K261021 · Winter Innovations, Inc. · GAT · Jul 1, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261021
Device Name
EasyWhip Family
Applicant
Winter Innovations, Inc.
Product Code
GAT · General, Plastic Surgery
Decision Date
Jul 1, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic, PCCP
Indications for Use
The EasyWhip Family is indicated for use in approximation and/or ligation of soft tissues, including the use of allograft tissue for orthopedic surgeries.
Device Story
EasyWhip Family is a non-absorbable surgical suture system featuring a specialized two-part needle; designed for orthopedic soft tissue stitching. System consists of a needle with a receiver and an insert that temporarily connects to the needle; allows both suture tails to pass through tissue via a single needle hole. Enables creation of whip, WhipLock, and hybrid stitch patterns. Used by surgeons in clinical settings to standardize and simplify manual suturing; facilitates controlled, accurate stitch placement. Benefits include reduced variation in stitch patterns and improved efficiency in tissue approximation. Provided sterile for single use.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including suture tensile strength (USP <861>), needle attachment strength (USP <871>), and diameter (USP <881>), alongside biocompatibility testing per ISO 10993-1.
Technological Characteristics
Two-part needle system (stainless steel) and braided UHMWPE non-absorbable suture. Needle components: 300/400 series stainless steel. Suture: USP sizes 0-3&4, undyed or dyed (chromium-cobalt-aluminum oxide, D&C Black No. 4). Sterilization: Ethylene Oxide. Meets USP requirements for tensile strength and needle attachment; diameter deviates from USP standards. MR Safe.
Indications for Use
Indicated for approximation and/or ligation of soft tissues, including allograft tissue, in orthopedic surgeries (e.g., Achilles, ACL, biceps, peroneal).
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
July 01, 2026
Winter Innovations, Inc.
Lia Winter
CEO
2450 Ej Chapman Dr., Suite 114
Knoxville, TN 37996
Re: K261021
Trade/Device Name: EasyWhip Family
Regulation Number: 21 CFR 878.5000
Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture
Regulatory Class: Class II
Product Code: GAT, GAB
Dated: May 29, 2026
Received: May 29, 2026
Dear Lia Winter:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261021 - Lia Winter
Page 2
premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}
K261021 - Lia Winter
Page 3
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N.
LAMICHHANE -S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261021
Device Name
EasyWhip Family
Indications for Use (Describe)
The EasyWhip Family is indicated for use in approximation and/or ligation of soft tissues, including the use of allograft tissue for orthopedic surgeries.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# 510(k) SUMMARY
### **Sponsor/Applicant**
Winter Innovations Inc.
2450 EJ Chapman Drive, Suite 114
Knoxville, TN 37996
(865) 205-5728
### **Contact Person:**
Chief Executive Officer
**Date of Preparation:** June 30, 2026
**Trade Name:** EasyWhip® Family
**Common Name:** non-absorbable surgical suture and two-part needle system
**Product Code:** GAT
**Classification Name:** Suture, Non-Absorbable, Synthetic, Polyethylene
**Regulation Number:** 21 CFR § 878.5000
**Device Class:** II
**Review Panel:** General and Plastic Surgery
**Predicate Device:** K210675, EasyWhip®
### **Device Description & Intended Use**
The EasyWhip® Family is a non-absorbable surgical suture system with a specialized two-part needle intended for soft tissue stitching in orthopedic procedures (Achilles, ACL, biceps, peroneal, etc.). The two-part needle system is designed to facilitate easy, fast, and accurate stitch placement with less variation. It enables stitch patterns that capture both sides of the tissue using a single needle pass.
The devices can be used to create multiple stitch patterns, including a whip stitch, a WhipLock™ stitch (which combines a whip stitch with a locking stitch similar to a Krackow stitch), and hybrid stitch configurations according to individual user needs and preferences. The design simplifies and standardizes existing manual suturing techniques by facilitating controlled stitch placement as a convenience to surgeons.
Devices are provided sterile and are intended for single use only. The devices meet applicable USP requirements for non-absorbable surgical sutures, except for diameter.
### **Indications for Use**
The EasyWhip® Family is indicated for use in approximation and/or ligation of soft tissues, including the use of allograft tissue for orthopedic surgeries.
Winter Innovations, Inc.
EasyWhip Family 510(k)
1
510(k) Summary
{5}
# Technological Characteristics
The two-part needle system consists of (1) a Needle with a sharp tip and a receiver at the back end, and (2) an Insert that slides into the receiver of the Needle to temporarily connect the two components. The Needle and Insert are attached to opposite ends of a length of suture. When disconnected, the suture is straight; when connected, the suture forms a loop that allows both suture tails to pass through tissue with a single needle hole.
The suture is braided, non-absorbable ultra-high molecular weight polyethylene (UHMWPE) that is either undyed or dyed using color additives that are approved and comply with applicable FDA requirements of 21 CFR 70.5(c), and it is equivalent to FDA-cleared suture material. The needle components are fabricated from stainless steel and are comparable to materials used in predicate devices.
At a high level, the subject and predicate devices share the same fundamental technological characteristics, with only minor differences in device configurations as summarized in the table below.
| Technological Characteristic | EasyWhip® Family (Subject Device) | EasyWhip® (Predicate - K210675) |
| --- | --- | --- |
| Product Code, Regulation | GAT, 21 CFR 878.5000 | Same |
| Indications for Use | ... indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. | Same |
| Principles of Operation | Uses a two-part needle attached to suture to place stitches in soft tissue for approximation and/or ligation. | Same |
| Technical Characteristics | Two-part needle attached to ends of suture. Two-parts temporarily connectable. | Same |
| Patient Contact Duration | Mixed: Needle – limited (≤24 hours), Suture – permanent (>30 days) | Same |
| Biocompatibility | Meets ISO 10993 requirements | Same |
| Sterilization | Sterile, Ethylene Oxide | Same |
| Device Materials | UHMWPE non-absorbable suture, stainless steel needle components, silicone, cyanoacrylate | Same |
| Needle Materials | Stainless Steel (300 and 400 series) | Stainless Steel (300 series) |
| Two-Part Needle Configuration | Needle components attached to both ends of suture | Same |
| Needle Length | 1.78" – 2.56" | 2.04" |
| Needle Geometry | Straight – Curved | Straight |
| Needle Body | Solid | Hollow |
| Needle Tip | Taper-Point | Tri-beveled |
| Suture Size | USP Size 0 – Size 3&4, deviate for diameter | USP Size 2, oversize diameter |
| Suture Braid | Round – Flat | Round |
| Suture Length | 30" – 40" total (15" – 20" working) | 40" total (20" working) |
| Suture Dye | chromium-cobalt-aluminum oxide, D&C Black No. 4 | D&C Black No. 4 |
| Suture Absorption | Non-Absorbable | Same |
| Performance | Meets USP surgical suture requirements, except for diameter | Same |
| Suture Diameter USP <861> | Round: 13.8 mils – 33 mils deviate from USP diameter Flat: 1.35 – 1.7 mm differs from USP requirements | 19.7-24.8 mils Oversized for USP |
| Sutures-Needle Attachment USP <871> | Meets USP | Same |
| Tensile Strength USP <881> | Meets USP | Same |
Winter Innovations, Inc.
EasyWhip Family 510(k)
2
510(k) Summary
{6}
## Non-clinical Testing
The EasyWhip® Family meets the requirements of the *Class II Special Controls Guidance: Surgical Sutures; Guidance for Industry and FDA* (June 2003).
Performance testing was conducted or referenced to support the safety and effectiveness at time zero and over the duration of the labeled shelf life, including evaluation of suture tensile strength (USP <861>), needle attachment strength (USP <871>), and diameter (USP <881>). Deviations in diameter are identified in labeling.
The suture component has been evaluated and demonstrated to be “MR Safe,” posing no known hazards in MR environments.
Biocompatibility safety testing was performed in accordance with guidance document *Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and Testing within a Risk Management Process* (September 2023).
Testing supports that the EasyWhip® Family is as safe and effective as the predicate device.
## Clinical Testing
This submission did not rely on clinical testing to demonstrate device performance and substantial equivalence.
## Predetermined Change Control Plan (PCCP)
This submission includes a Predetermined Change Control Plan (PCCP) for the EasyWhip® Family. The PCCP establishes a framework for managing pre-planned modifications:
| Planned Modification | Test Methods and Validation Activities | Communication to Users, as needed |
| --- | --- | --- |
| Size Changes size range expanded for additional suture sizes ranging from 5-0 to 5 (5-0, 4-0, 3-0, 2-0, 0, 2, 3&4, and 5) | Suture diameter conformance to USP <861>, suture tensile strength conformance to USP <881>, and suture-needle attachment strength testing to USP <871> will be performed or verified, as applicable, in accordance with the Modification Protocol to confirm that the device meets applicable physical and mechanical performance requirements. | Each device configuration will have device-specific labeling according to Surgical Sutures - Class II Special Controls Guidance Document for Industry and FDA Staff (June 3, 2003), which will serve as the primary mechanism for communication of the implemented modification to users. |
| Color Changes FDA-compliant and cleared color additives Green Pigment 7 and D&C Violet No. 2 | Biocompatibility evaluation, including cytotoxicity testing in accordance with ISO 10993-5 and FDA guidance, will be performed in accordance with the Modification Protocol to confirm that the modified device meets applicable cytotoxicity requirements. | Each device configuration will have device-specific labeling according to Surgical Sutures - Class II Special Controls Guidance Document for Industry and FDA Staff (June 3, 2003), which will serve as the primary mechanism for communication of the implemented modification to users. |
Winter Innovations, Inc.
EasyWhip Family 510(k)
3
510(k) Summary
{7}
| Combined Size and Color Changes (outlined above) | Suture diameter conformance to USP <861>, suture tensile strength conformance to USP <881>, and suture-needle attachment strength testing to USP <871> will be performed or verified, as applicable, in accordance with the Modification Protocol to confirm that the device meets applicable physical and mechanical performance requirements. Biocompatibility evaluation, including cytotoxicity testing in accordance with ISO 10993-5 and FDA guidance, will be performed in accordance with the Modification Protocol to confirm that the modified device meets applicable cytotoxicity requirements. | Each device configuration will have device-specific labeling according to Surgical Sutures - Class II Special Controls Guidance Document for Industry and FDA Staff (June 3, 2003), which will serve as the primary mechanism for communication of the implemented modification to users. |
| --- | --- | --- |
The PCCP describes the testing methods, validation activities, and acceptance criteria to be performed prior to implementation of modifications while ensuring the EasyWhip® Family continues to meet applicable performance and safety requirements and remains as safe and effective as the predicate device. The suture component of the device, including all sizes and colors in the PCCP, has been previously cleared by FDA and is obtained directly from the corresponding 510(k) holder with a letter of reference for performance and biocompatibility testing data. The performance evaluation activities referenced from the suture 510(k) clearance include suture diameter verification in accordance with USP <861> and suture tensile strength verification in accordance with USP <881>. Please note that the scope of the PCCP will be limited to attaching a needle to the previously cleared suture sizes (5-0, 4-0, 3-0, 2-0, 0, 2, 3&4, and 5) for the same indications mentioned above, and using suture colors (Green Pigment 7 and D&C Violet No. 2) that have also been previously cleared by the Agency. No new suture sizes or colors that the FDA has not previously reviewed will be added under the PCCP framework.
Users will be informed of modifications implemented under the PCCP through updated labeling specific to the applicable device modifications.
The PCCP includes a description of modifications, a modification protocol, traceability from modifications to the associated verification and validation activities, and an impact assessment.
## Conclusion
The EasyWhip® Family is substantially equivalent to the legally marketed predicate EasyWhip® K210675. Any differences between the EasyWhip® Family and the predicate were evaluated with non-clinical testing, which demonstrated substantial equivalence to the predicate.
Winter Innovations, Inc.
EasyWhip Family 510(k)
4
510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.