomyces vulgaris Thermoactinomyces vulgaris. m53 Microspora canis Microsporum canis (Microspora canis ). m54 Aspergillus flavus Aspergillus flavus. m63 Helminthosporium intersemin Helminthosporium intersemin. m66 Mucor plumbeus Mucor plumbeus. m67 Mycogone Mycogone perniciosa. m68 Nigrospora oryzae Nigrospora oryzae. m69 Rhodotorula Rhodotorula rubra (Rhodotorula mucilaginosa ). m70 Malassezia furfur (Pityrosporum orbiculare ) Malassezia furfur (Pityrosporum orbiculare ). m71 Spondylocladium Spondylocladium spp. m72 Epidermophyton Epidermophyton floccosum. m73 Epicoccum nigrum Epicoccum purpurascens (Epicoccum nigrum ). m80 Staphylococcal enterotoxin A (Sta a SEA ) Staphylococcus aureus. m80 Helminthosporium spp Helminthosporium spp. m81 Staphylococcal enterotoxin B (Sta a SEB ) Staphylococcus aureus. m88 Stemphylium solani Stemphylium solani. m93 Gliocladium fimbriatum Gliocladium fimbriatum. m94 Phycomyces blakesleeanus Phycomyces blakesleeanus. m201 Tilletia tritici (Ustilago nuda, Ustilago tritici ) (Barley smut) Tilletia tritici (Ustilago nuda, Ustilago tritici ). m202 Acremonium kiliense (Cephalosporium acremonium ) Acremonium kiliense (Cephalosporium acremonium ). m203 Trichosporon pullulans Trichosporon pullulans. m204 Ulocladium chartarum Ulocladium chartarum. m205 Trichophyton rubrum Trichophyton rubrum. m207 Aspergillus niger Aspergillus niger. m208 Chaetomium globosum Chaetomium globosum. m209 Penicillium frequentans Penicillium glabrum (Penicillium frequentans ). m209 Stachybotrys chartarum Stachybotrys chartarum (Stachybotrys atra ). m210 Trichophyton mentagrophytes var. goetzii Trichophyton mentagrophytes var. goetzii. m211 Trichophyton mentagrophytes var. interdigitale Trichophyton mentagrophytes var. interdigitale. m211 Oat smut Ustilago avenae. m212 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m212 Geotrichum candidum Geotrichum candidum. m213 Bermuda grass smut Ustilago cynodontis. m214 Johnson grass smut Sphacelotheca cruenta. m215 Corn smut Ustilago maydis. m218 Asp f 1.0101 Aspergillus fumigatus. a3050 Asp r 1 Aspergillus restrictus. m219 Asp f 2 Aspergillus fumigatus. m220 Asp f 3.0101 Aspergillus fumigatus. m221 Asp f 4 Aspergillus fumigatus. m222 Asp f 6.0101 Aspergillus fumigatus. m223 Staphylococcal enterotoxin C (Sta a SEC ) Staphylococcus aureus. m224 Staphylococcal enterotoxin D (Sta a SED ) Staphylococcus aureus. m226 Staphylococcal enterotoxin TSST (Sta a TSST ) Staphylococcus aureus. m227 Malassezia spp. (Pityrosporum spp.) Malassezia spp. (Pityrosporum spp.). m228 Aspergillus flavus. m229 Alt a 1.0101 Alternaria alternata (Alternaria tenuis ). m230 Alt a 6.0101 Alternaria alternata (Alternaria tenuis ). m231 Cla h 8.0101 Cladosporium herbarum (Hormodendrum ). m300 Eurotium spp Eurotium spp. m304 Aspergillus oryzae Aspergillus oryzae. m305 Penicillium brevicompactum Penicillium brevicompactum. m309 Aspergillus terreus Aspergillus terreus. m310 Aspergillus nidulans Aspergillus nidulans. m311 Aspergillus flavus Aspergillus flavus. m312 Aspergillus clavatus Aspergillus clavatus. Epidermal & Animal e6 Guinea pig epithelium Cavia porcellus. e7 Pigeon droppings Columba palumbus ,Columba livia. e25 Chicken serum Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e26 Parrot serum Psittacoidea spp. e62 Camel Camelus dromedaries. e70 Goose feathers Anser anser. e71 Mouse epithelium Mus musculus (Mus spp.). e73 Rat epithelium Rattus norvegicus. e74 Rat urine proteins Rattus norvegicus ,Rattus rattus. e75 Rat serum proteins Rattus norvegicus ,Rattus rattus. e76 Mouse serum proteins Mus musculus (Mus spp.). e77 Budgerigar droppings Melopsittacus undulatus. e78 Budgerigar feathers Melopsittacus undulatus. e79 Budgerigar serum proteins Melopsittacus undulatus. e80 Goat epithelium Capra hircus. e81 Sheep epithelium Ovis aries (Ovis spp.). e82 Rabbit epithelium Oryctolagus cuniculus (Oryctolagus spp.). e83 Swine epithelium Sus scrofa (Sus scrofa domesticus; Sus spp.). e84 Hamster epithelium Cricetus cricetus, Mesocricetus auratus, and Phodopus sungorus. e85 Chicken feathers Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e86 Duck feathers Anas platyrhynchos. e87 Rat epithelium, serum proteins, and urine proteins Rattus norvegicus Rattus rattus. e88 Mouse epithelium, serum proteins, and urine proteins (mouse) Mus musculus (Mus spp.). e89 Turkey feathers Meleagris gallopavo. e90 Budgerigar serum proteins, feathers, and droppings Melopsittacus undulatus. e91 Pigeon serum proteins, feathers, and droppings Streptopelia roseogrisea ,Psittacidae spp. e92 Parrot serum proteins, feathers, and droppings Ara spp. e93 Pigeon serum proteins Streptopelia roseogrisea. e94 Fel d 1.0101 Felis domesticus. a345 Fel d 1 Felis domesticus. e98 Parrot droppings Psittacoidea spp. e101 Can f 1.0101 Canis familiaris (Canis domesticus ). a174 Can f 1 Canis familiaris (Canis domesticus ). e102 Can f 2.0101 Canis familiaris (Canis domesticus ). e196 Parakeet feathers Nymphicus hollandicus. e197 Parakeet droppings Nymphicus hollandicus. e198 Parakeet serum Nymphicus hollandicus. e199 Canary bird serum Serinus canarius. e200 Canary bird droppings Serinus canarius. e201 Canary bird feathers (Canary feathers) Serinus canarius. e202 Reindeer epithelium Rangifer tarandus. e203 Mink epithelium Mustela spp. e204 Bos d 6 Bos domesticus (Bos taurus; Bos spp.). e205 Horse, serum proteins Equus caballus (Equus spp.). e206 Rabbit, serum proteins Oryctolagus cuniculus (Oryctolagus spp.). e208 Chinchilla epithelium Chinchilla laniger. e209 Gerbil epithelium Meriones unguiculatus. e210 Fox epithelium Vulpes vulpes. e211 Rabbit, urine proteins Oryctolagus cuniculus (Oryctolagus spp.). e212 Swine, urine proteins Sus scrofa (Sus scrofa domesticus; Sus spp.). e213 Parrot feathers Ara spp. e214 Finch feathers Lonchura domestica. e215 Pigeon feathers Streptopelia roseogrisea (Streptopelia spp.),Columbia spp. e216 Deer epithelium Dama dama. e217 Ferret epithelium Mustela putorius. e218 Chicken droppings Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e219 Chicken, serum proteins Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e220 Fel d 2, Cat serum albumin Felis domesticus. e221 Can f 3 Canis familiaris (Canis domesticus ) (Dog serum albumin ). e222 Swine serum albumin (Sus s PSA) Sus scrofa (Sus scrofa domesticus; Sus spp.). e225 Lovebird feathers Psittacoidea agapomis. e226 Can f 5.0101 Canis familiaris. e227 Equ c 1.0101 Equus caballus. e228 Fel d 4.0101 Felis domesticus. e230 Equ c 3 Equus caballus. e231 Mus m 1 Mus musculus. Food f9 Rice Oryza sativa. f12 Pea (green pea) Pisum sativum. f15 White bean Phaseolus vulgaris. f19 Cayenne pepper Capsicum frutescens (Capsicum annum ). f21 Sugar cane Saccharum officinarum. f22 Raspberry Rubus idaeus. f26 Pork Sus scrofa (Sus scrofa domesticus; Sus spp.). f29 Watermelon Citrullus lanatus (Citrullus vulgaris ). f31 Carrot Daucus carota. f32 Oyster mushroom Pleurotus ostreatus. f33 Orange Citrus sinensis. f35 Potato Solanum tuberosum. f43 Mother's milk Homo sapiens. f44 Strawberry Fragaria vesca (Fragaria spp.). f45 Yeast, baker's Saccharomyces cerevisiae. f46 Pepper, Red Capsicum annuum. f47 Garlic Allium sativum. f48 Onion Allium cepa. f49 Apple Malus x domestica (Malus spp.). f51 Bamboo shoot Phyllostachys pubescens. f52 Cacao/chocolate Theobroma cacao. f54 Sweet potato Ipomoea batatas. f55 Common millet Panicum miliaceum. f56 Foxtail millet Setaria italica. f57 Japanese millet Echinochloa crus-galli. f58 Pacific squid Todarodes pacificus. f59 Octopus Octopus vulgaris (Octopus spp.). f63 Kefir NA. f67 Parmesan cheese NA. f81 Cheese, cheddar type NA. f82 Cheese, mold type NA. f83 Chicken Gallus domesticus (Gallus gallus domesticus; Gallus spp.). f86 Parsley Petroselinum crispum. f87 Melon Cucumis melo Cucumis melo + Citrullus lanatus. f88 Mutton (lamb) Ovis aries (Ovis spp.). f90 Malt Hordeum vulgare. f92 Banana Musa spp. f93 Cacao Theobroma cacao. f94 Pear Pyrus communis (Pyrus spp.). f97 Yam Dioscorea spp.Dioscorea opposita. f97 Chamomile tea Matricaria chamomilla. f98 Gliadin Triticum aestivum (Triticum spp.). f102 Cantaloupe Cucumis melo var. cantalupensis. f105 Chocolate Theobroma cacao. f109 Cottonseed Gossypium hirsutum. f110 Giant radish Raphanus sativus. f118 Zucchini Cucurbita pepo. f119 Radish Raphanus sativus. f120 Venison Capreolus capeolus. f121 Pinto bean Phaseolus vulgaris. f122 Cheese, American NA. f127 Black-eyed pea Vigna unguiculata. f131 Black Olive Olea europaea. f136 Red beet Beta vulgaris var. conditiva. f139 Goat's Cheese Capra aegagrus. f140 Bran NA. f141 Corn (vegetables) Zea mays. f152 Green bell pepper Capsicum annuum. f155 Brewer's yeast Saccharomyces carlsbergensis. f157 Duck Anas domesticus. f158 Goose Anser anser. f160 Camembert cheese NA. f162 Nectarine Prunus persica var. nucipersica. f163 Kohlrabi Brassica oleracea var. gongylodes. f65 Perch f166 Leek Allium porrum. f170 Cheese (Switzerland) (Swiss cheese) NA. f174 Fig Ficus carica. f177 Cranberry Vaccinium macrocarpon. f179 Raisin Vitis spp. f182 Lima bean Phaseolus lunatus. f198 Flaxseed (bruised grain) Linum usitatissimum. f199 Untreated native milk Bos domesticus (Bos taurus; Bos spp.). f208 Lemon Citrus limon. f209 Grapefruit Citrus paradisi. f210 Pineapple Ananas comosus. f211 Blackberry Rubus fruticosus. f212 Mushroom (champignon) Agaricus hortensis (Agaricus spp.). f213 Rabbit Oryctolagus cuniculus (Oryctolagus spp.). f214 Spinach Spinacia oleracea. f215 Lettuce Lactuca sativa. f216 Cabbage Brassica oleracea var. capitata. f217 Brussels sprouts Brassica oleracea var. gem. f218 Paprika, sweet pepper Capsicum annuum. f219 Fennel seed Foeniculum vulgare. f219 Sage Salvia officinalis. f220 Cinnamon Cinnamomum spp. f221 Coffee Coffea spp. f222 Tea Camellia sinensis. f223 Green olive Olea europaea. f225 Summer squash, pumpkin Cucurbita pepo. f225 Pumpkin Cucurbita maxima. f226 Pumpkin seed Cucurbita pepo. f227 Sugar-beet seed Beta vulgaris. f229 Safflower Seed Carthamus tinctorius. f231 Milk, boiled Bos domesticus (Bos taurus; Bos spp.). f234 Vanilla Vanilla planifolia. f237 Apricot Prunus armeniaca. f241 Gouda cheese NA. f242 Cherry Prunus avium. f244 Cucumber Cucumis sativus. f246 Guar, guar gum Cyamopsis tetragonoloba. f247 Honey NA. f248 Rosemary Rosmarinus officinalis. f254 Plaice Pleuronectes platessa. f255 Plum Prunus domestica ,Prunus americana. f258 Squid Loligo spp. f259 Grape Vitis vinifera (Vitis spp.). f260 Broccoli Brassica oleracea var. italica (Brassica oleracea var. cultivar ). f261 Asparagus Asparagus officinalis. f262 Aubergine, eggplant Solanum melongena. f263 Green pepper Piper nigrum ,Capsicum annuum. f264 Eel Anguilla anguilla. f265 Caraway Carum carvi. f265 Cumin Cuminum cyminum. f266 Mace Myristica fragrans. f267 Cardamon Elettaria cardamomum. f268 Clove Syzygium aromaticum. f269 Basil Ocimum basilicum. f270 Ginger Zingiber officinale. f271 Anise Pimpinella anisum. f272 Tarragon Artemisia dracunculus. f273 Thyme Thymus vulgaris. f274 Marjoram Origanum majorana. f275 Lovage Levisticum officinale. f276 Fennel, fresh Foeniculum vulgare. f277 Dill Anethum graveolens. f278 Bay leaf Laurus nobilis. f279 Chili pepper Capsicum frutescens. f280 Black pepper Piper nigrum. f281 Curry (Santa Maria) NA. f282 Nutmeg Myristica fragrans. f283 Oregano Origanum vulgare. f284 Turkey meat Meleagris gallopavo. f285 Elk/moose meat Alces spp. f286 Mare's milk Equus caballus (Equus spp.). f287 Red kidney bean Phaseolus vulgaris. f288 Blueberry Vaccinium myrtillus (Vaccinium spp.). f289 Date Phoenix dactylifera. f291 Cauliflower Brassica oleracea var. botrytis. f292 Guava Psidium guajava. f293 Papaya Carica papaya. f294 Passion fruit, Maracuja Passiflora edulis (Passiflora spp.). f295 Carambola Averrhoa carambola. f296 Carob Ceratonia siliqua. f297 Gum Arabic Acacia senegal (Acacia spp.). f298 Tragacanth Astragalus spp. f299 Sweet chestnut (chestnut) Castanea sativa. f300 Pinto bean Phaseolus spp. f301 Persimmon (kaki fruit, sharon) Diospyros kaki. f302 Mandarin (tangerine, clementine, satsumas) Citrus reticulata. f305 Fenugreek Trigonella foenum-graecum. f306 Lime Citrus aurantifolia. f307 Hake Merluccius merluccius. f308 Sardine (pilchard) Sardina pilchardus. f310 Blue vetch Lathyrus sativus. f311 Megrim Lepidorhombus whiffiagonis. f315 Green bean Phaseolus vulgaris. f316 Rape seed Brassica napus. f317 Coriander Coriandrum sativum. f318 Jack fruit Artocarpus heterophyllus. f319 Beetroot Beta vulgaris. f320 Crayfish Astacus astacus. f321 Horse meat Equus caballus (Equus spp.). f322 Red currant Ribes sylvestre. f324 Hop (fruit cone) Humulus lupulus. f325 Saffron Colchicum autumnale. f328 Fig Ficus carica. f329 Watermelon Citrullus lanatus. f330 Rose hip Rosa spp. f331 Saffron Crocus sativus. f332 Mint Mentha piperita. f333 Linseed Linum usitatissimum. f336 Jujube Ziziphus jujuba. f336 Wine vinegar Vitis vinifera (Vitis spp.). f337 Sole Solea solea. f337 English sole Parophrys vetulus. f338 Wine, white Vitis vinifera (Vitis spp.). f339 Allspice Pimenta dioica. f339 Wine, red Vitis vinifera (Vitis spp.). f341 Cranberry Vaccinium oxycoccus ,Vaccinium macrocarpon. f342 Olive (black, fresh) Olea europaea. f343 Raspberry Rubus idaeus. f344 Sage Salvia officinalis. f346 Chives Allium schoenoprasum. f347 Quinoa Chenopodium quinoa. f348 Litchi Litchi chinensis. f349 Chum salmon roe Oncorhynchus keta. f358 Artichoke Cynara scolymus. f360 Yogurt NA. f368 Black bass Micropterus dolomieu (Micropterus dolomieui ). f374 Karaya gum Sterculia urens. f375 Horseradish Armoracia rusticana. f377 Maple syrup NA. f379 Okra Abelmoschus esculentus. f382 Beet, sugar Beta vulgaris var. altissima. f401 Loquat Eriobotrya japonica. f402 Fig Ficus carica. f403 Brewer's yeast Saccharomyces cerevisiae. f405 Mint Mentha spp. f406 Arugula Eruca vesicaria. House Dust h1 Greer Labs., Inc NA. h2 Hollister-Stier Labs NA. h6 Japan NA. Venoms & Insects i7 Midge Chironomus yoshimatsui. i8 Moth Bombyx mori ,Heterocera spp. i47 Water flea Daphnia spp. i49 Deer fly Chrysops spp. i51 Black ant Camponotus pennsylvanicus. i54 Flea mix (dog/cat), common flea Ctenocephalides spp. i71 Mosquito Aedes communis ,Aedes spp.and Culex spp. i72 Green nimitti Cladotanytarsus lewisi. i73 Blood worm Chironomus thummi ,Chironomusri parius ,Chironomus spp. i75 European hornet Vespa crabro. i76 Berlin beetle Trogoderma angustum. i77 European paper wasp Polistes dominulus. i78 Fly Musca domestica. i80 Bumblebee Bombus pennsylvanicus. i201 Horse bot fly Gasterophilus intestinalis. i202 Grain weevil Sitophilus granarius. i203 Mediterranean flour moth Ephestia kuehniella (Anagasta kuehniella ). i204 Horse fly Tabanus spp. i205 Bumblebee Bombus terrestris. i208 Api m 1.0101 Apis mellifera. a45 Api m 1 Apis mellifera. i209 Ves v 5.0101 Vespula vulgaris. a670 Ves v 5 Vespula vulgaris. i210 Pol d 5.0101 Polistes dominulus. i211 Ves v 1.0101 Vespula vulgaris. i213 Api m 4 Apis mellifera. i214 Api m 2 Apis mellifera. i215 Api m 3 Apis mellifera. i216 Api m 5 Apis mellifera. i217 Api m 10 Apis mellifera. i220 Bla g 1.0101 Blattella germanica. i221 Bla g 2.0101 Blattella germanica. i222 Bla g 5.0101 Blattella germanica. i223 Bla g 7 Blattella germanica. a46 Api m 2 Apis mellifera. Miscellaneous o1 Cotton, crude fibers Gossypium spp. o3 Cotton (treated) Gossypium spp. o70 Seminal fluid Homo sapiens. o71 Staphylococcus aureus Staphylococcus aureus. o72 Pichia pastoris crude extract customer specific Pichia pastoris. o72 Sperm-sediment Homo sapiens. o73 Pichia pastoris crude extr. vector customer specific Pichia pastoris. o74 Pichia pastoris with vector customer specific Pichia pastoris. o201 Tobacco leaf, tobacco dust Nicotiana tabacum. o202 Artemia salina, fish feed Artemia salina. o203 Tetramin, fish feed NA. o207 Daphnia, fish feed Daphnia spp. o211 Mealworm Tenebrio molitor. o212 Streptavidin Streptomyces avidini. o213 MBP (maltose binding protein) Escherichia coli. o214 CCD; MUXF3 from bromelain Ananas comosus. o72 Enterotoxin A (Sta a SEA) Staphylococcus aureus. o73 Enterotoxin B (Sta a SEB) Staphylococcus aureus. Parasites p1 Ascaris Ascaris suum. p2 Echinococcus Echinococcus granulosus. p3 Schistosoma Schistosoma mansoni. p4 Anisakis (Herring Worm) Anisakis simplex (Anisakis spp.). p5 Toxocara canis Toxocara canis. p10 Ani s 3.0101 Anisakis simplex (Anisakis spp.). p11 Ani s 1 Anisakis simplex (Anisakis spp.). Occupational k4 Threshing dust NA. k5 Flax NA. k7 Hay Dust NA. k8 Hop (hops) Humulus lupulus. k12 Grain mill dust NA. k14 Kapok NA. k20 Sheep's wool (treated) (wool) Ovis aries (Ovis spp.). k21 Sheep's wool (Untreated) Ovis aries (Ovis spp.). k23 Straw Dust NA. k33 Oak NA. k70 Green coffee bean Coffea spp. k71 Castor bean Ricinus communis. k72 Ispaghula Plantago psyllium/Plantago ovata. k73 Silk waste NA. k74 Silk Bombyx mori. k75 Isocyanate TDI (Toluene diisocyanate) NA. k76 Isocyanate MDI (Diphenylmethane diisocyanate) NA. k77 Isocyanate HDI (Hexamethylen diisocyanate) NA. k78 Ethylene oxide NA. k79 Phthalic anhydride NA. k80 Formaldehyde/Formalin NA. k81 Ficus Ficus benjamina (Ficus spp.). k83 Cotton seed Gossypium hirsutum. k84 Sunflower seed Helianthus annuus. k85 Chloramin T NA. k86 Trimellitic anhydride, TMA NA. k87 Asp o 21, alpha-amylase Aspergillus oryzae. k89 Orris root Iris florentina. k99 HSA (Human Serum Albumin) (Hom s HSA) Homo sapiens. k201 Car p 1, Papain Carica papaya. k202 Ana c 2, Bromelain Ananas comosus. k204 Maxatase Bacillus licheniformis. k205 Alcalase Bacillus spp. k206 Savinase, Protease 1 (Bac l Subtilisin) Bacillus spp. k208 Gal d 4, Lysozyme Gallus domesticus (Gallus gallus domesticus; Gallus spp.). k209 Hexahydrophtalic anhydrid NA. k210 Maleic anhydride NA. k211 Methyltetrahydrophtalic anhydrid NA. k212 Abachi wood dust Triplochiton scleroxylon. k213 Pepsin (Sus s Pepsin) Sus scrofa (Sus scrofa domesticus; Sus spp.). k213 TCPA NA. k214 Bougainvillea Bougainvillea spp. k225 Horse radish peroxidase (Arm r HRP) Armoracia rusticana. k226 Ascorbate oxidase (Cuc p ascorbate oxidase) Cucurbita pepo. k301 Flour dust Triticum spp. k501 Savinase customer specific Proprietary knowledge of customer. k502 Lipolase customer specific Proprietary knowledge of customer. k503 Termamyl customer specific Proprietary knowledge of customer. k504 Clazinase customer specific Proprietary knowledge of customer.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
# **510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY**
# **I Background Information:**
# **A 510(k) Number**
K261027
# **B Applicant**
Hycor Biomedical
# **C Proprietary and Established Names**
NOVEOS Specific IgE (sIgE), Capture Reagent F014, Soybean (*Glycine max*)
NOVEOS Specific IgE (sIgE), Capture Reagent F017, Hazelnut (*Corylus avellana*)
NOVEOS Specific IgE (sIgE), Capture Reagent F352, Ara h 8 PR-10, Peanut
NOVEOS Specific IgE (sIgE), Capture Reagent F423, Ara h 2, Peanut
NOVEOS Specific IgE (sIgE), Capture Reagent F447, Ara h 6, Peanut
# **D Regulatory Information**
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DHB | Class II | 21 CFR 866.5750 - Radioallergosorbent (RAST) Immunological Test System | IM - Immunology |
# **II Submission/Device Overview:**
# **A Purpose for Submission:**
Five new devices
# **B Measurand:**
Allergen specific IgE to Soybean - F014, *Glycine max*)
Allergen specific IgE to Hazelnut - F017, *Corylus avellana*
Allergen specific IgE to Ara h 8 PR-10, Peanut - F352, Ara h 8
Allergen specific IgE to Ara h 2, Peanut - F423, Ara h 2
Allergen specific IgE to Ara h 6, Peanut - F447, Ara h 6
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
# C Type of Test:
Chemiluminescent, Quantitative
# III Intended Use/Indications for Use:
# A Intended Use(s):
See Indications for Use below.
# B Indication(s) for Use:
The NOVEOS Specific IgE assay is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum. NOVEOS Specific IgE assay is to be used with the NOVEOS Immunoassay Analyzer. It is intended for use as an in vitro diagnostic aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings and is to be used in clinical laboratories.
# C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
# D Special Instrument Requirements:
For use on the NOVEOS Immunoassay Analyzer
# IV Device/System Characteristics:
# A Device Description:
The NOVEOS Specific IgE assay consists of the following reagents:
• IgE Common Kit:
○ Diluent A (human serum albumin in buffer with preservative)
○ Conjugate IgE (horse radish peroxidase labeled mouse monoclonal anti-IgE antibody with preservative)
○ Substrate A
○ Substrate B
○ Fluo Beads (streptavidin coated magnetic particles with preservative)
• Capture Reagent Pack
○ Soybean - F014, Glycine max
○ Hazelnut - F017, Corylus avellana
○ Ara h 8 PR-10, Peanut - F352, Ara h 8
K251234 - Page 2 of 16
{2}
○ Ara h 2, Peanut - F423, Ara h 2
○ Ara h 6, Peanut - F447, Ara h 6
• IgE Calibrator Set (six levels: 0.07, 0.35, 0.70, 3.5, 17.5, 100 kU/L)
• IgE Calibrator Antibody Pack (mouse monoclonal anti-human IgE with preservative)
• IgE Positive Control Pack (known to be >0.35 kU/L)
• IgE Negative Control Packs (known to be <0.35 kU/L)
• Others:
○ Probe Wash Pack (phosphate buffered citric acid solution with preservative)
○ Wash Buffer Concentrate Pack (10X phosphate buffer saline solution with preservative)
○ Cuvette Wash Pack (citric acid solution with preservative)
# **B Principle of Operation:**
The NOVEOS Specific IgE assay is an immunometric, chemiluminescent procedure for the quantitative determination of IgE of known specificity in human serum samples. It employs fluorescent labelled magnetic, streptavidin coated microparticles which are incubated with a biotinylated allergenic capture reagent, patient sample and monoclonal anti-human IgE antibody labeled with horseradish peroxidase (Conjugate IgE). The beads are collected by using a magnetic field to remove the liquid and then undergo a wash step to remove unbound biotinylated allergen prior to incubation with the sample. If present in the sample, sIgE binds to the captured biotinylated allergen. After incubation, the beads are washed and subsequently incubated with an horseradish peroxidase (HRP)-labeled anti-IgE monoclonal antibody (Conjugate IgE) to form an antibody-conjugate complex that is bound to the captured biotinylated allergen. After washing away the non-bound conjugate, a chemiluminescent substrate is added to the assay mixture resulting in light generation in proportion to the amount of antibody-conjugate that has been bound to the beads. The fluorescence is used to correct for any bead loss during the assay, and the (corrected) chemiluminescence is compared to a calibration curve to provide quantification of bound sIgE. The higher the value of chemiluminescent signal detected by the instrument, the higher the amount of sIgE detected in the sample tested. The concentration of allergen specific IgE is determined from a standard curve, which is traceable to the World Health Organization (WHO) reference reagent serum Immunoglobulin E (IgE) 11/234.
# **V Substantial Equivalence Information:**
# **A Predicate Device Name(s):**
ImmunoCAP Specific IgE Assay
# **B Predicate 510(k) Number(s):**
K051218
K251234 - Page 3 of 16
{3}
# C Comparison with Predicate(s):
| Device & Predicate Device(s): | K261027 (Candidate Device) | K051218 (Predicate Device) |
| --- | --- | --- |
| Device Trade Name | NOVEOS Specific IgE (sIgE) | ImmunoCAP Specific IgE |
| **General Device Characteristic Similarities** | | |
| Intended Use/ Indications For Use | The NOVEOS Specific IgE assay is an *in vitro* quantitative assay for the measurement of allergen specific IgE in human serum. NOVEOS Specific IgE assay is to be used with the NOVEOS Immunoassay Analyzer. It is intended for use as an *in vitro* diagnostic aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings and is to be used in clinical laboratories. | ImmunoCAP Specific IgE is an *in vitro* quantitative assay for the measurement of allergen specific IgE in human serum or plasma (EDTA or Na-Heparin). ImmunoCAP Specific IgE is to be used with instruments Phadia 100, Phadia 250, Phadia 1000, Phadia 2500 and Phadia 5000. It is intended for *in vitro* diagnostic use as an aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings, and is to be used in clinical laboratories. |
| Traceability | World Health Organization (WHO) reference reagent serum Immunoglobulin E (IgE) 11/234 | Same |
| Assay Type | Quantitative | Same |
| Calibration Method | Heterologous interpolation based on Total IgE calibration curve | Same |
| Reaction Temperature | 37°C | Same |
| Limit of Quantification | 0.10 kU/L | Same |
| **General Device Characteristic Differences** | | |
| Solid Phase | Magnetic microparticles | Cellulose derivative |
| Assay Principle | Chemiluminescent assay | Fluoroenzyme-immunoassay |
| Instrument(s) | NOVEOS Immunoassay Analyzer | Phadia 100, Phadia 250/1000/2500/5000 |
| Specimen Type | Serum | Serum or plasma (EDTA, Na-Heparin) |
| Specimen Volume | 4 μL | 40 μL |
| Detection Antibody | HRP-conjugated mouse anti-human IgE monoclonal antibody | β-Galactosidase-anti-human IgE (mouse monoclonal antibody) |
| Calibrator levels | 6 levels: 0.07, 0.35, 0.7, 3.5, 17.5, 100 kU/L | 6 levels: 0, 0.35, 0.7, 3.5, 17.5, 100 kU/L |
| Time to First Result | 1 hour and 45 minutes | 1 hour and 45 minutes to 2 hours and 30 minutes depending on instrument model |
K251234 - Page 4 of 16
{4}
# **VI Standards/Guidance Documents Referenced:**
The following Clinical and Laboratory Standards Institute (CLSI) guidelines were used:
- CLSI I/LA20-A3: Performance Characteristics, Quality Assurance, and Clinical Utility of Immunological Assays for Human Immunoglobulin E Antibodies and Defined Allergen Specificities, Approved Guideline – Third Edition
- CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition.
- CLSI EP06, 2nd ed.: Evaluation of Linearity of Quantitative Measurement, Approved Guideline – Second Edition
- CLSI EP07, 3rd ed.: Interference Testing in Clinical Chemistry; Approved Guideline –Third Edition
- CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
- CLSI EP25, 2nd ed: Evaluation of Stability of In Vitro Medical Laboratory Test Reagent Reagents– Second Edition
- CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline – Third Edition
- CLSI EP37, Supplemental Tables for Interference Testing in Clinical Chemistry – First Edition
# **VII Performance Characteristics (if/when applicable):**
# **A Analytical Performance:**
# 1. Precision/Reproducibility:
# a. Within-Laboratory imprecision
The within-laboratory precision study was conducted per the CLSI EP05-A3 for each of four NOVEOS Specific IgE assays. A panel of four to five human sera (one to two negative and two to three positive patient samples) were tested in at least duplicates per run, two runs per day for a minimum of 20 days using one reagent lot on one or two NOVEOS Immunoassay Analyzers. The standard deviation (SD) and %CV of the within-run, between-run, between-day, and total within-laboratory imprecision were calculated for each sample. The 20-day within-laboratory precision results for each NOVEOS Specific IgE assay are summarized in the following tables:
K251234 - Page 5 of 16
{5}
| NOVEOS Specific IgE, F014, Soybean | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (kU/L) | N | Within-Run | | Between-Run | | Between-Day | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.16 | 80 | 0.01 | 6.7 | 0.01 | 7.3 | 0.01 | 5.0 | 0.02 | 11.1 |
| 2 | 0.27 | 82* | 0.02 | 7.7 | 0.01 | 3.3 | 0.01 | 3.7 | 0.02 | 9.1 |
| 3 | 4.06 | 80 | 0.13 | 3.2 | 0.11 | 2.7 | 0.10 | 2.5 | 0.20 | 4.8 |
| 4 | 92.69 | 80 | 3.83 | 4.1 | 5.51 | 5.9 | 3.54 | 3.8 | 7.59 | 8.2 |
*Sample 2 had two outliers removed on Day 12 and one additional day was added
| NOVEOS Specific IgE, F017, Hazelnut | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (kU/L) | N | Within-Run | | Between-Run | | Between-Day | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.23 | 80 | 0.01 | 4.5 | 0.01 | 4.4 | 0.01 | 5.2 | 0.02 | 8.2 |
| 2 | 0.31 | 80 | 0.01 | 3.0 | 0.03 | 10.3 | 0.01 | 3.6 | 0.04 | 11.3 |
| 3 | 1.37 | 80 | 0.04 | 2.8 | 0.05 | 3.9 | 0.08 | 6.1 | 0.11 | 7.8 |
| 4 | 5.48 | 80 | 0.17 | 3.0 | 0.27 | 5.0 | 0.35 | 6.4 | 0.47 | 8.6 |
| 5 | 88.97 | 82* | 7.35 | 8.3 | 5.64 | 6.3 | 0.00 | 0.0 | 9.26 | 10.4 |
*An extra day was tested due to two dropped replicates (incorrect sample loading).
| NOVEOS Specific IgE, F352, Ara h 8 PR-10, Peanut | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (kU/L) | N | Within-Run | | Between-Run | | Between-Day | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.17 | 80 | 0.01 | 5.4 | 0.00 | 2.7 | 0.02 | 10.5 | 0.02 | 12.1 |
| 2 | 0.27 | 80 | 0.02 | 5.8 | 0.01 | 4.7 | 0.02 | 5.9 | 0.03 | 9.5 |
| 3 | 1.60 | 80 | 0.07 | 4.4 | 0.01 | 0.9 | 0.07 | 4.2 | 0.10 | 6.2 |
| 4 | 6.45 | 80 | 0.19 | 2.9 | 0.13 | 2.0 | 0.21 | 3.2 | 0.31 | 4.7 |
| 5 | 81.42 | 92* | 6.32 | 7.8 | 5.25 | 6.4 | 2.23 | 2.7 | 8.52 | 10.5 |
*Sample 5 was tested in five replicates in the first run and two replicates in the second run on each day for four days, and in two replicates per run and two runs per day for the remaining 16 days
| NOVEOS Specific IgE, F423, Ara h 2, Peanut | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (kU/L) | N | Within-Run | | Between-Run | | Between-Day | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.14 | 80 | 0.01 | 9.9 | 0.01 | 5.7 | 0.01 | 6.0 | 0.02 | 12.9 |
| 2 | 0.32 | 80 | 0.03 | 8.0 | 0.00 | 0.0 | 0.00 | 0.0 | 0.03 | 8.0 |
| 3 | 2.75 | 80 | 0.15 | 5.5 | 0.06 | 2.0 | 0.11 | 4.0 | 0.19 | 7.1 |
| 4 | 77.35 | 80 | 4.15 | 5.4 | 3.29 | 4.3 | 2.37 | 3.1 | 5.80 | 7.5 |
| NOVEOS Specific IgE, F447, Ara h 6, Peanut | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (kU/L) | N | Within-Run | | Between-Run | | Between-Day | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.19 | 80 | 0.01 | 6.2 | 0.01 | 4.7 | 0.01 | 3.5 | 0.02 | 8.5 |
| 2 | 0.50 | 89* | 0.02 | 4.0 | 0.02 | 4.2 | 0.04 | 8.1 | 0.05 | 10.0 |
| 3 | 4.49 | 83** | 0.34 | 7.6 | 0.00 | 0.0 | 0.14 | 3.2 | 0.37 | 8.2 |
| 4 | 76.23 | 80 | 5.03 | 6.6 | 3.84 | 5.0 | 3.89 | 5.1 | 7.43 | 9.7 |
* Sample 2 was tested in five replicates in the first run and two replicates in the second run on each day for 3 days, and in two replicates per run and two runs per day for the remaining 17 days
**Sample 3 had one outlier removed on Day 19 and one additional day was added
K251234 - Page 6 of 16
{6}
# b. Lot-to-lot precision
A panel of four to six serum samples tested with three different lots of each NOVEOS sIgE allergen. The samples were tested in replicates of five per run, one run per day, for five to six days, to generate a total of 75 replicates per sample unless otherwise indicated for all three lots. The results are summarized in the following tables:
| NOVEOS Specific IgE, F014, Soybean | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (kU/L) | N | Within-Run | | Between-Day | | Between-Lot | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.14 | 75 | 0.02 | 11.8 | 0.01 | 10.4 | 0.00 | 0.0 | 0.02 | 15.7 |
| 2 | 0.46 | 75 | 0.02 | 5.3 | 0.01 | 3.0 | 0.00 | 0.6 | 0.03 | 6.2 |
| 3 | 1.35 | 75 | 0.04 | 2.8 | 0.05 | 3.4 | 0.02 | 1.6 | 0.06 | 4.7 |
| 4 | 4.08 | 75 | 0.16 | 3.9 | 0.11 | 2.7 | 0.00 | 0.0 | 0.19 | 4.7 |
| 5 | 72.06 | 75 | 2.94 | 4.1 | 2.16 | 3.0 | 0.00 | 0.0 | 3.65 | 5.1 |
| NOVEOS Specific IgE, F017, Hazelnut | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (kU/L) | N | Within-Run | | Between-Day | | Between-Lot | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.15 | 75 | 0.01 | 8.7 | 0.02 | 11.1 | 0.00 | 0.0 | 0.02 | 14.1 |
| 2 | 0.46 | 75 | 0.02 | 5.2 | 0.04 | 8.4 | 0.00 | 0.0 | 0.05 | 9.9 |
| 3 | 1.49 | 75 | 0.06 | 3.8 | 0.10 | 6.8 | 0.00 | 0.0 | 0.12 | 7.8 |
| 4 | 6.04 | 75 | 0.30 | 4.9 | 0.43 | 7.1 | 0.18 | 2.9 | 0.55 | 9.1 |
| 5 | 60.28 | 75 | 2.66 | 4.4 | 2.30 | 3.8 | 0.00 | 0.0 | 3.52 | 5.8 |
| 6 | 88.21 | 79* | 3.71 | 4.2 | 5.92 | 6.7 | 0.00 | 0.0 | 6.99 | 7.9 |
*Sample 6 had one outlier removed on Day 5 and one additional day (1 run with 5 replicates) was added
| NOVEOS Specific IgE, F352, Ara h 8 PR-10, Peanut | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (kU/L) | N | Within-Run | | Between-Day | | Between-Lot | | Total | |
| | | | SD | 0.15 | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.18 | 75 | 0.01 | 6.0 | 0.01 | 4.6 | 0.01 | 2.9 | 0.01 | 8.1 |
| 2 | 0.31 | 75 | 0.02 | 5.1 | 0.02 | 4.9 | 0.01 | 4.3 | 0.03 | 8.3 |
| 3 | 0.56 | 75 | 0.04 | 6.5 | 0.00 | 0.0 | 0.04 | 7.9 | 0.06 | 10.2 |
| 4 | 1.51 | 75 | 0.05 | 3.5 | 0.02 | 1.4 | 0.04 | 2.7 | 0.07 | 4.6 |
| 5 | 5.93 | 75 | 0.21 | 3.6 | 0.08 | 1.4 | 0.07 | 1.1 | 0.24 | 4.0 |
| 6 | 78.85 | 75 | 4.84 | 6.1 | 4.06 | 5.2 | 0.00 | 0.0 | 6.32 | 8.0 |
| NOVEOS Specific IgE, F423, Ara h 2, Peanut | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (kU/L) | N | Within-Run | | Between-Day | | Between-Lot | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.18 | 75 | 0.01 | 5.4 | 0.01 | 3.2 | 0.00 | 0.0 | 0.01 | 6.2 |
| 2 | 0.27 | 75 | 0.02 | 5.8 | 0.01 | 4.3 | 0.00 | 0.0 | 0.02 | 7.2 |
| 3 | 0.52 | 75 | 0.02 | 3.6 | 0.03 | 5.5 | 0.00 | 0.0 | 0.03 | 6.6 |
| 4 | 1.15 | 75 | 0.05 | 4.5 | 0.06 | 5.5 | 0.00 | 0.0 | 0.08 | 7.1 |
| 5 | 93.56 | 75 | 3.13 | 3.3 | 4.00 | 4.3 | 0.00 | 0.0 | 5.07 | 5.4 |
K251234 - Page 7 of 16
{7}
| NOVEOS Specific IgE, F447, Ara h 6, Peanut | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (kU/L) | N | Within-Run | | Between-Day | | Between-Lot | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.20 | 79* | 0.01 | 5.5 | 0.01 | 5.9 | 0.00 | 0.7 | 0.02 | 8.1 |
| 2 | 0.47 | 75 | 0.02 | 4.1 | 0.03 | 5.5 | 0.01 | 1.7 | 0.03 | 7.1 |
| 3 | 9.98 | 75 | 0.40 | 4.0 | 0.56 | 5.6 | 0.00 | 0.0 | 0.69 | 6.9 |
| 4 | 75.92 | 75 | 3.98 | 5.2 | 2.64 | 3.5 | 0.00 | 0.0 | 4.78 | 6.3 |
*Sample 1 had one outlier removed on Day 5 and one additional day (1 run with 5 replicates) was added
## 2. Linearity:
Linearity of each NOVEOS Specific IgE assay was evaluated in accordance with the CLSI guideline I/LA20-A3. Three sets of linearity dilution panels including six levels were prepared by using three positive human serum samples (low, mid, and high). The samples were serially diluted with one negative serum sample to create overlapping dilution panels. Testing was performed using one lot of each NOVEOS Specific IgE Assay in four replicates of each sample. The linearity data analysis was performed individually for each NOVEOS Specific IgE assay in accordance with the CLSI guideline EP06, 2nd Edition. For each sample level in the dilution panels, the mean value of the measured values, the predicted value and the %deviation from linearity were calculated. The linear range and the claimed analytical meaning range for each NOVEOS Specific IgE are shown in the table below:
| NOVEOS Specific IgE | Tested Linear Range (kU/L) | Analytical Measuring Range (kU/L) | Regression Equation |
| --- | --- | --- | --- |
| F014, Soybean | 0.07–124.26 | 0.10 – 100.0 | y = 1.02x + 0.01 |
| F017, Hazelnut | 0.08–111.87 | 0.10 – 100.0 | y = 0.99x – 0.02 |
| F352, Ara h 8 PR-10, Peanut | 0.06–100.93 | 0.10 – 100.0 | y = 0.93x + 0.01 |
| F423, Ara h 2, Peanut | 0.09–99.09 | 0.10 – 99.0 | y = 0.97x + 0.00 |
| F447, Ara h 6, Peanut | 0.07–118.95 | 0.10 – 100.0 | y = 0.94x – 0.01 |
## 3. Analytical Specificity/Interference:
### a. Inhibition Study
Immunological specificity of each NOVEOS Specific IgE assay was verified through competitive inhibition study according to the CLSI guideline I/LA20-A3. Positive samples with specific IgE concentration used in the study include: 1.39 kU/L for F014, Soybean; 7.94 kU/L for F017, Hazelnut; 5.77 kU/L for F352, Ara h 8 PR-10, Peanut; 6.48 kU/L for F423, Ara h 2, Peanut; and 6.83 kU/L for F447, Ara h 6, Peanut.
The allergen solution (inhibitor) was diluted 1:1 into the positive sample and subsequently serially diluted five times in two-fold increments. Each diluted sample was evaluated for dose-dependent inhibition in replicates of four using one lot of NOVEOS Specific IgE assay for F014 (Soybean), F017 (Hazelnut), F352 (Ara h 8 PR-10, Peanut), F423 (Ara h 2, Peanut), and F447 (Ara h 6, Peanut). Greater than 85% inhibition was achieved for F017 (Hazelnut) at 25 μg/mL (87% inhibition) and F447 (Ara h 6, Peanut) at
K251234 - Page 8 of 16
{8}
3.125 μg/mL (88% inhibition). Substantial inhibition was also observed at the following inhibitor concentrations: 84% inhibition for F014 (Soybean) at 25 μg/mL, 83% inhibition for F352 (Ara h 8 PR-10, Peanut) at 50 μg/mL, and 85% inhibition for F423 (Ara h 2, Peanut) at 3.125 μg/mL.
For the NOVEOS Specific IgE Assays F014 (Soybean), F017(Hazelnut), F352 (Ara h 8 PR-10, Peanut), F423 (Ara h 2, Peanut), and F447(Ara h 6, Peanut) the related and unrelated allergens tested were as follows:
- For F014 (Soybean), the related allergen F018 (Brazil Nut) and the unrelated allergens K082 (Latex), M005 (Candida albicans) and W013 (Cocklebur Pollen) were tested as potential inhibitors at a final concentration of at least ten times the highest concentration of F014 at which 84% inhibition was achieved. All allergen inhibitors were spiked into the positive serum sample at a 1:1 dilution and evaluated with one lot of NOVEOS Specific IgE Assay, F014. The results of the single-dose inhibition studies using one related allergen (F018) and three unrelated allergens showed ≤15% inhibition at a test concentration of 2180 μg/mL (F018), 975 μg/mL (K082), 1235 μg/mL (M005) and 1080 μg/mL (W013). The inhibition studies indicate that the NOVEOS Specific IgE Assay, F014 contains the immunologically relevant allergen.
- For F017 (Hazelnut), the related allergen F338 (Sea Scallops) and the unrelated allergens D201 (Blomia tropicalis), E070 (Goose Feathers), I006 (German Cockroach), I070 (Fire Ant), M003 (Aspergillus fumigatus), and M012 (Aureobasidium pullulans) were tested as potential inhibitors at a final concentration of at least ten times the highest concentration of F017 that achieved 87% inhibition. All allergen inhibitors were spiked into the positive serum sample at a 1:1 dilution and evaluated with one lot of NOVEOS Specific IgE Assay, F017. The results of the single-dose inhibition studies using one related allergen (F338) and six unrelated allergens showed ≤15% inhibition at test concentrations of 250 μg/mL (F338, D201, E070, I006, I070, M003, M012). The inhibition studies indicate that the NOVEOS Specific IgE Assay, F017 contains the immunologically relevant allergen.
- For F352 (Ara h 8 PR-10, Peanut), the related allergen F077 (Bos d 5 β-lactoglobulin, Milk) and the unrelated allergens E221 (Can f 3 Dog Serum Albumin, Dog), M218 (Asp f 1, Aspergillus fumigatus), and T215 (Bet v 1 PR-10, Birch) were tested as potential inhibitors at a final concentration of at least ten times the highest concentration of F352 that achieved 83% inhibition. All allergen inhibitors were spiked into the positive serum sample at a 1:1 dilution and evaluated with one lot of NOVEOS Specific IgE Assay, F352. The results of the single-dose inhibition studies using one related allergen (F077) and three unrelated allergens showed ≤15% inhibition at test concentrations of 3520 μg/mL (F077), 1970 μg/mL (E221), 660 μg/mL (M218) and 1200 μg/mL (T215). The inhibition studies indicate that the NOVEOS Specific IgE Assay, F352 contains the immunologically relevant allergen.
- For F423 (Ara h 2, Peanut), the related allergen F427 (Ara h 9 LTP, Peanut) and the unrelated allergens E221 (Can f 3 Dog Serum Albumin), M218 (Asp f 1, Aspergillus fumigatus), and T215 (Bet v 1 PR-10, Birch) were tested as potential inhibitors at a final concentration of at least ten times the highest concentration of F423 that
K251234 - Page 9 of 16
{9}
achieved 85% inhibition. All allergen inhibitors were spiked into the positive serum sample at a 1:1 dilution and evaluated with one lot of NOVEOS Specific IgE Assay, F423. The results of the single-dose inhibition studies using one related allergen (F427) and three unrelated allergens showed ≤15% inhibition at test concentrations of 31.25 μg/mL (F427, E221, M218 and T215). The inhibition studies indicate that the NOVEOS Specific IgE Assay, F423 contains the immunologically relevant allergen.
- For F447 (Ara h 1, Peanut), the related allergen F076 (Bos d 4, α-lactoglobulin, Milk) and the unrelated allergens E221 (Can f 3, Dog Serum Albumin), G215 (Phl p 5b, Timothy Grass), and M220 (Asp f 3, Aspergillus fumigatus) were tested as potential inhibitors at a final concentration of at least ten times the concentration of F447 at which 88% inhibition was achieved. All allergen inhibitors were spiked into the positive serum sample at a 1:1 dilution and evaluated with one lot of NOVEOS Specific IgE Assay, F447. The results of the single-dose inhibition studies using one related allergen (F076, Bos d 4, α-lactoglobulin, Milk) and three unrelated allergens showed ≤15% inhibition at test concentrations of 1460 μg/mL (F076), 1970 μg/mL (E221), 1500 μg/mL (G215) and 1220 μg/mL (M220). The inhibition studies indicate that the NOVEOS Specific IgE Assay, F447 contains the immunologically relevant allergen.
# b. Interference:
# i. Endogenous Substance Interference:
The effect of the presence of elevated levels of human serum albumin, hemoglobin, triglycerides, conjugated bilirubin, and unconjugated bilirubin in serum samples was evaluated by testing three serum sample pools (one negative above LoQ, one near the cut-off, and one positive) spiked with varying levels of each interferent. Testing was conducted with seven replicates in one assay run using one lot each of the NOVEOS sIgE assays for F014 (Soybean), F017 (Hazelnut), F352 (Ara h 8 PR-10, Peanut), F423 (Ara h 2, Peanut), and F447 (Ara h 6, Peanut). The % Interference for each sample spiked with the potential interfering substance was calculated by comparing its result to that of the corresponding control sample spiked with an equal volume of the solvent without the interfering substance. No interference was noted for samples containing human serum albumin up to 121 g/L, hemoglobin up to 200 mg/dL, intralipid up to 3000 mg/dL, conjugated bilirubin up to 30 mg/dL, and unconjugated bilirubin up to 20 mg/dL. Interference testing with rheumatoid factor (RF) was previously performed for NOVEOS sIgE allergens and found no significant interference at 513 IU/mL (refer to K200825).
# ii. Exogenous Interference:
Refer to K200825
# c. Cross-reactivity:
The potential cross-reactivity of non-IgE (IgA, IgD, IgG, and IgM) to the NOVEOS Specific IgE assays was evaluated in K200825.
K251234 - Page 10 of 16
{10}
# 4. Assay Reportable Range:
The assay reportable range is the same as the claimed analytical measuring range for each NOVEOS Specific IgE as shown in the table below:
| NOVEOS Specific IgE Assay | Assay Reportable Range/ Analytical Measuring Range |
| --- | --- |
| F014, Soyabean | 0.10 – 100.00 kU/L |
| F017, Hazelnut | 0.10 – 100.00 kU/L |
| F352, Ara h 8 PR-10, Peanut | 0.10 – 100.00 kU/L |
| F423, Ara h 2, Peanut | 0.10 – 99.00 kU/L |
| F447, Ara h 6, Peanut | 0.10 – 100.00 kU/L |
# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
# a. Traceability
The IgE calibrators are traceable to the World Health Organization (WHO) third International Standard 11/234 of Human Serum Immunoglobulin E.
# b. Kit Stability:
The stability study was conducted in accordance with CLSI EP25-A2.
*Shelf-life stability:* A real-time stability study for the NOVEOS Specific sIgE Capture Reagent Pack of F014 (Soybean), F017 (Hazelnut), F352 (Ara h 8 PR-10, Peanut), F423 (Ara h 2, Peanut), and F447 (Ara h 6, Peanut) stored at 2–8°C is ongoing using three lots. The testing was performed using two positive samples and one negative sample. In addition, an accelerated stability study was conducted with two positive samples and one negative sample using three lots of each Capture Reagent stored at 37°C to predict a shelf-life stability of 36 months when stored as 2–8°C. All other reagents required to perform specific IgE testing on the NOVEOS system are not allergen specific and the shelf-life stability of these reagents has been previously established per K182479 and K191510 (for the Fluo Beads reagent).
*On-board stability:* A study was performed on one instrument using three samples (two positive and one negative or low-level sample) using one lot of each NOVEOS Specific IgE Capture Reagent Pack of F014 (Soybean), F017 (Hazelnut), F352 (Ara h 8 PR-10, Peanut), F423 (Ara h 2, Peanut), and F447 (Ara h 6, Peanut). The testing was performed for each sample in three replicates at Day 0 and scheduled timepoints of 15, 28, and 29 days. The result supports that the allergen specific capture reagent is stable for 28 days stored on-board. The result also supports that the allergen specific Capture Reagent Packs once opened and stored at 2-8°C are stable up to 28 days. The on-board stability for all the other assay components has been previously established per K182479.
K251234 - Page 11 of 16
{11}
Open Vial Stability: Both an on-going real-time stability study and an accelerated stability study were performed using one lot NOVEOS Specific IgE Capture Reagent Pack of F014 (Soybean), F017 (Hazelnut), F352 (Ara h 8 PR-10, Peanut), F423 (Ara h 2, Peanut), and F447 (Ara h 6, Peanut). The studies evaluated opened capture reagent vials filled to 100% and 50% fill volume, stored at 37°C for up to 10 days to simulate the stress of real-time aging. The allergen specific Capture Reagent Packs are stable for 15 days once opened and stored at 2-8°C.
# 6. Detection Limit
The limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) of each NOVEOS Specific IgE assay, i.e., F014 (Soybean), F017(Hazelnut), F352 (Ara h 8 PR-10, Peanut), F423 (Ara h 2, Peanut), and F447(Ara h 6, Peanut) on the NOVEOS Immunoassay Analyzer were determined according to CLSI EP17-A2.
The LoB was determined by testing four analyte-free human serum samples in replicates of five per run, over three runs, with no more than one run per day, using two lots of each capture reagent on one NOVEOS Immunoassay Analyzer. The LoB was estimated as the 95th percentile of the 60 measurements for each of the two lots tested.
The LoD was determined by testing protocol summarized in the table below and calculated based on CLSI EP17-A2:
| NOVEOS sIgE | No. of Samples | No. of Lots | No. of Analyzers | Testing Protocol |
| --- | --- | --- | --- | --- |
| F014 | 5 | 2 | 1 | 5 samples x 5 replicates/run x 3 runs = 75 replicates/lot |
| F017 | 6 | 2 | 1 | 4 samples x 5 replicates/run x 6 runs + 2 samples x 5 replicates/run x 3 runs = 150 replicates/lot |
| F352 | 4 | 2 | 1 | 4 samples x 5 replicates/run x 3 runs = 60 replicates/lot |
| F423 | 4 | 2 | 1 | 4 samples x 5 replicates/run x 3 runs = 60 replicates/lot |
| F447 | 4 | 2 | 1 | 4 samples x 5 replicates/run x 3 runs = 60 replicates/lot |
The LoQ was determined by testing four to six low IgE samples with five replicates per run over three to six testing days using two lots of each capture reagent on one NOVEOS Immunoassay Analyzer resulting in a total of 60 to 150 replicates across all samples per lot. Data analysis was performed according to the precision profile method using within-lab precision results. The LoQ is defined as the mean value of the lowest sample which fulfills the specification for the total within-laboratory imprecision (<20%CV).
The claimed LoB, LoD, and LoQ are based on the highest value obtained from the two lots tested and are summarized in the table below:
| NOVEOS Specific IgE | LoB (kU/L) | LoD (kU/L) | LoQ (kU/L) |
| --- | --- | --- | --- |
| F014, Soybean | 0.07 | 0.09 | 0.10 |
| F017, Hazelnut | 0.07 | 0.08 | 0.10 |
| F352, Ara h 8 PR-10, Peanut | 0.07 | 0.08 | 0.10 |
| F423, Ara h 2, Peanut | 0.07 | 0.09 | 0.10 |
| F447, Ara h 6, Peanut | 0.07 | 0.08 | 0.10 |
K251234 - Page 12 of 16
{12}
# 7. Assay Cut-Off:
See clinical cut-off
# **B Comparison Studies:**
# 1. Method Comparison with Predicate Device:
Refer to clinical study section below.
# 2. Matrix Comparison:
Not applicable
# **C Clinical Studies:**
# 1. Clinical Sensitivity and Specificity
For each NOVEOS Specific IgE assay, the clinical performance was evaluated by comparing the test results from a cohort of atopic and non-atopic samples to the clinical diagnosis of allergy. The atopic samples were from different U.S. and EU vendors with a clinical history of allergy-like symptoms and identified as positive by oral food challenge (OFC) or diagnosed positive by a physician. The non-atopic samples were from donors with no reported allergy and were deemed negative by ImmunoCAP testing (<0.35 kU/L). The sIgE test results are expressed as positive or negative using a cut-off of 0.35 kU/L. The clinical sensitivity and clinical specificity for each NOVEOS Specific IgE Assay are summarized in the following tables:
| | Clinical Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Atopic | Non-atopic | Total |
| NOVEOS Specific IgE, F014, Soybean | Positive | 35 | 0 | 35 |
| | Negative | 4 | 131 | 135 |
| | Total | 39 | 131 | 170 |
| Clinical Sensitivity: 89.7% (35/39) (95% CI: 76.4 – 95.9%) | | | | |
| Clinical Specificity: 100.0% (131/131) (95% CI: 97.2 – 100.0%) | | | | |
| | Clinical Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Atopic | Non-atopic | Total |
| NOVEOS Specific IgE, F017, Hazelnut | Positive | 47 | 2 | 49 |
| | Negative | 3 | 124 | 127 |
| | Total | 50 | 126 | 176 |
| Clinical Sensitivity: 94.0% (47/50) (95% CI: 83.8 – 97.9%) | | | | |
| Clinical Specificity: 98.4% (124/126) (95% CI: 94.4 – 99.6%) | | | | |
K251234 - Page 13 of 16
{13}
| | Clinical Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Atopic | Non-atopic | Total |
| **NOVEOS Specific IgE, F352, Ara h 8 PR-10, Peanut** | **Positive** | 18 | 0 | 18 |
| | **Negative** | 36 | 132 | 168 |
| | **Total** | 54 | 132 | 186 |
| Clinical Sensitivity: 33.3% (18/54) (95% CI: 22.2 – 46.6%) | | | | |
| Clinical Specificity: 100.0% (132/132) (95% CI: 97.2 – 100.0%) | | | | |
| | Clinical Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Atopic | Non-atopic | Total |
| **NOVEOS Specific IgE, F423, Ara h 2, Peanut** | **Positive** | 42 | 0 | 43 |
| | **Negative** | 14 | 129 | 143 |
| | **Total** | 56 | 129 | 185 |
| Clinical Sensitivity: 75.0% (42/56) (95% CI: 62.3 – 84.5%) | | | | |
| Clinical Specificity: 100.0% (129/129) (95% CI: 97.1 – 100.0%) | | | | |
| | Clinical Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Atopic | Non-atopic | Total |
| **NOVEOS Specific IgE, F447, Ara h 6, Peanut** | **Positive** | 40 | 0 | 40 |
| | **Negative** | 16 | 126 | 142 |
| | **Total** | 56 | 126 | 182 |
| Clinical Sensitivity: 71.4% (40/56) (95% CI: 58.5 – 81.6%) | | | | |
| Clinical Specificity: 100.0% (126/126) (95% CI: 97.0 – 100.0%) | | | | |
The following literature that supports the observed sensitivity of the NOVEOS Specific IgE assays F352 (Ara h 8 PR-10, Peanut), F423 (Ara h 2, Peanut), and F447 (Ara h 6, Peanut) are presented in the table below.
| NOVEOS sIgE | Literature Citation(s) | Prevalence | Reported Performance |
| --- | --- | --- | --- |
| F422, Ara h 1, Peanut Sensitivity: 33.3% Specificity: 100.0% | Warren C., *et al.* J Allergy Clin Immunol. 2021;147(6):2263-2270.e5 | 1.3% - 2.9% of U.S. adults have a reported peanut allergy | not reported |
| | Calamelli E., *et al.* BioMed Research International. 2013, Article ID 170452, | Among children sensitized to peanut, only 35% had IgE reactivity to Ara h 8. | |
| | Vereda A., *et al.* J Allergy Clin Immunol. 2011;127:603-7. | Among patients with peanut allergy, only 25% had IgE reactivity to Ara h 8. | |
K251234 - Page 14 of 16
{14}
| NOVEOS sIgE | Literature Citation(s) | Prevalence | Reported Performance |
| --- | --- | --- | --- |
| F423, Ara h 2, Peanut Sensitivity: 75.0% Specificity: 100.0% | Warren C., *et al.* J Allergy Clin Immunol. 2021;147(6):2263-2270.e5 | 1.3% - 2.9% of U.S. adults have a reported peanut allergy | |
| | Calamelli E., *et al.* BioMed Research International. 2013, Article ID 170452, | Among children sensitized to peanut, only 27% had IgE reactivity to Ara h 2. | |
| | Vereda A., *et al.* J Allergy Clin Immunol. 2011;127:603-7. | Among patients with peanut allergy, only 64% had IgE reactivity to Ara h 2. | |
| F423, Ara h 2, Peanut Sensitivity: 75.0% Specificity: 100.0% | Warren C., *et al.* J Allergy Clin Immunol. 2021;147(6):2263-2270.e5 | 1.3% - 2.9% of U.S. adults have a reported peanut allergy | |
| | Calamelli E., *et al.* BioMed Research International. 2013, Article ID 170452, | Among children sensitized to peanut, only 27% had IgE reactivity to Ara h 2. | |
| | Vereda A., *et al.* J Allergy Clin Immunol. 2011;127:603-7. | Among patients with peanut allergy, only 64% had IgE reactivity to Ara h 2. | |
| F447, Ara h 6, Peanut Sensitivity: 71.4% Specificity: 100.0% | Warren C., *et al.* J Allergy Clin Immunol. 2021;147(6):2263-2270.e5 | 1.3% - 2.9% of U.S. adults have a reported peanut allergy | |
| | Kaur N., *et al.* J Allergy Clin Immunol Pract. 2021;9(1):245-53. | Among children referred for peanut OFC, 37% had IgE reactivity to Ara h 6. Among children found to be allergic to peanut by OFC, 75% had IgE reactivity to Ara h 6. | |
# 2. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not Applicable
# **D Clinical Cut-Off**
The clinical cut-off of the assay is 0.35 kU/L. The interpretation of results of the assay uses the following classification system:
K251234 - Page 15 of 16
{15}
| Class | Concentration (kU/L) | Interpretation |
| --- | --- | --- |
| 0 | <0.35 | Negative |
| I | 0.35 to <0.70 | Positive with increasing sIgE concentration |
| II | 0.70 to <3.50 | |
| III | 3.50 to <17.50 | |
| IV | 17.50 to <50.00 | |
| V | 50.00 to <100.00 | |
| VI | ≥100 | |
### E Expected Values/Reference Range:
The expected value is <0.35 kU/L (negative) for a specific allergen in an apparently healthy (non-atopic) individual. Each laboratory should establish its own expected value/reference range.
The reference range of NOVEOS Specific IgE assay for F014 (Soybean), F017 (Hazelnut), F352 (Ara h 8, PR-10, Peanut), F423 (Ara h 2, Peanut), and F447 (Ara h 6, Peanut) in the normal population was evaluated based on the data collected in clinical studies which tested samples from apparently healthy subjects including: 131 non-atopic samples for F014; 126 non-atopic samples for F017; 132 non-atopic samples for F352; 129 non-atopic samples for F423; and 126 non-atopic samples for F447. Except for two out of 126 samples evaluated with F017 (Hazelnut), all samples tested with the NOVEOS Specific IgE assays were below 0.35 kU/L.
### VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
### IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K251234 - Page 16 of 16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.