K260998 · Ceribell, Inc. · OMB · Jun 24, 2026 · Neurology
Device Facts
Record ID
K260998
Device Name
Ceribell Neurology Review Software
Applicant
Ceribell, Inc.
Product Code
OMB · Neurology
Decision Date
Jun 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K260998 · Jun 24, 2026
Ceribell Neurology Review Software
Ceribell, Inc.
Retrospective clinical EEG recordings from hospital-based continuous EEG monitoring
Retrospective clinical EEG datasets were used to validate the performance of the Artifact Reduction and Epileptiform Abnormality Detection algorithms against an expert-reviewed reference standard.
1,362 EEG recordings from patients 1 year of age and older
3 (expert neurologists)
Indications for Use
The Ceribell Neurology Review Software is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EEG. This device is intended to be used by qualified clinical practitioners who will exercise professional judgment in using the information. The Artifact Reduction component is intended to reduce artifact in EEG recordings. The Artifact Reduction component does not remove the entire artifact signal and may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified clinical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms. The Epileptiform Abnormality Detection component is intended to mark previously acquired sections of the patient's EEG recordings that may correspond to epileptiform abnormalities in order to assist qualified clinical practitioners in the assessment of EEG traces. The Epileptiform Abnormality Detection component is intended to be used in patients at least one year old.
Device Story
Software for review, monitoring, and analysis of EEG recordings; processes data from external EEG devices. Features two primary components: Artifact Reduction and Epileptiform Abnormality Detection. Artifact Reduction module mitigates noise (EKG, EMG, eye blinks) to improve signal quality; Epileptiform Abnormality Detection module identifies and marks potential epileptiform activity. Used in professional healthcare facilities by neurologists and qualified clinical practitioners. Output provided as marked EEG traces for clinician review; assists in clinical assessment of brain activity. Clinicians must exercise professional judgment and refer to original waveforms for final diagnosis. Benefits include improved efficiency in EEG review and identification of potential abnormalities.
Clinical Evidence
Bench and clinical performance validation. Artifact Reduction validated on 120 EEG recordings; non-inferiority demonstrated against Persyst algorithm (SNR improvement 1.12 dB, distortion -2.40%). Epileptiform Abnormality Detection validated on 1,362 patients (age 1+); compared against expert neurologist consensus. Results: PPA 98.29% (Frequent), 98.95% (Abundant), 94.37% (Continuous); False Positive rates 0.286, 0.142, and 0.052 FP/hr respectively. All metrics met pre-specified acceptance criteria.
Technological Characteristics
Software-based EEG analysis tool. Operates on standard EEG data inputs. Utilizes signal processing for artifact reduction and automated detection algorithms for epileptiform abnormalities. Designed for professional healthcare facility use. Compliant with software verification and validation standards per FDA guidance.
Indications for Use
Indicated for review, monitoring, and analysis of EEG recordings in patients aged 1 year and older to aid neurologists in assessment. Includes artifact reduction and epileptiform abnormality detection components. Requires use by qualified clinical practitioners.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 24, 2026
Ceribell, Inc.
Raymond Woo
CTO
360 N. Pastoria Ave.
Sunnyvale, California 94085
Re: K260998
Trade/Device Name: Ceribell Neurology Review Software
Regulation Number: 21 CFR 882.1400
Regulation Name: Electroencephalograph
Regulatory Class: Class II
Product Code: OMB, OLT
Dated: March 26, 2026
Received: March 26, 2026
Dear Raymond Woo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260998 - Raymond Woo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K260998 - Raymond Woo
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Patrick Antkowiak -S
for
Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K260998
Device Name
Ceribell Neurology Review Software
Indications for Use (Describe)
The Ceribell Neurology Review Software is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EEG. This device is intended to be used by qualified clinical practitioners who will exercise professional judgment in using the information.
The Artifact Reduction component is intended to reduce artifact in EEG recordings. The Artifact Reduction component does not remove the entire artifact signal and may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified clinical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms.
The Epileptiform Abnormality Detection component is intended to mark previously acquired sections of the patient's EEG recordings that may correspond to epileptiform abnormalities in order to assist qualified clinical practitioners in the assessment of EEG traces. The Epileptiform Abnormality Detection component is intended to be used in patients at least one year old.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
ceribell®
# 510(k) Summary – K260998
1. SUBMITTER
Ceribell, Inc.
360 North Pastoria Avenue
Sunnyvale, California 94085
Contact Person: Raymond Woo, PhD
Chief Technical Officer
Telephone: (800) 436-0826
E-mail: ray@ceribell.com
Alternative Contact: Tom McDougal
Director, Regulatory Affairs
E-mail: tom.mcdougal@ceribell.com
Date Prepared: March 23, 2026
2. DEVICE
Trade Name: Ceribell Neurology Review Software
Common Name: Automatic Event Detection Software for Full-Montage Electroencephalograph
Classification: Electroencephalograph (21 CFR 882.1400)
Device Class: II
Product Code: OMB, OLT
3. PREDICATE DEVICES
Primary: encevis, K240993
Secondary: autoSCORE, K243743
4. DEVICE DESCRIPTION
The Ceribell Neurology Review Software is intended for analyzing EEG data acquired from legally marketed EEG devices to aid qualified clinical practitioners in the review of EEG data. In particular, the subject device is intended to reduce artifact and identify sections of EEG that may correspond to epileptiform abnormalities.
1
{5}
ceribell®
## 5. INDICATIONS FOR USE
The Ceribell Neurology Review Software is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EEG. This device is intended to be used by qualified clinical practitioners who will exercise professional judgment in using the information.
The Artifact Reduction component is intended to reduce artifact in EEG recordings. The Artifact Reduction component does not remove the entire artifact signal and may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified clinical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms.
The Epileptiform Abnormality Detection component is intended to mark previously acquired sections of the patient’s EEG recordings that may correspond to epileptiform abnormalities in order to assist qualified clinical practitioners in the assessment of EEG traces. The Epileptiform Abnormality Detection component is intended to be used in patients at least one year old.
## 6. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Compared to the predicate devices, the subject device has the same intended use. The following table summarizes the substantial equivalence comparison between the subject device and the predicate devices.
| Attribute | Primary Predicate Device encevis (K240993) | Secondary Predicate Device autoSCORE (K243743) | Subject Device Ceribell Neurology Review Software | Comparison |
| --- | --- | --- | --- | --- |
| Product Codes | OMB, OLT, OMA | OMB | OMB, OLT | Same |
| Indications for Use | (Only relevant portions of the Indications for Use are replicated here) 1. encevis is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the | (Only relevant portions of the Indications for Use are replicated here) 1. autoSCORE is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid | The Ceribell Neurology Review Software is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the | The subject device and both predicates are indicated for the review, monitoring, and analysis of EEG recordings. The Indications for Use of the Artifact Reduction |
2
{6}
ceribell®
| Attribute | Primary Predicate Device encevis (K240993) | Secondary Predicate Device autoSCORE (K243743) | Subject Device Ceribell Neurology Review Software | Comparison |
| --- | --- | --- | --- | --- |
| | assessment of EEG. This device is intended to be used by qualified medical practitioners who will exercise professional judgment in using the information. 7. encevis PureEEG (Artifact Reduction) is intended to reduce EMG and electrode artifacts in a standard 10-20 EEG recording. PureEEG does not remove the entire artifact signal and is not effective for other types of artifacts. PureEEG may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified medical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms. | neurologists in the assessment of EEG. This device is intended to be used by qualified medical practitioners who will exercise professional judgment in using the information. 3. autoSCORE is intended to assess the probability that previously acquired sections of EEG recordings contain abnormalities, and classifies these into pre-defined types of abnormalities, including epileptiform and non-epileptiform abnormalities. autoSCORE does not have a user interface. autoSCORE sends this information to the EEG reviewing software to indicate where markers indicating abnormality are to be placed in the EEG. autoSCORE also provides the probability that EEG recordings include abnormalities and the type of abnormalities. The user is required to review the EEG and exercise their clinical judgement to independently make a conclusion supporting or not supporting brain disease. | assessment of EEG. This device is intended to be used by qualified clinical practitioners who will exercise professional judgment in using the information. The Artifact Reduction component is intended to reduce artifact in EEG recordings. The Artifact Reduction component does not remove the entire artifact signal and may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified clinical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms. The Epileptiform Abnormality Detection component is intended to mark previously acquired sections of the patient's EEG recordings that may correspond to epileptiform abnormalities in order to assist qualified clinical practitioners in the assessment of EEG traces. The Epileptiform Abnormality Detection component is intended to be used in patients at least one year old. | component of the subject device is shared with encevis and the Indications for Use of the Epileptiform Abnormality Detection component of the subject device is shared with autoSCORE. |
| Intended Patient Population | Adults | > 3 months | Ages 1+ | As demonstrated by clinical performance data, age range for |
3
{7}
ceribell®
| Attribute | Primary Predicate Device encevis (K240993) | Secondary Predicate Device autoSCORE (K243743) | Subject Device Ceribell Neurology Review Software | Comparison |
| --- | --- | --- | --- | --- |
| | | | | the subject device does not raise any new or different questions of safety or effectiveness. |
| Intended Location of Use | Professional healthcare facilities | Professional healthcare facilities | Professional healthcare facilities | Same |
## 7. NON-CLINICAL TESTING
Software verification and validation testing was conducted, and documentation provided as recommended by FDA Guidance for Industry and FDA Staff, “*Content of Premarket Submissions for Device Software Functions.*”
Software verification and validation activities support the safety and effectiveness of the Ceribell Neurology Review Software.
## 8. PERFORMANCE DATA
The following clinical performance data were submitted to support a determination of substantial equivalence:
### Artifact Reduction
The Artifact Reduction module is validated using a dataset of 120 EEG recordings collected from three geographically diverse clinical sites. This dataset represents real-world clinical use cases without any inclusion/exclusion criteria applied.
A Clinical Reference Standard was established by three expert EEG reviewers who identified segments containing common artifacts (EKG, eye blinks, EMG, and electrical noise). Consensus was defined as agreement between at least two of the three reviewers.
The Ceribell Artifact Reduction algorithm was compared against a legally marketed artifact reduction algorithm (Persyst) using two primary metrics: Signal-to-Noise Ratio (SNR) Improvement and Waveform Distortion. A non-inferiority analysis was performed to ensure the Ceribell algorithm effectively reduces artifact while maintaining the integrity of the underlying EEG signal.
4
{8}
ceribell®
| Performance Measure | Paired Difference [Ceribell - Persyst] [95% CI] | Result |
| --- | --- | --- |
| SNR Improvement (dB) | 1.12 [0.51, 1.74] | Pass (Non-Inferior) |
| Waveform Distortion (%) | -2.40 [-3.92, -0.88] | Pass (Non-Inferior) |
This analysis was also performed at each of the three contributing sites to ensure generalizable performance. This analysis is provided below.
| Site | SNR Improvement (Ceribell - Persyst) | Distortion (Ceribell - Persyst) |
| --- | --- | --- |
| Temple University Hospital (n=14) | Ceribell: 7.38 Persyst: 5.37 Difference: 2.01 dB [0.75, 3.26] | Ceribell: 11.6% Persyst: 14.1% Difference: -2.49% [-5.61%, 0.64%] |
| Stanford Hospital (n=14) | Ceribell: 8.06 Persyst: 6.81 Difference: 1.25 dB [0.30, 2.20] | Ceribell: 8.66% Persyst: 11.3% Difference: -2.67% [-4.26%, -1.07%] |
| Massachusetts General Hospital (n=12) | Ceribell: 8.35 Persyst: 8.30 Difference: 0.05 dB [-1.00, 1.10] | Ceribell: 13.6% Persyst: 15.5% Difference: -1.92% [-5.55%, 1.71%] |
| All sites | Ceribell: 7.95 Persyst: 6.82 Difference: 1.12 dB [0.505, 1.74] | Ceribell: 11.0% Persyst: 13.4% Difference: -2.40% [-3.92%, -0.88%] |
5
{9}
ceribell®
## Epileptiform Abnormality Detection
The Epileptiform Abnormality Detection module is validated by evaluating the performance of the epileptiform abnormality detection algorithm on a dataset of EEG recordings representative of the intended patient population. The validation dataset consisted of EEG recordings obtained from patients 1 year of age and older (n=1,362) who received continuous EEG monitoring within the hospital environment. There were no patient inclusion or exclusion criteria applied. To form the reference standard for epileptiform abnormalities, the EEG recordings were retrospectively reviewed by a panel of three expert neurologists. A two-thirds majority agreement was required to form a determination of epileptiform abnormalities.
The reviewing neurologists did not have access to any of the outputs from the Epileptiform Abnormality Detection module; the experts were fully blinded. Importantly, none of the data in the validation dataset were used for training of the epileptiform abnormality detection algorithm; the validation dataset is completely independent.
### Acceptance Criteria
Performance of the Epileptiform Abnormality Detection algorithm was assessed by evaluating the positive percent agreement (PPA) and the false positive rate per hour (FP/hr) of the algorithm compared to the expert reviewer reference standard:
- PPA: For each threshold of Epileptiform Abnormality Burden activity (Frequent, Abundant, Continuous) **Lower bound of the 95% confidence interval ≥ 70% PPA**
- FP/hr: For each threshold of Epileptiform Abnormality Burden activity (Frequent, Abundant, Continuous) **Upper bound of the 95% confidence interval ≤ 0.446 FP/hr**
### Device Performance
Performance against the acceptance criteria was assessed. In the overall dataset, the acceptance criteria were met and the Epileptiform Abnormality Detection algorithm passes. The detailed results for PPA and FP/hr are shown in the following table:
6
{10}
ceribell®
| Activity Category | Positive Percent Agreement (PPA) | 95% Confidence Interval | False Positive Rate (FP/hr) | 95% Confidence Interval | Pass / Fail |
| --- | --- | --- | --- | --- | --- |
| Epileptiform Abnormality Episodes with Epileptiform Abnormality Burden ≥ 10% (meeting ACNS definition of 'Frequent' activity) | 98.29% | [93.35, 99.44] | 0.286 | [0.270, 0.302] | Pass |
| Epileptiform Abnormality Episodes with Epileptiform Abnormality Burden ≥ 50% (meeting ACNS definition of 'Abundant' activity) | 98.95% | [94.48, 100.00] | 0.142 | [0.130, 0.154] | Pass |
| Epileptiform Abnormality Episodes with Epileptiform Abnormality Burden ≥ 90% (meeting ACNS definition of 'Continuous' activity) | 94.37% | [86.62, 98.59] | 0.052 | [0.045, 0.060] | Pass |
| Acceptance Criteria: PPA: Lower bound of the 95% confidence interval ≥ 70% PPA FP/hr: Upper bound of the 95% confidence interval ≤ 0.446 FP/hr | | | | | |
Performance was also evaluated across subject age. The results for PPA and FP/hr are shown in the following table:
| Activity Category | Age (yrs) | Positive Percent Agreement (PPA) | 95% Confidence Interval | False Positive Rate (FP/hr) | 95% Confidence Interval |
| --- | --- | --- | --- | --- | --- |
| Epileptiform Abnormality Episodes with Burden ≥10% (meeting ACNS definition of 'Frequent' activity) | 1-11 | 98.15% | [90.20, 100.00] | 0.350 | [0.315, 0.386] |
| | 12-17 | 96.3% | [81.48, 100.00] | 0.265 | [0.234, 0.297] |
| | 18+ | 98.94% | [94.38, 100.00] | 0.268 | [0.248, 0.289] |
| | Overall | 98.29% | [93.35, 99.44] | 0.286 | [0.270, 0.302] |
| | 1-11 | 100.0% | [100, 100] | 0.178 | [0.151, 0.206] |
7
{11}
ceribell®
| Epileptiform Abnormality Episodes with Burden ≥50% (meeting ACNS definition of ‘Abundant’ activity) | 12-17 | 100.0% | [100, 100] | 0.099 | [0.081, 0.123] |
| --- | --- | --- | --- | --- | --- |
| | 18+ | 98.51% | [92.31, 100] | 0.146 | [0.130, 0.163] |
| | Overall | 98.95% | [94.48, 100] | 0.142 | [0.130, 0.154] |
| Epileptiform Abnormality Episodes with Burden ≥90% (meeting ACNS definition of ‘Continuous’ activity) | 1-11 | 100.0% | [100, 100] | 0.071 | [0.054, 0.092] |
| | 12-17 | 100.0% | [100, 100] | 0.038 | [0.025, 0.056] |
| | 18+ | 92.16% | [81.63, 98.04] | 0.050 | [0.040, 0.062] |
| | Overall | 94.37% | [86.62, 98.59] | 0.052 | [0.045, 0.060] |
The below table presents the epileptiform abnormality episode count across each burden level and age subgroup:
| Activity Category | 1-11 | 12-17 | 18+ | Total |
| --- | --- | --- | --- | --- |
| ≥10% burden | 54 | 27 | 94 | 175 |
| ≥50% burden | 20 | 8 | 67 | 95 |
| ≥90% burden | 11 | 9 | 51 | 71 |
## 9. SUMMARY
The Ceribell Neurology Review Software has the same intended use as both predicate devices. In addition, it has similar technological characteristics, clinical workflow, and underlying operating principles. Differences within the Neurology Review Software have been validated through performance testing. Therefore, the Ceribell Neurology Review Software is substantially equivalent to the cleared predicate devices.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.