The °M Warmer System with °M Station is indicated for use to warm blood, blood products, colloids and crystalloid solutions prior to parenteral administration. It is intended to be used by healthcare professionals in hospital, clinical, and field environments to help prevent hypothermia. The field environment includes road ambulances.
Device Story
The °M Warmer System is an in-line fluid warming device. It consists of a sterile, single-use disposable warmer unit and the °M Station, a reusable AC mains power supply. The warmer features a parylene-coated aluminum heating chamber with resistive heating elements. It is inserted into an infusion line to warm fluids (blood, colloids, crystalloids) to 39±3°C at flow rates up to 150ml/min. The device is operated by healthcare professionals in clinical or field settings (road ambulances). Electronics and software within the warmer monitor temperature via two redundant sensors and control the heating process. Visual LED indicators provide alerts for overheat, under-heat, and low battery conditions. The system helps prevent hypothermia during fluid administration in medical emergencies or surgeries.
Clinical Evidence
Bench testing only. Testing included electrical safety and EMC compliance per IEC 60601-1 (Ed 3.2), IEC 60601-1-2 (Ed 4.1), and IEC 60601-1-12 (Ed 1.1). Performance data confirmed the system maintains fluid temperature at 39±3°C at 150ml/min flow rates, consistent with the predicate device.
Technological Characteristics
In-line fluid warmer; resistive heating elements; parylene-coated anodized aluminum heating chamber; silicone gaskets; plastic tubing. Complies with ASTM 2172:2002. Powered by AC mains (via °M Station) or rechargeable battery. Software-controlled temperature regulation. Visual LED status/alarm indicators. Two redundant temperature sensors for alarm monitoring. Sterilized disposable unit.
Indications for Use
Indicated for warming blood, blood products, colloids, and crystalloid solutions prior to parenteral administration in patients requiring normothermic fluid therapy. Intended for use by healthcare professionals in hospital, clinical, and road ambulance environments to prevent hypothermia.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 24, 2026
MEQU A/S
Lene Margrete Moesby
Senior Regulatory Affairs Specialist
Fruebjergvej 3
2100 Ø
København Ø
Denmark
Re: K260988
Trade/Device Name: °M Warmer System with °M Station
Regulation Number: 21 CFR 880.5725
Regulation Name: Infusion Pump
Regulatory Class: Class II
Product Code: LGZ, BSB
Dated: March 25, 2026
Received: March 25, 2026
Dear Lene Margrete Moesby:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260988 - Lene Margrete Moesby
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260988 - Lene Margrete Moesby
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jake K. Lindstrom -S
Jake Lindstrom, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260988 | |
| Device Name °M Warmer System with °M Station | |
| Indications for Use (Describe) The °M Warmer System with °M Station is indicated for use to warm blood, blood products, colloids and crystalloid solutions prior to parenteral administration. It is intended to be used by healthcare professionals in hospital, clinical, and field environments to help prevent hypothermia. The field environment includes road ambulances. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1 of 5
# K260988 510(k) Summary
Submitter: MEQU A/S
Address: Fruebjergvej 3
København Ø
2100
Denmark
Phone number: +45 3110 5487
Contact person: Mrs. Lene Margrete Moesby
Date prepared: 22 Apr 2026
Trade name: °M Warmer System with °M Station
Common name: In-line infusion fluid warmer
Classification name: Warmer, Thermal Infusion Liquid
(21 CFR 880.5725, product code LGZ)
Warmer, Blood, Non-Electromagnetic Radiation
(21 CFR, 864.9205, product code BSB)
Substantial equivalence is claimed to: °M Warmer System (K232107)
{5}
Page 2 of 5
# Purpose of 510(k)
The purpose of this 510(k) is to add the °M Station as an additional AC mains power source to the °M Warmer in the °M Warmer System.
# Indications for use
The °M Warmer System with °M Station is indicated for use to warm blood, blood products, colloids and crystalloid solutions prior to parenteral administration. It is intended to be used by healthcare professionals in hospital, clinical, and field environments to help prevent hypothermia.
The field environment includes road ambulances.
# Device Description
The °M Warmer System with °M Station consists of:
1. °M Warmer – A single-use, sterile, disposable in-line fluid warming unit
2. °M Station – A new multi-patient, reusable AC mains power supply providing continuous power to the °M Warmer during use.
The warmer contains a sterile fluid path with standard Luer lock connectors allowing it to be connected in the infusion fluid line. The fluid path includes a parylene coated aluminum heating chamber in which fluids passing through the warmer are heated. Heat is generated using resistive heating elements, using power supplied by the rechargeable battery pack.
The warmer contains the electronics and software to control the temperature of the chamber and thus the temperature of the outgoing fluid.
The °M Station is a transportable accessory designed to be used as an AC mains power supply for the °M Warmer in hospital/clinic settings or in ground-based ambulances.
# Substantial equivalence
The subject device °M Warmer System with °M Station is substantially equivalent to the cleared °M Warmer System. Both the cleared device and the °M Warmer System with °M Station is a portable IV fluid warming system.
The differences between the predicate and subject device do not raise any concerns regarding safety and effectiveness and equivalent performance have been demonstrated. Thus, °M Warmer System with M Station is substantially equivalent.
A detailed comparison between the predicate and the subject device is provided in the table below:
{6}
| | °M Warmer System with °M Station (Subject device) | °M Warmer System (K232107) | Comparison |
| --- | --- | --- | --- |
| Indications for Use | The °M Warmer System with °M Station is indicated for use to warm blood, blood products, colloids and crystalloid solutions prior to parenteral administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.
The field environment includes road ambulances. | The °M Warmer System is indicated for use to warm blood, blood products, colloids and crystalloid solutions prior to parenteral administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.
The field environment includes road, rotary and fixed-wing ambulances. | Difference: °M Warmer System with °M Station is more specific in the supported field environment in the indications for use. |
| Intended use | Medical emergencies or surgeries where warm fluid administration is required to treat the patient.
Whenever parenteral introduction of normothermic fluid is desired or indicated | Medical emergencies or surgeries where warm fluid administration is required to treat the patient.
Whenever parenteral introduction of normothermic fluid is desired or indicated | Same |
| User population | Healthcare professional (i.e. paramedic, nurse doctor etc.) | Healthcare professional (i.e. paramedic, nurse doctor etc.) | Same |
| User Interface / Notifications | Visual (LED) | Visual (LED) | Same |
| Notification Types | Overheat
Under heat
Low battery | Overheat
Under heat
Low battery | Same |
| Alarm Sensor | Two redundant temperature sensors. | Two redundant temperature sensors. | Same |
| Alarm Condition / Alarm indicators | Visual (RED LED) instantaneous activated from 43 °C. | Visual (RED LED) instantaneous activated from 43 °C. | Same |
| Usage Environment | Clinic and Field. The field environment includes road ambulances. | Clinic and Field. The field environment includes road, rotary and fixed-wing ambulances. | Difference: °M Warmer System with °M Station is more specific in the supported field environment. |
Page 3 of 5
{7}
| | °M Warmer System with °M Station (Subject device) | °M Warmer System (K232107) | Comparison |
| --- | --- | --- | --- |
| Components | - Disposable warmer with integrated fluid path - Power source (AC mains) | - Disposable warmer with integrated fluid path - Power source (rechargeable battery pack) - Charger | Difference: The °M Warmer System with °M Station is operated with AC mains power supply. |
| Infusion temperature | 39±3°C @ 150ml/min at 5°C input temperature | 39±3°C @ 150ml/min at 5°C input temperature | Same |
| Flow Rate | Up to 150ml/min at 5°C input temperature | Up to 150ml/min at 5°C input temperature | Same |
| Operating Environment | 15 % RH to 95 % RH (relative humidity). Temperature range of 0°C to +40°C; non-condensing. | 15 % RH to 95 % RH (relative humidity). Temperature range of 0°C to +40°C; non-condensing. | Same |
| Heating technology | Software-controlled resistive heating | Software-controlled resistive heating | Same |
| Fluid path | Sterile fluid path consisting of plastic tubing, silicone gasket and biocompatible parylene coated anodized aluminum. | Sterile fluid path consisting of plastic tubing, silicone gasket and biocompatible parylene coated anodized aluminum. | Same |
| Warmer type | Inline | Inline | Same |
| Power Source | AC mains power supply | Rechargeable battery | Difference: AC mains power supply was added as an additional power source for the °M Warmer |
| Biocompatibility | The fluid path is made of biocompatible parylene coated anodized aluminum. | The fluid path is made of biocompatible parylene coated anodized aluminum. | Same |
| Software | The software control the heating process and the operation of the device. | The software control the heating process and the operation of the device. | Same |
| Sterility | The disposable unit is provided sterile for single patient use. | The disposable unit is provided sterile for single patient use. | Same |
| Product specific Standard with which the Device Complies | ASTM 2172:2002, Standard specification for blood/Intravenous Fluid Irrigation Fluid Warmers | ASTM 2172:2002, Standard specification for blood/Intravenous Fluid Irrigation Fluid Warmers | Same |
Performance data from non-clinical studies
{8}
The following non-clinical testing was conducted to demonstrate substantial equivalence to the predicate device:
## Bench testing
Electrical Safety and EMC testing was performed in accordance with
- IEC 60601-1 Edition 3.2 2020-08 "Medical electrical equipment - Part 1: General requirements for basic safety and essential performance"
- IEC 60601-1-2 Edition 4.1 2020-09 "Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests"
- IEC 60601-1-12 Edition 1.1 2020-07 "Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment"
In all instances, the subject device functioned as intended and demonstrated equivalent performance to the predicate device.
## Conclusion
The modifications to the °M Warmer System by adding a °M Station (AC mains power supply) do not raise different questions of safety and effectiveness and are supported by risk management activities. Based on the equivalence to the predicate device, and the results of the non-clinical performance data, MEQU A/S find that the °M Warmer System with °M Station is as safe, as effective, and performs as well as the predicate device cleared under K232107.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.