MobileLink Acetabular Cup System – 40mm ID Inserts (various)
K261009 · Waldemar Link GmbH & Co. KG · LZO · Jul 6, 2026 · Orthopedic
Device Facts
Record ID
K261009
Device Name
MobileLink Acetabular Cup System – 40mm ID Inserts (various)
Applicant
Waldemar Link GmbH & Co. KG
Product Code
LZO · Orthopedic
Decision Date
Jul 6, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
General Indications: – Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur which cannot be treated by conservative or osteosynthetic procedures Indications: – Primary and secondary osteoarthritis – Rheumatoid arthritis – Correction of functional deformities – Avascular necrosis – Femoral neck fractures – Revision Surgery Dual Mobility Insert (in addition to the indications and general indications): – Dislocation risk The MobileLink Acetabular Shells are intended for uncemented fixation The TrabecuLink Augments are intended for uncemented fixation to the bone and for cemented fixation to the mating shell. Additional indications specific to the TrabecuLink Augments: – Primary or revision surgery for hip replacement as a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies.
Device Story
MobileLink Acetabular Cup System expanded to include 40mm ID inserts for larger acetabular shells (OD 54-80mm). Inserts manufactured from highly crosslinked polyethylene (with or without Vitamin E); feature standard and 5mm shoulder variants. Fixation achieved via snap mechanism, conical clamping, and rim notches for rotational stability. Used by orthopedic surgeons in clinical settings for total hip replacement. Output is a prosthetic hip component; restores joint function and stability; reduces dislocation risk in dual mobility configurations.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing: acetabular construct disassembly (ASTM F1820), range of motion, wear testing (ISO 14242, ASTM F1877), impingement testing (ASTM F2582), and biocompatibility evaluation.
Technological Characteristics
Highly crosslinked polyethylene (with/without Vitamin E). Modular acetabular insert; snap/conical clamping fixation. Dimensions: 40mm ID for 54-80mm OD shells. Standards: ASTM F1820 (disassembly), ISO 14242/ASTM F1877 (wear), ASTM F2582 (impingement).
Indications for Use
Indicated for adults (22+ years) with mobility-limiting hip/proximal femur disease, fractures, or defects requiring hip replacement (osteoarthritis, rheumatoid arthritis, functional deformities, avascular necrosis, femoral neck fractures, revision surgery). Dual mobility inserts indicated for patients at risk of dislocation. TrabecuLink augments indicated for segmental acetabular deficiencies.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 6, 2026
Waldemar Link GmbH & Co. KG
% Luisa Moncada
Regulatory Affairs
LinkBio Corp.
69 King St.
Dover, New Jersey 07801
Re: K261009
Trade/Device Name: MobileLink Acetabular Cup System - 40mm ID Inserts (various)
Regulation Number: 21 CFR 888.3353
Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis
Regulatory Class: Class II
Product Code: LZO, MEH, OQG, LPH
Dated: June 8, 2026
Received: June 9, 2026
Dear Luisa Moncada:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261009 - Luisa Moncada
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261009 - Luisa Moncada
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# RYAN TROMBETTA -S
For: Limin Sun, Ph.D.
Assistant Director
DHT6A: Division of Joint
Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
K261009 Page 1 of 1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261009 | ? |
| Please provide the device trade name(s). | | ? |
| MobileLink Acetabular Cup System – 40mm ID Inserts (various) | | |
| Please provide your Indications for Use below. | | ? |
| General Indications: – Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur which cannot be treated by conservative or osteosynthetic procedures Indications: – Primary and secondary osteoarthritis – Rheumatoid arthritis – Correction of functional deformities – Avascular necrosis – Femoral neck fractures – Revision Surgery Dual Mobility Insert (in addition to the indications and general indications): – Dislocation risk The MobileLink Acetabular Shells are intended for uncemented fixation The TrabecuLink Augments are intended for uncemented fixation to the bone and for cemented fixation to the mating shell. Additional indications specific to the TrabecuLink Augments: – Primary or revision surgery for hip replacement as a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
K261009
# 510(k) Summary
510(k)
Submitter:
Waldemar Link GmbH & Co. KG
Barkhausenweg 10
22339 Hamburg, Germany
Phone: +49-40-539950
Facility Registration: 300317426
Contact Person:
LinkBio Corp.
Luisa Moncada (Regulatory Affairs)
69 King Street
Dover, NJ 07801
Phone: 973-625-1333 x 103
Fax: 973-625-4445
E-Mail: l.moncada@linkbio.com
Date Prepared:
2026-07-02
Trade Name:
MobileLink Acetabular Cup System – 40mm ID Inserts (various)
Common Name:
Total hip replacement system
Classification Name:
Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.; 21 CFR §888.3353, product code LZO
Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.; 21 CFR §888.3353, product code MEH
Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis. 21 CFR §888.3358, product code OQG
Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis. 21 CFR §888.3358, product code LPH
Classification and Panel:
Class II, Orthopedic / 87
Predicate Devices:
| Subject Device Components | Predicate Devices | 510(k) Number |
| --- | --- | --- |
| 40mm Inserts | 1. MobileLink Actabular Cup System [Primary Predicate] | K182321 |
| | 2. MobileLink Actabular Cup System – Line Extension [Additional Predicate] | K241636 |
| Subject Device Components | Reference Devices | 510(k) Number |
| --- | --- | --- |
| 40mm Inserts | 1. Zimmer Biomet G7 Acetabular System [Reference] | K140669 |
Reason for Submission
New acetabular inserts that can accomodate 40mm heads in the larger acetabular shell OD sizes.
Device Description:
The MobileLink Acetabular Cup System is a versatile cup system, designed to provide several options for surgeons and patients within one system. This 510k adds acetabular insert components to the MobileLink Acetabular Cup System previously cleared in K182321, K192559, K200607, K222066, and K241636.
40mm Inserts: The 40mm Inserts are manufactured from highly crosslinked with Vitamin E (E-Dur) and highly crosslinked without Vitamin E (X-LINKed)
Page 1 of 2
{5}
K261009
polyethylene material. The designs include standard and 5mm shoulder variants which currently exist in the MobileLink Acetabular Cup System cleared in K182321 and K241636. The 40mm inserts are available only for the larger acetabular shell sizes (OD sizes 54-80mm). The inserts are fixed using a clamping system consisting of a combination of snap mechanism and conical clamping. Notches (tabs/recesses) at the rim further contribute to the rotational stability of the PE inserts.
# **Intended Use:**
General Indications:
- Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur which cannot be treated by conservative or osteosynthetic procedures
Indications:
- Primary and secondary osteoarthritis
- Rheumatoid arthritis
- Correction of functional deformities
- Avascular necrosis
- Femoral neck fractures
- Revision Surgery
Dual Mobility Insert (in addition to the indications and general indications):
- Dislocation risk
The MobileLink Acetabular Shells are intended for uncemented fixation
The TrabecuLink Augments are intended for uncemented fixation to the bone and for cemented fixation to the mating shell.
Additional indications specific to the TrabecuLink Augments:
- Primary or revision surgery for hip replacement as a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies.
# **Comparison to Predicate Device:**
The subject MobileLink 40mm Inserts are substantially equivalent to the previously cleared MobileLink E-Dur and X-LINKed inserts (K182321 and K241636) in terms of design, intended use, manufacturing method, sterilization, and packaging, and have an inner diameter similar to the reference device.
# **Performance Testing:**
Non-clinical performance testing and analysis were provided, including:
- Evaluation of acetabular construct disassembly testing (Push-out, lever-out, torque out for the acetabular modular connections) conducted per ASTM F1820
- Range of Motion evaluation
- Wear testing under standard conditions per ISO 14242 and ASTM F1877
- Evaluation of impingement testing conducted per F2582
- Biocompatibility evaluation
The results of non-clinical performance testing demonstrate Substantial Equivalence to the predicate.
# **Clinical Testing:**
Clinical performance testing was not required to demonstrate the substantial equivalence of this device.
# **Conclusion:**
The subject MobileLink 40mm Inserts are substantially equivalent to the predicate devices identified in this premarket notification.
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.