Codman® Distal Peritoneal Catheter, Barium Open End (82-1682); Codman® HOLTER® Peritoneal Catheter (SALMON Design) (82-1684)
K261004 · Integra Lifesciences Production Corporation · JXG · Jun 23, 2026 · Neurology
Device Facts
Record ID
K261004
Device Name
Codman® Distal Peritoneal Catheter, Barium Open End (82-1682); Codman® HOLTER® Peritoneal Catheter (SALMON Design) (82-1684)
Applicant
Integra Lifesciences Production Corporation
Product Code
JXG · Neurology
Decision Date
Jun 23, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Codman® Distal Peritoneal Catheter, Barium Open End is indicated for use with a proximal valve in the treatment of hydrocephalus where ventriculoperitoneal shunting is the procedure of choice. The HOLTER Peritoneal Catheter (SALMON Design) is indicated for use with a HOLTER Valve in the treatment of hydrocephalus where the ventriculoperitoneal shunting technique is the procedure of choice.
Device Story
Sterile, single-use, radiopaque silicone peritoneal drainage catheters; used as part of CSF shunting system. Catheters attach to distal end of proximal valve; distal end implanted into patient's peritoneal cavity. Excess CSF regulated by valve drains through catheter into peritoneal cavity for reabsorption into bloodstream. Used by neurosurgeons in clinical settings. Subject device updates include MR Conditional labeling, updated symbols per ISO 15223-1, and addition of Patient Implant Card/Patient Information Leaflet per FDA guidance.
Clinical Evidence
No clinical studies were required. Substantial equivalence supported by bench testing, including MRI safety testing per ASTM F2052, ASTM F2213, ASTM F2182, and ASTM F2119.
Technological Characteristics
Radiopaque silicone tubing; sterile, single-use. MR Conditional. Complies with ISO 15223-1 for symbols. Bench testing performed per ASTM F2052, ASTM F2213, ASTM F2182, and ASTM F2119.
Indications for Use
Indicated for patients requiring ventriculoperitoneal shunting for the treatment of hydrocephalus.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
Codman® Distal Peritoneal Catheter, Barium Open End (Preamendment)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 23, 2026
Integra Lifesciences Production Corporation
Leigh Anne Diener
Regulatory Specialist
11 Cabot Blvd.
Mansfield, Massachusetts 02048
Re: K261004
Trade/Device Name: Codman® Distal Peritoneal Catheter, Barium Open End (82-1682);
Codman® HOLTER® Peritoneal Catheter (SALMON Design) (82-1684)
Regulation Number: 21 CFR 882.5550
Regulation Name: Central Nervous System Fluid Shunt And Components
Regulatory Class: Class II
Product Code: JXG
Dated: March 26, 2026
Received: March 26, 2026
Dear Leigh Anne Diener:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261004 - Leigh Anne Diener
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261004 - Leigh Anne Diener
Page 3
Sincerely,
YEN-CHIH
LIN -S
Digitally signed by YEN-
CHIH LIN -S
Date: 2026.06.23 15:44:13
-04'00'
for Jaime Raben, Ph.D.,
Division Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K261004 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| Codman® Distal Peritoneal Catheter, Barium Open End (82-1682); Codman® HOLTER® Peritoneal Catheter (SALMON Design) (82-1684) | | | |
| Please provide your Indications for Use below. | | | ? |
| Codman® Distal Peritoneal Catheter, Barium Open End is indicated for use with a proximal valve in the treatment of hydrocephalus where ventriculoperitoneal shunting is the procedure of choice. The HOLTER Peritoneal Catheter (SALMON Design) is indicated for use with a HOLTER Valve in the treatment of hydrocephalus where the ventriculoperitoneal shunting technique is the procedure of choice. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☐ Adults (22 years old and greater) | | ? |
{4}
K261004 Page 1 of 3
## **510(k) Summary**
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 807.92(a) (1) Submitter Information | |
| --- | --- |
| Name and Address | Integra LifeSciences Production Corporation 11 Cabot Boulevard Mansfield, MA 02048 |
| Telephone number | +1 781-971-5601 |
| Primary Contact | Leigh Anne Diener |
| Date Summary Prepared | January 28, 2026 |
| 807.92(a) (2) Name of Device | |
| Trade or Proprietary Name | Codman® Distal Peritoneal Catheter, Barium Open End Codman® HOLTER® Peritoneal Catheter (SALMON Design) |
| Common Name | Hydrocephalus Shunt System |
| Classification Name | Central Nervous System Fluid Shunt and Components (21 CFR 882.5550) |
| Device Class | II |
| Product Code | JXG |
| 807.92(a) (3) Predicate Information | |
| Predicate Device | Codman® Distal Peritoneal Catheter, Barium Open End - Preamendment Codman® HOLTER® Peritoneal Catheter (SALMON Design) - Preamendment |
| 807.92(a) (4) Device Description | |
| The Codman® Distal Peritoneal Catheter, Barium Open End and Codman® HOLTER® Peritoneal Catheter (SALMON Design) are sterile, single use, peritoneal drainage catheters that are used as part of a CSF shunting system to treat hydrocephalus. Both types of catheters may be attached to the distal end of a valve shunting system and the distal end of the catheter is then implanted into the Peritoneal cavity of the patient. The excessive CSF which is regulated by the valve is then drained through the peritoneal drainage catheter into the peritoneal cavity where it is reabsorbed into the bloodstream. The Codman® Distal Peritoneal Catheter, Barium Open End and Codman® HOLTER® Peritoneal Catheter (SALMON Design) are made of radiopaque silicone tubing. | |
| 807.92(a) (5) Indications for Use | |
| Codman® Distal Peritoneal Catheter, Barium Open End is indicated for use with a proximal valve in the treatment of hydrocephalus where ventriculoperitoneal shunting is the procedure of choice. The HOLTER Peritoneal Catheter (SALMON Design) is indicated for use with a HOLTER Valve in the treatment of hydrocephalus where the ventriculoperitoneal shunting technique is the procedure of choice. | |
| 807.92(a) (6) Technological Characteristics Compared to Predicate | |
| The proposed Codman® Distal Peritoneal Catheter, Barium Open End and Codman® HOLTER® Peritoneal Catheter (SALMON Design) have the same intended use, design, materials, sterility, and principles of operation as the predicate devices. The proposed changes to labeling do not impact the technological characteristics of the devices. The changes do not raise any new questions of safety and effectiveness. | |
{5}
K261004 Page 2 of 3
| Document, Section (if applicable) | Update | Rationale |
| --- | --- | --- |
| IFU, Throughout Document | Formatting, clerical changes, and addition of new sections. | Proposed updates drive additional consistency across and within global product IFUs and in support of EU Medical Device Regulation compliance. |
| IFU and Labeling, Magnetic Resonance Imaging (MRI) Safety Information | MRI safety information added and include an “MR Conditional” claim. | Testing provided in submission to support an “MR Conditional” claim for the catheters. |
| IFU, Patient Implant Card, and Patient Information Leaflet | Addition of instructions for healthcare describing how to complete the card and additional information on the card and leaflet that should be advised to the patient. PIC and PIL are new documents provided to customer in packaging. | Patient Implant Card and Patient Information Leaflet are to be provided to patient per FDA guidance *Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment*, issued on October 10, 2023, as the catheters are MR conditional devices. |
| IFU and Labeling, Symbols | Symbols have been updated to include harmonized symbols per ISO 15223-1. | Symbols now align with the latest version of ISO 15223-1. |
#### 807.92(b) 1-2: Summary of Nonclinical and Clinical Testing Performed
The following performance testing has been conducted in support of the substantial equivalence determination. The testing utilized well-established methods, including those from FDA consensus standards. All testing was performed on production equivalent devices.
| Performance Bench Test Results | |
| --- | --- |
| Test | Conclusion |
| MRI Safety Testing per ASTM F2052, ASTM F2213, ASTM F2182 and ASTM FF2119 | Pass |
#### Sterilization/Cleaning
There are no changes in sterility as a result of the proposed changes.
#### Shelf Life
There are no changes in shelf life as a result of the proposed changes.
#### Animal Studies
No animal studies were required as appropriate verification of the subject devices was achieved based on the comparison to the predicate devices and from the results of the bench testing and analysis.
#### Clinical Studies
No clinical studies were required as appropriate verification of the subject devices was achieved based on the comparison to the predicate devices and from the results of the bench testing and analysis.
{6}
K261004 Page 3 of 3
# **807.92(b) (3) Conclusion**
Based upon the intended use, design, comparison to the predicate device, and testing performed, Integra LifeSciences believes that the proposed modifications to the Codman® Distal Peritoneal Catheter, Barium Open End and Codman® HOLTER® Peritoneal Catheter (SALMON Design) do not raise any new questions of safety and effectiveness, and is therefore, substantially equivalent to the predicate Codman® Distal Peritoneal Catheter, Barium Open End and Codman® HOLTER® Peritoneal Catheter (SALMON Design).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.