K260962 · Taewoong Medical Co., Ltd. · KNS · May 21, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K260962
Device Name
Sphincterotome
Applicant
Taewoong Medical Co., Ltd.
Product Code
KNS · Gastroenterology, Urology
Decision Date
May 21, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sphincterotome is indicated for use in the selective cannulation of the Common Bile Ducts (CBD) and the transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. The Sphincterotome can also be used to inject contrast medium.
Device Story
Endoscopic electrosurgical accessory; used for selective cannulation of Common Bile Duct and sphincterotomy of Papilla of Vater/Sphincter of Oddi. Device comprises three-lumen polymer sheath, cutting wire (SUS304), and handle. Operates via physician manipulation of handle to rotate wire and application of monopolar radio frequency current from external ESU to perform incision. Sheath lumens facilitate guidewire introduction and contrast agent/saline injection. Used in clinical endoscopic settings by physicians. Provides controlled tissue incision; benefits patient by enabling minimally invasive biliary access and therapeutic intervention.
Clinical Evidence
Bench testing only. Testing included visual inspection, dimensional analysis, tensile strength, leakage, electrical resistance, endoscopy compatibility, device operability, cutting wire function, HF dielectric strength, connector retention, rotation, and thermal performance. Biocompatibility verified per ISO 10993 series. Electrical safety and EMC verified per IEC 60601 series.
Technological Characteristics
Materials: SUS304 cutting wire, PTFE sheath, ABS handle. Energy: Monopolar radio frequency current (max 700V peak). Dimensions: 1.6mm distal tip O.D., 1950mm length. Connectivity: Compatible with standard endoscope working channels (>=2.8mm) and guidewires (0.021"-0.035"). Sterilization: Ethylene Oxide (EO), SAL 10^-6. Standards: ISO 10993-1, IEC 60601-1, IEC 60601-2-2, IEC 60601-2-18.
Indications for Use
Indicated for patients requiring selective cannulation of the Common Bile Duct and transendoscopic sphincterotomy of the Papilla of Vater or Sphincter of Oddi; includes capability for contrast medium injection.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
Autotome™ Pro RX 39 Sphincterotome/Autotome™ Pro RX 44 Sphincterotome (K232162)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 21, 2026
Taewoong Medical Co., Ltd.
% Matthew Krueger
Principal Consultant
Biologics Consulting Group, Inc.
100 Daingerfield Rd.
Suite 101
Alexandria, Virginia 22314
Re: K260962
Trade/Device Name: Sphincterotome
Regulation Number: 21 CFR 876.4300
Regulation Name: Endoscopic Electrosurgical Unit And Accessories
Regulatory Class: Class II
Product Code: KNS
Dated: March 23, 2026
Received: March 23, 2026
Dear Matthew Krueger:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260962 - Matthew Krueger
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260962 - Matthew Krueger
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANTHONY LEE -S
Anthony Lee, Ph.D., MBA
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260962 | ? |
| Please provide the device trade name(s). | | ? |
| Sphincterotome | | |
| Please provide your Indications for Use below. | | ? |
| Sphincterotome is indicated for use in the selective cannulation of the Common Bile Ducts (CBD) and the transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. The Sphincterotome can also be used to inject contrast medium. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
Taewoong Medical
K260962
Page 1 of 7
# 1. SUBMITTER INFORMATION
Applicant: Taewoong Medical
Contact: Yongjin Jeff Kim
Phone: +82 70 4649 1543
Email: jinjeff@stent.net
Address: 14, Gojeong-ro, Wolgot-myeon, Gimpo-si, Gyeonggi-do, 10022, Republic of Korea
# 2. CORRESPONDENT INFORMATION
Contact: Matthew Krueger
Title: Principal Consultant
Firm: Biologics Consulting Group
Phone: (667) 352-2578
Email: mkrueger@biologicsconsulting.com
Address: 100 Daingerfield Rd., Suite 101
Alexandria, VA 22314
# 3. DATE PREPARED: MARCH 20, 2026
# 4. DEVICE INFORMATION
Device Name: Sphincterotome
Regulation Number: 21 CFR 876.4300
Regulation Name: Endoscopic electrosurgical unit and accessories.
Product Code: KNS
Regulatory Class: Class II
# 5. PREDICATE DEVICE INFORMATION
Primary Predicate Device Name: Autotome™ Pro RX 39 Sphincterotome/Autotome™ Pro RX 44 Sphincterotome
510(k) Number: K232162
Manufacturer: Boston Scientific Corporation
Page 1 of 7
{5}
510(k) Summary
Taewoong Medical
K260962
Page 2 of 7
Secondary Predicate Device Name: Autotome™ RX Model
510(k) Number: K013153
Manufacturer: Boston Scientific Corporation
## 6. DEVICE DESCRIPTION
The Sphincterotome is an endoscopic electrosurgical accessory and consists of three parts: a cutting wire, a sheath, and a handle. The cutting wire is insulated with a non-conductive polymer tube and features a pre-curved and tapered tip. The sheath is made of a three-lumen polymer tube: one lumen for a cutting wire that transmits high-frequency current to the target site, another lumen for delivering contrast agents or saline, and the third lumen for introducing a guidewire. The proximal end of the sheath and cutting wire assembly terminates in a handle. The handle includes three parts: a plug that transmits high frequency electric current from the ESU to the cutting wire, an injection port, and a guidewire port. Physicians can manipulate the cutting wire by rotating the handle during the procedure.
## 7. INDICATIONS FOR USE
Sphincterotome is indicated for use in the selective cannulation of the Common Bile Ducts (CBD) and the transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. The Sphincterotome can also be used to inject contrast medium.
## 8. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Page 2 of 7
{6}
510(k) Summary
Taewoong Medical
K260962
Page 3 of 7
Table 1 provides the technological comparison.
Page 3 of 7
{7}
510(k) Summary
Taewoong Medical
K260962
Page 4 of 7
Table 1: Technological Comparison to Predicate Devices
| | Subject Device | Primary Predicate Device (K232162) | Secondary Predicate Device (K013153) |
| --- | --- | --- | --- |
| Device Name | Sphincterotome | Autotome™ Pro RX 39 Sphincterotome Autotome™ Pro RX 44 Sphincterotome | Autotome™ RX Model |
| Common Name | Sphincterotome | Sphincterotome | Sphincterotome |
| Manufacturer | Taewoong Medical | Boston Scientific | Boston Scientific |
| Clinical Condition | An incision of papilla | An incision of papilla | An incision of papilla |
| Application | Common Bile Duct, Papilla of Vater and/or Sphincter of Oddi | Common Bile Duct, Papilla of Vater and/or Sphincter of Oddi | Papilla of Vater and/or Sphincter of Oddi |
| Procedure | Endoscopic | Endoscopic | Endoscopic |
| Size Specification | - Distal tip O.D.: 1.6 mm - Tip length: 7 mm - Cutting wire length: 20/25/30 mm - Length: 1950 mm | - Distal tip O.D.: 1.47 (4.4Fr) mm - Tip length: 5 mm - Cutting wire length: 25 mm - Length: 2000 mm | - Distal tip O.D.: 1.3 (3.9Fr) / 1.47 (4.4Fr) / 1.63 (4.9Fr) mm - Tip length: 5 mm - Cutting wire length: 20 / 30 mm - Length: 2000 mm |
| Materials | - Cutting wire: SUS304 - Sheath: PTFE tube - Handle: ABS | - Cutting wire: Metal - Sheath: Polymer | - Cutting wire: Metal - Sheath: Polymer |
| Injection | Yes | Yes | Yes |
| Compatible endoscopy working channel | 2.8 mm or larger | 2.8 mm or larger | 2.8 mm or larger |
| Compatible guidewire | Traditional guidewire (4.5m) or short guidewire (2.6m) with 0.021” ~ 0.035” diameter | Traditional guidewire (4.5m) or short guidewire (2.6m) with 0.021” ~ 0.035” diameter | Traditional guidewire (4.5m) or short guidewire (2.6m) with 0.021” ~ 0.035” diameter |
| Energy Used | Monopolar Radio Frequency Current | Monopolar Radio Frequency Current | Monopolar Radio Frequency Current |
| Maximum Voltage Rating | 700 V peak (1400 V peak-to-peak) | 750 V peak (1500 V peak-to-peak) | 750 V peak (1500 V peak-to-peak) |
| Insulation | Yes | Yes | No |
| Packaging | Single-use EO sterilized pouch with one device per pouch | Single-use EO sterilized pouch with one device per pouch | Single-use EO sterilized pouch with one device per pouch |
| Shelf Life | 3 years | 3 years | 3 years |
Page 4 of 7
{8}
510(k) Summary
Taewoong Medical
K260962
Page 5 of 7
| | Subject Device | Primary Predicate Device (K232162) | Secondary Predicate Device (K013153) |
| --- | --- | --- | --- |
| Sterilization method / SAL | EO Sterilization, SAL of 10^{-6} | EO Sterilization, SAL of 10^{-6} | EO Sterilization, SAL of 10^{-6} |
| Biocompatibility | Conforms with ISO 10993-1 | Conforms with ISO 10993-1 | Conforms with ISO 10993-1 |
| Electromagnetic compatibility and Electrical Safety | Conforms with IEC 60601-1 | Conforms with IEC 60601-1 | Conforms with IEC 60601-1 |
Page 5 of 7
{9}
510(k) Summary
Taewoong Medical
K260962
Page 6 of 7
# 9. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
## Biocompatibility Testing
Biocompatibility testing was conducted to the following standards:
- ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation
- ISO 10993-10:2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- USP 43 <85>, Bacterial Endotoxin Testing
- USP 43 <161>, Bacterial Endotoxin Testing
## Electrical Safety
The subject device was tested against the following standard for electrical safety:
- IEC 60601-1:2005/AMD1:2012/AMD2:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
## Electromagnetic Compatibility (EMC)
The subject device was tested against the following standards for electromagnetic compatibility:
- IEC 60601-1-2:2014/AMD1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-2:2017/AMD1:2023, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- IEC 60601-2-18:2009, Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
## Software
Not applicable as the subject device does not contain software.
## Performance Testing
The following performance testing was conducted:
- Visual inspection
- Dimension
- Tensile Strength
{10}
510(k) Summary
Taewoong Medical
K260962
Page 7 of 7
- Leakage
- Resistance
- Packaging compatibility
- Endoscopy compatibility
- Device Operability
- Cutting wire function
- HF Dielectric Strength
- Interconnection conditions for HF applications
- Detachable connector retention
- Rotation
- Thermal performance
## 10. CONCLUSION
The results of the performance testing described above demonstrate that the Sphincterotome performs equivalently as compared to the primary and secondary predicate devices and supports a determination of substantial equivalence.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.